EscapeRecalls & reported problems across all models
Boston Scientific Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 50
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 0
We found 2 models of Escape by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 50 adverse event reports in the FDA's MAUDE database, received between Mar 3, 2017 and Jul 16, 2026 — classified as 9 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Break” (27 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Escape problems
Device problems
- Break27
- Material Split, Cut or Torn10
- Detachment of Device or Device Component9
- Difficult to Open or Close8
- Adverse Event Without Identified Device or Use Problem7
- Unsealed Device Packaging4
- Device Contamination with Chemical or Other Material2
- Material Deformation2
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- No Consequences Or Impact To Patient16
- No Patient Involvement4
- Foreign Body In Patient2
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
- Perforation1
Escape models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Escape M0063902010DISLODGER, STONE, BASKET, URETERAL, METAL49 reports
- Escape M0063902000DISLODGER, STONE, BASKET, URETERAL, METAL1 report
Recent Escape adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Detachment of Device or Device Component; Difficult to Open or Close
IT WAS REPORTED THAT DURING THE PROCEDURE, WHEN ATTEMPTING TO CLOSE TO CAPTURE THE STONE, THE DEVICE FAILS TO FUNCTION PROPERLY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS; HOWEVER, INVESTIGATION OF THE RETURNED DEVICE REVEALED THAT THE PULL WIRE DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25841749 ↗MalfunctionFeb 16, 2026Nonstandard Device; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A FLEXIBLE URETEROSCOPIC LITHOTRIPSY PROCEDURE, AN ESCAPE BASKET WAS INTENDED TO BE USED. WHEN THE PHYSICIAN OPENED THE PACKAGE, IT WAS FOUND THAT THE TIP OF THE BASKET WAS MISSING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24352020 ↗InjuryJan 15, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT AN ESCAPE BASKET DEVICE WAS USED DURING A FLEXIBLE TRANSURETHRAL URETEROLITHOTRIPSY (F-TUL) PROCEDURE. DURING THE PROCEDURE. A WIRE ON THE SHEATH PART GOT DISCONNECTED FROM THE HANDLE. REPORTEDLY, THE REMAINING PART WAS REMOVED INSIDE THE PATIENT USING A FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24086256 ↗MalfunctionDec 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN ESCAPE BASKET WAS USED DURING A FLEXIBLE URETEROSCOPY WITH LASER LITHOTRIPSY PROCEDURE, DURING THE PROCEDURE, THE BASKET BROKE OFF INSIDE OF THE URETEROSCOPE. REPORTEDLY, THE SURGEON PUSHED A GUIDE WIRE THROUGH THE WORKING CHANNEL TO REMOVE SHEATH, AFTER THE SCOPE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23710216 ↗MalfunctionSep 30, 2025Break; Difficult to Open or Close; Material Split, Cut or Torn; Break; Difficult to Open or Close; Material Split, Cut or Torn
IT WAS REPORTED THAT AN ESCAPE RETRIEVAL BASKET WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. DURING THE PROCEDURE, IT WAS NOTICED THAT THE PROXIMAL END OF THE SHAFT AND THE OUTER SHEATH NEAR THE HANDLE WAS FRAYING, MAKING IT HARD TO OPEN AND CLOSE THE BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23184631 ↗MalfunctionSep 23, 2025Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT AN ESCAPE BASKET DEVICE WAS ABOUT TO BE USED DURING A RIGID URETEROSCOPIC LITHOTRIPSY PROCEDURE. DURING UNPACKING, IT WAS FOUND THAT THE TIP OF THE SHEATH AT THE DISTAL PART OF THE BASKET WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23127274 ↗MalfunctionSep 18, 2025Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING PREPARATION OF A FLEXIBLE URETEROSCOPIC LITHOTRIPSY PROCEDURE, IT WAS NOTED AFTER UNPACKING THAT THE FRONT PART OF THE ESCAPE BASKET WAS FRACTURED. THE PICTURE PROVIDED BY THE CUSTOMER SHOWED BASKET DETACHMENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23085275 ↗MalfunctionJul 24, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A ESCAPE BASKET WAS ABOUT TO BE USED DURING A RIGID URETEROSCOPIC LITHOTRIPSY AND URETEROLITHOTOMY PROCEDURE. DURING PREPARATION, UPON OPENING THE PACKAGE IT WAS FOUND THAT THE TIP AND TAIL END OF THE BASKET WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22588339 ↗InjuryJul 23, 2025Break; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Break; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT A ESCAPE BASKET DEVICE WAS USED DURING A FLEXIBLE TRANSURETHRAL URETEROLITHOTRIPSY (F-TUL) PROCEDURE. DURING THE PROCEDURE, A WIRE GOT DAMAGE ON THE SHEATH AND DISCONNECTED FROM THE HANDLE. ADDITIONALLY, THE OUTER SHEATH GOT SEPARATED AROUND THE MIDDLE PART. REPORTEDLY, THE PART OF THE OUTER SHEATH REMAINED INSIDE THE PATIENT AND WAS EVENTUALLY REMOVED USING A FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22577281 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT A ESCAPE BASKET WAS ABOUT TO BE USED DURING A FLEXIBLE URETEROSCOPIC LITHOTRIPSY PROCEDURE. DURING PREPARATION, IT WAS FOUND THAT THE BASKET WAS BROKEN WHEN THE PACKAGE WAS OPEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22543561 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT A ESCAPE BASKET DEVICE WAS ABOUT TO BE USED DURING A RIGID URETEROSCOPIC LITHOTRIPSY PROCEDURE. DURING PREPARATION, IT WAS FOUND THAT THE TIP OF THE BASKET WAS FRACTURED AFTER UNPACKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22543358 ↗InjuryJul 3, 2025Break; Break
IT WAS REPORTED THAT AN ESCAPE BASKET DEVICE WAS USED DURING A FLEXIBLE URETEROSCOPY PROCEDURE. DURING THE PROCEDURE, THE PHYSICIAN WAS HOLDING THE STONE WITH THE ESCAPE BASKET; HOWEVER, ONE OF THE BASKET WIRES SUDDENLY BROKE AND DETACHED FROM THE BASKET. REPORTEDLY, THEY TRIED TO RELEASE THE STONE, BUT THE BASKET WAS NOT WORKING, AND BASKET HAD TO BE DISABLED. THE DETACHED WIRE WAS RETRIEVED WITH ANOTHER BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22398594 ↗MalfunctionJun 4, 2025Break; Break
IT WAS REPORTED THAT AN ESCAPE DEVICE WAS TO BE USED IN A RIGID URETEROSCOPIC LITHOTRIPSY PROCEDURE. DURING UNPACKING, IT WAS FOUND THAT THE TIP OF THE BASKET WAS FRACTURED. THE BASKET WIRES WERE BROKEN AND ONE WIRE WAS COMPLETELY DETACHED FROM THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22144500 ↗MalfunctionMay 19, 2025Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT AN ESCAPE BASKET DEVICE WAS ABOUT TO BE USE DURING A URETEROSCOPIC HOLMIUM LASER LITHOTRIPSY AND STONE REMOVAL PROCEDURE. REPORTEDLY, DURING UNPACKING IT WAS FOUND THAT THE BASKET WAS FRACTURED AND DETACHED. ADDITIONALLY, IT WAS NOTICED THAT THE SHEATH WAS TORN AT THE MIDDLE PART. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22048446 ↗InjuryMay 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN ESCAPE RETRIEVAL BASKET DEVICE WAS USED DURING A URETEROSCOPY PROCEDURE. DURING THE PROCEDURE, THE BASKET DETACHED FROM THE DEVICE INSIDE THE PATIENT. REPORTEDLY, THE EXACT DEVICE TO RETRIEVE THE DETACHED BASKET IS UNKNOWN. IT WAS FURTHER REPORTED THAT POSSIBLY FLEXIBLE GRASPER OR ANOTHER ESCAPE WAS USED TO RETRIEVE THE DETACHED BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22027610 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page