Lotus Introducer SetRecalls & reported problems across all models
Creganna Unlimited Company · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 32
- Deaths reported
- 3
- Injuries reported
- 29
- Recalls
- 0
We found 2 models of Lotus Introducer Set by Creganna Unlimited Company in the FDA's device registry (GUDID). Together they are named in 32 adverse event reports in the FDA's MAUDE database, received between Apr 9, 2018 and Sep 3, 2020 — classified as 3 deaths, 29 injuries.
The most frequently reported specific device problem is “Difficult to Insert” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported Lotus Introducer Set problems
Device problems
- Adverse Event Without Identified Device or Use Problem28
- Difficult to Insert4
- Difficult to Remove3
- Difficult to Advance2
- Device-Device Incompatibility1
- Failure to Advance1
- Leak/Splash1
- Physical Resistance/Sticking1
Patient problems
- Vascular Dissection16
- Perforation of Vessels8
- Death5
- Hemorrhage/Blood Loss/Bleeding5
- Blood Loss3
- Hematoma2
- Exsanguination1
- Hemostasis1
Lotus Introducer Set models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Lotus™ Introducer Set H749NTR200Introducer, Catheter1 death24 reports
- Lotus™ Introducer Set H749NTR180Introducer, Catheter2 deaths8 reports
Recent Lotus Introducer Set adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: INITIALLY INSERTED INTO PATIENTS FEMORAL ARTERY, FOLLOWED BY A 23MM LOTUS EDGE VALVE. PHYSICIANS CONVEYED SIGNIFICANT RESISTANCE IN DELIVERING THE VALVE THRU THE SHEATH, AND AFTER CONSIDERABLE EFFORT THE DECISION WAS MADE TO REMOVE THE ISLEEVE SHEATH AND CONVERT TO THE LIS SHEATH INSTEAD. THIS PROVED TO WORK WELL, AND ALLOWED THE 23MM VALVE TO PASS THRU UNINHIBITED. FOLLOWING THE SUCCESSFUL DEPLOYMENT OF THE VALVE AND REMOVAL OF THE SHEATH, IT WAS RECOGNIZED THAT THE EXTERNAL ILIAC ARTERY HAS A SLIGHT DISSECTION. PHYSICIAN DECIDED TO BALLOON AND THEN STENT THE DISSECTION. SUCCESSFUL RESULT AND THE PATIENT WAS DISCHARGED THE FOLLOWING DAY. EVENT DATE: (B)(6) 2020
Patient problems: Vascular Dissection
MAUDE report 10486769 ↗InjuryAug 26, 2020Difficult to Insert; Adverse Event Without Identified Device or Use Problem; Difficult to Insert; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; EVENT DESCRIPTION: PER CNF, IT WAS REPORTED THAT: IT WAS PLANNED TO PERFORM TAVR ON THE PATIENT FOR SEVERE AORTIC VALVE STENOSIS. CALCIFICATION FOLLOWING LVOT WAS OBSERVED, SO TAVR BY LOTUS EDGE WAS JUDGED TO BE APPROPRIATE. ANNULUS WAS WITHIN THE TREATMENT RANGE OF 23 MM VALVE, BUT THE POSITION OF LVOT -4 MM WAS SLIGHTLY SMALLER THAN THE 23 MM RANGE, BUT IT WAS JUDGED THAT IT COULD BE TREATED BY PLACING IT HIGHER TO MINIMIZE PVL WHILE TAKING PMI RISK. REGARDING ACCESS, THERE WAS POSSIBILITY THAT IT MIGHT BE DIFFICULT TO INSERT THE LIS BECAUSE THERE WAS SEVERE CALCIFICATION IN BOTH SIDES EIA AND CIA. THE PHYSICIAN INSTRUCTED TO START PREPARING THE LOTUS VALVE AFTER THE LIS HAD PASSED THROUGH. DURING THE PROCEDURE, THE FIRST SURGEON FELT RESISTANCE WHEN TRYING TO PASS THE SHEATH TIP TO THE RIGHT EIA, SO THE PROCEDURE WAS TEMPORARILY INTERRUPTED AND DISCUSSIONS WERE HELD WITH OTHER PHYSICIANS AS TO WHETHER OR NOT TO CONTINUE. AFTER THAT, WHEN THE SECOND SURGEON TRIED TO INSERT THE LIS AGAIN, THE SHEATH PASSED THROUGH THE EIA AND THE TIP COULD BE ADVANCED INTO THE ABDOMINAL AORTA, EVEN THOUGH SLIGHT RESISTANCE
