Bard Flip-Flo Catheter ValveRecalls & reported problems across all models
C. R. Bard, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls
- 0
We found 2 models of Bard Flip-Flo Catheter Valve by C. R. Bard, Inc. in the FDA's device registry (GUDID). Together they are named in 21 adverse event reports in the FDA's MAUDE database, received between Jul 24, 2018 and Sep 14, 2026 — classified as 5 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Component Missing” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Bard Flip-Flo Catheter Valve problems
Device problems
- Component Missing3
- Fracture3
- Restricted Flow rate3
- Adverse Event Without Identified Device or Use Problem2
- Fluid/Blood Leak2
- Material Twisted/Bent2
- Complete Blockage1
- Decrease in Pressure1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
- Urinary Tract Infection2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Brain Injury1
- Exposure to Body Fluids1
- Head Injury1
Bard Flip-Flo Catheter Valve models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Bard® Flip-Flo™ Catheter Valve BFF5COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER12 reports
- Bard® Flip-Flo™ Catheter Valve BFF20COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER9 reports
Recent Bard Flip-Flo Catheter Valve adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 14, 2026Fracture
Patient problems: Unspecified Nervous System Problem
MAUDE report 50765286 ↗MalfunctionJun 25, 2026Complete Blockage; Difficult to Open or Close
IT WAS REPORTED THAT HEALTH PROFESSIONAL HAD BEEN RECEIVING A SIGNIFICANT NUMBER OF COMPLAINTS FROM OUR UROLOGY DEPARTMENT OVER THE PAST MONTH REGARDING THE FLIPFLO VALVES. THE VALVES FREQUENTLY BECOME BLOCKED, CAUSING THE LEVER TO BREAK OR BEND DURING HANDLING. AS A RESULT, OUTPATIENT PATIENTS OFTEN NEED TO RETURN MULTIPLE TIMES FOR A REPLACEMENT VALVE. ATTACHED, YOU WILL FIND A PHOTO OF A DEFECTIVE VALVE ALONG WITH THE CORRESPONDING LOT NUMBER. WE HAVE NOT ENCOUNTERED THESE ISSUES IN THE PAST. REPLACEMENT TO BE DISCUSSED WITH THE CUSTOMER - PLEASE REACH OUT TO THE RESPONSIBLE AM LIEVE VIVIER FOR THIS (CURRENTLY ON VACATION UNTIL JUNE 28TH) PLEASE ALSO VERIFY OTHER REPORTED COMPLAINTS, BY OTHER CUSTOMERS OR OTHER COUNTRIES AND SHARE YOUR FEEDBACK. INCREASED COMPLAINTS ABOUT FLIPFLO VALVES OVER THE PAST MONTH DUE TO FREQUENT BLOCKAGE AND BREAKAGE OF THE LEVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25630815 ↗MalfunctionAug 12, 2025Fracture; Fracture
IT WAS REPORTED THAT THE FLIP FLOW CATHETER VALVE HAD CRACKS IN THE HUB THAT WAS INSERTED INTO THE CATHETER AND PICTURE IN THE ATTACHMENT WITH BATCH NGJW3053. DUE TO THE VALUE OF THE PRODUCT, THE CUSTOMER WOULD PREFER NOT TO REQUEST A CREDIT NOTE OR SIMILAR FOR THIS, BUT WANTED TO REPORT IT IN CASE THIS WAS A MANUFACTURING DEFECT IN THIS BATCH. CRACKS IN THE HUB THAT IS INSERTED INTO THE CATHETER (SEE PICTURE IN THE ATTACHMENT). BATCH: NGJW3053.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22781994 ↗MalfunctionJul 2, 2025Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT THE SOMETIMES WHEN USING THE FLIP FLOW CATHETER VALVES THE URINE WAS TRICKLING OUT AND NOT FLOWING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22384682 ↗MalfunctionMar 28, 2025Fracture; Fracture
IT WAS REPORTED THAT A BATCH OF BARD FLIP FLOW CATHETER VALVE, THAT HAD CRACKS IN THEM. THEY WERE RETURNED BY A PATIENT AS WHEN PATIENT HAD BEEN USING, THEY HAVE HAD LEAKED, AND PATIENT HAD NOTICED THESE CRACKS. 2 BOXES OF BATCH NGJU3636 WHERE ALL THE TAPS WERE AFFECTED. PATIENT WAS EXPOSED TO HAZARDOUS, BLOOD, OR BODILY FLUIDS. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED.
