CAMINORecalls & reported problems across all models
Integra Lifesciences Corporation · 6 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 0
We found 6 models of CAMINO by Integra Lifesciences Corporation in the FDA's device registry (GUDID). Together they are named in 32 adverse event reports in the FDA's MAUDE database, received between Nov 16, 2017 and May 26, 2021 — classified as 4 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Output Problem” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported CAMINO problems
Device problems
- Output Problem7
- Unable to Obtain Readings7
- Incorrect, Inadequate or Imprecise Result or Readings6
- Device Operational Issue3
- Device Issue2
- Device Sensing Problem2
- Break1
- Degraded1
Patient problems
- No Known Impact Or Consequence To Patient28
- No Consequences Or Impact To Patient5
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- Test Result1
CAMINO models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CAMINO® 110-4BDevice, Monitoring, Intracranial Pressure15 reports
- CAMINO® 110-4BTDevice, Monitoring, Intracranial Pressure5 reports
- CAMINO® 110-4GDevice, Monitoring, Intracranial Pressure5 reports
- CAMINO® 110-4LDevice, Monitoring, Intracranial Pressure3 reports
- CAMINO® 110-4HMTDevice, Monitoring, Intracranial Pressure2 reports
- CAMINO® 110-4HMDEVICE, MONITORING, INTRACRANIAL PRESSURE2 reports
Recent CAMINO adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2021Degraded
OLM INTRACRANIAL PRESSURE MONITORING KIT - CLIENT CLAIMS THAT THE BUTTON WAS LOCKED, APPARENTLY IT IS RUSTY. DELAY IN PROCEDURE, RESOLVED BY CATHETER CHANGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11885351 ↗MalfunctionDec 23, 2019Output Problem; Output Problem
CAM02 MONITOR IS UNABLE TO DETECT 1104B CATHETER AFTER TRANSFERRING THE PATIENT TO THE OR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9512590 ↗MalfunctionDec 20, 2019Output Problem; Output Problem
IMPLANTED CATHETER HAS FAILED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9504687 ↗MalfunctionDec 3, 2019Output Problem; Output Problem
CATHETER HAS FAILED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9402914 ↗MalfunctionDec 3, 2019Output Problem; Output Problem
CATHETER HAS FAILED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9402892 ↗InjuryNov 26, 2019Low Test Results; Low Test Results
LICOX ICP METER WAS INSERTED TO PATIENT FOR TWO DAYS WITH A READING OF 2, HOWEVER PATIENT GOT WORSE. WHEN UNIT REPLACED WITH ANOTHER, READING WAS 32.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9376128 ↗InjuryNov 21, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
CATHETER DETACHED DURING PROCEDURE, LEAVING A PART BEHIND IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 9355047 ↗MalfunctionSep 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
CATHETER SUDDENLY STOPPED MONITORING 3 DAYS AFTER IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9097785 ↗MalfunctionSep 4, 2019Output Problem; Output Problem
CATHETER GAVE HIGH READINGS COMPARED TO EVD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8962508 ↗MalfunctionSep 4, 2019Output Problem; Output Problem
CATHETER GAVE HIGH READINGS COMPARED TO EVD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8962499 ↗MalfunctionAug 14, 2019Output Problem; Output Problem
CATHETER STOPPED WORKING AFTER IMPLANTATION ONCE CUSTOMER WAS TRANSPORTED FROM OR TO ICU.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8894067 ↗MalfunctionAug 14, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CATHETER GAVE DIFFERENT WAVEFORM READINGS AFTER IMPLANTATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8894061 ↗MalfunctionJun 27, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CATHETER WOULD NOT INITIALIZE OR REGISTER. HAD TO BE REPLACED WITH SECOND CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8739228 ↗MalfunctionApr 24, 2019Premature Separation; Premature Separation
OUTER HUB DID NOT PROPERLY SECURE TO THE PROBE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8544759 ↗MalfunctionApr 19, 2019Unable to Obtain Readings; Unable to Obtain Readings
OUTPUT VALUE DID NOT REGISTER DURING RECONNECTION AFTER TRANSFER OF PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8533479 ↗
Other Integra Lifesciences Corporation products
All Integra Lifesciences Corporation devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page