Patient problems: Blood Loss; Hemostasis; Perforation of Vessels
MAUDE report 10453407 ↗InjuryAug 20, 2020Difficult to Insert; Adverse Event Without Identified Device or Use Problem; Difficult to Insert; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT (B)(6) MEDICAL; EVENT DESCRIPTION: PRODUCT NAME: LOTUS. REFERENCE: LIS-S, LOT NUMBER: 0000007302, DATE OF KNOWLEDGE OF THE CLAIM BY COMPLAINANT: 7/20/20. EVENT INFORMATION: AT THE TIME OF IMPLANTATION, THE RIGHT FEMORAL ACCESS WAS CONSIDERED THE MOST SUITABLE. ONCE THE DISSECTION WAS DONE, AN ATTEMPT WAS MADE TO UPLOAD THE LOTUS INTRODUCER BUT WAS UNSUCCESSFUL. THE ARTERY HAD TO BE REPAIRED BECAUSE WHEN THE INTRODUCER WAS REMOVED, IT WAS DAMAGED. IT IS DECIDED TO GO ON THE LEFT SIDE. AFTER DISSECTION AN ATTEMPT IS MADE TO CLIMB THE LOTUS. FAILING TO DO SO, SMALLER DILATORS ARE USED TO MAKE WAY. A 14 FR AND 18 FR INTRODUCER FROM MEDTRONIC IS USED. WHEN THE LOTUS IS STILL UNABLE TO ASCEND, AN ATP IS PERFORMED WITH 7 AND 9 MM BALLS. WHEN CONTINUING WITHOUT THE POSSIBILITY OF ACCESSING, A STENT WAS PLACED WITH THE INTENTION OF MAINTAINING THE DIAMETER OF THE ARTERY THAT HAD BEEN ABLE TO EXPAND, WHEN ATTEMPTING TO RAISE THE LOTUS, THE ARTERY RUPTURED AND CONSEQUENTLY 2 LINED STENTS. ONCE THE ARTERY IS REPAIRED, THE PROCEDURE IS TERMINATED. IT IS IMPORTANT TO NOTE THAT FROM THE TIME THE OFFICIAL REPORT WAS RECEIVED (WHICH RECOMMEN
Patient problems: Perforation of Vessels
MAUDE report 10430160 ↗InjuryJul 23, 2020Difficult to Insert; Adverse Event Without Identified Device or Use Problem; Difficult to Insert; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: LEFT FEMORAL ACCESS WAS ATTEMPTED VIA STANDARD TECHNIQUE. PHYSICIAN WAS UNABLE TO ADVANCE LOTUS INTRODUCER SHEATH MORE THAN A FEW CENTIMETERS. ATTEMPTED TO DILATE UP THE ARTERIOTOMY USING MULTIPLE DILATORS THAT GRADUALLY INCREASED IN SIZE. WE WERE STILL UNABLE TO ADVANCE THE LOTUS SHEATH. ATTEMPTED A 24F GORE SHEATH AND WAS UNABLE TO INTRODUCE THAT AS WELL. MOVED TO THE RIGHT FEMORAL AND ATTEMPTED ACCESS THERE. JUST LIKE THE LEFT, UNABLE TO OBTAIN ACCESS DESPITE SAME ATTEMPTS AT UPSIZING DILATOR AND ATTEMPTING GORE SHEATH. IT WAS DECIDED TO ABORT THE PROCEDURE ALTOGETHER DUE TO ACCESS ISSUES. PATIENT WAS STABLE THROUGHOUT AND NO ADVERSE EVENTS WERE ENCOUNTERED. THE PATIENT DEVELOPED A LARGE HEMATOMA IN THE RIGHT GROIN POST PROCEDURE. VASCULAR SURGERY ASSESSED THE PATIENT AND IT WAS DETERMINED THE PATIENT A RT. FEMORAL ARTERIAL? TEAR?. REPAIRED BY VASCULAR SURGEON. UNABLE TO DETERMINE WHICH SHEATH (LIS OR GORE) THAT CAUSED THE TEAR. PATIENT REPORTED STABLE POST OP. EVENT DATE: (B)(6) 2020.
