We found 2 models of CinchLock by Stryker Corporation in the FDA's device registry (GUDID). Together they are named in 125 adverse event reports in the FDA's MAUDE database, received between May 25, 2018 and Jun 19, 2026 — classified as 14 injuries, 111 malfunctions.
The most frequently reported specific device problem is “Fracture” (37 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CinchLock problems
Device problems
- Fracture37
- Expiration Date Error30
- Break28
- Material Fragmentation15
- Activation, Positioning or Separation Problem5
- Adverse Event Without Identified Device or Use Problem4
- Detachment of Device or Device Component4
- Material Integrity Problem4
Patient problems
- No Clinical Signs, Symptoms or Conditions89
- No Known Impact Or Consequence To Patient20
- Insufficient Information15
- Foreign Body In Patient9
- Device Embedded In Tissue or Plaque3
- Pain3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Fluid Discharge1
CinchLock models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CinchLock CAT02462FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE124 reports
- CinchLock CAT02463BIT, DRILL1 report
Recent CinchLock adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJun 19, 2026Material Separation
IT WAS REPORTED THAT FOR THE INVOLVED CINCHLOCK SS ANCHOR. AFTER ANCHOR INSERTION, WHILE TENSIONING THE TAPE, THE TAPE PHYSICALLY BROKE THROUGH THE SIDE OF THE ANCHOR. THE SUTURE ITSELF REMAINED INTACT, BUT THE ANCHOR FAILED. AFTER THESE FAILURES, THE SURGEON MADE THE DECISION TO ABANDON THE LABRAL REPAIR AND INSTEAD DEBRIDE AND REMOVE THE PATIENT¿S LABRUM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25565571 ↗MalfunctionMay 11, 2026Fracture; Separation Problem
IT WAS REPORTED THAT WHEN THE CINCHLOCK ANCHOR WAS DEPLOYED, THE DEPLOYMENT WIRE SNAPPED. AS A RESULT, THE ANCHOR COULD NOT BE SUCCESSFULLY DEPLOYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25137868 ↗MalfunctionMar 6, 2026Material Fragmentation
IT WAS REPORTED THAT THE IMPLANT BROKE DURING INSERTION. A SPARE DEVICE WAS AVAILABLE. ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24533628 ↗MalfunctionDec 12, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR WAS THREADED NORMALLY AND PLACED IN THE HOLE. WHEN TIGHTENING/ADJUSTING THE THREAD TENSION, THE THREAD WAS TORN OUT OF THE ANCHOR. THE ANCHOR WAS THEREFORE NOT TRIGGERED AND WAS PULLED OUT OF THE HOLE. THE ANCHOR REMAINED IN THE BONE; THE TRIGGER MECHANISM WAS PULLED OUT. FURTHER INFO REPORTED: THE THREAD WAS TORN FROM THE ANCHOR. THIS MEANS THAT EITHER THE PROXIMAL PART OF THE ANCHOR IS BROKEN OR THE THREAD IS TORN. THE THREAD WAS STILL INTACT AND COULD BE USED FOR THE NEXT ANCHOR. WE COULDN'T SEE THE ANCHOR BECAUSE IT WAS EMBEDDED IN THE BONE. BASED ON OUR OBSERVATIONS, I HAVE TO ASSUME THAT THE ANCHOR IS BROKEN. WE DIDN'T SEE THE ANCHOR COME OUT OF THE HOLE. IT IS POSSIBLE THAT PART OF IT WAS DRILLED OUT FOR THE REST OF THE PROCEDURE (PLACEMENT OF ADDITIONAL ANCHORS). THE NEXT STEP WAS TO DRILL A NEW HOLE IN THE BONE TO REFIX THE LABRUM WITH A NEW ANCHOR. ESTIMATED DELAY: 5 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23794392 ↗MalfunctionOct 23, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY LABRAL REPAIR, A CINCHLOCK SS ANCHOR WAS USED AND INSERTED INTO THE DRILLED HOLE. AS REPORTED, THE WIRE GOT STUCK IN THE ANCHOR WHEN ANCHOR WAS DEPLOYED. THE SURGEON CUT THE ANCHOR SHORT TO BE FLUSH WITH THE ANCHOR AND REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23367905 ↗MalfunctionSep 25, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE ANCHOR FAILED AND PEEK ANCHOR SNAPPED. ANCHOR WAS ABLE TO BE PULLED OUT AND REPLACED WITH ANOTHER CINCHLOCK SUCCESSFULLY. THERE WAS NO KNOWN IMPACT OF THIS FAILED DEVICE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23150431 ↗MalfunctionSep 23, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING THE PROCEDURE WHEN THE CINCHLOCK SS ANCHOR WAS DEPLOYED IN THE JOINT SPACE THE WIRE THAT GOES THROUGH THE CENTER OF THE ANCHOR DID NOT DISENGAGE AS NORMAL. THE ENTIRE LENGTH OF THE WIRE WAS LEFT IN THE PATIENT¿S JOINT AT FIRST AND THE INNER PEEK COMPONENT OF THE ANCHOR WAS LOOSE. THE SURGEON HAD TO WORK TO REMOVE THE INNER COMPONENT, AND WIRE TOGETHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23128289 ↗MalfunctionJul 15, 2025Fracture; Fracture
