Luna G3 APAPRecalls & reported problems across all models
BMC Medical Co., Ltd. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Luna G3 APAP model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls
- 1
We found 2 models of Luna G3 APAP by BMC Medical Co., Ltd. in the FDA's device registry (GUDID). Together they are named in 10 adverse event reports in the FDA's MAUDE database, received between May 26, 2024 and Jan 29, 2026 — classified as 2 deaths, 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Device Emits Odor” (5 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Luna G3 APAP recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jul 15, 2026Open, Classified
Retroactive reporting for 2024 rework conducted at the 3B/React Health warehouse
Z-2979-2026 · Root cause: No Marketing Application · Recall details · FDA record ↗
Most reported Luna G3 APAP problems
Device problems
- Device Emits Odor5
- Adverse Event Without Identified Device or Use Problem2
- Thermal Decomposition of Device2
- Electrical /Electronic Property Problem1
- Output Problem1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- High Blood Pressure/ Hypertension2
- Insufficient Information2
- Alteration in Body Temperature1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bronchial Hemorrhage1
- Dyspnea1
Luna G3 APAP models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Luna® G3 APAP LG3600Ventilator, Non-Continuous (Respirator)1 recall2 deaths9 reports
- Luna® G3 APAP LG3600Ventilator, Non-Continuous (Respirator)1 report
Recent Luna G3 APAP adverse event reports
The 10 most recent reports across all models, in the reporter's own words as published by the FDA.
DeathJan 29, 2026Adverse Event Without Identified Device or Use Problem
INFORMATION FROM THE INITIAL MEDWATCH REPORT ¿ 3B MEDICAL, INC. DBA REACT HEALTH WAS CONTACTED BY A DURABLE MEDICAL EQUIPMENT (DME) PROVIDER ADVISING THAT A PATIENT HAD PASSED AWAY WHILE USING A LUNA G3 APAP DEVICE. THE DME HAD BEEN CONTACTED BY THE PATIENT'S REHABILITATION FACILITY REQUESTING ASSISTANCE WITH OBTAINING THE DEVICE¿S ELECTRONIC DATA FOR REVIEW AS PART OF THEIR INVESTIGATION. ACCORDING TO THE DME, THE PATIENT HAD BEEN USING THIS DEVICE FOR ABOUT A YEAR, ALONG WITH A RESMED AIRFIT F20 FULL FACE MASK (LARGE). THE PATIENT WAS NOT USING THE CPAP AS A LIFE-SUPPORTING DEVICE. ON THE DAY THAT THE PATIENT PASSED AWAY, THE FACILITY ADVISED THAT THERE WERE NO ABNORMALITIES AS FAR AS TREATMENTS, MEDICATION ADMINISTRATION, NOR WERE THERE ANY ABNORMAL MANIFESTATIONS OF HIS PHYSICAL SYMPTOMS. THE REHABILITATION FACILITY CONFIRMED THAT THE MASK WAS ON THE PATIENT ON THE NIGHT OF (B)(6) 2025 AND INTO THE MORNING OF (B)(6) 2025, HOWEVER THE FACILITY COULD NOT CONFIRM WHETHER OR NOT THE DEVICE WAS POWERED ON. THE PATIENT WAS NOT TRANSPORTED TO THE HOSPITAL FOR TREATMENT. EMS WAS CONTACTED AND THEY PRONOUNCED HIS DEATH AT THE FACILITY. THE DME DOWNLOADED THE DEVICE¿S SD CARD AND PROVIDE
Patient problems: Insufficient Information
MAUDE report 24212268 ↗DeathJan 28, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, THE MANUFACTURER RECEIVED A COMPLAINT FROM THE IMPORTER. THE CONTENT OF THE COMPLAINT WAS AS FOLLOWS: CUSTOMER REPORTED: MY TEAM RECEIVED A CALL FROM (B)(4) ASKING FOR A DOWNLOAD ON A CPAP MACHINE. THEY NOTIFIED US THAT THEY WERE CONTACTED BY (B)(4) ON A MUTUAL PATIENT AND WERE NOTIFIED THAT THE PATIENT PASSED WHILE USING THE DEVICE. AFTER SPEAKING WITH (B)(4) - THEY ARE WANTING TO DOUBLECHECK THAT A MANUAL DOWNLOAD MATCHES WHAT WAS PULLED ON REACT CONNECT. THEY WILL BE BRINGING THE DEVICE TO US TO PULL A DL ON THE SD CARD. PER THE REACT CONNECT DL, THE PATIENT DID NOT USE THE DEVICE THE NIGHT OF HIS PASSING ON (B)(6) 2025. WE WILL QUARANTINE THE DEVICE AFTER PULLING THE SD CARD. THIS DEVICE IS (B)(4) OWNED BUT WAS PROVIDED BY FOCUS. THERE IS NO ALLEGATION OF THE DEVICE CAUSING OR CONTRIBUTING TO THE DEATH. ON (B)(6) 2026, MANUFACTURER BMC RECEIVED THE 3500A REPORT THAT IMPORTER REACT HEALTH WAS ABOUT TO SUBMIT. THE CONTENT OF THE REPORT IS AS FOLLOWS:3B MEDICAL, INC. DBA REACT HEALTH WAS CONTACTED BY A DURABLE MEDICAL EQUIPMENT (DME) PROVIDER ADVISING THAT A PATIENT HAD PASSED AWAY WHILE USING A LUNA G3 APAP DEVICE. THE DME HAD BEEN CONTACTED BY THE PATIENT'S REHAB
Patient problems: Insufficient Information
MAUDE report 24197303 ↗MalfunctionNov 23, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER THAT A PATIENT STATED HIS MACHINE LOST SOME POWER AND HE WOKE UP SEVERAL TIMES WITH SHORTNESS OF BREATH. HIS BLOOD PRESSURE WAS 181/101 (HYPERTENSIVE CRISIS/URGENCY) AND HIS HAND WAS ICE COLD. THE PATIENT DID NOT SEEK MEDICAL ATTENTION. THE MANUFACTURER HAS RECEIVED THE DEVICE AND INVESTIGATED.