Patient problems: Hematoma; Perforation of Vessels
MAUDE report 10314018 ↗InjuryJun 18, 2020Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Difficult to Remove; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT (B)(6); EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: SHEATH WAS IN LEFT GROIN, ON CT, SIZE SHOWED 5.6/5.7; SUGGESTED RT GROIN ACCESS BUT WENT WITH LEFT WITH IDEA OF NO TORTUOUS NOTED; THEY SLOWLY DILATED THE VESSEL BEFORE ADVANCING LIS SHEATH TO LT GROIN, WAS NOT MENTION OF DIFFICULTIES ADVANCING THE SHEATH. AFTER THE VALVE DEPLOYED, DELIVERY SYSTEM CAME OUT WITHOUT ANY ISSUES. TRYING TO REMOVE THE LIS SHEATH, UNABLE TO; AFTER MUCH ATTEMPT, PART OF THE LIS SHEATH BROKE OFF, HAD TO CONVERT TO CUTDOWN AND REMOVED THE REMAKING LIS SHEATH AND PT RECEIVED BYPASS GRAFT FROM EXTERNAL ILIAC TO COMMON FEMORAL. EVENT DATE: (B)(6) 2020.
Patient problems: No Information
MAUDE report 10169363 ↗InjuryJun 3, 2020Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Difficult to Remove; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; EVENT DESCRIPTION: PER CNF, IT WAS REPORTED THAT: (B)(6) 2020 AFTER SUCCESSFUL TAVI PROCEDURE, RT EIA TEAR WAS CONFIRMED DURING DEVICE SHEATH REMOVAL. RT EIA OCCLUSION AND STENTING WERE PERFORMED BY PTA, TEARS WERE HEMOSTATIC, BUT PUNCTURE SITE BLEEDING OF THE RT FEMORAL HEAD INFERIOR BORDER CONTINUED, AND RT IIA & CFA BLEEDING CONTROL SURGERY WAS PERFORMED. AFTER THAT, THE BP DROP SHOWED RBC 1UNIT TRANSFUSION AND NE SUPPORT, AND THE CONDITION IMPROVED. AFTER TREATMENT, THE CONDITION IMPROVED AND SHE WAS DISCHARGED ON (B)(6). THE RT EIA TEAR WAS DEFINITELY RELATED TO STUDY PROCEDURE AND REPORTED BY SAE. EVENT DATE: (B)(6) 2020. ADDITIONAL INFORMATION RECEIVED 07-MAY-2020 WHAT BSC DEVICE CAUSED THE RT EIA TEAR? -> LOTUS INTRODUCER SET LARGE. WAS ANY RESISTANCE ENCOUNTERED WHEN INSERTING OR WITHDRAWING THE SHEATH? -> YES, WHEN WITHDRAWING THE SHEATH. IF YES, PLEASE EXPLAIN: -> THE OPERATOR SLOWLY AND CAREFULLY TRIED TO REMOVED THE SHEATH BUT HE FELT RESISTANCE AS HE WAS DOING SO AND REALIZED THERE WAS A TEAR. WAS ANY RESISTANCE ENCOUNTERED WHEN INSERTING THE LOTUS EDGE VALVE SYSTEM INTO THE SHEATH? -> NO. WAS
Patient problems: Low Blood Pressure/ Hypotension; Vascular Dissection
MAUDE report 10114533 ↗InjuryMay 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; EVENT DESCRIPTION: PER CNF, IT WAS REPORTED THAT: QUALITY EVENT YES PATIENT OUTCOME SUCCESSFUL MAJOR VASCULAR COMPLICATIONS YES PVL GRADE NONE QUALITY EVENT DESCRIPTION: ILIAC PERFORATION REQUIRING SURGICAL REPAIR. SHEATH WAS ADVANCED WITHOUT ISSUE. 25 LOTUS WAS IMPLANTED SUCCESSFULLY. WHEN TAKING OUT THE LIS, THE ILIAC VESSEL TORE REQUIRING SURGICAL REPAIR. ADDITIONAL INFORMATION RECEIVED. THE DOCTOR SAID IT WAS A LITTLE TIGHT BUT VERY SMOOTH ADVANCING THE SHEATH. THERE WAS RESISTANCE WITHDRAWING THE SHEATH. THE VALVE DELIVERY SYSTEM WAS REMOVED FROM THE PATIENT WHEN THE ILIAC TORN. THE CAUSED THE PERFORATION WAS THE VESSEL ADHERING TO THE SHEATH. THE PHYSICIAN STATES THE PATIENT IS DOING WELL. AS PER THE CNF THE DEVICE WAS DISCARDED POST PROCEDURE AND SO WILL NOT BE RETURNING. THE INCIDENT HAD NOTHING TO DO WITH THE VALVE. THE PHYSICIAN REITERATED THAT IT WAS A LITTLE TIGHT PUTTING IN THE SHEATH BUT WENT SMOOTHLY. THE EVENT DESCRIPTION STATES IT WAS A '6.0MM VESSEL WITH NO CALCIUM' THERE WAS NO TORTUOSITY OR CALCIFICATION. AN ILIAC CONDUIT WAS PLACED.
