CoaguChek XS PT stripsRecalls & reported problems across all models
Roche Diagnostics GmbH · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 632
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls
- 0
We found 2 models of CoaguChek XS PT strips by Roche Diagnostics GmbH in the FDA's device registry (GUDID). Together they are named in 632 adverse event reports in the FDA's MAUDE database, received between Jul 11, 2018 and Sep 16, 2026 — classified as 6 injuries, 626 malfunctions.
The most frequently reported specific device problem is “High Test Results” (505 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CoaguChek XS PT strips problems
Device problems
- High Test Results505
- Non Reproducible Results89
- Incorrect, Inadequate or Imprecise Result or Readings35
- Low Test Results27
- Inaccurate Information3
- Circuit Failure2
- Device Difficult to Setup or Prepare2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions336
- No Consequences Or Impact To Patient288
- Insufficient Information4
- Thrombosis/Thrombus2
- Coagulation Disorder1
- Easy Bruising1
- Hematuria1
- Hemorrhage/Blood Loss/Bleeding1
CoaguChek XS PT strips models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CoaguChek XS PT strips 4625315160TEST, TIME, PROTHROMBIN604 reports
- CoaguChek XS PT strips 7797826160TEST, TIME, PROTHROMBIN28 reports
Recent CoaguChek XS PT strips adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788039 ↗MalfunctionMay 20, 2026Device Difficult to Setup or Prepare; Non Reproducible Results
THERE WAS AN ALLEGATION OF DISCREPANT INR RESULTS FOR 1 PATIENT TESTED ON A COAGUCHEK XS SYSTEM. THE RESULTS WERE 3.2 INR AND 4.4 INR. THE TIMEFRAME BETWEEN TESTS WAS NOT PROVIDED. THE PATIENT'S THERAPEUTIC RANGE WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232975 ↗MalfunctionApr 21, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE INR RESULT FROM THE COAGUCHEK XS PLUS METER. THE FIRST METER RESULT WAS 5.0 INR. THE SECOND METER RESULT WITHIN 4 HOURS WAS 3.5 INR. THE PATIENT¿S THERAPEUTIC RANGE WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24945243 ↗MalfunctionFeb 9, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS PROFESSIONAL METER COMPARED TO A PATIENT'S ROCHE METER (TYPE OF METER UNSPECIFIED). THE PATIENT'S METER (METER A) RESULT WAS 3.2 INR. DUE TO THE HIGH RESULT, THE PATIENT WAS TESTED AT THE HOSPITAL. THE LABORATORY RESULT WAS 3.2 INR. THE RESULT ON THE HOSPITAL'S COAGUCHEK XS PROFESSIONAL METER (METER B) WAS 5.4 INR. THIS RESULT WAS QUESTIONED. THE TIME LAPSE BETWEEN TESTS IS UNKNOWN. DUE TO THE LABORATORY RESULT MATCHING THE RESULT FROM METER A, THE RESULT FROM METER A WAS NOT QUESTIONED. THE PATIENT¿S THERAPEUTIC RANGE WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24296996 ↗MalfunctionJan 12, 2026Adverse Event Without Identified Device or Use Problem; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK VANTUS METER. ON (B)(6) 2025 AT 7:30 AM, THE RESULT WAS 3.2 INR. AT 7:35 AM, THE RESULT WAS 2.1 INR. ON (B)(6) 2025 AT 7:05 AM, THE RESULT WAS 1.5 INR. AT 7:10 AM, THE RESULT WAS 1.9 INR. AT 7:15 AM, THE RESULT WAS 2.2 INR. ON (B)(6) 2025 AT 7:15, THE RESULT WAS 2.6 INR. AT 7:20, THE RESULT WAS 2.5 INR. AT 8:20, THE RESULT WAS 2.0 INR. THE THERAPEUTIC RANGE WAS 1.8 TO 2.0 INR. THE PATIENT "DOES A ROUTINE WEEKEND VEGETABLE MEAL TO MAINTAIN NORMAL VITAMIN K LEVELS".
