BD EpiCenter Data Management SystemRecalls & reported problems across all models
Becton, Dickinson And Company · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one BD EpiCenter Data Management System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 3
We found 3 models of BD EpiCenter Data Management System by Becton, Dickinson And Company in the FDA's device registry (GUDID). Together they are named in 21 adverse event reports in the FDA's MAUDE database, received between Feb 17, 2021 and Jun 5, 2024 — classified as 21 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (17 reports).
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
BD EpiCenter Data Management System recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 31, 2024Open, Classified
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Z-0586-2025 · Root cause: Software design · Recall details · FDA record ↗
- Oct 31, 2024Open, Classified
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Z-0585-2025 · Root cause: Software design · Recall details · FDA record ↗
- Oct 31, 2024Open, Classified
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Z-0584-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported BD EpiCenter Data Management System problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings17
- False Positive Result3
- False Negative Result2
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- No Consequences Or Impact To Patient1
BD EpiCenter Data Management System models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- BD EpiCenter™ Data Management System 441007CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE1 recall21 reports
- BD EpiCenter™ Data Management System 440887Calculator/data processing module, for clinical use1 recall0 reports
- BD EpiCenter™ Data Management System 440981Calculator/data processing module, for clinical use1 recall0 reports
Recent BD EpiCenter Data Management System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHEN USING THE BD EPICENTER¿ SINGLE USER SOFTWARE, DUE TO A NURSE'S ERROR, THE ACCESS NUMBER WAS INCORRECTLY LINKED TO THE WRONG PATIENT. NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19467137 ↗MalfunctionOct 13, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING USE WITH THE BD EPICENTER¿ SINGLE USER SOFTWARE, A PATIENT RESULT WAS MISASSOCIATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER COMPLAINTS THEIR PATIENT MRN IN EPICENTER IS BELONGS TO DIFFERENT PATIENT VIA THEIR LIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17935947 ↗MalfunctionOct 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING USE WITH THE BD EPICENTER¿ SINGLE USER SOFTWARE, PATIENT RESULTS WERE MISASSOCIATED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM ITALIAN: PROBLEM ASSOCIATING ACCESS NUMBER (B)(4) WITH THE RIGHT REGISTRY. EPICENTER CONTINUES TO SHOW A PATIENT FROM 2022 WITH THE ACCESSION NUMBER MENTIONED ABOVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17917007 ↗MalfunctionSep 13, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED HAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE THERE WAS MISASSOCIATION PATIENT TO SPECIMEN MISASSOCIATION SPECIMEN TO TEST MISASSOCIATION OF DATA TEST TO RESULT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. IT WAS REPORTED BY THE CUSTOMER THAT EPICENTER RECEIVED OLD PATIENT DEMOGRAPHICS FOR A RECYCLED ACCESSION NUMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17740527 ↗MalfunctionAug 22, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE, MISASSOCIATED DATA OF ONE SAMPLE. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ERROR 21220: THE SAME ORGANISM ID FOR ISOLATE NUMBER 3 ASSIGNED TO SAMPLE X COLLECTED ON (B)(6) 2023 WAS ISOLATED FROM THE SAME PATIENT ON (B)(6) 2023 FOR SAMPLE Y. THE TEST HAS ALREADY BEEN REPORTED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17597672 ↗MalfunctionAug 16, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE, THERE IS A MISASSOCIATION OF DATA LINKED TO ONE SPECIMEN. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE DOCTOR REPORTS AN ANOMALY WHEREBY, FOR EXAMPLE, IF A BOTTLE IS POSITIVE, IT IS DISCHARGED AS POSITIVE BY THE INSTRUMENT - AND EPICENTER ALSO REPORTS IT AS POSITIVE. BUT THEN AS SOON AS OTHER VIALS LINKED TO THE SAME ACCESS NUMBER/PATIENT ARE FINALIZED, EPICENTER REPORTS ALL THE VIALS, INCLUDING THE ONE PREVIOUSLY DISCHARGED AND POSITIVE, AS NEGATIVE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17558837 ↗MalfunctionAug 3, 2023False Positive Result; False Positive Result