Patient problems: Perforation of Vessels
MAUDE report 10023943 ↗DeathApr 2, 2020Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Difficult to Remove; Adverse Event Without Identified Device or Use Problem
EVENT DESCRIPTION: PER CNF, IT WAS REPORTED THAT: LOTUS 23MM VALVE SUCCESSFULLY IMPLANTED WITH NO PVL. WHEN THE PHYSICIAN ATTEMPTED TO TAKE OUT THE LIS, IT WAS FROZEN IN THE VESSEL. VASOSPASM SUSPECTED. SHEATH WAS PARTIALLY REMOVED BUT EVIDENCE OF VESSEL DISRUPTION WAS APPARENT. VIABAHN COVERED STENTS WERE PLACED IN THE EXTERNAL AND COMMON ILIAC ARTERIES. PATIENT CONTINUED TO BLEED. PATIENT MOVED TO THE OR FOR SURGICAL REPAIR OF THE ILIACS WHERE PATIENT EXPIRED. EVENT DATE: (B)(6) 2020. ADDITIONAL INFORMATION RECEIVED FROM DISTRIBUTOR ON 12TH OF MARCH 2020: OPERATOR 2 WAS RUSSELL STAHL. LOTUS INTRODUCER LOT NUMBER: 0000012694. WAS THERE DIFFICULTY ADVANCING THE LOTUS EDGE? NO. WAS THERE DIFFICULTY REMOVING THE LOTUS EDGE? NO. WAS THE LOTUS EDGE ABLE TO BE REMOVED THROUGH THE LOTUS INTRODUCER? YES. WAS LOTUS EDGE REMOVED BEFORE THE VASOSPASM? YES. WHERE WAS THE SPASM? LEFT ILIACS. CAN YOU FURTHER DESCRIBE THE ''VESSEL DISRUPTION'? (PERFORATION, DISSECTION): APPEARED TO BE EVULSION OF THE ILIACS. WAS THE 'VESSEL DISRUPTION' A RESULT OF THE VASOSPASM AROUND THE LOTUS INTRODUCER SHEATH? WE BELIEVE, A PHYSICIAN WILL NEED TO COMMENT ON THAT. DID THE LOTUS EDGE VALVE CAUSE OR CONTRIBUTE T
Patient problems: Death; Hemorrhage/Blood Loss/Bleeding; Perforation
MAUDE report 9917757 ↗InjuryMar 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: LATER IT WAS DECIDED TO PERFORM INTERVENTION DUE TO VASCULAR INJURY TO RIGHT ILIAC AND/OR COMMON FEMORAL ARTERY. AS PER SOURCE THERE WAS POSSIBLE VASCULAR COMPLICATION DUE TO TAVR ACCESS SHEATH. WARFARIN WAS RECOMMENDED TO DISCONTINUE. PER EDC, SURGERY WAS PERFORMED TO TREAT THE EVENT (RIGHT COMMON FEMORAL ENDARTERECTOMY WITH PATCH ANGIOPLASTY. STENT PLACEMENT TO COVER DISSECTION OF RIGHT COMMON FEMORAL AND EXTERNAL ILIAC ARTERIES). ON (B)(6) 2020, THE EVENT WAS CONSIDERED TO BE RECOVERED. PER CTSENR, IT WAS REPORTED THAT: REPRISE IV 0784R42002 (ECG "SINUS PAUSES" ; NEW LEFT BUNDLE BRANCH BLOCK) EVENT DESCRIPTION: NEW LEFT BUNDLE BRANCH BLOCK (002- ADE)/ ECG "SINUS PAUSES" (003- NEW PERMANENT PACEMAKER IMPLANTATION RESULTING FROM NEW OR WORSENED CONDUCTION DISTURBANCES): ON (B)(6) 2020, SAME DAY OF THE INDEX PROCEDURE, POST TAVR, THE SUBJECT DEVELOPED NEW LEFT BUNDLE BRANCH BLOCK. OF NOTE, BALLOON VALVULOPLASTY WAS NOT PERFORMED. PER EDC NO DIAGNOSTIC TESTS WERE PERFORMED AND NO ACTION WAS TAKEN TO TREAT THE EVENT. ON (B)(6) 2020, 1 DAY POST INDEX PROCEDURE, THE SUBJECT DEVELOPED ECG "SINUS PAUSES". THE EVENT WAS INITIALLY TREATED
