GRYPHON PROKNOT PERMACORDRecalls & reported problems across all models
Medos International Sàrl · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 126
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls
- 0
We found 2 models of GRYPHON PROKNOT PERMACORD by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 126 adverse event reports in the FDA's MAUDE database, received between Jun 18, 2018 and May 4, 2026 — classified as 10 injuries, 116 malfunctions.
The most frequently reported specific device problem is “Break” (66 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported GRYPHON PROKNOT PERMACORD problems
Device problems
- Break66
- Migration17
- Difficult to Advance12
- Material Frayed9
- Adverse Event Without Identified Device or Use Problem4
- Detachment of Device or Device Component4
- Migration or Expulsion of Device4
- Loosening of Implant Not Related to Bone-Ingrowth3
Patient problems
- No Clinical Signs, Symptoms or Conditions72
- No Consequences Or Impact To Patient37
- Not Applicable7
- Foreign Body In Patient3
- No Known Impact Or Consequence To Patient3
- Pain3
- Unspecified Tissue Injury3
- Injury2
GRYPHON PROKNOT PERMACORD models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- GRYPHON PROKNOT PERMACORD 210820FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE109 reports
- GRYPHON PROKNOT PERMACORD 210821FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE17 reports
Recent GRYPHON PROKNOT PERMACORD adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2026Migration
THIS IS 2 OF 2 ASSOCIATED WITH (B)(4). IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY, AFTER SUCCESSFULLY USING TWO GRYPHON PROKNOTS WITH DIFFERENT LOTS, THEY TRIED USING TWO GRYPHON PROKNOTS THAT PULLED DIRECTLY OUT OF BONE WHEN TYING THE KNOTS. THE SURGEON PERFORMED THE EXACT SAME PROCEDURE AND STEPS WITH THE ANCHORS AS WITH THE PREVIOUS TWO DEVICES. THE BONE QUALITY WAS GREAT. FOR SOME REASON THESE TWO GRYPHON PROKNOTS PULLED OUT WITH VERY LITTLE TENSION PUT ON THEM. NO PATIENT HARM OCCURRED. A DELAY OF 25 MINUTES IN THE CASE WAS A RESULT, AND SURGEON USED A COMPETITIVE PRODUCT TO COMPLETE THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25067748 ↗MalfunctionMay 4, 2026Migration
THIS IS 1 OF 2 ASSOCIATED WITH (B)(4). IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY, AFTER SUCCESSFULLY USING TWO GRYPHON PROKNOTS WITH DIFFERENT LOTS, THEY TRIED USING TWO GRYPHON PROKNOTS THAT PULLED DIRECTLY OUT OF BONE WHEN TYING THE KNOTS. THE SURGEON PERFORMED THE EXACT SAME PROCEDURE AND STEPS WITH THE ANCHORS AS WITH THE PREVIOUS TWO DEVICES. THE BONE QUALITY WAS GREAT. FOR SOME REASON THESE TWO GRYPHON PROKNOTS PULLED OUT WITH VERY LITTLE TENSION PUT ON THEM. NO PATIENT HARM OCCURRED. A DELAY OF 25 MINUTES IN THE CASE WAS A RESULT, AND SURGEON USED A COMPETITIVE PRODUCT TO COMPLETE THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25067746 ↗MalfunctionFeb 13, 2026Difficult to Advance
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, ALTHOUGH THE GRYPHON BR W/ PROKNOT WAS USED ACCORDING TO THE PROCEDURE MANUAL, THE KNOT DID NOT PASS THROUGH THE POST THREAD, MAKING IT IMPOSSIBLE TO SUTURE. AN ADDITIONAL PRODUCT FROM THE SAME LOT WAS USED, AND THE SURGERY WAS COMPLETED WITHOUT INCIDENT. THERE WAS A 30 MINUTE SURGICAL DELAY AND NO HARM TO THE PATIENT. THE DEVICE WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335190 ↗MalfunctionDec 24, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE GRYPHON BR W/ PROKNOT DEVICE WAS BEING USED WHEN THE ANCHOR FRACTURED DOWN THE CENTER WHILST "MALLETING" IN. IT IS UNKNOWN WHETHER ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23887889 ↗MalfunctionAug 26, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR (ARCR) SURGICAL PROCEDURE, WHEN THE SURGEON ATTEMPTED TO INSERT THE GRYPHON BR W/ PROKNOT ANCHOR DEVICE INTO THE BONE HOLE IT WAS OBSERVED THAT TIP OF THE ANCHOR BROKE OFF. ANOTHER PRODUCT WAS USED, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. IT WAS REPORTED THAT NO PIECES OF THE DEVICE WERE LEFT IN THE BODY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT, OR SURGICAL DELAY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22894169 ↗MalfunctionAug 22, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE; AFTER CREATING A BURR HOLE, THE SURGEON ATTEMPTED TO INSERT THE ANCHOR OF THE GRYPHON BR W/ PROKNOT DEVICE INTO THE BURR HOLE, BUT THE TIP OF THE ANCHOR BROKE OFF. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22873718 ↗MalfunctionJul 18, 2025Migration; Migration
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGERY, THE GRYPHON PEEK W/PROKNOT DEVICE ANCHOR WAS PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22540720 ↗MalfunctionJul 2, 2025Break; Device Damaged Prior to Use; Break; Device Damaged Prior to Use
