INVIVERecalls & reported problems across all models
Boston Scientific Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 680
- Deaths reported
- 1
- Injuries reported
- 613
- Recalls
- 0
We found 2 models of INVIVE by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 680 adverse event reports in the FDA's MAUDE database, received between Jan 8, 2015 and Oct 2, 2025 — classified as 1 death, 613 injuries, 66 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (481 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported INVIVE problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings481
- Premature Discharge of Battery161
- Pacing Problem101
- Over-Sensing74
- Inappropriate or Unexpected Reset48
- Defective Device44
- Interrogation Problem30
- Signal Artifact/Noise29
Patient problems
- No Clinical Signs, Symptoms or Conditions401
- Syncope/Fainting65
- No Code Available64
- Device Overstimulation of Tissue59
- Insufficient Information40
- Asystole32
- No Known Impact Or Consequence To Patient27
- Dizziness25
INVIVE models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- INVIVE™ V173Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death670 reports
- INVIVE™ V172Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)10 reports
Recent INVIVE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryOct 2, 2025Incorrect, Inadequate or Imprecise Result or Readings; Data Problem; Incorrect, Inadequate or Imprecise Result or Readings; Data Problem
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXHIBITED AN ALERT MESSAGE INDICATING THAT BATTERY REPLACEMENT INDICATOR HAD BEEN REACHED ON 17JUL2014. TECHNICAL SERVICES DISCUSSED THAT THIS WAS ASSOCIATED WITH UNEXPECTED HARDWARE RESETS AND THE DEVICE DATA WAS CONSISTENT WITH A HIGH IMPEDANCE BATTERY AND RECOMMENDED DEVICE REPLACEMENT. THIS DEVICE WAS SUBSEQUENTLY EXPLANTED AND WAS NOT RETURNED. OTHER THAN THE SURGICAL PROCEDURE, NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23200736 ↗InjurySep 3, 2025Signal Artifact/Noise; Failure to Capture; Inaccurate Synchronization; Signal Artifact/Noise; Failure to Capture; Inaccurate Synchronization
IT WAS REPORTED THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WENT INTO VENTRICULAR FIBRILLATION (VF) DURING A GENERATION CHANGE PROCEDURE. TECHNICAL SERVICES (TS) REVIEWED DEVICE DATA AND NOTED A PAUSE IN HEART RATE PRIOR TO ELECTROCAUTERY NOISE ON BOTH THE RIGHT ATRIAL AND RIGHT VENTRICULAR CHANNELS. TS NOTED THAT THE PATIENT'S RHYTHM WAS LIKELY ALREADY IN AN ARRHYTHMIA, HOWEVER RECOMMENDED PHYSICIAN REVIEW TO CONFIRM, BUT THE CAUSE WAS NOT DETERMINED. THE PATIENT WAS CONTINUOUSLY MONITORED AT THE HOSPITAL TO RESOLVE. THIS DEVICE WAS SUBSEQUENTLY EXPLANTED AND REPLACED DUE TO NORMAL BATTERY DEPLETION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Bradycardia; Ventricular Fibrillation
MAUDE report 22954379 ↗InjuryAug 12, 2025Pacing Problem; Incorrect, Inadequate or Imprecise Result or Readings; Pacing Problem; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS PACEMAKER REVERTED TO SAFETY MODE AND THE BRADY RATE WAS NOT AS EXPECTED. THIS PACEMAKER WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS PRODUCT HAS NOT BEEN RECEIVED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784027 ↗InjuryMay 1, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) ENTERED SAFETY MODE. THIS CRT-P WAS EXPLANTED AND REPLACED WITH A NEW CRT-P. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21942351 ↗InjuryApr 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) FELT THUMPING ON THEIR CHEST. TECHNICAL SERVICES (TS) WAS CONSULTED AND IT WAS DISCOVERED THIS CRT-P ENTERED SAFETY MODE, WITH THE SAFETY MODES UNIPOLAR PACING PROBABLY CAUSING THE THUMPING FELT BY THE PATIENT. THIS CRT-P WAS EXPLANTED. A NEW DEVICE WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Chest Pain; Unspecified Heart Problem
