INCEPTA ICDRecalls & reported problems across all models
Boston Scientific Corporation · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1,746
- Deaths reported
- 3
- Injuries reported
- 1,322
- Recalls
- 0
We found 4 models of INCEPTA ICD by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 1,746 adverse event reports in the FDA's MAUDE database, received between Jul 8, 2013 and Sep 30, 2026 — classified as 3 deaths, 1,322 injuries, 421 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (894 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported INCEPTA ICD problems
Device problems
- Premature Discharge of Battery894
- Device Displays Incorrect Message435
- High impedance248
- Noise, Audible176
- Adverse Event Without Identified Device or Use Problem167
- Inappropriate/Inadequate Shock/Stimulation141
- Pacing Problem123
- Over-Sensing111
Patient problems
- No Clinical Signs, Symptoms or Conditions647
- No Known Impact Or Consequence To Patient497
- No Code Available277
- Unspecified Infection141
- No Consequences Or Impact To Patient138
- Electric Shock93
- Shock from Patient Lead(s)52
- Sepsis50
INCEPTA ICD models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- INCEPTA™ ICD E160Implantable pulse generator, pacemaker (non-CRT)2 deaths601 reports
- INCEPTA™ ICD E162Implantable pulse generator, pacemaker (non-CRT)570 reports
- INCEPTA™ ICD E163Implantable pulse generator, pacemaker (non-CRT)1 death339 reports
- INCEPTA™ ICD E161Implantable pulse generator, pacemaker (non-CRT)236 reports
Recent INCEPTA ICD adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Inappropriate/Inadequate Shock/Stimulation
Patient problems: Shock from Patient Lead(s)
MAUDE report 50955017 ↗InjurySep 16, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation
It was reported that the health care professional (HCP) requested a review of reports for this implantable cardioverter defibrillator (ICD). The HCP suspected this ICD delivered anti-tachycardia pacing (ATP) and shocks for what appeared to be supraventricular tachycardia (SVT) with rapid ventricular response (RVR). Technical Services (TS) reviewed the data and noted the rhythm did appear to be SVT with RVR and recommended the patient follow up with cardiology. No additional adverse patient effects were reported. This ICD remains in service.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50786653 ↗MalfunctionSep 16, 2026Signal Artifact/Noise; Over-Sensing
It was reported that this implantable cardioverter defibrillator (ICD) had a magnet applied during a procedure. Following the procedure, Technical Services (TS) was consulted to evaluate device performance. TS reviewed the available data and confirmed that the device appeared to be functioning as programmed and expected. TS noted the presence of stored noise events that correlated with magnet application; however, no device-related issues or complications were identified. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50783366 ↗MalfunctionSep 4, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602345 ↗InjurySep 4, 2026High impedance; Inappropriate/Inadequate Shock/Stimulation
It was reported that the patient with this implantable cardioverter defibrillator (ICD) visited the emergency room (ER) for evaluation after receiving shocks. Technical Services (TS) was called for consult review of device data. The device was interrogated however, due to device being in End of Life (EOL) battery status, the episodes were unable to be reviewed to determine shocks were appropriate or not because device does not store episodes when in EOL status. TS reviewed the stored daily threshold and impedance measurements trend reports which showed high out of range shock impedances measurements on the right ventricular (RV) lead that appeared to have been gradually increasing since 2018. Subsequently, later on, a generator replacement procedure was performed. This ICD was explanted due to normal battery depletion (NBD) and replaced with a new device successfully. The RV lead remains in service. No additional adverse patient effects were reported.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50601337 ↗InjurySep 3, 2026Inappropriate/Inadequate Shock/Stimulation; Data Problem
It was reported that this implantable cardioverter defibrillator (ICD) was at End of Life (EOL) and the patient was shocked five times. It appeared that the rate was controlled after the shocks were delivered. There was supposedly a note that device tachy mode was off and it was not determined whether the shocks were appropriate. No additional adverse patient effects were reported. This ICD remains in service.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50588062 ↗InjuryAug 20, 2026Pacing Problem; Defibrillation/Stimulation Problem; Inappropriate/Inadequate Shock/Stimulation; Failure to Read Input Signal
It was reported that this implantable cardioverter defibrillator (ICD) delivered five shocks and several rounds of anti-tachycardia pacing (ATP) for what appeared to be atrial fibrillation (AF) with rapid ventricular response (RVR). Technical Services (TS) was contacted and after reviewing the data noted there was an irregular morphology present and that it appeared to become faster after a shock. This ICD remains in service. No additional adverse patient effects were reported.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50360078 ↗MalfunctionAug 19, 2026Premature Discharge of Battery; Delayed Charge Time; Battery Problem
It was reported that this implantable cardioverter defibrillator (ICD) was suspected of premature battery depletion (PBD) as the battery has depleted from 7 months down to reaching Elective Replacement Indicator state. The Boston Scientific Technical Services consultant evaluated the transmission data which demonstrated a charge time of 15.7 seconds, representing a secondary criterion for recommending a device replacement while the system continues to operate normally. Subsequently, this ICD was explanted and replaced. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50313817 ↗MalfunctionAug 18, 2026Signal Artifact/Noise; High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291255 ↗MalfunctionAug 17, 2026Low impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50287267 ↗InjuryAug 14, 2026Pacing Problem; Defibrillation/Stimulation Problem; Inappropriate/Inadequate Shock/Stimulation
It was reported that this implantable cardioverter defibrillator (ICD) delivered inappropriate shocks and anti-tachycardia pacing (ATP) that induced ventricular tachycardia. Additionally, inappropriate continued to be delivered resulting in the rhythm accelerating from ventricular tachycardia to ventricular fibrillation. Technical Services (TS) was consulted and confirmed four bursts of ATP and two shocks; all being deemed inappropriate. TS additionally confirmed that the third burst of ATP accelerating the rhythm as previously mentioned. TS recommended increasing the ventricular tachycardia rate cutoff, rate controlling the patient, or turning on additional features for the ventricular tachycardia zone. No additional adverse patient effects were reported. This ICD remains in service.
Patient problems: Tachycardia; Ventricular Fibrillation
MAUDE report 50254250 ↗InjuryAug 5, 2026Insufficient Device Problem Information
It was reported that the battery of this implantable cardioverter defibrillator (ICD) reached a completely depleted state beyond End of Life (EOL). Technical Services (TS) assessed the device and indicated that sinus bradycardia and ejection fraction (EF) of 20 percent were listed as the implant indications. To date, this ICD remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50150340 ↗InjuryJul 30, 2026Inappropriate/Inadequate Shock/Stimulation
It was reported that the patient with this implantable cardioverter defibrillator (ICD) received two inappropriate shocks for what appeared to be atrial fibrillation with rapid ventricular response. The patient remained under continued monitoring. This device remains in service.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50091378 ↗MalfunctionJul 27, 2026Communication or Transmission Problem; Interrogation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50036230 ↗MalfunctionJul 15, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25829094 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.