IH-Card ABO/D(DVI-)+Rev A1,BRecalls & reported problems across all models
Bio-Rad Medical Diagnostics GmbH · 4 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 137
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 4 models of IH-Card ABO/D(DVI-)+Rev A1,B by Bio-Rad Medical Diagnostics GmbH in the FDA's device registry (GUDID). Together they are named in 137 adverse event reports in the FDA's MAUDE database, received between Oct 23, 2018 and Apr 28, 2026 — classified as 137 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (110 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported IH-Card ABO/D(DVI-)+Rev A1,B problems
Device problems
- False Positive Result110
- False Negative Result18
- Incorrect, Inadequate or Imprecise Result or Readings16
- Output Problem7
- Defective Device6
- Defective Component4
- Incorrect Measurement2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions128
- No Known Impact Or Consequence To Patient27
- Reaction3
- Test Result3
- Hemolysis1
- Insufficient Information1
- No Code Available1
- No Consequences Or Impact To Patient1
IH-Card ABO/D(DVI-)+Rev A1,B models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- 127 reports
- 6 reports
- 2 reports
- 2 reports
Recent IH-Card ABO/D(DVI-)+Rev A1,B adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2026Display or Visual Feedback Problem
USER REPORTED MULTIPLE CARDS WITHIN THE SAME LOT THAT HAVE A LOW GEL-LEVEL ON THE ANTI-D WELL ONLY. IT SEEMS THAT THE AGGLUTINATION IS LOWER THAN EXPECTED DUE TO THE LOW GEL VOLUME. THUS, IH-COM IS INTERPRETING AS 2+, WHEN THE INTERPRETATION SHOULD BE EITHER 3+ OR 4+.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019743 ↗MalfunctionMar 21, 2026False Positive Result
THE USER INDICATED GETTING A FALSE POSITIVE WITH THE ANTI-D WELL WHEREAS THEY HAD TYPED THE PATIENT AS B RHD NEGATIVE. INVESTIGATION: - DATA REVIEW: THE RHD NEGATIVE RESULT WAS CONFIRMED AFTER SEVERAL REPEATS IN THE SAME WORK CONDITIONS, AND ALSO IN MANUAL WITH TUBE TECHNIQUE. - TREND ANALYSIS : ON THE (B)(6)., 6 COMPLAINTS ON THE IH-CARD ABO/D(DVI-)+REV A1B PN 813112100 PRODUCT REFERENCE. BUT THIS CASE IN THE UNIQUE RECORD ON THE LOT 9546010. - BATCH RECORDS : THE MANUFACTURING QUALITY RECORDS FOR THE REPORTED PRODUCTS ARE ALL IN COMPLIANCE AND THAT ALL SPECIFICATIONS WERE MET AT BATCH RELEASE. - TRACES FILES ANALYSIS : SEVERAL POTENTIAL EXPLANATIONS WERE CONSIDERED, INCLUDING (DRIED GEL IN THE ANTI-DVI- WELL, FIBRIN IN THE RBC DILUTION, CARRYOVER BETWEEN THE WELLS, OR INTER-SAMPLE CARRYOVER FROM THE PRIOR SAMPLE "(B)(6)," WHICH IS OF O BLOOD TYPE AND CONTAINS HIGH TITERS OF ANTI-A AND ANTI-B ANTIBODIES). UNFORTUNATELY, NONE OF THESE HYPOTHESES HAVE PROVEN TO BE DEMONSTRABLE OR VALID, DUE TO THE UNAVAILABILITY OF THE GEL CONTROL IMAGE, AND PRIMARILY BECAUSE THE DISPENSING SEQUENCE WENT FROM THE CTL WELL TO THE ANTI-A WELL, PASSING THROUGH THE ANTI-DVI- AND ANTI-B WELLS. AND WE CAN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24656606 ↗MalfunctionFeb 17, 2026False Negative Result
