HeartSine Samaritan PAD 450PRecalls & reported problems across all models
Heartsine Technologies Ltd · 5 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name at least one HeartSine Samaritan PAD 450P model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 23
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls
- 3
We found 5 models of HeartSine Samaritan PAD 450P by Heartsine Technologies Ltd in the FDA's device registry (GUDID). Together they are named in 23 adverse event reports in the FDA's MAUDE database, received between Aug 20, 2021 and May 6, 2026 — classified as 2 deaths, 21 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (11 reports).
3 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
HeartSine Samaritan PAD 450P recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 8, 2024Class IIOpen, Classified
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.
Z-1820-2024 · Root cause: Process control · names 4 of these models · Recall details · FDA record ↗
- Feb 14, 2024Class IIOpen, Classified
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
Z-1141-2024 · Root cause: Process control · Recall details · FDA record ↗
- Jan 13, 2022Class IIITerminated
Semi-automatic defibrillators shipped out by Stryker without battery and electrode packs.
Z-0745-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported HeartSine Samaritan PAD 450P problems
Device problems
- Failure to Power Up11
- No Audible Prompt/Feedback7
- Failure to Sense3
- Battery Problem1
- Defective Component1
- Device Sensing Problem1
- Electrical Power Problem1
- Failure to Analyze Signal1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- Cardiac Arrest1
HeartSine Samaritan PAD 450P models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- HeartSine Samaritan PAD 450P SAM450PAutomated external defibrillators (non-wearable)2 recalls2 deaths23 reports
- HeartSine Samaritan PAD 450P SAM450PAutomated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM450PAutomated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM 450PAutomated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM450PAutomated external defibrillators (non-wearable)1 recall0 reports
Recent HeartSine Samaritan PAD 450P adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Failure to Power Up
A CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE'S POWER BUTTON DOES NOT FUNCTION. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25091199 ↗MalfunctionApr 1, 2026No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DOES NOT PROVIDE AUDIBLE INSTRUCTION WHEN POWERED ON. WITHOUT VOICE PROMPTS, A LAY USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE TO DELIVER DEFIBRILLATION THERAPY TO A PATIENT. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24754462 ↗MalfunctionSep 8, 2025Failure to Power Up; Device Sensing Problem; Failure to Power Up; Device Sensing Problem
HEARTSINE CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO ANALYZE RHYTHM DURING AN INTERVENTION. THERE WAS NO HARM TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22994127 ↗MalfunctionAug 25, 2025No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE WOULD NOT PROVIDE AUDIO VOICE PROMPTS. WITHOUT VOICE PROMPTS, A LAY USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE ON A PATIENT WHICH COULD DELAY THERAPY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22889957 ↗MalfunctionJul 23, 2025Failure to Sense; Failure to Sense
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE HAD FAILED TO DETECT A PATIENT IMPEDANCE DURING A PATIENT INVOLVED EVENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22574224 ↗MalfunctionJun 19, 2025Failure to Analyze Signal; Failure to Analyze Signal
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DID NOT ANALYZE WHEN CONNECTED TO A PATIENT. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THE PATIENT DID NOT SURVIVE.
Patient problems: Cardiac Arrest
MAUDE report 22294185 ↗MalfunctionJun 5, 2025Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22154775 ↗MalfunctionJan 14, 2025No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE FAILED TO ISSUE AUDIO PROMPTS AT POWER UP. WITHOUT VOICE PROMPTS, A LAY USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE ON A PATIENT WHICH COULD DELAY THERAPY.THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21147115 ↗MalfunctionJan 10, 2025Program or Algorithm Execution Problem; Program or Algorithm Execution Problem
HEARTSINE RECALLED THE DEVICE UNDER FA301. UPON RECEIPT AT HEARTSINE, THE INVESTIGATION DISCOVERED THE DEVICE HAD BEEN INCORRECTLY PROGRAMMED DURING THE MANUFACTURING PROCESS. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER THERAPY TO SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21124415 ↗MalfunctionOct 23, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT BE POWERED ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20512362 ↗MalfunctionSep 17, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20241921 ↗MalfunctionSep 10, 2024No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE FAILED TO DELIVER VOICE PROMPTS AT POWER UP. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20186823 ↗MalfunctionJul 23, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT BE POWERED ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19809793 ↗DeathJun 27, 2024No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE ISSUED NO VOICE PROMPTS DURING A PATIENT INVOLVED EVENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THE PATIENT INVOLVED IN THE REPORTED EVENT IS DECEASED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19630619 ↗MalfunctionJun 26, 2024No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE DID NOT PROVIDE ANY VOICE PROMPTS. WITHOUT VOICE PROMPTS, A LAY USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE ON A PATIENT. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19613475 ↗
Other Heartsine Technologies Ltd products
All Heartsine Technologies Ltd devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page