CERENOVUS ENTERPRISERecalls & reported problems across all models
Medos International Sàrl · 16 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 478
- Deaths reported
- 5
- Injuries reported
- 53
- Recalls
- 0
We found 16 models of CERENOVUS ENTERPRISE by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 478 adverse event reports in the FDA's MAUDE database, received between Sep 21, 2016 and Sep 15, 2026 — classified as 5 deaths, 53 injuries, 420 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (241 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CERENOVUS ENTERPRISE problems
Device problems
- Failure to Advance241
- Premature Activation197
- Activation Failure49
- Adverse Event Without Identified Device or Use Problem38
- Retraction Problem20
- Difficult to Advance11
- Material Twisted/Bent11
- Material Separation5
Patient problems
- No Clinical Signs, Symptoms or Conditions392
- No Consequences Or Impact To Patient43
- Ischemia Stroke6
- Infarction, Cerebral5
- Stenosis4
- Thrombosis4
- Transient Ischemic Attack4
- Aneurysm3
CERENOVUS ENTERPRISE models
16 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CERENOVUS ENTERPRISE ENF402312intracranial neurovascular stent1 death90 reports
- CERENOVUS ENTERPRISE ENF402300intracranial neurovascular stent2 deaths66 reports
- CERENOVUS ENTERPRISE ENF403012intracranial neurovascular stent1 death42 reports
- CERENOVUS ENTERPRISE ENF452212intracranial neurovascular stent41 reports
- CERENOVUS ENTERPRISE ENF401612intracranial neurovascular stent36 reports
- CERENOVUS ENTERPRISE ENF401600intracranial neurovascular stent36 reports
- CERENOVUS ENTERPRISE ENF452812intracranial neurovascular stent31 reports
- CERENOVUS ENTERPRISE ENF403000intracranial neurovascular stent29 reports
- CERENOVUS ENTERPRISE ENF452200intracranial neurovascular stent25 reports
- CERENOVUS ENTERPRISE ENF453712intracranial neurovascular stent16 reports
- CERENOVUS ENTERPRISE ENF452800intracranial neurovascular stent16 reports
- CERENOVUS ENTERPRISE ENF403900intracranial neurovascular stent16 reports
- CERENOVUS ENTERPRISE ENF403912intracranial neurovascular stent1 death15 reports
- CERENOVUS ENTERPRISE ENF451412intracranial neurovascular stent8 reports
- CERENOVUS ENTERPRISE ENF451400intracranial neurovascular stent8 reports
- CERENOVUS ENTERPRISE ENF453700intracranial neurovascular stent3 reports
Recent CERENOVUS ENTERPRISE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Failure to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50776612 ↗MalfunctionMay 20, 2026Premature Activation
IT WAS REPORTED THAT THE 4.0MM DIA X 16MM ENTERPRISE®2 (ENF401612 / 8302888) WAS PREMATURELY DEPLOYED WITHIN THE CATHETER HUB THEREFORE, RENDERED IT UNUSABLE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON (B)(6) 2026, ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE STENT WAS PREMATURELY DETACHED AND WAS NO LONGER ON THE DELIVERY WIRE. THE INFORMATION CONFIRMED THERE WAS NO ADVERSE OUTCOME TO THE PATIENT. BASED ON THE ADDITIONAL INFORMATION RECEIVED ON 12-MAY-2026, THE EVENT HAS BEEN DETERMINED TO BE USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25236313 ↗MalfunctionMay 20, 2026Premature Activation
IT WAS REPORTED THAT THE 4.0MM DIA X 23MM ENTERPRISE®2 (ENF402312 / 10021142) WAS PREMATURELY DEPLOYED AT THE HUB AND SLIPPED OUT. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 12-MAY-2026, ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE STENT WAS PREMATURELY DETACHED AND WAS NO LONGER ON THE DELIVERY WIRE. THE INFORMATION CONFIRMED THERE WAS NO ADVERSE OUTCOME TO THE PATIENT. BASED ON THE ADDITIONAL INFORMATION RECEIVED ON 12-MAY-2026, THE EVENT HAS BEEN DETERMINED TO BE USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232265 ↗MalfunctionMay 14, 2026Premature Activation
IT WAS REPORTED THAT THE 4.0MM DIA X 16MM WITH NO TIP ENTERPRISE®2 (ENF401600 / 8697625) WAS PREMATURELY DEPLOYED AT THE HUB AND SLIPPED OUT. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 12-MAY-2026, ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE STENT WAS PREMATURELY DETACHED AND WAS NO LONGER ON THE DELIVERY WIRE. THE INFORMATION CONFIRMED THERE WAS NO ADVERSE OUTCOME TO THE PATIENT. BASED ON THE ADDITIONAL INFORMATION RECEIVED ON 12-MAY-2026, THE EVENT HAS BEEN DETERMINED TO BE USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25175294 ↗MalfunctionApr 27, 2026Premature Activation
A HEALTHCARE PROFESSIONAL REPORTED THAT AN ENTERPRISE2 4MMX23MM NO TIP VASCULAR RECONSTRUCTION DEVICE (PRODUCT CODE: ENF402300, LOT NUMBER: UNKNOWN) WAS SEPARATED FROM THE WIRE DURING THE PROCEDURE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25002199 ↗MalfunctionApr 10, 2026Failure to Advance
