MarksmanTMRecalls & reported problems across all models
Micro Therapeutics, Inc. · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 146
- Deaths reported
- 0
- Injuries reported
- 21
- Recalls
- 0
We found 5 models of MarksmanTM by Micro Therapeutics, Inc. in the FDA's device registry (GUDID). Together they are named in 146 adverse event reports in the FDA's MAUDE database, received between Nov 22, 2017 and Aug 17, 2026 — classified as 21 injuries, 125 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (87 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported MarksmanTM problems
Device problems
- Physical Resistance/Sticking87
- Break41
- Retraction Problem30
- Material Deformation26
- Deformation Due to Compressive Stress18
- Unintended Movement9
- Adverse Event Without Identified Device or Use Problem6
- Burst Container or Vessel3
Patient problems
- No Clinical Signs, Symptoms or Conditions109
- No Known Impact Or Consequence To Patient14
- Foreign Body In Patient11
- Intracranial Hemorrhage4
- Device Embedded In Tissue or Plaque3
- Obstruction/Occlusion3
- Vasoconstriction2
- Aneurysm1
MarksmanTM models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MarksmanTM FA-55150-1030Catheter, continuous flush109 reports
- MarksmanTM FA-55160-1030Catheter, continuous flush32 reports
- MarksmanTM FA-55150-1030Catheter, continuous flush3 reports
- MarksmanTM FA-55160-1030Catheter, continuous flush1 report
- MarksmanTM FA-55135-1030Catheter, continuous flush1 report
Recent MarksmanTM adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionAug 17, 2026Material Integrity Problem
Additional information was received reporting that the patients vessel tortuosity was normal. The patient is recovering good. No patient symptoms or further complications were reported as a result of this event. The Pipeline had not been placed in a vessel bend when it failed to open. No additional steps or other devices were attempted to open the Pipeline. The inner liner coating was not observed to be from the Marksman catheter.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50278788 ↗MalfunctionMay 6, 2026Retraction Problem; Deformation Due to Compressive Stress
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR AMORPHOUS, UNRUPTURED MULTIPLE ANEURYSMS AT THE LEFT C7 SEGMENT ANEURYSM WITH A MAX DIAMETER OF 5MM AND A 4 MM NECK DIAMETER. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE LANDING ZONE WAS 3.4 MM DISTALLY AND 4.15 MM PROXIMALLY. DUAL ANTIPLATELET TREATMENT (DAPT) WAS ADMINISTERED. THE PRU LEVEL WAS NORMAL. IT WAS REPORTED THAT THE MIDDLE SEGMENT OF THE PED-425-18 STENT FAILED TO OPEN AND REMAINED UNABLE TO OPEN DESPITE REPEATED ATTEMPTS. ANOTHER PED-400-20WAS USED, BUT THE DISTAL TIP END OF THE SECOND STENT FAILED TO OPEN, AND THE REST OF THE STENT COULD NOT BE RETRIEVED. IT HAD TO BE WITHDRAWN . THE ANGIOGRAPHIC RESULT POST PROCEDURE SHOWED BLOOD FLOW WAS UNOBSTRUCTED, WITH SIGNIFICANT CONTRAST STAGNATION WITHIN THE ANEURYSM. THE PIPELINE WAS USED FOR AN INDICATION THAT IS APPROVED (ON-LABEL). THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE A U-TRACK 5F SHEATH, MARKSMAN MICROCATHETER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25104534 ↗MalfunctionMay 4, 2026Retraction Problem; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT OF A MARKSMAN, ECHELON, AND PHENOM THAT ENCOUNTERED RESISTANCE WITH A PIPELINE THAT ALSO FAILED TO OPEN. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, RUPTURED ANEURYSM WITH A MAX DIAMETER OF 5MM AND A 4MM NECK DIAMETER. THE LANDING ZONE (ARTERY SIZE) WAS 1.8MM DISTAL AND 2.3MM PROXIMAL. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE FEMORAL ARTERY WAS THE ACCESS VESSEL WITH A 8-9MM DIAMETER. DAPT (DUAL ANTIPLATELET TREATMENT) WAS ADMINISTERED AND THE PLATELET REACTIVITY UNITS (PRU) LEVEL MET THE STANDARD. THE ANGIOGRAPHIC RESULT POST PROCEDURE WAS A MIDDLE CEREBRAL ARTERY M2 SEGMENT ANEURYSM. IT WAS REPORTED THAT THE OPERATOR PLANNED TO PARTIALLY DEPLOY THE PED, THEN PLACE COILS, AND FINALLY FULLY DEPLOY THE PED. AFTER ANGIOGRAPHY, A NAVIEN WAS ADVANCED. THE MARKSMAN WAS DELIVERED FIRST; AFTER THE MICROCATHETER REACHED THE M1 SEGMENT, RESISTANCE INCREASED. THE OPERATOR OVERCAME THE RESISTANCE AND CONTINUED ADVANCING THE MICROCATHETER UNTIL IT WAS IN POSITION. THEN THE OPERATOR STARTED TO DELIVER ECHELON. A 250-18 STENT WAS THEN SELECTED FOR DEPLOYMENT. THE TIP END OPENED SMOOTHLY AND WAS ANCHORED. THE STENT WAS FURTHER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25073109 ↗MalfunctionMar 31, 2026Retraction Problem; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A PIPELINE ENCOUNTERED RESISTANCE IN A MARKSMAN CATHETER. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AMORPHOUS, UNRUPTURED ANEURYSM LOCATED IN THE RIGHT INTERNAL CAROTID ARTERY WITH A MAX DIAMETER OF 8 MM AND A 5 MM NECK DIAMETER. LANDING ZONE: DISTAL: 4 MM AND PROXIMAL: 4.3 MM. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL AND ALSO NOTED NORMAL. THE ACCESSED VESSEL WAS THE FEMORAL ARTERY. DUAL ANTIPLATELET TREATMENT WAS ADMINISTERED. THE PRU LEVEL WAS NORMAL. THE ANGIOGRAPHIC RESULT POST PROCEDURE WAS NORMAL. IT WAS REPORTED THAT PIPELINE 4.50-35 WAS STUCK INSIDE MARKSMAN MICROCATHETER AND REFUSED TO BE REMOVED FROM OR OUTSIDE THE MARKSMAN. IT WAS REPORTED THE PIPELINE WAS STUCK (LOCKED UP) IN THE CATHETER OR RESISTANCE IN THE MIDDLE OF THE CATHETER OCCURRED. IT WAS REPORTED THE MARKSMAN AND THE PIPELINE WERE STUCK TOGETHER WERE COULD NOT BE DETACHED. IT WAS REPORTED THE PIPELINE WAS NOT ABLE TO OPEN. THE CATHETER WAS FLUSHED CONTINUOUSLY WITH HEPARINIZED SALINE. THE PIPELINE BECAME STUCK IN THE MIDDLE SECTION DURING DEPLOYMENT. THE PHYSICIAN RELEASED THE LOAD (SLACK) IN THE SYSTEM IN AN ATTEMPT TO RESOLVE THE ISSUE, BUT IT W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24742810 ↗MalfunctionMar 30, 2026Retraction Problem
ADDITIONAL INFORMATION WAS RECEIVED. THE PIPELINE FAILED TO OPEN AT THE PROXIMAL END OF THE STENT AND WAS NOT PLACED IN A VESSEL BEND AT THE TIME OF THE FAILURE TO OPEN. NO ADDITIONAL STEPS OR OTHER DEVICES WERE SUCCESSFULLY USED TO OPEN THE PIPELINE, AS THE STENT COULD NOT BE RETRIEVED INTO THE CATHETER DESPITE AN ATTEMPT BEING MADE. THE CATHETER DID NOT ENCOUNTER RESISTANCE DURING RETRIEVAL OF THE STENT. THE CAUSE OF THE EVENT WAS NOT DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24734397 ↗MalfunctionMar 23, 2026Break; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED. ¿THE PROXIMAL END