EasyFuseRecalls & reported problems across all models
Wright Medical Technology, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of EasyFuse by Wright Medical Technology, Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Apr 24, 2023 and Apr 24, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
EasyFuse recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 10, 2022Terminated
The drill and adjustable drill guide can jam/bind intraoperatively resulting in the drill guide breaking.
Z-0853-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported EasyFuse problems
Device problems
- Break1
- Fracture1
Patient problems
- Insufficient Information1
EasyFuse models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EasyFuse FFSP1530Orthopedic manual surgical instrument1 recall0 reports
- EasyFuse™ FFS4MTPSStaple, fixation, bone1 report
Recent EasyFuse adverse event reports
The 1 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 24, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THERE WAS A REVISION SURGERY. THE MTP STAPLE WAS IMPLANTED BUT NEEDED TO BE REPOSITIONED. WHEN TAKEN OUT AND RELOADED ON TO IMPLANT CARTRIDGE, THE CARTRIDGE SNAPPED AND BECAME UNABLE TO USE.
Patient problems: Insufficient Information
MAUDE report 16800826 ↗
Other Wright Medical Technology, Inc. products
- INFINITY51 models · 733 reports · 3 recalls
- INBONE56 models · 365 reports · 4 recalls
- INFINITY ADAPTIS8 models · 219 reports
- ORTHOLOC 3Di23 models · 28 reports · 2 recalls
- INFINITY BIOFOAM 3D5 models · 118 reports
- INFINITY EVERLAST20 models · 111 reports
- EVOLVE14 models · 28 reports · 1 recall
- SWANSON12 models · 17 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page