THE CUSTOMER REPORTED A FALSE POSITIVE ANTI-B RESULT WITH IH-CARD ABO/D(DVI-)+REV A1,B WHEN TESTING ON THEIR IH-1000 INSTRUMENT. A PATIENT SAMPLE WAS TESTED IN DUPLICATE. THE ANTI-B WAS INITIALLY 1 POSITIVE BUT NEGATIVE WHEN REPEATED USING THE SAME INSTRUMENT AND REAGENTS. THE TRACE FILES OF THE AFFECTED INSTRUMENT WERE COLLECTED FOR INVESTIGATION.THE SNAPSHOTS PROVIDED DIDN'T SHOWS ANY ISSUE CONCERNING THIS SAMPLES. THE USER PERFOMED TESTING WITH THE YELLOW WELL AND NOTICED THE SAME FALSE-POSITIVE RESULTS.THE ISSUE COULD NOT BE REPRODUCED BY THE MANUFACTURER. WE ALSO CHECKED THE IH-LISS RETENTION SAMPLES IN DREIEICH, AND AMONG THE 34 BOXES CURRENTLY IN THE REFERENCE STOCK, NO WELLS WITH ABNORMAL COLORATION WERE OBSERVED. THIS LEADS US TO BELIEVE THAT THIS IS AN ISOLATED CASE, AND WE ARE NOT ABLE TO LINK IT TO MANUFACTURING. RESOLUTION: FOR IH-LISS, IT IS SPECIFICALLY INDICATED NOT TO USE THE PRODUCT IN CASE OF TURBIDITY; THIS ALSO APPLIES WHEN THERE ARE VISIBLE COLOR DIFFERENCES IN THE WELLS. IN THIS CONTEXT, OUR RECOMMENDATION IS TO VISUALLY CHECK THE IH-]LISS BEFORE LOADING IT ONTO THE DEVICES, AND IF ANY TURBIDITY OR DISCOLORATION IS OBSERVED, TESTING WITH THAT RACK SHOULD NOT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24362188 ↗MalfunctionFeb 3, 2026False Positive Result
CUSTOMER REPORTED FALSE POSITIVE ANTI-B ON 2 SAMPLES. TESTING WAS PERFORMED USING IH-1000 WITH IH-CARD ABO/D(DVI-)+REV A1,B LOT 9535010, SAMPLES (B)(6) WERE TESTED IN DUPLICATE. THE FALSE POSITIVE ANTI-B REACTION INTERPRETATIONS APPEAR JUSTIFIED. BOTH SAMPLES WERE REPEATED ON THE SAME INSTRUMENT AND ALSO TESTED BY TUBE METHOD. REPEAT TESTING FOR BOTH SAMPLES WAS O POSITIVE. THE SHIPMENT FOR THE ABO CARDS WAS RECEIVED 12/1/2025 AND THERE WERE (B)(4) BOXES. THEY HAVE 9 BOXES REMAINING. THE CUSTOMER REMOVED THE ABO CARDS FROM THE INSTRUMENT AND DID SEE SOME REAGENT SPLASHES, SPECIFICALLY IN THE ANTI-D WELLS. THEY SPUN THEM DOWN AND THEY LOOKED BETTER, SO THEY ARE BACK IN USE. CUSTOMER IS A LONG TIME USER THAT IS AWARE OF IFU INSTRUCTIONS FOR STORAGE AND HANDLING OF CARDS. INVESTIGATION: NO OTHER COMPLAINTS FOR THIS LOT REGISTERED. THE PRODUCTION BATCH RECORDS WERE CHECKED, AND ALL MANUFACTURING/QC RECORDS ARE COMPLIANT, AND THERE ARE NO MANUFACTURING-RELATED QUALITY CONCERNS. THE CARDS OF THE RETENTION SAMPLE WERE VISUALLY CHECKED FOR THE SEALING FOIL, THE PRESENCE OF SUPERNATANT, AND GEL HOMOGENEITY, AND SHOWED NO ABNORMALITIES. SEROLOGICAL TESTS WERE PERFORMED USING KNOWN A+ AND B+
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24244754 ↗MalfunctionAug 25, 2025False Negative Result; False Negative Result
