EIT ALIF InserterRecalls & reported problems across all models
EIT Emerging Implant Technologies GmbH · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 0
We found 3 models of EIT ALIF Inserter by EIT Emerging Implant Technologies GmbH in the FDA's device registry (GUDID). Together they are named in 25 adverse event reports in the FDA's MAUDE database, received between Jan 17, 2020 and Mar 29, 2024 — classified as 2 injuries, 23 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported EIT ALIF Inserter problems
Device problems
- Device-Device Incompatibility15
- Material Deformation5
- Material Twisted/Bent4
- Mechanical Jam2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- No Consequences Or Impact To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Code Available1
- Tissue Damage1
- Unspecified Tissue Injury1
EIT ALIF Inserter models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EIT ALIF Inserter AET30101Intervertebral Fusion Device With Bone Graft, Lumbar19 reports
- EIT ALIF Inserter AET20101Intervertebral Fusion Device With Bone Graft, Lumbar4 reports
- EIT ALIF Inserter AET30100Intervertebral Fusion Device With Bone Graft, Lumbar2 reports
Recent EIT ALIF Inserter adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2024Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, DURING THE CASE, THE SPINE SURGEON WAS TRYING TO DISTRACT THE DISC SPACE FOR ALIF CASE. THE DISTRACTOR THREAD AT THE DISTAL END OF THE INSTRUMENT WAS DAMAGED (LOOKED BENT) SO IT WOULD NOT THREAD/LOCK PROPERLY FOR DISTRACTION. THEY MANAGED THIS BY HAVING THE ACCESS SURGEON TO HOLD AND TIGHTEN THE DISTRACTOR MANUALLY WITHOUT THREADING THE LOCKING MECHANISM. DURING THIS SAME CASE, WHEN THE SURGEON WAS INSERTING THE CAGE WITH THE CAGE INSERTER. THE CAGE WAS MALLETED IN AS PER USUAL PROCESS. WHEN THE SURGEON WENT TO UNDO THE INSERTER FROM THE CAGE TO DISENGAGE. THE INSERTER SPINDLE AT THE DISTAL END WOULD NOT TURN. AFTER MULTIPLE ATTEMPTS, THE SURGEON REMOVED THE CAGE ALTOGETHER WITH THE INSERTER. WE HAD ANOTHER INSERTER IN THE SET AS WELL AS ANOTHER SAME SIZE IMPLANT. CASE WAS COMPLETED WITH 5 MINUTES DELAY WITH NO ADVERSE TO PATIENT REPORTED. THIS REPORT INVOLVES ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19004634 ↗MalfunctionSep 13, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, DURING THE CASE, THE SPINE SURGEON WAS TRYING TO DISTRACT THE DISC SPACE FOR ALIF CASE. THE DISTRACTOR THREAD AT THE DISTAL END OF THE INSTRUMENT WAS DAMAGED (LOOKED BENT) SO IT WOULD NOT THREAD/LOCK PROPERLY FOR DISTRACTION. THEY MANAGED THIS BY HAVING THE ACCESS SURGEON TO HOLD AND TIGHTEN THE DISTRACTOR MANUALLY WITHOUT THREADING THE LOCKING MECHANISM. DURING THIS SAME CASE, WHEN THE SURGEON WAS INSERTING THE CAGE WITH THE CAGE INSERTER. THE CAGE WAS MALLETED IN AS PER USUAL PROCESS. WHEN THE SURGEON WENT TO UNDO THE INSERTER FROM THE CAGE TO DISENGAGE. THE INSERTER SPINDLE AT THE DISTAL END WOULD NOT TURN. AFTER MULTIPLE ATTEMPTS, THE SURGEON REMOVED THE CAGE ALTOGETHER WITH THE INSERTER. WE HAD ANOTHER INSERTER IN THE SET AS WELL AS ANOTHER SAME SIZE IMPLANT. CASE WAS COMPLETED WITH 5 MINUTES DELAY WITH NO ADVERSE TO PATIENT REPORTED. THIS REPORT INVOLVES ONE (1) TRIAL INSERTER SH CONNECTION. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17736734 ↗MalfunctionJul 27, 2023Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam
THE PROCEDURE WAS COMPLETED SUCCESSFULLY. AND A GOOD OUTCOME WAS ACHIEVED. NO OTHER MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17405443 ↗MalfunctionJun 5, 2023Mechanical Jam; Mechanical Jam
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, A DEMONSTRATION WAS DONE WITH THE CONDUIT LLIF, AND THE IMPLANT INTRODUCED GOT JAMMED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17059207 ↗MalfunctionMay 2, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT A CUSTOMER REPORTED, THE INSTRUMENTS CAN NO LONGER BE MOUNTED AND THUS CAN NO LONGER BE USED FOR SURGICAL PROCEDURES. THIS WAS DETERMINED OUTSIDE OF AN OPERATION, SO NO DAMAGE WAS CAUSED. NO PATIENT INVOLVEMENT THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16852462 ↗MalfunctionFeb 9, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2023 , THAT DURING TRAIL IMPLANT INSERTION THE TRIAL INSERTER INSTRUMENT X 1 REF AET20101 BENT/DEFORMED AND DURING IMPLANT INSERTION THE IMPLANT INSERTER INSTRUMENT REF AET30101 X1. THE SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER COMPLICATIONS. THIS REPORT IS FOR ONE (1) TRIAL INSERTER SH CONNECTION. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16335349 ↗MalfunctionFeb 9, 2023