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT WHEN THE GRYPHON BR W/ PROKNOT DEVICE WAS OPENED, THE TIP OF THE ANCHOR SEEMED TO BE CHIPPED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE IS NO POSSIBILITY OF FRAGMENTS BEING LEFT IN THE BODY SINCE THE ANCHOR IN QUESTION WAS NOT INSERTED IN THE BODY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22383830 ↗MalfunctionMay 16, 2025Use of Device Problem; Device-Device Incompatibility; Use of Device Problem; Device-Device Incompatibility
THIS IS REPORT 2 OF 2 OF (B)(4). IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY BANKART SURGICAL PROCEDURE THE GRYPHON BR W/ PRO-KNOT DEVICE TWO IMPLANTS COULD NOT BE DEPLOYED DUE TO THE DAMAGED IN THE KNOT/ SUTURE STRAND. THE SURGEON THEN USED THE GRYPHON P WHICH WORKED PROPERLY IN THE PATIENT. THERE WAS ABOUT A TEN DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22039207 ↗MalfunctionMay 16, 2025Use of Device Problem; Device-Device Incompatibility; Use of Device Problem; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY BANKART SURGICAL PROCEDURE THE GRYPHON BR W/ PROKNOT DEVICE TWO IMPLANTS COULD NOT BE DEPLOYED DUE TO THE DAMAGED IN THE KNOT/ SUTURE STRAND. THE SURGEON THEN USED THE GRYPHON P WHICH WORKED PROPERLY IN THE PATIENT. THERE WAS ABOUT A TEN DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22039161 ↗MalfunctionFeb 25, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC BANKART REPAIR PROCEDURE FOR SHOULDER INSTABILITY IT WAS OBSERVED THAT WHEN THE SURGEON WAS PULLING THE FREE END OF THE SUTURE TO PASS THROUGH THE LABRUM, IT LOOKED AS IF THE INNER PART OF THE GRYPHON BR W/ PROKNOT ANCHOR BROKE, EFFECTIVELY UNLOADING THE ANCHOR. IT WAS REPORTED THAT TWO ADDITIONAL GRYPHON BR W/ PROKNOT ANCHORS WERE USED, AND THE SAME PROBLEM OCCURRED. IT WAS REPORTED THAT THE ANCHOR BODY REMAINED EMBEDDED IN GLENOID BONE BUT THERE WAS NO MORE SUTURE RUNNING THROUGH IT. IT WAS REPORTED THAT ANOTHER COMPETITOR¿S ANCHOR WAS SUCCESSFULLY USED INSTEAD. IT WAS REPORTED THAT THERE WAS A 15-MINUTE CASE DELAY AND NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21459356 ↗MalfunctionFeb 25, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC BANKART REPAIR PROCEDURE FOR SHOULDER INSTABILITY IT WAS OBSERVED THAT WHEN THE SURGEON WAS PULLING THE FREE END OF THE SUTURE TO PASS THROUGH THE LABRUM, IT LOOKED AS IF THE INNER PART OF THE GRYPHON BR W/ PROKNOT ANCHOR BROKE, EFFECTIVELY UNLOADING THE ANCHOR. IT WAS REPORTED THAT TWO ADDITIONAL GRYPHON BR W/ PROKNOT ANCHORS WERE USED, AND THE SAME PROBLEM OCCURRED. IT WAS REPORTED THAT THE ANCHOR BODY REMAINED EMBEDDED IN GLENOID BONE BUT THERE WAS NO MORE SUTURE RUNNING THROUGH IT. IT WAS REPORTED THAT ANOTHER COMPETITOR¿S ANCHOR WAS SUCCESSFULLY USED INSTEAD. IT WAS REPORTED THAT THERE WAS A 15-MINUTE CASE DELAY AND NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21459270 ↗MalfunctionFeb 25, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC BANKART REPAIR PROCEDURE FOR SHOULDER INSTABILITY IT WAS OBSERVED THAT WHEN THE SURGEON WAS PULLING THE FREE END OF THE SUTURE TO PASS THROUGH THE LABRUM, IT LOOKED AS IF THE INNER PART OF THE GRYPHON BR W/ PROKNOT ANCHOR BROKE, EFFECTIVELY UNLOADING THE ANCHOR. IT WAS REPORTED THAT TWO ADDITIONAL GRYPHON BR W/ PROKNOT ANCHORS WERE USED, AND THE SAME PROBLEM OCCURRED. IT WAS REPORTED THAT THE ANCHOR BODY REMAINED EMBEDDED IN GLENOID BONE BUT THERE WAS NO MORE SUTURE RUNNING THROUGH IT. IT WAS REPORTED THAT ANOTHER COMPETITOR¿S ANCHOR WAS SUCCESSFULLY USED INSTEAD. IT WAS REPORTED THAT THERE WAS A 15-MINUTE CASE DELAY AND NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21459204 ↗MalfunctionDec 10, 2024Detachment of Device or Device Component; Difficult to Advance; Detachment of Device or Device Component; Difficult to Advance
IT WAS REPORTED THAT DURING AN UNKNOWN SURGICAL PROCEDURE WHILE USING 3X GRYPHON P BR DS ANCHOR W/OC DEVICE AND INFLOW TUBING FMS VUE 24PK, THE GRYPHON DEVICE WAS DAMAGED. THE ANCHORING OF THE DEVICE WHEN KNOTTING CAME OUT OF THE BONE OF THE PTE. THE FENESTRATED GUIDE WAS DAMAGED THERE WAS AN UNKNOWN AMOUNT OF DELAY TO THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO OTHER MEDICAL PROCEDURES WERE REQUIRED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 PF THE SAME EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20894555 ↗MalfunctionNov 18, 2024Break; Break
REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE EYELETS OF TWO (2) GRYPHON BR W/ PROKNOT DEVICES BROKE WHILE TYING THE KNOT ON THE PROKNOT ANCHORS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20704801 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page