THE CUSTOMER REPORTED DISCREPANT RHD RESULTS BETWEEN BIO-RAD AND ORTHO GEL. THE SAME PATIENT WAS TESTED ON (B)(6) 2024 (PLEASE SEE REPORT 9610824-2025-000027) AND THEN AGAIN ON (B)(6) 202 (THIS REPORT), USING TWO DIFFERENT LOTS OF IH-CARD ABO(D(DVI-)+REVA1,B ON TWO DIFFERENT IH-1000 INSTRUMENTS THE RHD WAS NEGATIVE EACH TIME IN (B)(6) 2025 THE CUSTOMER WAS NOTIFIED BY THE PATIENT'S PHYSICIAN THAT THE PATIENT WAS TESTED AT ANOTHER FACILITY USING AN UNKNOWN ORTHO REAGENT AND YIELDED A 2+ REACTION. THE CUSTOMER DID NEITHER PROVIDE A PATIENT SAMPLE NOR A PRODUCT SAMPLE OF THE ALLEGEDLY DEFECTIVE IH-CARD FOR INVESTIGATIONAL TESTING. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLES OF BOTH LOTS OF THE ALLEGEDLY DEFECTIVE IH-CARDS. FIRST, ALL CARDS FROM THE RETENTION SAMPLES WERE VISUALLY CHECKED REGARDING THE SEALING FOIL, THE PRESENCE OF SUPERNATANT, THE GEL HOMOGENEITY AND THE ABSENCE OF SPATTER IN THE REACTION CHAMBERS. THE SUPERNATANT WAS REDUCED ON ALL CARDS. THE CARDS WERE THEN TESTED ON THE IH-1000. TWO WEEK RHD POSITIVE SPECIFICITY, FOUR HETEROZYGOUS RHD POSITIVE SPECIFICITY AND ONE RHD NEGATIVE SPECIFICITY OF THE RETENTION SAMPLE WERE TESTED ON THE IH-100
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881977 ↗MalfunctionAug 25, 2025False Negative Result; False Negative Result
THE CUSTOMER REPORTED DISCREPANT RHD RESULTS BETWEEN BIO-RAD AND ORTHO GEL. THE SAME PATIENT WAS TESTED ON (B)(6) 2024 (THIS REPORT) AND THEN AGAIN ON (B)(6) 2025 (PLEASE REFER TO REPORT 9610824-2025-000026), USING TWO DIFFERENT LOTS OF IH-CARD ABO(D(DVI-)+REVA1,B ON TWO DIFFERENT IH-1000 INSTRUMENTS THE RHD WAS NEGATIVE EACH TIME IN (B)(6) 2025 THE CUSTOMER WAS NOTIFIED BY THE PATIENT'S PHYSICIAN THAT THE PATIENT WAS TESTED AT ANOTHER FACILITY USING AN UNKNOWN ORTHO REAGENT AND YIELDED A 2+ REACTION. THE CUSTOMER DID NEITHER PROVIDE A PATIENT SAMPLE NOR A PRODUCT SAMPLE OF THE ALLEGEDLY DEFECTIVE IH-CARD FOR INVESTIGATIONAL TESTING. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLES OF BOTH LOTS OF THE ALLEGEDLY DEFECTIVE IH-CARDS. FIRST, ALL CARDS FROM THE RETENTION SAMPLES WERE VISUALLY CHECKED REGARDING THE SEALING FOIL, THE PRESENCE OF SUPERNATANT , THE GEL HOMOGENEITY AND THE ABSENCE OF SPATTER IN THE REACTION CHAMBERS. THE SUPERNATANT WAS REDUCED ON ALL CARDS. THE CARDS WERE THEN TESTED ON THE IH-1000. TWO WEEK RHD POSITIVE SPECIFICITY, FOUR HETEROZYGOUS RHD POSITIVE SPECIFICITY AND ONE RHD NEGATIVE SPECIFICITY OF THE RETENTION SAMPLE WERE TESTED ON THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881972 ↗MalfunctionMar 6, 2025False Positive Result; False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE REACTION IN THE ANTI-B WELL FOR ONE PATIENT SAMPLE OF BLOOD TYPE A RHD POSITIVE. TESTING WAS CONDUCTED ON THE CUSTOMERS IH-1000 INSTRUMENT. THE CUSTOMER DID NOT PROVIDE A SAMPLE OF THE ALLEGEDLY DEFECTIVE PRODUCT FOR INVESTIGATIONAL TESTING NOR THE PATIENT SAMPLE THAT HAD CAUSED THE FALSE POSITIVE TEST RESULT. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE OF THE AFFECTED LOT. FIRST, ALL CARDS FROM THE RETENTION SAMPLE WERE VISUALLY CHECKED REGARDING THE SEALING FOIL, THE PRESENCE OF SUPERNATANT AND THE GEL HOMOGENEITY AND SHOWED NO ABNORMALTY. FOR THE SEROLOGICAL TESTING, 5 DONORS (BLOOD GROUP A) WITH NEGATIVE SPECIFICITY IN THE ANTI-B CAVITY, AS WELL AS ONE DONOR (BLOOD GROUP B) WITH POSITIVE SPECIFICITY IN THE ANTI-B WERE TESTED ON THE IH-1000. ADDITIONALLY, TWO SAMPLES OF THE IH_BASIC QC SET THAT THE CUSTOMER HAD USED DURING HIS TESTS WERE ALSO USED FOR THIS TEST. ALL RESULTS WERE CORRECT. NO FALSE POSITIVE REACTIONS IN THE ANTI-B CAVITY OCCURRED, ALL SIX DONOR BLOODS REACTED CORRECTLY. NO MANUFACTURING-RELATED QUALITY ISSUE WITH THE CARDS WERE NOTICED. TRACE FILES OF THE AFFECTED IH-1000 INSTRUMENT WERE ANALYZED A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21534487 ↗MalfunctionJan 9, 2025False Positive Result; False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE REACTION OF ONE PATIENT SAMPLE WITH THE ANTI-B OF IH-CARD ABO/D(DVI-)+REV. A1, B ON THE IH-1000 DURING FORWARD TYPING AND IN THE DIRECT COOMBS TEST. THE PATIENT IS A POSITIVE. THE PATIENT SAMPLE WAS RE-TESTED ON THE SAME DAY AND ON THE SAME INSTRUMENT, USING THE SAME REAGENTS AND ACCURATE RESULTS CONSISTENT WITH THE PATIENT'S HISTORY WERE OBTAINED. TRACE FILES OF THE AFFECTED IH-1000 INSTRUMENT WERE ANALYZED WITH THE FOLLOWING RESULTS: FOR THE SAMPLE IN QUESTION, THREE TESTS WERE REQUESTED AT THE SAME TIME: BLOOD GROUP TEST (BGR), DIRECT ANTIGLOBULIN TEST (DAT), AND ANTIBODY SCREENING. THE ANTIBODY SCREENING TEST RESULTED NEGATIVE WHILE THE BLOOD GROUP TEST AND THE DAT TEST SHOWED UNEXPECTED POSITIVE REACTIONS. THE BGR FORWARD TEST WAS DISPENSED FROM THE CONTROL WELL TO THE ANTI-A WELL, THE DAT WELL 1 WAS SUBSEQUENTLY DISPENSED USING THE SAME DILUTION. ALTHOUGH THE GEL CONTROL WAS NOT ACTIVATED, WE CANNOT SUSPECT INTER-WELL CARRY-OVER CAUSED BY THE PRESENCE OF SUPERNATANT DROPLETS IN THE IH-CARD REACTION CHAMBERS AS SUCH A HYPOTHESIS WOULDN'T EXPLAIN WHY THE CONTROL WELL SHOWED A POSITIVE REACTION SINCE IT WAS FIRST DISPENSED IN THE SEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21109726 ↗MalfunctionAug 13, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS ON THE ANTI-B WELL OF 0 IH-CARD ABO/D(DVI-)+REV A1,B WHEN USED ON THE IH-1000. THE SAMPLE WAS FALSE POSITIVE 1+ ON ANTI-B WITH IH-CARD ABO/D(DVI-)+REV A1,B WHILE A REPEAT TESTING FOUR DAYS LATER ON THE SAME INSTRUMENT YIELDED AS NEGATIVE ON ANTI-B. THE CUSTOMER ANNOUNCED TO SEND IN THE PATIENT SAMPLES FOR INVESTIGATIONAL TESTING. WHILE OUR QUALITY CONTROL LAB WAITED FOR THESE SAMPLES, THE STAFF TESTED THEIR RETENTION