Patient problems: Easy Bruising; Hematuria
MAUDE report 24037653 ↗MalfunctionDec 19, 2025Non Reproducible Results; Non Reproducible Results
IT WAS REPORTED THAT A PATIENT RECEIVED DISCREPANT RESULTS WHEN TESTING WITH THEIR DOCTOR'S COAGUCHEK XS METER (SERIAL NUMBER UNKNOWN). A SAMPLE FROM THE PATIENT WAS TESTED USING THE DOCTOR'S METER, RESULTING IN A VALUE OF 2.9 INR. WITHIN 30 MINUTES OF THE INITIAL TEST, A SAMPLE FROM THE PATIENT WAS TESTED USING THE DOCTOR'S METER, RESULTING IN A VALUE OF 4.0 INR. A VENOUS SAMPLE COLLECTED FROM THE PATIENT AT AN UNKNOWN TIME ON (B)(6) 2025, RESULTED IN A LABORATORY VALUE OF 3.2 INR. THE PATIENT'S THERAPEUTIC RANGE WAS 3.0 - 3.5 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23854241 ↗MalfunctionDec 17, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS METER. AT 01:29 PM, THE RESULT WAS 4.7 INR. AT 01:31 PM, THE RESULT WAS 3.6 INR. THE THERAPEUTIC RANGE WAS 2.0-3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23826658 ↗MalfunctionDec 9, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS PLUS METER. AT 10:56 AM THE METER RESULT WAS 7.3 INR. AT 11:08 AM THE METER RESULT FROM ANOTHER COAGUCHEK XS METER WAS 3.5 INR. THE CUSTOMER'S THERAPEUTIC RANGE IS 2.0-3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23756582 ↗MalfunctionAug 5, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE INR RESULTS FROM THE COAGUCHEK XS METER. THE METER RESULT WAS 4.4 INR. THE METER RESULT WAS 3.1 INR. THE TESTS WERE PERFORMED WITHIN 4 HOURS OF EACH OTHER. THE PATIENT'S THERAPEUTIC RANGE IS 2.8-3.8 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22720644 ↗MalfunctionJun 23, 2025Component Missing; Component Missing
WE RECEIVED AN ALLEGATION THAT THE COAGUCHEK XS PT TEST 1X24 STRIPS BOX AND BOTTLE EXPIRATION DATES, AND THE TEST STRIP BOX AND THE CODE ON THE TEST STRIP DO NOT MATCH. THE REPORTER STATED THAT THE BOX'S EXPIRATION DATE IS 31-JUL-2025, AND THE BOTTLE'S IS 30-APR-2025; THE TEST STRIP CODE ON THE BOX IS 138, AND THE CODE ON THE TEST STRIP IS 117. THE REPORTER DID NOT USE THE TEST STRIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22307149 ↗MalfunctionApr 1, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS METER, SERIAL NUMBER (B)(6) COMPARED TO A PATIENT'S COAGUCHEK XS METER WITH (B)(6) SERIAL NUMBER. AT 12:30 PM THE METER RESULT FROM THE PATIENT'S COAGUCHEK XS METER (B)(6) WAS 4.6 INR. THE PATIENT WAS USING TEST STRIP LOT 76900023 WITH EXPIRATION DATE 31-OCT-2025. AT 12:30 PM THE METER RESULT WAS 2.4 INR. THE CUSTOMER'S THERAPEUTIC RANGE IS 2.0-3.0 INR. CUSTOMER TESTS ON A WEEKLY BASIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21739200 ↗MalfunctionMar 6, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS PLUS METER, SERIAL NUMBER UNKNOWN COMPARED TO A PATIENT'S COAGUCHEK VANTUS METER SERIAL NUMBER (B)(6). AT 12:30 PM THE COAGUCHEK XS PROFESSIONAL METER RESULT WAS 2.6 INR. AT 4:04 PM THE COAGUCHEK VANTUS METER RESULT WAS 1.6 INR. THE PATIENT WAS USING STRIP LOT 77503323 WITH EXPIRATION 31-OCT-2025. THE CUSTOMER'S THERAPEUTIC RANGE IS 2.0-3.0 INR. CUSTOMER TESTS ON A WEEKLY BASIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21535255 ↗MalfunctionNov 12, 2024Non Reproducible Results; Non Reproducible Results
IT WAS REPORTED THAT A PATIENT RECEIVED DISCREPANT RESULTS WHEN TESTING WITH HIS COAGUCHEK XS METER SERIAL NUMBER (B)(6) AND A SECOND COAGUCHEK XS METER (UNKNOWN SERIAL NUMBER) USED AT THE DOCTOR'S OFFICE. THIS MEDWATCH WILL APPLY TO THE TEST STRIPS USED ON THE DOCTOR'S OFFICE METER. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE TEST STRIPS USED ON THE PATIENT'S METER. AT 10:30 A.M., A SAMPLE FROM THE PATIENT WAS TESTED USING THE PATIENT'S METER, RESULTING IN A VALUE OF 4.4 INR. AT 11:30 A.M., A SAMPLE FROM THE PATIENT WAS TESTED USING THE DOCTOR'S OFFICE METER, RESULTING IN A VALUE OF 3.1 INR. THE PATIENT'S THERAPEUTIC RANGE IS 2.6 - 3.6 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20669512 ↗MalfunctionOct 18, 2024Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS PRO METER. THE RESULT FROM THE PROFESSIONAL METER WAS 2.7 INR. THE RESULT FROM THE PATIENT'S METER WAS 1.8 INR THE THERAPEUTIC RANGE WAS 2.0-3.0 INR AND THE TESTING FREQUENCY VARIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20485464 ↗MalfunctionSep 10, 2024Non Reproducible Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION OF QUESTIONABLE INR RESULTS FOR 1 PATIENT TESTED WITH A COAGUCHEK XS PRO METER (FACILITY'S METER) WHEN COMPARED TO A DIFFERENT COAGUCHEK XS METER (PATIENT'S METER). ON (B)(6) 2024 AT 6:35 PM THE PATIENT WAS TESTED WITH THE FACILITY'S METER AND HAD A METER RESULT OF 2.4 INR WHILE AT 6:48 PM THE PATIENT WAS TESTED WITH HIS METER AND HAD A METER RESULT OF 3.4 INR. THE PATIENT WAS TESTED USING 2 DIFFERENT INDEX FINGERS. THE PATIENT¿S THERAPEUTIC RANGE WAS 2.0-3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20183979 ↗
Other Roche Diagnostics GmbH products
- cobas e 411 Immunoassay Analyzer2 models · 1,218 reports
- cobas e 801 Immunoassay analyzer3 models · 713 reports · 1 recall
- CoaguChek XS Patient Self Testing Care Kit2 models · 346 reports
- Elecsys FT3 III4 models · 283 reports
- Elecsys Estradiol III2 models · 162 reports
- Elecsys TSH4 models · 155 reports
- Sodium electrode3 models · 150 reports
- Elecsys HIV Duo3 models · 114 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page