Patient problems: Alteration in Body Temperature; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Bronchial Hemorrhage; High Blood Pressure/ Hypertension
MAUDE report 23628032 ↗InjuryOct 17, 2025Output Problem; Output Problem
3B MEDICAL, INC. DBA REACT HEALTH WAS CONTACTED BY A DURABLE MEDICAL EQUIPMENT (DME) PROVIDER ADVISING THAT THEIR PATIENT REPORTED THAT OVER THE LAST FEW WEEKS HE'S BEEN WAKING UP WITH INCREASED BLOOD PRESSURE AND "FEELINGS OF SUFFOCATION¿ USING THE DEVICE. THE DME PROVIDER ADVISED REACT HEALTH THAT HE HAD REVIEWED THE DATA FROM THE PATIENT¿S DEVICE AND NOTICED INCREASES IN PRESSURE, BUT THERE WERE NO APPARENT OBSTRUCTIONS OCCURRING (OR AT LEAST NOTHING BEING RECORDED BY THE DEVICE). ADDITIONALLY, HE ADVISED THAT THE LEAK LEVEL WAS SHOWING NO DEVIATION AT ALL ¿ IT WAS AT 0. THE DEVICE WAS THEN RETURNED TO, AND EVALUATED BY, REACT HEALTH WHO VERIFIED THE REPORTED ISSUES THAT THE DEVICE WAS INCREASING IN PRESSURE AND THAT THE DATA WAS NOT BEING RECORDED. DURING TESTING, REACT HEALTH ALSO OBSERVED THAT THE FLOW WAS ABNORMALLY LOW. REACT HEALTH DETERMINED THAT THE ROOT CAUSE OF THE REPORTED ISSUES WAS THE FLOW SENSOR ON THE MAIN PCBA. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER.
Patient problems: Dyspnea; High Blood Pressure/ Hypertension
MAUDE report 23326312 ↗MalfunctionSep 8, 2024Thermal Decomposition of Device; Device Emits Odor; Sparking; Thermal Decomposition of Device; Device Emits Odor; Sparking
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER STATING A PATIENT REPORTED THAT THE POWER SUPPLY ON THEIR CPAP SPARKED AND CREATED A BURNING SMELL. PATIENT REPORTED THAT THE POWER BRICK PORTION OF THE POWER CORD WAS HOT AND THE PLASTIC HOUSING HAD MELTED. THERE IS NO HARM OR INJURY REPORTED. THE DEVICE HAS BEEN RETURNED TO THE IMPORTER. THE MANUFACTURER HAS RECEIVED THE DEVICE AND INVESTIGATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20165838 ↗MalfunctionJun 24, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL PROVIDER STATING THAT A DEVICE HAD EVIDENCE OF MELTING WHERE POWER CORD PLUGS IN. THE DEVICE HAS BEEN RECEIVED BY REACT HEALTH.THE MANUFACTURER HAS RECEIVED THE DEVICE AND INVESTIGATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19592234 ↗MalfunctionJun 12, 2024Device Emits Odor; Device Emits Odor
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER STATING THAT A DEVICE HAD A BURNING SMELL. THERE WAS NO PATIENT HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. A FOLLOW UP REPORT WILL BE FILED ONCE THE INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19525644 ↗MalfunctionJun 4, 2024Device Emits Odor; Device Emits Odor
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER STATING THAT A DEVICE HAD A BURNING SMELL.THERE WAS NO PATIENT HARM OR INJURY REPORTED.THE DEVICE HAS BEEN RETURNED TO THE IMPORTER BUT THE MANUFACTURER HAS NOT TAKEN POSSESSION.A FOLLOW UP REPORT WILL BE FILED ONCE THE INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19456822 ↗MalfunctionJun 3, 2024Device Emits Odor; Device Emits Odor
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER STATING THAT AN OUT OF BOX FAILURE OCCURRED AND THAT THE DEVICE HAD A BURNING SMELL.THERE WAS NO PATIENT HARM OR INJURY.THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER.A FOLLOW UP REPORT WILL BE FILED ONCE THE INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19445361 ↗MalfunctionMay 26, 2024Device Emits Odor; Device Emits Odor
REACT HEALTH RECEIVED A COMPLAINT FROM A DURABLE MEDICAL EQUIPMENT PROVIDER STATING THAT THE PRESSURE WAS PULSING AND THE DEVICE WAS EMITTING A BURNING SMELL. NO PATIENT HARM OR INJURY WAS REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE IMPORTER. A FOLLOW UP REPORT WILL BE FILED ONCE THE INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19398863 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page