Patient problems: Exposure to Body Fluids
MAUDE report 21721835 ↗MalfunctionMar 26, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE CUSTOMER COULD NOT CONNECT THE FLIP-FLO TO AN OVERNIGHT BAG OR A LEG BAG. IT WAS STATED THAT THE CUSTOMER HAD BIG ISSUE WAS WITH THE NIGHT BAG AND CONNECTION TO THE FLIP-FLO VALVE COULD BEND AND BLOCK OFF THEIR BLADDER COMPLETELY AND IT DID NOT ALLOW VOIDING/EMPTYING PROPERLY, ALSO STATED THAT CUSTOMER WAS HAVING TO TAPED IT TO ENSURE IT STAY ON AND STIFF. THE REPRESENTATIVE DID OFFER THE SOLUTION OF REMOVING THE FLIP-FLO VALVE AND CONNECTING DIRECTLY TO THE DRAINAGE, BUT THEY STATED IT WOULD STILL NOT FIX THE ISSUE WITH THE VALVE BENDING OVERALL. PER FOLLOW UP VIA PHONE ON 15FEB2025, IT WAS REPORTED THAT THE CUSTOMER PROVIDED SUGGESTIONS ON HOW TO MAKE THE DRAIN BAGS BETTER WITH THE FLIP FLO VALVES. OFFERED TO SEND IN A DRAWING ON HOW TO MAKE THEY BETTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21698683 ↗MalfunctionMar 14, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE PRODUCT BARD FLIP-FLO CATHETER VALVE BXBFF5 WITH LOT NUMBER NGHZ1048 AND EXPIRATION DATE (B)(6) 2028, HAD REGULAR PROBLEMS AS THEY DO NOT GO THE WAY THEY ARE SUPPOSED TO GO, THEY ARE STIFF HARD TO GET CLOSED AFTER 1 OR 2 WEEKS. ALSO, THEY BROUGHT IT UP WITH THE UROLOGY NURSE THEY WERE AWARE. IT WAS GETTING EXPENSIVE AS THE PATIENT NEEDED 2 OR 3 IN 6 WEEKS. AMONG WHICH ONE ALREADY BROKEN OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21605301 ↗MalfunctionAug 27, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT PREVIOUS DAY THE PATIENT HAD CHANGED THE FLIP FLOW VALVE. WOKE UP TO WET SHEETS ETC. AS THE VALVE WAS LEAKING FROM THE TURNING POINT, IT WAS IN THE CLOSED POSITION ALL NIGHT, X1 AFFECTED OUT OF X5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20075138 ↗MalfunctionApr 24, 2024Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT THE FLIP FLOW CATHETER VALVE WAS FAULTY. IT WOULD NOT WORK. PER ADDITIONAL INFORMATION VIA EMAIL FROM IBC ON 01APR2024, URINE WOULD NOT FLOW THROUGH, AND URINE WAS NOT FLOWING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19171979 ↗InjuryAug 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT THE PATIENT EXPERIENCED URINARY TRACT INFECTION (UTI) PRIOR TO THE DEVICE PLACEMENT AND SYMPTOMATIC URINARY TRACT INFECTION (UTI) COMPLICATIONS WERE DIRECTLY ATTRIBUTABLE TO THE DEVICE OCCURRED 4 DAYS AFTER THE DEVICE PLACEMENT AND NO MEDICAL OR SURGICAL INTERVENTION WAS NEEDED IN RELATION TO BFF20 - FLIP-FLO CATHETER VALVE WITH 180° LEVER TAP. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED FOR URINARY TRACT INFECTION.