OF THE STENT MICROCATHETER BROKE¿ WAS CLARIFIED TO MEAN THAT THE CATHETER BROKE AT THE PROXIMAL END. THE BREAK WAS SUSPECTED TO BE DUE TO VASCULAR TORTUOSITY AND EXCESSIVE TENSION. NO CAUSES OR CONTRIBUTING FACTORS FOR THE RESISTANCE WERE IDENTIFIED. THE RESISTANCE OCCURRED IN THE PROXIMAL PORTION OF THE CATHETER. NO DAMAGE TO THE PIPELINE PUSHWIRE WAS OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24669656 ↗MalfunctionMar 23, 2026Retraction Problem; Deformation Due to Compressive Stress; Material Deformation
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE EVENT WAS DUE TO THE ACCESS PATH BEING EXCESSIVELY TORTUOUS, AND MICROCATHETER DAMAGE RESULTING IN STENT DAMAGE. THE STENT COULD NOT BE RETRIEVED INTO THE MICROCATHETER. DURING DELIVERY THE OPERATOR REPORTED THAT THERE WAS INCREASED RESISTANCE WHICH MADE ADVANCEMENT DIFFICULT. THE RESISTANCE OCCURRED AT THE DISTAL SECTION OF THE CATHETER. IT WAS STATED THAT THE VESSEL WAS TORTUOUS, AND THE CAVERNOUS SINUS WAS TYPE IV. THE OVERALL PROCEDURE WAS COMPLETED WITH A A NEW SYSTEM BEING USED INSTEAD, AND THE MICROCATHETER BEING REPLACED WITH A PHENOM 2. IT WAS CLARIFIED THAT PREVIOUSLY REPORTED "ACCORDION-LIKE DEFORMATION WAS OBSERVED AT THE DISTAL END." WAS REFERRING TO CATHETER DAMAGE DUE TO THE ELONGATION OF THE DISTAL SEGMENT OF THE MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24664909 ↗MalfunctionMar 19, 2026Retraction Problem; Material Deformation
ADDITIONAL INFORMATION WAS RECEIVED. AFTER REMOVING THE STENT DELIVERY GUIDEWIRE TOGETHER WITH THE MICROCATHETER, THE ACCESS ROUTE WAS RE-ESTABLISHED AND THE SUBSEQUENT PROCEDURAL STEPS WERE COMPLETED TO FINISH THE PROCEDURE. THERE WERE NO CAUSES OR CONTRIBUTING FACTORS FOR THE RESISTANCE OR IRREGULAR SHAPE THAT WERE IDENTIFIED. IT WAS CLARIFIED THAT "IRREGULAR SHAPE" MEANS THAT CURVATURE AND DEFORMATION WERE OBSERVED AT THE DISTAL END OF THE MICROCATHETER. NO SHAPING HAD BEEN PERFORMED. THE CATHETER WAS CONTINUOUSLY FLUSHED THROUGHOUT THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24639214 ↗MalfunctionFeb 6, 2026Retraction Problem
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE TIP OF THE SOLITAIRE SHEATH WAS SECURED INTO THE HUB OF THE MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24278336 ↗MalfunctionFeb 5, 2026Retraction Problem
ADDITIONAL INFORMATION RECEIVED. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY OR POSITIONING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24271936 ↗MalfunctionJan 23, 2026Retraction Problem; Deformation Due to Compressive Stress; Material Deformation; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS DIFFICULTY DURING DELIVERY, AS THE PIPELINE COULD NOT BE PUSHED OUT OF THE MARKSMAN CATHETER. RESISTANCE WAS ENCOUNTERED IN THE DISTAL SEGMENT OF THE CATHETER. THE CATHETER WAS FLUSHED AND FULLY HYDRATED AS INDICATED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24161199 ↗InjuryJan 13, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT HAD MULTIPLE INTRACRANIAL ANEURYSMS OF THE RIGHT INTERNAL CAROTID ARTERY, INCLUDING A LARGE 10MM ANEURYSM AT THE POSTERIOR COMMUNICATING ARTERY SEGMENT AND A SMALL 4MM ANEURYSM AT THE OPHTHALMIC ARTERY SEGMENT. THE OPERATOR PLANNED TO IMPLANT A PIPELINE FOR TREATMENT. DURING