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16335322 ↗MalfunctionSep 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022 THAT AFTER POSITIONING THE CAGE IN THE INTERVERTEBRAL DISC SPACE, THE INSERTION COULD NOT BE DETACHED FROM THE CAGE. THIS CAUSED A SURGICAL DELAY OF 60 MINUTES. THIS REPORT INVOLVES ONE IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15395749 ↗InjuryAug 2, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, AFTER THE ALIF CAGE WAS PLACED IN THE L5/S1 SEGMENT (ACCESS ANTEROLATERAL FROM THE LEFT), SURGEON REMOVED THE IMPLANT HOLDER FROM THE CAGE. AFTER CONTROL, THE CONDUIT ALIF NEEDED TO BE REPOSITIONED, BECAUSE THE POSITION WAS NOT SATISFACTORY AS PER X-RAY. AGAIN, SURGEON CONNECTED IMPLANT HOLDER TO THE CAGE FOR REPOSITIONING OF THE CAGE. AFTER THIS PROCEDURE THE IMPLANT HOLDER WAS NOT POSSIBLE TO BE REMOVED FROM THE CAGE. IT WAS NO LONGER POSSIBLE TO TURN THE KNOB TO THE RIGHT OR LEFT SIDE (OPEN /CLOSED). SURGEON REMOVED THE CAGE HOLDER AND CONNECTED CAGE FROM THE DISC SPACE L5/S1 AND DID ALTERNATIVE IMPLANTATION OF A TM400 COMPETITOR CAGE. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Tissue Injury
MAUDE report 15153395 ↗MalfunctionJul 19, 2022Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN UNKNOWN PROCEDURE, THE IMPLANT INSERTER IN QUESTION GOT STUCK ON THE CAGE BEING IMPLANTED AND WOULD NOT COME APART. THE CAGE WAS ABLE TO BE REMOVED FROM THE INSERTER, BUT THE END OF THE DEVICE WAS CROSS-THREADED. IT WAS ULTIMATELY DECIDED THAT THE CAGE SHOULD NOT BE IMPLANTED. THERE IS NO REPORTED ADVERSE PATIENT OUTCOME. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15058272 ↗MalfunctionJul 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE SURGEON COULD NOT DETACH THE CAGE FROM THE INSERTER. THE CAGE WAS ALREADY CORRECTLY POSITIONED IN THE LIGAMENT SHEATH SPACE, SO THE SURGEON HAD TO COMPLETELY REMOVE THE CAGE WITH INSERTION AND USE ANOTHER SYSTEM. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A 40 MINUTE DELAY. THIS REPORT IS FOR AN IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14932612 ↗MalfunctionJan 31, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, AFTER INSERTING THE IMPLANT, THE SURGEON COULD NOT DETACH THE INSERTER FROM THE IMPLANT. IT HAD TO BE COMPLETELY REMOVED AGAIN AND CANCELED. THERE WAS A TWENTY FIVE (25) MINUTE SURGICAL DELAY. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 2 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13391894 ↗MalfunctionDec 21, 2021Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. DURING THE SURGERY, WHEN INSERTING AN IMPLANT, THE INSTRUMENT WAS BENT. IT WAS UNKNOWN IF THE SURGERY COMPLETED SUCCESSFULLY. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICE REPORTED. ALIF UI H 12MM 10DEG S (PART# AUI31234, LOT# UNKNOWN, QTY1) THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) IMPLANT INSERTER SH CONNECTION THIS REPORT IS 1 OF 1 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13053769 ↗MalfunctionNov 3, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, AFTER IMPLANTATION OF THE CAGE, THE PLACEMENT INSTRUMENT ((B)(4)) COULD NO LONGER BE DETACHED FROM THE IMPLANT ((B)(4)). THE IMPLANT HAD TO BE REMOVED AND A NEW ONE WHICH WAS INSERTED WITH AN ALTERNATIVE PLACEMENT INSTRUMENT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH FIFTEEN(15) MINUTES DELAY. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12745574 ↗MalfunctionAug 24, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, DURING THE PROCEDURE THE SURGEON WANTED TO REMOVE THE IMPLANT INSERTER FROM THE IMPLANT, THE KNOB AT THE BACK OF THE BLACK HANDLE COULD NOT BE TURNED AND GOT STUCK. THE IMPLANT COULD NOT BE RELEASED FROM THE IMPLANT INSERTER AND HAD TO BE REMOVED FROM THE PATIENT. AT A SECOND TRY TO RELEASE THE IMPLANT, OUTSIDE OF THE PATIENT ON A SIDE TABLE, THE INNER SHAFT OF THE IMPLANT INSERTER BROKE OFF AND A SHORT PIECE OF THE INNER SHAFT STAYED STUCK WITHIN THE METAL KNOB. THE IMPLANT WAS RELEASED FROM THE HOLDER AND HAD THEN TO BE IMPLANTED FREEHAND, BECAUSE THERE WAS NO SECOND IMPLANT INSERTER OR ANY OTHER ALTERNATIVE ON THE SET. THERE WAS A SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE (1) IMPLANT INSERTER SH CONNECTION. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12362319 ↗
Other EIT Emerging Implant Technologies GmbH products
All EIT Emerging Implant Technologies GmbH devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page