Patient problems: Urinary Tract Infection
MAUDE report 17624102 ↗MalfunctionJul 28, 2023Inaccurate Flow Rate; Inaccurate Flow Rate
IT WAS REPORTED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT NO, THEY WERE NOT ABLE TO SUCCESSFULLY USE THE DEVICE FOR URINE DRAINAGE BECAUSE IT HAD TROUBLE DRAINING THE URINE IN RELATION TO BFF20 ¿ FLIP-FLO CATHETER VALVE WITH 180° LEVER TAP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17418891 ↗InjuryJul 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IN AN ONLINE SURVEY A PHYSICIAN STATED THAT THE PATIENT EXPERIENCED UTI (URINARY TRACT INFECTION) PRIOR TO THE DEVICE PLACEMENT AND 15 DAYS AFTER THE DEVICE PLACEMENT NON-SYMPTOMATIC UTI COMPLICATIONS WERE DIRECTLY ATTRIBUTABLE TO THE DEVICE, AND PLACEMENT OF WIRED CATHETER WAS NEEDED IN RELATION TO BFF20 ¿ FLIP-FLO CATHETER VALVE WITH 180° LEVER TAP. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED FOR URINARY TRACT INFECTION.
Patient problems: Urinary Tract Infection
MAUDE report 17349446 ↗MalfunctionJun 15, 2023Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE NURSE COMPLAINED THAT THE FOLEY CATHETER SLIPPED OUT AUTOMATICALLY. THE NURSE WENT TO THE BEDSIDE TO CHECK AND FOUND THAT THE BLADDER OF THE FOLEY CATHETER WAS DEFLATED, AND CHECKED THAT THE PATIENT'S URETHRAL OPENING WAS NOT DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17133864 ↗MalfunctionMay 11, 2023Component Missing; Component Missing
IT WAS REPORTED THAT THE PATIENT RECEIVED 5 BFF5 BUT 2 OF THE BOXES WERE EMPTY. ALL 5 BOXES WERE SEALED, BUT PATIENT ONLY RECEIVED 3 VALVES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16911225 ↗MalfunctionApr 4, 2023Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT THE CUSTOMER HAD BROUGHT SOME FLIP FLOW CATHETER VALVES AND STATED THAT THEY WERE NOT DRAINING PATIENT'S BLADDER UNLESS ATTACHED TO A BAG. STATED THAT THEY HAD PROBLEMS WITH THEIR CURRENT CATHETER VALVES AS IT WAS NOT EMPTYING UNLESS ATTACHED TO A LEG OR NIGHT BAG. WHEN THE TAP WAS OPENED, THERE WAS NO FLOW BUT, AS SOON AS THEY ATTACHED IT TO A BAG, THEIR BLADDER EMPTIED. STATED THAT THEY PREVIOUSLY HAD VALVES WHICH WOULD EMPTY THEIR BLADDER VIA THE CATHETER DIRECTLY JUST BY TURNING THE LEVER. THE CUSTOMER ASKED WHETHER THERE WAS SOMETHING WRONG WITH THE VALVE OR WAS THERE ANOTHER POSSIBLE REASON FOR THIS. THE CUSTOMER HAD TRIED SEVERAL WITH THE SAME RESULT. THE REPRESENTATIVE GAVE THEM THE NURSE ADVICE LINE AND ADVISED THAT THEY MIGHT BE ABLE TO HELP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16675158 ↗
Other C. R. Bard, Inc. products
- Bardex I.C. Foley Catheter26 models · 3,607 reports
- Bardex Lubricath Foley Catheter38 models · 1,670 reports · 1 recall
- Bardex Lubri-Sil I.C. All-Silicone Foley Catheter26 models · 1,518 reports
- Magic3 Coude Intermittent Catheter with Sure-Grip Sleeve5 models · 1,466 reports
- Magic3 Intermittent Catheter with Sure-Grip Sleeve14 models · 1,310 reports
- Bardia Foley Catheter Silicone Coated19 models · 1,283 reports
- PureWick Female External Catheter5 models · 1,184 reports
- Bardex All-Silicone Foley Catheter12 models · 1,170 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page