SAMPLES OF THE SUPPOSEDLY DEFECTIVE IH-CARD ABO/D(DVI-)+REV A1,B. THE LAB TESTED ALSO LOT 9346030 WHICH THE CUSTOMER HAD ALSO USED BUT WHICH HAD NOT RESULTED IN A FALSE POSITIVE RESULT. THE CARDS WERE VISUALLY INSPECTED AND SHOWED NO ABNORMALITIES. THERE WAS SUFFICIENT GEL IN ALL WELLS. INHOMOGENEITY OF THE GEL COLUMNS, AS WELL AS GEL SPLASHES IN THE REACTION CHAMBERS WERE NOT DETECTED. THE SEALING OF THE CARDS WERE FLAWLESS. IN ADDITION, 3 CARDS WERE OPENED FROM EACH BATCH OF CARDS. THERE WASN`T ANY ANTISERUM ON THE BOTTOM OF THE FOIL. TO CHECK THE SPECIFICITY OF THE COMPLAINED CARD BATCH ON THE IH-1000, 14 DIFFERENT EDTA DONOR BLOOD ( 6 X A RHD POS, 5 X O RHD POS, 1 X O RHD NEG, 1 X B RHD POS, 1 X B RHD NEG ) WITH KNOWN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19971683 ↗MalfunctionJul 16, 2024False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED DISCREPANT RESULTS IN THE BACK TYPE OF ONE PATIENT SAMPLE WITH THE BLOOD GROUP O RHD POSITIVE WHEN USING IH-CARD ABO/D(DVI-)+REV A1,B. THE SAMPLE WAS TESTED ON FOUR DIFFERENT INSTRUMENTS, TWO IH-1000 AND TWO IH-500. ONE IH-1000 AND ONE IH-500 SHOWED THE CORRECT REVERSE TYPE, THE OTHER INSTRUMENTS NOT. THE CUSTOMER HAS NOT YET PROVIDED ANY DAILY JOURNALS, THEREFORE IT IS NOT CLEAR WHAT THE DISCREPANCY EXACTLY IS. THE CUSTOMER PROVIDED 12 CARDS OF IH-CARD ABO/D(DVI-)+REV A1,B, IH-CELL A1&B AND PATIENT MATERIAL OF SAMPLE NO (B)(6) FOR INVESTIGATIONAL TESTING. THE PROVIDED MATERIAL OF IH-CELL A1&B HAD ALREADY BEEN EXPIRED WHEN THE COMPLAINT WAS FILED, THEREFORE LOT 9422011 WAS USED FOR INVESTIGATIONAL TESTING. AT FIRST, OUR QUALITY CONTROL LABORATORY INVESTIGATED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT IH-CARD ABD/D(DVI-)+REV A1,B. THE VISUAL INSPECTION FOR INTACT SEALING, HOMOGENEOUS GEL, CORRECT FILLING HEIGHT AND SPLASHES IN THE REACTION CHAMBER WAS PASSED WITHOUT ANY IRREGULARITIES. WE DID NOT OBSERVE ANY SPLASHES IN THE REACTION CHAMBER. IN THE 12 CARDS IH-CARD ABO/D(DVI-)+REV A1,B PROVIDED BY THE CUSTOMER, WE IDENTIFIED GEL/ANTISERUM SPLASHES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19748867 ↗MalfunctionJul 4, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF TWO PATIENT SAMPLES WITH THE ANTI-B WELL OF IH-CARD ABO/D(DVI-)+REV. A1, B, LOT: 9336010 ON THEIR IH-1000 INSTRUMENT. A 2+ RESPECTIVELY 3+ REACTION WAS OBTAINED IN THE ANTI-B WELL, BUT REVERSE TYPE INDICATED BLOOD GROUP A. AFTER REPETITION ON THE SAME INSTRUMENT THE CORRECT BLOOD TYPE (BLOOD GROUP A) WAS OBTAINED. THE CUSTOMER PROVIDED US FOUR (4) PATIENT SAMPLES, IH-CELL A1&B LOT 9418011, AND 12 CARDS OF IH-CARD ABD/D(DVI-)+REV A1,B, LOT 9336010 FOR INVESTIGATIONAL TESTING. AT FIRST OUR QUALITY CONTROL LABORATORY CHECKED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT IH-CARD ABD/D(DVI-)+REV A1,B. THE VISUAL INSPECTION FOR INTACT