Patient problems: Ischemia; Vascular Dissection
MAUDE report 9903241 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EVENT DESCRIPTION: IT WAS REPORTED THAT: TRANSFEMORAL ACCESS WITHOUT PROBLEMS WITH LIS, INSERTING BALLOON WITHOUT PROBLEMS, INSERTING LOTUS EDGE WITHOUT PROBLEMS, REMOVING LOTUS EDGE WITHOUT PROBLEMS, ANGIO ILLIACA COMMUNIS AFTER REMOVE LIS, DISSECTION ILLIACA COMMUNIS RIGHT, SUPPLYING THE DISSECTION WITH STENT, THEN CONTROL ANGIO, NO MORE DISSECTION SMALL VIDEO RECORDED WAS BAV PERFORMED? YES ASSOCIATED WITH LABELED USE? YES PATIENT STATUS POST PROCEDURE/EVENT: FINE. EVENT DATE: (B)(6) 2020.
Patient problems: Vascular Dissection
MAUDE report 9706164 ↗InjuryFeb 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; PER EMAIL, IT WAS REPORTED THAT: PLEASE NOTE THAT, PER EDC, MULTIPLE PER CLOSURE DEVICES ATTEMPTED WITH BALLOON ANGIOPLASTY ACROSS FEMORAL PERFORATION TO TREAT THE EVENT AND THE EVENT WAS RESOLVED WITH MANUAL COMPRESSION AND TR BAND. PER CTSENR, IT WAS REPORTED THAT: (B)(6) 0110R4204 (NEW COMPLETE HEART BLOCK) EVENT DESCRIPTION: NEW COMPLETE HEART BLOCK (002- NEW PERMANENT PACEMAKER IMPLANTATION RESULTING FROM NEW OR WORSENED CONDUCTION DISTURBANCES): ON (B)(6) 2019, 3 DAYS OF POST INDEX PROCEDURE, THE SUBJECT DEVELOPED NEW COMPLETE HEART BLOCK. OF NOTE, NO CONDUCTION DISTURBANCE WAS NOTED POST BAV. -ELECTROPHYSIOLOGY STUDY WAS PERFORMED AND A NEW PERMANENT PACEMAKER WAS RECOMMENDED DUE TO THE COMPLETE HEART BLOCK. -SUBJECT WAS UNDER PROLONGED HOSPITALIZATION -ON AN UNKNOWN DAY, A NEW PERMANENT PACEMAKER WAS IMPLANTED. ON (B)(6) 2019, THE EVENT WAS CONSIDERED RECOVERED/RESOLVED WITH SEQUELAE. ON (B)(6) 2019, THE SUBJECT WAS DISCHARGED ON ASPIRIN AND CLOPIDOGREL. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS MOMENT, HOWEVER SITE HAS BEEN QUERIED FOR SOURCE DOCUMENT AND BLANK RELATIONSHIP TO STUDY DEVICE AND
Patient problems: Perforation of Vessels