IT WAS REPORTED THAT THE WIRE BROKE OFF IN IMPLANT WITH A CINCHLOCK SS ANCHOR. THE WIRE WAS REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22515805 ↗MalfunctionJun 6, 2025Fracture; Fracture
IT WAS REPORTED THAT THE CINCHLOCK ANCHOR DEPLOYED WITH A WIRE TAIL REMAINING BEHIND THAT NEED TO BE CLIPPED AND REMOVED. THERE WAS HEIGHTENED BLEEDING DUE REMOVAL OF WIRE. THE WIRE WAS COMPLETELY REMOVED FROM THE PATIENT. NO MEDICAL INTERVENTION REQUIRED AS A RESULT OF THE HEIGHTENED BLEEDING DURING THE REMOVAL OF WIRE. THERE WERE NO FURTHER ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161285 ↗MalfunctionApr 25, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING HIP LABRAL REPAIR, THE CINCHLOCK SS ANCHOR WAS INSERTED AND TENSIONED WITHOUT ISSUE. WHEN THE SURGEON WENT TO LOCK THE ANCHOR AND REMOVE THE INSERTER, THE PULL WIRE REMAINED INSIDE THE ANCHOR EXTENDING OUT THROUGH THE TRANSPORT CANNULA OUTSIDE OF THE JOINT/BODY. THE PULL WIRE WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21914485 ↗MalfunctionApr 25, 2025Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT THE ANCHOR WAS INSERTED TO THE LASERLINE. UPON TENSIONING, THE SUTURE TAPE BROKE THROUGH THE ANCHOR. AFTER THIS, THE SURGEON PULLED ON THE INSERTER TO REMOVE THE IMPLANT. THIS WAS IDENTIFIED DURING THE PROCEDURE WHEN IT COULD BE SEEN THE NITINOL ROD WAS STILL ATTACHED TO THE ANCHOR. ALL PIECES WERE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21913685 ↗MalfunctionApr 25, 2025Fracture; Fracture
IT WAS REPORTED THAT WHEN THE HANDLE FOR THE ANCHOR DEPLOYMENT WAS SQUEEZED, THE WIRE THAT PULLS THE DISTAL TIP OF THE ANCHOR INTO THE MAIN BODY OF THE ANCHOR, SNAPPED, AND WAS LEFT IN THE ANCHOR. THE WIRE REMAINED AT SIGNIFICANT LENGTH STILL ATTACHED TO THE ANCHOR. THE SURGEON HAD TO ACT TO REMOVE THE WIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21911947 ↗MalfunctionApr 22, 2025Fracture; Fracture
IT WAS REPORTED THAT THE CINCHLOCK SS ANCHOR WAS DEPLOYED DURING PROCEDURE AND SOME OF THE WIRE BROKE OFF FROM THE ANCHOR AND INSERTER. THE WIRE RUNNING THROUGH THE CINCHLOCK SS IMPLANT BROKE OFF IN THE JOINT SPACE AS THE WIRE WAS NOT COMPLETELY RELEASED FROM THE IMPLANT AS IT NORMALLY IS. THE WIRE WAS SUCCESSFULLY ABLE TO BE REMOVED FROM THE PATIENT AFTER ADDITIONAL EFFORT DURING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21886757 ↗MalfunctionApr 22, 2025Fracture; Fracture
IT WAS REPORTED THAT WHEN THE HANDLE WAS DEPRESSED, THE PULL WIRE DID NOT COME OUT WITH THE INSERTER AND WAS LEFT STICKING OUT OF THE MIDDLE OF THE ANCHOR ALL THE WAY OUT OF THE PATIENT. HAD TO MANUALLY PULL THE WIRE OUT OF THE ANCHOR. ANCHOR REMAINED IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21885737 ↗MalfunctionApr 22, 2025Fracture; Fracture
IT WAS REPORTED THAT WHEN THE HANDLE WAS DEPRESSED, THE PULL WIRE DID NOT COME OUT WITH THE INSERTER AND WAS LEFT STICKING OUT OF THE MIDDLE OF THE ANCHOR ALL THE WAY OUT OF THE PATIENT. HAD TO MANUALLY PULL THE WIRE OUT OF THE ANCHOR. ANCHOR REMAINED IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21884403 ↗
Other Stryker Corporation products
- RemB7 models · 3,073 reports
- Prime Series21 models · 1,548 reports · 2 recalls
- TARGET114 models · 1,077 reports
- ProCuity17 models · 368 reports · 13 recalls
- Spetzler-Malis20 models · 7 reports · 19 recalls
- SYNCHRO 28 models · 313 reports · 9 recalls
- Neptune E-SEP19 models · 412 reports · 7 recalls
- StrykeFlow4 models · 552 reports · 4 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page