HydroSet XTRecalls & reported problems across all models
Stryker Corporation · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls
- 2
We found 4 models of HydroSet XT by Stryker Corporation in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Jul 25, 2019 and Feb 28, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (3 reports).
2 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
HydroSet XT recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Nov 15, 2018Terminated
Incorrect expiry date of 9-Feb-20 Correct expiry date is 30-Aug-19
Z-0678-2019 · Root cause: Labeling mix-ups · Recall details · FDA record ↗
- Nov 15, 2018Terminated
Incorrect expiry date of 17-Dec-19 Correct expiry date is 11-Aug-19
Z-0677-2019 · Root cause: Labeling mix-ups · Recall details · FDA record ↗
Most reported HydroSet XT problems
Device problems
- Use of Device Problem3
- Crack2
- Fracture2
- Defective Device1
- Insufficient Device Problem Information1
Patient problems
- No Consequences Or Impact To Patient2
- Ambulation or Postural Difficulties1
- No Known Impact Or Consequence To Patient1
- Pain1
HydroSet XT models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- HydroSet XT 897003FILLER, BONE VOID, CALCIUM COMPOUND1 recall1 report
- HydroSet XT 897005FILLER, BONE VOID, CALCIUM COMPOUND1 recall1 report
- HydroSet XT 897010FILLER, BONE VOID, CALCIUM COMPOUND1 report
- HydroSet XT 897015FILLER, BONE VOID, CALCIUM COMPOUND1 report
Recent HydroSet XT adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryFeb 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A PATIENT WAS IMPLANTED WITH A VARIAX PLATE, VARIAX SCREWS, A 2.5 CC VITOSS BA2X FOAM PACK, AND 15 CC HYDROSET XT IN THEIR FOOT FOLLOWING REMOVAL OF A BONE MASS. BEGINNING APPROXIMATELY FOUR-MONTHS POST-OPERATIVELY, THE PATIENT REPORTED EXPERIENCING "SEVERE FOOT PAIN" AND DIFFICULTY WALKING. THIS REPORT CAPTURES THE 15 CC HYDROSET XT.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 18804141 ↗MalfunctionJul 25, 2019Use of Device Problem; Crack; Fracture; Use of Device Problem
IT WAS REPORTED THAT DURING A PRIMARY PROCEDURE FOR A LEFT CALCANEUS FRACTURE, THAT HYDROSET XT (10 CC) ¿HARDENED SOLID¿ AFTER APPROXIMATELY 90 TO 105 SECONDS AND "A LOT" OF FORCE WAS NEEDED TO GET THE LIQUID OUT OF ITS CONTAINER AND INTO THE POWDER. THE PROCEDURE CALLED FOR A PLATE AND HYDROSET XT WAS INTENDED TO FILL ANY REMAINING CRACKS; HOWEVER, IT WAS DECIDED BY THE SURGEON TO DISCONTINUE USE OF HYDROSET XT DUE TO PREMATURE HARDENING OF THE COMPONENT. THIS RESULTED IN A SURGICAL DELAY OF 4-5 MINUTES. NO ADVERSE CONSEQUENCES TO THE PATIENT AND SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826120 ↗MalfunctionJul 25, 2019Use of Device Problem; Crack; Fracture; Use of Device Problem
IT WAS REPORTED THAT DURING A PRIMARY PROCEDURE FOR A LEFT CALCANEUS FRACTURE, THAT HYDROSET XT (3 CC) ¿HARDENED SOLID¿ AFTER APPROXIMATELY 90 TO 105 SECONDS AND "A LOT" OF FORCE WAS NEEDED TO GET THE LIQUID OUT OF ITS CONTAINER AND INTO THE POWDER. THE PROCEDURE CALLED FOR A PLATE AND HYDROSET XT WAS INTENDED TO FILL ANY REMAINING CRACKS; HOWEVER, IT WAS DECIDED BY THE SURGEON TO DISCONTINUE USE OF HYDROSET XT DUE TO PREMATURE HARDENING OF THE COMPONENT. THIS RESULTED IN A SURGICAL DELAY OF 4-5 MINUTES. NO ADVERSE CONSEQUENCES TO THE PATIENT AND SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8826117 ↗MalfunctionJul 25, 2019Use of Device Problem; Use of Device Problem; Defective Device
IT WAS REPORTED THAT DURING A PRIMARY PROCEDURE FOR A LEFT CALCANEUS FRACTURE, THAT HYDROSET XT (5 CC) ¿HARDENED SOLID¿ AFTER APPROXIMATELY 90 TO 105 SECONDS AND "A LOT" OF FORCE WAS NEEDED TO GET THE LIQUID OUT OF ITS CONTAINER AND INTO THE POWDER. THE PROCEDURE CALLED FOR A PLATE AND HYDROSET XT WAS INTENDED TO FILL ANY REMAINING CRACKS; HOWEVER, IT WAS DECIDED BY THE SURGEON TO DISCONTINUE USE OF HYDROSET XT DUE TO PREMATURE HARDENING OF THE COMPONENT. THIS RESULTED IN A SURGICAL DELAY OF 4-5 MINUTES. NO ADVERSE CONSEQUENCES TO THE PATIENT AND SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826116 ↗
Other Stryker Corporation products
- RemB7 models · 3,073 reports
- Prime Series21 models · 1,548 reports · 2 recalls
- TARGET114 models · 1,077 reports
- ProCuity17 models · 368 reports · 13 recalls
- Spetzler-Malis20 models · 7 reports · 19 recalls
- SYNCHRO 28 models · 313 reports · 9 recalls
- Neptune E-SEP19 models · 412 reports · 7 recalls
- StrykeFlow4 models · 552 reports · 4 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page