IT WAS REPORTED THAT THE PHYSICIAN FELT STRONG RESISTANCE DURING ADVANCING ENTERPRISE2 4MMX23MM NO TIP (ENF402300/ 9508489) THROUGH THE HUB OF AN UNSPECIFIED MICROCATHETER. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24845464 ↗MalfunctionApr 9, 2026Premature Activation
A HEALTHCARE PROFESSIONAL REPORTED THAT THERE WAS EARLY DEPLOYMENT AT THE CATHETER HUB OF AN ENTERPRISE2 4MMX30MM VASCULAR RECONSTRUCTION DEVICE (PRODUCT CODE: ENF403012, LOT NUMBER: 9664724). ADDITIONAL EVENT INFORMATION RECEIVED ON 24-MAR-2026 INDICATED THAT THE STENT PREMATURELY DETACHED AND NO LONGER ENGAGED ON DELIVERY WIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832824 ↗MalfunctionApr 7, 2026Premature Activation
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 4.0MM DIA X 23MM ENTERPRISE®2 (ENF402312 / 9584147) DEPLOYED EARLY. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 26-MAR-2026, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, DURING THE DEVICE PREPARATION, THE STENT COMPONENT BECAME PREMATURELY DETACHED AND WAS NO LONGER ENGAGED / ATTACHED TO THE DELIVERY WIRE. THE TIP OF THE INTRODUCER CORRECTLY FIT INTO THE ROTATING HEMOSTASIS VALVE (RHV) AND THE HUB OF THE MICROCATHETER. THE INTRODUCER WAS SECURELY PLACED IN THE HUB PRIOR TO THE ATTEMPT TO ADVANCE THE DELIVERY WIRE. THERE WAS NO DAMAGE NOTED ON THE ASSOCIATED MICROCATHETER. BASED ON THE ADDITIONAL INFORMATION RECEIVED ON 26-MAR-2026, THE EVENT HAS BEEN DETERMINED TO BE USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24806109 ↗MalfunctionMar 19, 2026Premature Activation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE STENT COMPONENT OF THE 4.0MM DIA X 23MM ENTERPRISE® 2 (ENF402312 / LOT# UNKNOWN) WAS PREMATURELY SEPARATED FROM THE DELIVERY WIRE IN THE INTRODUCER SHEATH. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 11-MAR-2026, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE LOT NUMBER OF THE 4.0MM DIA X 23MM ENTERPRISE® 2 (ENF402312) WAS 9854676. THE REPLACEMENT DEVICE WAS ANOTHER 4.0MM DIA X 23MM ENTERPRISE® 2 (ENF402312). THERE WAS NO NEGATIVE PATIENT IMPACT. ALL OTHER INFORMATION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24640665 ↗MalfunctionMar 10, 2026Material Separation
IT WAS REPORTED, VIA A HEALTHCARE PROFESSIONAL, THAT THE STENT AND WIRE WERE OF AN ENTERPRISE2 4MMX23MM NO TIP VASCULAR RECONSTRUCTION DEVICE (PRODUCT CODE: ENF402300, LOT NUMBER: UNKNOWN) WERE SEPARATED INSIDE THE ENCLOSED SHEATH, AND ONLY THE WIRE SLIPPED OUT. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559023 ↗MalfunctionFeb 26, 2026Retraction Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AN ENTERPRISE2 4MMX16MM NO TIP (PRODUCT CODE: ENF401600, LOT NUMBER: 9493904) RECAPTURE FAILED WHEN DEPLOYED AT THE M1 SEGMENT STENOSIS. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24455393 ↗MalfunctionDec 22, 2025Premature Activation; Premature Activation
A HEALTHCARE PROFESSIONAL REPORTED THAT AN ENTERPRISE2 4MMX23MM NO TIP VASCULAR RECONSTRUCTION DEVICE (PRODUCT CODE ENF402300, LOT NUMBER 9508489) PREMATURELY DEPLOYED OUT TO DISPENSER. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23863113 ↗MalfunctionDec 9, 2025Material Separation; Material Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE WIRE ON THE 4.0MM DIA X 23MM WITH NO TIP ENTERPRISE® 2 (ENF402300 / 8640233) WAS BROKEN BEFORE THE DEVICE WAS USED. THERE WAS NO REPORT OF ANY PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764318 ↗MalfunctionNov 7, 2025Premature Activation; Premature Activation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PROCEDURE ON (B)(6) 2025, A 4.0MM DIA X 30MM ENTERPRISE®2 (ENF403012 / 8994896) WAS USED, AND THE STENT WAS PREMATURELY DEPLOYED IN THE HUB OF THE CONCOMITANT PROWLER SELECT PLUS MICROCATHETER. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE PROCEDURE AND THE REPORTED DEVICE ISSUE WERE UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED AT A LATER DATE, THIS FILE WILL BE UPDATED ACCORDINGLY. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION AND ANALYSIS. THE PRODUCT INVESTIGATION COMPLETED ON 07-NOV-2025. BASED ON THE COMPLETED PRODUCT INVESTIGATION, THIS EVENT HAS BEEN DEEMED USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23503599 ↗MalfunctionNov 5, 2025Migration; Migration
A HEALTHCARE PROFESSIONAL REPORTED THAT AN ENTERPRISE2 4MMX16MM NO TIP VASCULAR RECONSTRUCTION DEVICE (ENF401600, LOT UNKNOWN) MIGRATED FROM THE DISTAL INTERNAL CAROTID ARTERY (ICA) TO THE PROXIMAL ICA, WHERE IT HAD BEEN INITIALLY DEPLOYED SUCCESSFULLY. AT 1 YEAR FOLLOW UP, IMAGING DEMONSTRATED THE MIGRATION OF THE STENT. THE PATIENT REMAINS ASYMPTOMATIC. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23477158 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page