BD Prevue II Peripheral Vascular Access SystemRecalls & reported problems across all models
Bard Access Systems, Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 158
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 6 models of BD Prevue II Peripheral Vascular Access System by Bard Access Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 158 adverse event reports in the FDA's MAUDE database, received between Apr 21, 2023 and May 29, 2026 — classified as 158 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (112 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported BD Prevue II Peripheral Vascular Access System problems
Device problems
- Poor Quality Image112
- No Apparent Adverse Event34
- Defective Component31
- Application Program Freezes or Fails to Launch13
- Break8
- Product Quality Problem8
- No Display/Image2
- Imprecision1
Patient problems
- No Clinical Signs, Symptoms or Conditions158
BD Prevue II Peripheral Vascular Access System models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- BD Prevue II Peripheral Vascular Access System 9770703System, image processing, radiological93 reports
- BD Prevue II Peripheral Vascular Access System 9770700Transducer, ultrasonic, diagnostic59 reports
- BD Prevue II Peripheral Vascular Access System 9770704Transducer, ultrasonic, diagnostic3 reports
- BD Prevue II Peripheral Vascular Access System 9770700RTransducer, ultrasonic, diagnostic1 report
- BD Prevue II Peripheral Vascular Access System 9770704RSystem, imaging, pulsed echo, ultrasonic1 report
- BD Prevue II Peripheral Vascular Access System 9770709System, imaging, pulsed echo, ultrasonic1 report
Recent BD Prevue II Peripheral Vascular Access System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2026Poor Quality Image
IT WAS REPORTED THAT PROBE IS GIVING VERY DARK IMAGES. TRIED PROBE ON 3 DIFFERENT CONSOLES, ISSUE FOLLOWED THE PROBE. NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333465 ↗MalfunctionMay 5, 2026Poor Quality Image
IT WAS REPORTED THAT DEVICE HAD POOR IMAGE QUALITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25088905 ↗MalfunctionMay 4, 2026Poor Quality Image
IT WAS REPORTED THAT THE DEVICE HAS POOR IMAGE QUALITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25069317 ↗MalfunctionApr 22, 2026Product Quality Problem
IT WAS REPORTED THAT PROBE FAILED ELEMENT TEST. (B)(6) 2026 IT WAS FOUND DURING AN INVESTIGATION THERE IS A DARK SPOT IN THE IMAGE DUE TO THE ELEMENT FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954291 ↗MalfunctionApr 6, 2026Break
IT WAS REPORTED THAT WIRES EXPOSED FROM STRAIN RELIEF. NO FURTHER DETAILS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24794107 ↗MalfunctionMar 24, 2026Poor Quality Image
IT WAS FOUND DURING EVALUATION AT BD SERVICE CENTER: THE PROBE ASSESSMENT TEST DISPLAYED VARYING NUMBERS FROM 1 TO UP TO 4 RED "X'S" INDICATING TRANSDUCER ELEMENT FAILURE. NUMEROUS TESTS WERE RUN TO VERIFY THESE FAILURES. THE ULTRASOUND IMAGE THAT DISPLAYS FROM THIS PROBE IS VERY DARK. THIS ISSUE IS INTERMITTENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24676962 ↗MalfunctionMar 23, 2026Application Program Freezes or Fails to Launch
IT WAS REPORTED UNIT FROZE DURING PROCEDURE. UNIT IS RUNNING ON SOFTWARE VERSION 1.1.0, SENDING UPDATE DRIVE TO SEE IF THE PROBLEM CONTINUES. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24669266 ↗MalfunctionMar 18, 2026Poor Quality Image
IT WAS REPORTED UNIT WAS DROPPED AND NOW HAS IMAGE ISSUES. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24629780 ↗MalfunctionMar 10, 2026Product Quality Problem
IT WAS REPORTED PROBE IS FAILING ASSESSMENT TEST. NO FURTHER INFORMATION WAS PROVIDED. 6 MARCH 2026 EVALUATION FOUND ELEMENT FAILURE RESULTED IN A DARK AREA ON THE RIGHT SIDE OF THE ULTRASOUND IMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24563155 ↗MalfunctionMar 9, 2026Poor Quality Image
IT WAS REPORTED PROBE IS GIVING NOISE IN IMAGE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547662 ↗MalfunctionMar 2, 2026Poor Quality Image
IT WAS REPORTED CUSTOMER HAVING NOISE ISSUES ON THEIR PREVUE II. THEY TRIED MULTIPLE PROBES; ALL HAVING THE SAME ISSUE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490042 ↗MalfunctionFeb 27, 2026Break
IT WAS REPORTED THAT THE CORD HAS BEEN COMPLETELY PULLED OUT AT THE STRAIN RELIEF/EXPOSING WIRING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24474228 ↗MalfunctionFeb 19, 2026Poor Quality Image
IT WAS REPORTED SHADOWS ON THE IMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24397169 ↗MalfunctionFeb 5, 2026Poor Quality Image; Product Quality Problem
IT WAS REPORTED THE PROBE IS GIVING BLACK LINES ON IMAGES. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24273745 ↗MalfunctionFeb 5, 2026Poor Quality Image; Product Quality Problem
IT WAS REPORTED THE PROBE IS GIVING BLACK LINES ON IMAGES. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24273456 ↗
Other Bard Access Systems, Inc. products
- Port Access Needle122 models · 3,744 reports · 3 recalls
- Groshong NXT ClearVue Catheter8 models · 3,266 reports
- PowerPort ClearVUE27 models · 1,993 reports · 15 recalls
- Site~Rite 8 Ultrasound System11 models · 1,965 reports
- PowerPort M.R.I., ChronoFlex22 models · 1,642 reports · 1 recall
- PowerPICC Provena116 models · 1,337 reports · 2 recalls
- PowerPort M.R.I., Groshong10 models · 1,285 reports
- Navarre15 models · 1,228 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.