Davol Female Cath KitRecalls & reported problems across all models
C. R. Bard, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls
- 0
We found 2 models of Davol Female Cath Kit by C. R. Bard, Inc. in the FDA's device registry (GUDID). Together they are named in 28 adverse event reports in the FDA's MAUDE database, received between Jul 17, 2017 and Jan 28, 2026 — classified as 5 injuries, 23 malfunctions.
The most frequently reported specific device problem is “Disconnection” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Davol Female Cath Kit problems
Device problems
- Disconnection7
- Material Fragmentation4
- Component Missing3
- Hole In Material3
- Component Misassembled2
- Fitting Problem2
- No Flow2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient7
- Foreign Body In Patient3
- No Information2
- No Patient Involvement2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
- Patient Problem/Medical Problem1
Davol Female Cath Kit models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Davol® Female Cath Kit 0035380COLLECTOR, URINE, PEDIATRIC, FOR INDWELLING CATHETER19 reports
- Davol® Female Cath Kit 0035720COLLECTOR, URINE, PEDIATRIC, FOR INDWELLING CATHETER9 reports
Recent Davol Female Cath Kit adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJan 28, 2026Disconnection
IT WAS REPORTED THAT THE SPECIMEN CATHETER OF THE FEMALE EXTERNAL CATHETER CAME UNATTACHED FROM THE SPECIMEN CONTAINER. THE CATHETER HAD TO BE RETRIEVED THE PATIENTS URETHRA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24206302 ↗MalfunctionJan 28, 2026Disconnection
IT WAS REPORTED THAT THE PLASTIC CATHETER PART WAS NOT STAYING IN THE SPECIMEN CONTAINER, AND THEY ACTUALLY HAD TO PULL OUT THE SMALL PLASTIC CATHETER PART FROM THE URETHRA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24206105 ↗MalfunctionDec 24, 2025Disconnection; Disconnection; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE FEMALE CATHETER THAT GOES INTO THE PATIENT WAS VERY LOOSE/SLIDING OUT OF THE HOLDER INSERTING OR GOING BACK INTO THE HOLDER INSTEAD OF INTO THE PATIENT. CLINIC HAD 2 LIKE THIS BUT MANAGED TO HOLD THEM IN ORDER TO USE THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23889046 ↗MalfunctionNov 5, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE HOSPITAL USER STATED THAT THEY WERE GIVEN THE PACKAGE OF A DEFECTIVE BARD FEMALE CATHETER KIT 8 FR., SKU 0035720 FROM THEIR EMERGENCY DEPARTMENT. THEY MENTIONED THAT THE CATHETER TUBE WAS BROKEN IN HALF, BUT THEY DID NOT KEEP THE PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23480137 ↗MalfunctionAug 11, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT GETTING MULTIPLE REPORTS FROM EMERGENCY ROOM ABOUT BENT AND CRIMPING TUBING ON OUR FEMALE CATHETER KITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22772225 ↗MalfunctionApr 3, 2025Use of Device Problem; Fitting Problem; Use of Device Problem; Fitting Problem
IT WAS REPORTED THAT THE CATHETER TUBE THAT GOES INTO THE PATIENT WAS LOOSER IN SOME DEVICES THAN OTHERS, CAUSING IT TO EASILY SLIDE OUT OF THE COLLECTIONS CONTAINER AND POTENTIALLY BE RETAINED IN A PATIENT. THEY HAVE TESTED 3 AND 2 WERE LOOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21761209 ↗InjuryMar 6, 2025Disconnection; Disconnection
IT WAS REPORTED THAT CATHETER WAS SEPARATED FROM FEMALE CATHETER KIT COLLECTION TUBING. CATHETER WAS STUCK IN PATIENT FOR MULTIPLE DAYS AND NEEDED TO BE SURGICALLY REMOVED.
Patient problems: Foreign Body In Patient
MAUDE report 21536181 ↗MalfunctionFeb 25, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE FEMALE CATHETER KIT TUBE THAT GONE INTO THE PATIENT WAS LOOSER IN SOME DEVICES THAN OTHERS, CAUSING IT TO EASILY SLIDE OUT OF THE COLLECTIONS CONTAINER AND POTENTIALLY BE RETAINED IN A PATIENT. THEY HAD TESTED THREE AND TWO WERE LOOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21462161 ↗MalfunctionJun 10, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE CATHETER TUBE IN THE FEMALE STRAIGHT CATHETER KIT SNAPPED IN HALF WHEN TRYING TO PULL THE CATHETER FURTHER OUT OF THE COLLECTION TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19498671 ↗InjuryDec 22, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT HOW MUCH OF THE FOLEY CATHETER SHOULD BE LEFT IN THE TUBE WHEN USING IT TO OBTAIN A URINE SPECIMEN. AND REPORTED THAT THE INTERMITTENT CATHETER SOMETIMES A SMALLER SIZE COULD BE USED DUE TO IT JUST BEING FOR SPECIMEN COLLECTION. THE POSSIBLE LEAKING AROUND THE CATHETER WAS NOT AS GREAT OF A CONCERN IF NOT BEING PLACED AS AN INDWELLING. AND REPORTED THAT 5 FRENCH CATHETER KITS WERE VERY SECURE IN THE COLLECTION TUBE. AND SOME RESISTANCE WHEN PULLING OUT THE CATHETER, WHILE THE 8 FRENCH FEMALE CATHETER KITS HAD VERY LITTLE RESISTANCE WHEN PULLING OUT THE CATHETER. IT DID NOT FALL OUT ON ITS OWN BUT TOOK VERY LITTLE PULL TO HAVE IT BECOME COMPLETELY DISLODGED. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 19JAN2024, THEY DID NOT HAVE THE LOT NUMBERS FOR THE DEVICES AT CELEBRATION WHERE THE PRODUCT THAT LEAD TO PATIENT HARM WAS USED. BUT THE PRODUCT THAT THEY REVIEWED AND EXPERIENCED EASY MOVEMENT OF THE CATHETER WITHIN THE COLLECTION TUBE WAS NGHN0341. MULTIPLE DIFFERENT LOT NUMBER OF THE INTERMITTENT 8 FRENCH CATH KIT HAVE BEEN LOOKED AT BETWEEN TWO CAMPUSES. THE SAME CONCERN WAS VOICED WITH THE EASE OF MOVEMENT WITH THE 8 FRENCH CATH WITHIN COLLECTION TUBE FO
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 18382951 ↗InjuryNov 17, 2023Disconnection; Material Fragmentation; Disconnection; Material Fragmentation
IT WAS REPORTED THAT CATHETER IN THE FEMALE CATHETER KIT BECAME DISLODGED FROM COLLECTION DEVICE WHEN REGISTER NURSE WAS INSERTING IT AND CATHETER ITSELF WAS STUCK IN URINARY MEATUS. WHEN COMPARED AFTER TO SIMILAR PRODUCTS, THE OTHER CATHETERS WERE MUCH MORE SECURE IN COLLECTION DEVICE AND HARDER TO REMOVE. WHEN THE NURSES WERE OBTAINING URINE VIA STRAIGHT CATHETER, THE URINE COLLECTION KIT CAME APART, AND THE CATHETER PORTION DISLODGED AND BECAME STUCK IN THE PATIENT. THE PROVIDER STAFF WAS UNABLE TO REMOVE THE CATHETER, AND THE CHILD WAS SENT TO THE OPERATING ROOM FOR UROLOGY TO REMOVE. THE CHILDREN EMERGENCY CENTER HAD 2 TYPES OF STRAIGHT CATHETER KITS. UPON EXAMINATION, THE TYPE OF KIT USED IN THIS CASE EASILY FELL APART, UNLIKE THE OTHER KIT THEY CARRY. THEY WERE REMOVED FROM THE UNIT TO PREVENT FURTHER ISSUES. PER ADDITIONAL CLINICAL INFORMATION RECEIVED VIA EMAIL ON 01NOV2023, THE NURSE MANAGER REPORTED THAT THE CATHETER DEVICE WAS BEING USED ON AN 8-YEAR-OLD MALE PATIENT WITH CHRONIC MEDICAL CONDITIONS EXPERIENCING LETHARGY, AND THERE WERE CONCERNS FOR URINARY TRACT INFECTION/SEPSIS. PRIOR TO USE, THERE WERE NO DEVICE ISSUES NOTED. DURING INSERTION, THE CATHETER PORTION OF
Patient problems: Foreign Body In Patient
MAUDE report 18159009 ↗MalfunctionMar 21, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED THAT CATHETER IN THE FEMALE STRAIGHT CATHETER KIT WAS VERY LOOSE INSIDE OF CAP. CATHETER USED TO BE HELD IN PLACE INSIDE CAP WHICH WOULD ALLOW IT TO BE PULLED FORWARD BUT WOULD PREVENT IT FROM SLIDING BACK IN TUBE OR FALLING OUT. STATED THAT THE DANGER WAS THAT THE CATHETER COULD BE ADVANCED TO FAR FORWARD INTO THE URETHRA AND BECOME DIFFICULT TO REMOVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16584468 ↗MalfunctionJan 3, 2023Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE SWAB STICKS WERE DRY AND WHITE, AS IF THE POVIDONE-IODINE SOLUTION HAD NEVER BEEN ADDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16090877 ↗MalfunctionNov 9, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT A PATIENT'S NURSE WAS COMPLETING IN AND OUT OF FEMALE CATHETER AND THE CATHETER BROKE WHILE STILL IN THE PATIENT. STATED THAT THE NURSE WAS ABLE TO HOLD ONTO BOTH PIECES TO AVOID THE BROKEN PIECE GOING UP INTO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 15758548 ↗MalfunctionSep 2, 2022Disconnection; Component Misassembled; Disconnection; Component Misassembled
IT WAS REPORTED THAT DAVOL FEMALE CATHETER KIT WAS FALLING OUT OF COLLECTION TUBE INTO PATIENT, FREQUENTLY THUS CAUSING THEM TO HAVE TO FISH OUT THE TUBING FROM THE PATIENT. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 29-AUG-2022, THERE WAS NO PATIENT IMPACT OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15347707 ↗
Other C. R. Bard, Inc. products
- Bardex I.C. Foley Catheter26 models · 3,607 reports
- Bardex Lubricath Foley Catheter38 models · 1,670 reports · 1 recall
- Bardex Lubri-Sil I.C. All-Silicone Foley Catheter26 models · 1,518 reports
- Magic3 Coude Intermittent Catheter with Sure-Grip Sleeve5 models · 1,466 reports
- Magic3 Intermittent Catheter with Sure-Grip Sleeve14 models · 1,310 reports
- Bardia Foley Catheter Silicone Coated19 models · 1,283 reports
- PureWick Female External Catheter5 models · 1,184 reports
- Bardex All-Silicone Foley Catheter12 models · 1,170 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page