IT WAS REPORTED THAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE AMIKACIN SUSCEPTIBILITY IS BEING INCORRECTLY INTERPRETED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPROTER: CUSTOMER IS REPORTING AMIKACIN SUSCEPTIBLITY IS BEING INCORRECTLY INTERPRETED BY EPICENTER. IT IS FALSELY RESULTING SUSCEPTIBLE ON RESULTS THAT SHOULD BE INTERMEDIATE TO PATIENT CHARTS, SO THIS IS VERY MUCH IMPACTING PATIENT CARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17455491 ↗MalfunctionJul 21, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE, THERE WAS MISASSOCIATION OF SEQUENCE NUMBER 449380113245. THERE WAS NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ERROR 4005 RELATED TO SEQUENCE NUMBER 449380113245. MADE A CHANGE OF ASSOCIATION OF SAMPLES AS THE ACCESS NUMBER HAD BEEN READ AND REGISTERED INCORRECTLY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17370874 ↗MalfunctionJul 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING USE WITH THE BD EPICENTER¿ SINGLE USER SOFTWARE, PATIENT DATA WAS ASSOCIATED INCORRECTLY. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM ITALIAN: INCORRECT ASSOCIATION BLOOD CULTURES MASTER DATA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17300983 ↗MalfunctionJul 10, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE THERE WAS MIS ASSOCIATION OF ONE PATIENT. THERE WAS NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ERROR 2612 ASSOCIATION ERROR ERROR MESSAGE APPEARED THE DAY BEFORE AT 9:50 P.M., PREVIOUS MESSAGE: THE LIS HAS CHANGED THE NAME OF PATIENT X".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17291840 ↗MalfunctionApr 12, 2023False Negative Result; False Negative Result
IT WAS REPORTED THAT BD EPICENTER¿ SINGLE USER SOFTWARE ANTIBIOTICS ARE NOT BEING REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED ACINETOBACTER THAT WAS RAN ON INSTRUMENT 1 AND MEROPENEM WAS NOT DONE BY PHOENIX. WE NEED TO CHECK AGAIN WHY INSTRUMENT 1 IS NOT DOING THE MEROPENEM BUT IT¿S DONE ON INSTRUMENT 2
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16729096 ↗MalfunctionApr 5, 2023False Negative Result; False Negative Result
IT WAS REPORTED THAT BD EPICENTER¿ SINGLE USER SOFTWARE REPORTS WILL NOT RECEIVE MIC FROM EPI. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THAT MEROPENEM SOMETIMES REPORTS AND DO NOT ON ACINETOBACTER FOR NMIC 311 NO MIC RESULT DUE INSTRUMENT CONFIGURATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16685909 ↗MalfunctionMar 2, 2023False Positive Result; False Positive Result
IT WAS REPORTED THAT BD EPICENTER¿ SINGLE USER SOFTWARE PATIENT RESULTS WERE IMPACTED BY INCORRECT RESULTS FOR TETRACYCLINE RESISTANCE. CUSTOMER DID NOT PROVIDE INFORMATION ON TREATMENT CHANGE TO PATIENT OR AMOUNT OF SAMPLES AFFECTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER RESPONDED THAT ALL TETRACYCLINE RESULTS WHERE THE RULE WAS FIRED BETWEEN (B)(6) 2022 AND (B)(6) 2022 IMPACTED PATIENT RESULTS. TETRACYCLINE FOR ESBL ISOLATES WAS CHANGED TO RESISTANT AS A RESULT OF RULE 1487 FOR PATIENTS FROM (B)(6) 2022 TO (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16475779 ↗MalfunctionSep 9, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAD PROBLEMS FINALIZING A PANEL ON BD EPICENTER¿ SINGLE USER SOFTWARE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: "PROBLEMS FINALIZING A PANEL (IT IS FINALIZED BUT EPICENTER DOES NOT ALLOW FLAGGING TO SAVE). URGENT APPLICATION SUPPORT REQUEST SENT TO SPECIALISTS."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15396909 ↗MalfunctionJun 7, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE BD EPICENTER¿ SINGLE USER SOFTWARE ACQUIRED MIS-ASSOCIATION DUE TO USE ERROR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WRONG ACCEPTANCE OF SAMPLES CLEAR PROBLEM/ EPICENTER FAILURE: NO. CAUSE OF EPICENTER FAILURE: EXTERNAL ERROR. RESOLUTION OF THE FAILURE/ WHAT WAS DONE TO RESOLVE THE EPICENTER FAILURE: THE NURSE WHO PERFORMED THE SAMPLING REVERSED THE ACCESS NUMBER LABELS ON THE BACTEC BOTTLES BY ENTERING THE RESERVATION OF THE AEROBIC BLOOD CULTURE ON THE ANAEROBIC BOTTLE AND THE ACCEPTANCE OF THE ANAEROBIC BLOOD CULTURE ON THE AEROBIC BOTTLE. HAS REVERSED THE ACCESS NUMBER LABELS ON THE BACTEC BOTTLES BY INSERTING THE RESERVATION OF THE AEROBIC BLOOD CULTURE ON THE ANAEROBIC BOTTLE AND THE ACCEPTANCE OF THE ANAEROBIC BLOOD CULTURE ON THE AEROBIC BOTTLE. SOLVED BY EXPLAINING TO THE CUSTOMER HOW TO ASSOCIATE THE RIGHT SAMPLE WITH THE RIGHT BOTTLE. VERIFIED WITH THE CUSTOMER PASSAGE OF THE RESISTANCE MARKERS AND NOTES TO THE REPORT. CHECKED FOR CORRECT OPERATION. INCORRECT LABELING BY NURSES OF BLOOD CULTURE SAMPLING."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14628481 ↗
Other Becton, Dickinson And Company products
- BD Vacutainer SST10 models · 4,985 reports · 3 recalls
- BD Luer-Lok48 models · 5,117 reports
- BD Insyte Autoguard33 models · 4,465 reports · 8 recalls
- BD SafetyGlide17 models · 2,185 reports
- BD Vacutainer Eclipse Blood Collection Needle4 models · 1,880 reports
- BD Vacutainer Push Button Blood Collection Set9 models · 1,831 reports
- BD Nexiva25 models · 1,614 reports · 1 recall
- BD BACTEC MGIT Mycobacteria Growth Indicator Tubes2 models · 1,554 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page