SEALING, HOMOGENEOUS GEL, CORRECT FILLING HEIGHT AND SPLASHES IN THE REACTION CHAMBER WAS PASSED WITHOUT ANY IRREGULARITIES. WE DID NOT OBSERVE ANY SPLASHES IN THE REACTION CHAMBER. THEN QUALITY CONTROL LABORATORY TESTED THE RETENTION SAMPLE WITH DIFFERENT DONOR SAMPLES AND WITH THE PATIENT SAMPLES PROVIDED BY THE CUSTOMER ON THE IH-1000. AS THE REASON FOR THE COMPLAINT WERE FALSE POSITIVE REACTIONS IN THE ANTI-B, THE FOCUS WAS ON THE TESTING OF BLOOD GROUP A AND O RHD-POSITIVE BLOOD SAMPLES. ALL PO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19674777 ↗MalfunctionJul 2, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF ONE PATIENT SAMPLE WITH THE ANTI-B AND CONTROL WELL OF IH-CARD ABO/D(DVI-)+REV. A1, B, LOT: 9336010 ON THEIR IH-1000 INSTRUMENT. A 2+ RESPECTIVELY 3+ REACTION WAS OBTAINED IN THE ANTI-B AND CONTROL WELL FOR A KNOWN BLOOD GROUP A PATIENT. THE REVERSE TYPE INDICATED THE BLOOD GROUP A. AFTER REPETITION ON THE SAME INSTRUMENT THE CORRECT BLOOD TYPE (BLOOD GROUP A) WAS OBTAINED. THE CUSTOMER PROVIDED US FIVE (5) PATIENT SAMPLES, IH-CELL A1&B LOT 9418011, AND 12 CARDS OF IH-CARD ABD/D(DVI-)+REV A1,B, LOT 9336010 FOR INVESTIGATIONAL TESTING. AT FIRST OUR QUALITY CONTROL LABORATORY CHECKED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT IH-CARD ABD/D(DVI-)+REV A1,B. THE VISUAL INSPECTION FOR INTACT SEALING, HOMOGENEOUS GEL, CORRECT FILLING HEIGHT AND SPLASHES IN THE REACTION CHAMBER WAS PASSED WITHOUT ANY IRREGULARITIES. WE DID NOT OBSERVE ANY SPLASHES IN THE REACTION CHAMBER. THEN THE QUALITY CONTROL LABORATORY TESTED THE RETENTION SAMPLE WITH DIFFERENT DONOR SAMPLES AND WITH THE PATIENT SAMPLES PROVIDED BY THE CUSTOMER ON THE IH-1000. AS THE REASON FOR THE COMPLAINT WERE FALSE POSITIVE REACTIONS IN THE ANTI-B AND THE CONTROL WELL,
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19653348 ↗MalfunctionJul 2, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF TWO PATIENT SAMPLES WITH THE ANTI-B WELL OF IH-CARD ABO/D(DVI-)+REV. A1, B, LOT: 9336010 ON THEIR IH-1000 INSTRUMENT. A 2+ RESPECTIVELY 3+ REACTION WAS OBTAINED IN THE ANTI-B WELL, BUT REVERSE TYPE INDICATED BLOOD GROUP A. AFTER REPETITION ON THE SAME INSTRUMENT THE CORRECT BLOOD TYPE (BLOOD GROUP A) WAS OBTAINED. THE CUSTOMER PROVIDED US FOUR (4) PATIENT SAMPLES, IH-CELL A1&B LOT 9418011, AND 12 CARDS OF IH-CARD ABD/D(DVI-)+REV A1,B, LOT 9336010 FOR INVESTIGATIONAL TESTING. AT FIRST OUR QUALITY CONTROL LABORATORY CHECKED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT IH-CARD ABD/D(DVI-)+REV A1,B. THE VISUAL INSPECTION FOR INTACT SEALING, HOMOGENEOUS GEL, CORRECT FILLING HEIGHT AND SPLASHES IN THE REACTION CHAMBER WAS PASSED WITHOUT ANY IRREGULARITIES. WE DID NOT OBSERVE ANY SPLASHES IN THE REACTION CHAMBER. THEN QUALITY CONTROL LABORATORY TESTED THE RETENTION SAMPLE WITH DIFFERENT DONOR SAMPLES