MAUDE report 21891680 ↗InjuryApr 14, 2025Pacing Problem; Incorrect, Inadequate or Imprecise Result or Readings; Premature Discharge of Battery; Pacing Problem; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS EXPLANTED DUE TO IT SWITCHED TO SAFETY MODE. A NEW DEVICE WAS SUCCESSFULLY IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Device Overstimulation of Tissue; Fever; No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 21836656 ↗InjuryMar 10, 2025Defective Device; Defective Device
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS EXPLANTED DUE TO A PRODUCT PERFORMANCE ISSUE. A NEW DEVICE WAS SUCCESSFULLY EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21560706 ↗InjuryFeb 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) ENTERED SAFETY MODE. THE CRT-P WAS EXPLANTED, AND ANOTHER DEVICE WAS SUCCESSFULLY PLACED. THIS DEVICE WAS RECEIVED FOR ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Inappropriate Device Stimulation of Tissue; No Clinical Signs, Symptoms or Conditions
MAUDE report 21428990 ↗InjuryJan 20, 2025Over-Sensing; Incorrect, Inadequate or Imprecise Result or Readings; Interrogation Problem; Over-Sensing; Incorrect, Inadequate or Imprecise Result or Readings; Interrogation Problem
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT P) WAS DIFFICULT TO INTERROGATE AND THE DEVICE WAS FOUND TO BE IN SAFETY MODE. THE DEVICE SWITCHED TO UNIPOLAR AND CAUSED POCKET STIMULATION. THIS CRTP WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Device Overstimulation of Tissue
MAUDE report 21193488 ↗InjuryDec 17, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) REVERTED TO SAFETY MODE. THIS DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20957145 ↗InjuryDec 10, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) VISITED THE EMERGENCY ROOM (ER) AFTER A SYNCOPE EPISODE. THE DEVICE WAS INTERROGATED AND FOUND TO BE IN SAFETY MODE. AN EMERGENT DEVICE REPLACEMENT PROCEDURE WAS PERFORMED, THE CRT-P WAS EXPLANTED AND REPLACED WITH A NEW DEVICE SUCCESSFULLY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Syncope/Fainting
MAUDE report 20895653 ↗InjuryDec 6, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS EXPLANTED AND REPLACED DUE TO ENTERING SAFETY MODE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. PRODUCT RETURN WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20872512 ↗MalfunctionNov 11, 2024High impedance; Over-Sensing; High Capture Threshold; High impedance; Over-Sensing; High Capture Threshold
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXHIBITED FAR-FIELD OVERSENSING IN THE RIGHT ATRIAL (RA) CHANNEL. WHICH RESULTED IN INAPPROPRIATE ATRIAL TACHY RESPONSE (ATR) EPISODES TO BE STORED. REPROGRAMMING OPTIONS AND SENSITIVITY ADJUSTMENTS WERE PERFORMED. SUBSEQUENTLY, THE LEFT VENTRICULAR (LV) ACTIVITY STARTED TO BE SENSED ON THE RA CHANNEL. TS RECOMMENDED TURNING OFF THE LV SENSITIVITY, BUT IT REMAINED AT THE PHYSICIAN DISCRETION. IN ADDITION, THIS CRT-P SYSTEM PRESENTED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS WITH HIGH PACING THRESHOLDS. THIS CRT-P REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20654091 ↗MalfunctionOct 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS FOUND TO BE IN SAFETY MODE. THE PATIENT EXPERIENCED PHRENIC NERVE STIMULATION. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) WAS CONSULTED AND DEVICE REPLACEMENT WAS RECOMMENDED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AT THIS TIME, THIS DEVICE REMAINS IN SERVICE.
Patient problems: Device Overstimulation of Tissue
MAUDE report 20409282 ↗InjuryOct 2, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THIS PACEMAKER RECORDED A 1010 ERROR CODE WHEN INTERROGATING THE DEVICE. THE PATIENT MENTIONED HAVING FELL AND THEN THE PACEMAKER WAS EXPLANTED AND RETURNED FOR ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Fall
MAUDE report 20350175 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page