MAUDE report 9681131 ↗InjuryJan 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING COMPLAINT EVENT DESCRIPTION WAS RECEIVED AT CREGANNA MEDICAL; IT WAS REPORTED THAT: PT HAD TO HAVE A FEM FEM BYPASS TO THE LEFT ILIAC. CONVERTED TO SURGERY. PACEMAKER IMPLANTED. PRE-DIL PERFORMED. EVENT DATE: (B)(6) 2019. WHY WAS THE FEM-FEM BYPASS IT PERFORMED? - PATIENT HAD A FEM FEM ON LEFT LEG, NOT SURE IF CLOSURE WAS THE CAUSE, LARGE PERFORATION IN THE ILIAC, WHEN WAS IT PERFORMED? IT WAS PERFORMED AFTER PATIENT WAS STABILIZED AND VALVE DEPLOYED. THE LOTUS INTRODUCER WAS USED DURING THE PROCEDURE. THE PATIENT PASSED AWAY ON (B)(6). THE PATIENT'S ANATOMY WAS MILD TO MOD CALCIFICATION, TORTUOUS. PATIENT WENT INTO CHB. IT IS UNKNOWN WHETHER THE PATIENT HAD ANY PRIOR CONDUCTION ISSUES. THE PERFORATION OCCURRED PAST INTRODUCER SHEATH OR WITHIN SHEATH RANGE/LENGTH. THE PERFORATION WAS IN THE LEFT ILIAC. 3 MENSIO DETAILS WERE SENT TO CREGANNA. USER EXPERIENCED A HARD TIME LOCKING THE VALVE, IT TOOK MULTIPLE TRIES AND OVER AN HOUR TO LOCK. THE REASON FOR THE DIFFICULTY LOCKING THE VALVE WAS THAT THE USER(S) COULD NOT FIND THE TWISTED POST AND HAD A HARD TIME LOCKING BUT EVENTUALLY DID LOCK. THERE WAS ONE FULL RE SHEATH DONE AND NUMEROUS PARTIAL RE-SHEATHING. USER NOT SUR
Patient problems: Death; Perforation of Vessels
MAUDE report 9647006 ↗InjuryJan 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
COMPLAINT DESCRIPTION RECEIVED AT CREGANNA MEDICAL IS AS FOLLOWS: EVENT DESCRIPTION: PER UPDATE RECEIVED ON CNF, IT WAS REPORTED THAT: UNABLE TO LOCK VALVE AND THEN UNABLE TO RESHEATH VALVE FULLY. REMOVED PARTIALLY FLOWERED DEVICE AND SHEATH AT THE SAME TIME. ILIAC PERFORATION AND EXTRAVASATION. ACCESS ROUTE TF LOCATION RIGHT TF ACCESS PERCUTANEOUS GUIDEWIRE USED SAFARI2 (XS) OTHER WIRE LUNDRQUIST SHEATH USED LIS OTHER SHEATH EMBOLIC PROTECTION USED NONE. PER EMAIL, IT WAS REPORTED THAT: SUMMARY: DIFFICULTY LOCKING AND RESHEATHING DEVICE. PATIENT DEATH. VALVE WAS PREPPED WITHOUT ISSUE AS PER DFU. PATIENT ANATOMY WAS CALCIFIED AND EXTREMELY TORTUOUS. STANDARD TAVR PROCEDURE. LOTUS LIS SHEATH IN RIGHT COMMON FEMORAL ARTERY. CONTRALATERAL ACCESS FOR LUNDERQUIST BUDDY WIRE TO FACILITATE DEVICE TRACKING THROUGH TORTUOSITY. DEVICE WAS DELIVERED TO AORTIC ANNULUS. CONSIDERABLE TORQUE WAS REQUIRED TO DELIVER THROUGH ANATOMY. DEPLOYMENT OF DEVICE: ? FIRST RESULT WAS HIGH, ASYMMETRIC UNSHEATHING? SECOND DEPLOYMENT WAS ATTEMPTED WITH LOWER STARTING POINT. PRIOR TO LOCKING, TWISTED POST WAS IDENTIFIED, RESHEATHING ATTEMPTED TO FIX. ? THIRD DEPLOYMENT TWISTED POST WAS AGAIN NOTICED. PHYSICIAN C
Patient problems: Death; Perforation of Vessels
MAUDE report 9622424 ↗InjuryDec 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