AND WITH THE PATIENT SAMPLES PROVIDED BY THE CUSTOMER ON THE IH-1000. AS THE REASON FOR THE COMPLAINT WERE FALSE POSITIVE REACTIONS IN THE ANTI-B, THE FOCUS WAS ON THE TESTING OF BLOOD GROUP A AND O RHD-POSITIVE BLOOD SAMPLES. ALL PO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19653278 ↗MalfunctionJun 26, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF PATIENT AND QC SAMPLES WITH THE ANTI-B AND THE CONTROL WELL OF IH-CARD ABO/D(DVI-)+REV. A1, B. DURING MANUAL TESTING IT WAS ALSO NOTICED THAT SPLASHES OF ANTI-A WERE FOUND BESIDES THE ANTI-A COLUMNS. THE CUSTOMER SAID SHE RECEIVED ALL THE BOXES IN ONE SHIPMENT. OUR QUALITY CONTROL LABORATORY INVESTIGATED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT. FIRST, OUR QUALITY CONTROL LABORATORY VISUALLY INSPECTED THE RETENTION SAMPLE FOR INTACT SEALING, HOMOGENEOUS GEL, CORRECT FILLING HEIGHT AND SPLASHES IN THE REACTION CHAMBER. ALL ACCEPTANCE CRITERIA WERE MET. WE DID NOT OBSERVE ANY SPLASHES IN THE REACTION CHAMBER. THEN THE QUALITY CONTROL LABORATORY TESTED THE RETENTION SAMPLE WITH DIFFERENT DONOR SAMPLES IN THE MANUAL TEST METHOD. AS THE REASON FOR THE COMPLAINT WERE FALSE POSITIVE REACTIONS IN THE ANTI-B, THE FOCUS WAS ON THE TESTING OF BLOOD GROUP A AND O SAMPLES. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION IN THE ANTI-B OR CONTROL WELL WITH THE RETENTION SAMPLE. THE CUSTOMER SENT IN 75 IH-CARD ABO/D(DVI-)+REV A1,B OF THE AFFECTED LOT FOR INVESTIGATIONAL TESTING. WE IDEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19611965 ↗MalfunctionMar 15, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE REACTION OF A PATIENT SAMPLE IN THE ANTI-B OF THE IH-CARD ABO/D(DVI-) + REV. A1, B ON IH-1000. THE PATIENT WAS HISTORICALLY TYPED AS BLOOD GROUP A. THE PATIENT SAMPLE ALSO SHOWED A POSITIVE REACTION IN THE CONTROL WELL OF THE IH-CARD ABO/D(DVI-)+REV. A1, B ON IH-1000. THE PATIENT SAMPLE SHOWED CORRECT RESULTS IN THE REVERSE TYPING ON THE IH-1000: THE REACTION WITH IH-CELL A1 WAS CORRECTLY NEGATIVE, WHILE THE REACTION WITH IH-CELL B WAS CORRECTLY POSITIVE. DUE TO THE POSITIVE REACTION IN THE CONTROL WELL AND THE DISCREPANCY BETWEEN FORWARD AND REVERSE TYPING NO BLOOD GROUP TEST RESULT WAS RELEASED BY THE IH-1000. FURTHERMORE, THE CUSTOMER STATED THAT THE PATIENT SAMPLE SHOWED A POSITIVE RESULT IN THE DAT (DIRECT ANTIGLOBULIN TEST). THE CUSTOMER DID NEITHER RETURN A PRODUCT SAMPLE FOR INVESTIGATIONAL TESTING NOR THE PATIENT SAMPLE THAT HAD CAUSED THE UNEXPECTED POSITIVE REACTIONS. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT. FIRST, OUR QUALITY CONTROL LABORATORY VISUALLY CHECKED THE RETENTION SAMPLE FOR INTACT SEALING, CORRECT POSITION OF THE ALUMINIUM FOIL FLAPS, HOMOGENEOUS GEL, CORRE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18911925 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.