THE PROCEDURE TO ESTABLISH ACCESS, IT WAS FOUND THAT THE MICROCATHETER COULD NOT ENTER THE LUMEN OF THE NAVIEN DUE TO SIGNIFICANT RESISTANCE. AFTER REPLACING WITH A NEW ONE, IT WENT THROUGH SUCCESSFULLY. THE MARKSMAN MICROCATHETER WAS POSITIONED, AND A PED-450-30 WAS DELIVERED TO THE DISTAL END OF THE MICROCATHETER. HOWEVER, THE STENT COULD NOT BE CAME OUT DESPITE BEING ABLE TO BE RETRIEVED. MULTIPLE ATTEMPTS WERE MADE WITHINCREASED PUSHING FORCE, BUT THE PROBLEM PERSISTED. SUBSEQUENTLY, POSSIBLY DUE TO REPEATED FORCEFUL MANIPULATION, THE ANEURYSM SUDDENLY RUPTURED DURING THE PROCEDURE. THE OPERATOR STATED THAT THE STENT HAD A QUALITY ISSUE AND WOULD NOT BE CHARGED FOR IT. THE DEVICE WAS STUCK (LOCKED UP) IN THE CATHETER/RESISTANCE IN THE CATHETER. THE CATHETER WAS FLUSHED CONTINUOUSLY WITH HEPARANIZED SALINE. THE PIPELINE BECAME STUCK IN THE DISTAL SECTION OF THE CA
Patient problems: Intracranial Hemorrhage; Ruptured Aneurysm
MAUDE report 24060658 ↗MalfunctionJan 13, 2026Retraction Problem; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR AN INTRACRANIAL ANEURYSM. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE RIGHT FEMORAL ARTERY, WITH 7MM DIAMETER. IT WAS REPORTED THAT AFTER THE CATHETER WAS POSITIONED BY THE OPERATOR, THE PED-400-20 STENT WAS DELIVERED TO THE POSITION AND DEPLOYED SMOOTHLY. WHEN RETRIEVING THE STENT PUSH ROD, THE DISTAL END OF THE PUSH ROD COULD NOT BE WITHDRAWN INTO THE CATHETER AT THE DISTAL AREA DESPITE MULTIPLE ATTEMPTS, LEAVING THE TIP/DISTAL RADIOPAQUE MARKER AND PROTECTIVE SLEEVE OUTSIDE THE CATHETER. AT THIS POINT, THE OPERATOR COULD ONLY REMOVE THE CATHETER AND THE PUSH ROD TOGETHER. THE OPERATOR STATED THAT THE STENT PUSH ROD MUST BE ABLE TO BE WITHDRAWN INTO THE CATHETER FOR REMOVAL TO ENSURE THE SAFETY AND SUCCESS OF THE PROCEDURE. THE OPERATOR REQUESTED TO FILE A COMPLAINT REGARDING THE CATHETER. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AND FLUSHED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24054083 ↗MalfunctionJan 12, 2026Retraction Problem
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE CAUSE OF THE EVENT WAS NOT DETERMINED. CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. THERE WAS NO OBSERVED DAMAGE TO THE PIPELINE PUSHWIRE OR CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24037263 ↗MalfunctionDec 30, 2025Burst Container or Vessel; Physical Resistance/Sticking; Burst Container or Vessel; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE RESISTANCE OCCURRED IN THE MIDSECTION OF THE CATHETER. THERE WAS CONTINUOUS SALINE FLUSH ADMINISTERED AND MAINTAINED AD INDICATED IN THE IFU. THERE WAS NO OBSERVED DAMAGE TO THE PIPELINE PUSHWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23922301 ↗
Other Micro Therapeutics, Inc. products
- PipelineTM Flex Embolization Device with Shield TechnologyTM79 models · 1,020 reports
- AxiumTM139 models · 832 reports · 1 recall
- PIPELINE114 models · 622 reports · 1 recall
- Concerto Detachable Coil System34 models · 595 reports
- Axium Prime Detachable Coil System70 models · 497 reports
- ConcertoTM70 models · 436 reports
- EchelonTM5 models · 225 reports · 2 recalls
- Axium Detachable Coil System69 models · 288 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.