COMPLAINT EVENT RECEIVED AS FOLLOWS; 'IT WAS REPORTED THAT: STUDY ID: (B)(6) PATIENT CURRENT CONDITION: STABLE DETAILS: PATIENT (B)(6) REQUIRED STENTING OF THE EXTERNAL ILIAC (7 X 110 GORE) WHICH WAS SOMEWHAT EXPECTED. THEY BALLOONED THE EXTERNAL ILIAC PRIOR TO THE STENT AND THEN PLACED THE STENT THEREAFTER. THE ACCESS VESSELS WERE SMALL IN BOTH LEGS, SO THERE WAS A HIGHER LIKELIHOOD THAT THIS WOULD HAPPEN. BOTH THE PROCTOR AND MYSELF SET THE EXPECTATION THAT VESSEL ACCESS WOULD BE CHALLENGING, BUT DOCTOR WANTED TO PROCEED WITH THE PROCEDURE. PRIOR TO IMPLANT, THE TEAM GAINED CONTRALATERAL ACCESS IN THE EVENT THERE WAS A VASCULAR COMPLICATION. THERE WAS A DISSECTION NOTED FOLLOWING THE IMPLANT. THE DISSECTION WAS CONTROLLED (VERY LITTLE BLEEDING), THE TEAM BALLOONED AND STENTED THE LEFT EXTERNAL ILIAC AND THE PATIENT WAS STABLE THROUGHOUT THE ENTIRE PROCESS.' ADDITIONAL INFORMATION RECEIVED FROM THE DISTRIBUTOR CONFIRMING THAT THE LOTUS INTRODUCER WAS THOUGHT TO HAVE BEEN INVOLVED IN THE DISSECTION. FURTHER DISCUSSIONS WERE HELD WITH THE DISTRIBUTOR ON 20-JAN-2020 AND ADDITIONAL INFORMATION RECEIVED (FOUND IN THE ATTACHMENT OF THIS FOLLOW UP MDR REPORT) WAS DISCUSSED AND SHARED. TH
Patient problems: Vascular Dissection
MAUDE report 9400061 ↗InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EVENT DESCRIPTION: IT WAS REPORTED THAT: DELIVERY SYSTEM KINKED DUE TO OVERLY TORTUOUS AORTIC ANATOMY. KINK OBSERVED UNDER FLUOROSCOPY ELECTED TO WITHDRAW SYSTEM. KINK CAUSED CHANGE IN DELIVERY SYSTEM PROFILE AND RESULTED IN INCREASED FRICTION WHILST WITHDRAWING THROUGH LARGE LOTUS INTRODUCER SHEATH (LIS). HAEMOSTATIC CAP ON LIS DETACHED DUE TO FRICTION. TEMPORARY EXSANGUINATION THROUGH SHEATH RESULTING IN HYPOTENSION. PATIENT STABILISED. LOTUS EDGE VALVE NOT IMPLANTED. BSC CONTACT PERSON SUPPORTING THE CASE (ALEX MONCADA) PATIENT STATUS/CONDITION: STABLE WHEN DID THE PROBLEM OCCUR? DURING TRACKING OF LOTUS EDGE DELIVERY SYSTEM THROUGH AORTA AND WITHDRAWAL OF DELIVERY SYSTEM FROM INTRODUCER SHEATH PROCEDURE OUTCOME WAS IT COMPLETED WITH THIS DEVICE? ANOTHER SAME? NO. NO DEVICE IMPLANTED. WHERE DID THE PROBLEM OCCUR? INSIDE THE PATIENT ANY OTHER INFO: PHOTOS AVAILABLE AS DEVICE DISPOSED OF. EVENT DATE: (B)(6) 2019. AS OF 13-NOV-2019, THE ADDITIONAL INFORMATION WAS RECEIVED. THE KINK WAS THE LOTUS EDGE. FLUID CHALLENGE WAS GIVEN TO COMPENSATE FOR THE HYPOVOLEMIA/VAGAL RESPONSE TO BLEEDING OUT THROUGH THE LIS. BLOOD WAS SOUGHT IMMEDIATELY AND TRANSFUSION WAS GIVEN. INTRODUCER SHEATH W
Patient problems: Exsanguination
MAUDE report 9340164 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page