mild Device KitRecalls & reported problems across all models
Stryker Corporation · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 32
- Recalls
- 0
We found 3 models of mild Device Kit by Stryker Corporation in the FDA's device registry (GUDID). Together they are named in 32 adverse event reports in the FDA's MAUDE database, received between Dec 27, 2018 and Apr 16, 2026 — classified as 32 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported mild Device Kit problems
Device problems
- Adverse Event Without Identified Device or Use Problem29
- Appropriate Device Problem Term/Code Not Available3
Patient problems
- Hematoma13
- Cerebrospinal Fluid Leakage7
- Pain5
- Neuropathy3
- Numbness3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Movement Disorder2
- Atrophy1
mild Device Kit models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- mild Device Kit MDK-0001Arthroscope29 reports
- mild Device Kit MDK-0002Arthroscope2 reports
- mild Device Kit MDK-0001Arthroscope1 report
Recent mild Device Kit adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryApr 16, 2026Adverse Event Without Identified Device or Use Problem
EVENT WAS REPORTED BY PATIENT VIA MEDWATCH. THE PATIENT STATED THAT DURING A MILD PROCEDURE, THE DOCTOR REMOVED OLD SCAR TISSUE FROM A PREVIOUS TWO-LEVEL SPINAL FUSION INSTEAD OF THE LIGAMENTUM FLAVUM, WHICH CAUSED NERVE ISSUES. THE PATIENT ALLEGED THAT THE SCAR TISSUE REMOVAL LED TO PERMANENT DISABLEMENT AND MAJOR MUSCLE ATROPHY IN EVERY MUSCLE THAT THE S1 INNERVATES. THE PATIENT STATED THIS ALSO CAUSED CAUDA EQUINA SYNDROME AND ARACHNOIDITIS. BASED ON THE INFORMATION PROVIDED, THERE WAS NO MALFUNCTION OF THE MILD DEVICE.
Patient problems: Atrophy; Neuropathy; Pain
MAUDE report 24904533 ↗InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING THE PROCEDURE, THE PATIENT REPORTED INCREASED PAIN IN THEIR LEGS AS WELL AS HYPERTENSION AND HEADACHES. THE PATIENT WAS INSTRUCTED TO GO THE ER WHERE IMAGING WAS USED TO DIAGNOSE A EPIDURAL HEMATOMA. THEY WERE SCHEDULED FOR A LAMINECTOMY. IT SHOULD BE NOTED THAT THE PATIENT IS DOING WELL AND RECOVERING NOW. PER SALES REPRESENTATIVE, THERE WAS NO ALLEGED DEVICE ISSUE DURING PROCEDURE.
Patient problems: Pain
MAUDE report 24877063 ↗InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING THE PROCEDURE THE PATIENT COULD NOT FEEL THEIR LEGS AND ALLEGEDLY EXPERIENCED PROLONGED WEAKNESS. THE PATIENT WAS GIVEN AN MRI WHICH SHOWED AIR OR FLUID IN THE SPINAL SPACE AND UNDERWENT AN ADDITIONAL SPINE SURGERY. THERE WAS NO ALLEGED DEVICE MALFUNCTION. PER SALES REPRESENTATIVE, THERE WAS NO ALLEGED DEVICE ISSUE DURING PROCEDURE.
Patient problems: Numbness
MAUDE report 24820502 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 10/31/2024, A VERTOS REPRESENTATIVE REPORTED THAT A FEMALE PATIENT, WHO WAS UNDER TREATMENT FOR A MULTI-LEVEL L3/L4 AND L4/L5 MILD PROCEDURE ON THE SAME DAY, EXHIBITED EVIDENCE OF A CSF LEAK DURING THE PROCEDURE. THE OPERATING PHYSICIAN CONFIRMED THAT THE DURA WAS PUNCTURED DURING TREATMENT WHILE USING THE MILD DEVICE KIT, RESULTING IN THE CSF LEAK. THE PATIENT DID NOT SHOW ANY SYMPTOMS BEYOND THE CSF LEAK WHILE SHE REMAINED LAYING DOWN ON THE OPERATING TABLE, BUT DID EXPERIENCE HEADACHES SITTING UP SOON AFTER THE PROCEDURE. A BLOOD PATCH WAS PERFORMED TO REMEDY THE DURA PUNCTURE FOLLOWING THE MILD PROCEDURE, AND THE PATIENT WAS DISCHARGED ABOUT AN HOUR AFTER THE PROCEDURE WAS COMPLETED. THE PATIENT WAS REPORTED TO HAVE A POST LUMBAR SPINAL FUSION AT THE L4/L5 LEVEL PRIOR TO THE MILD PROCEDURE, BUT OTHERWISE OBSERVED WITH NO OTHER UNUSUAL ANATOMY. THERE HAVE BEEN NO FURTHER POST-OPERATIVE SYMPTOMS FROM THE PATIENT REPORTED FOLLOWING HER DISCHARGE FROM THE HOSPITAL.
Patient problems: Cerebrospinal Fluid Leakage; Dural Tear
MAUDE report 20748208 ↗InjurySep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2019, A 73 YEAR-OLD FEMALE PATIENT HAD A MILD PROCEDURE PERFORMED ON L2-L3 AND L3-L4, FOLLOWED BY AN EPIDURAL STEROID INJECTION AT THE SAME LEVELS. IMMEDIATELY FOLLOWING THE PROCEDURE, SHE NOTED SEVERE LOWER BACK AND RIGHT LEG PAIN. AN MRI WAS DONE AND WAS INCONCLUSIVE. A CET WITH MYELOGRAM SHOWED AN L1-L2 EPIDURAL HEMATOMA. THE PATIENT HAD TWO DECOMPRESSION LAMINECTOMIES PERFORMED BY A NEUROSURGEON FOLLOWING THE MILD PROCEDURE AND WAS PLACED IN AN INPATIENT REHAB TO RECOVER. THE PATIENT HISTORY PRIOR TO THE MILD PROCEDURE INCLUDED A FUSION AT L4-L5 WITH POSTERIOR INSTRUMENTATION AND SPINAL CORD STIMULATOR WITH ANCHORS OVERLYING THE RIGHT L2-L3 AND L3-L4 INTERSPACE.
Patient problems: Hematoma
MAUDE report 20293200 ↗InjurySep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2019, AN 88 YEAR-OLD MALE PATIENT HAD A MILD PROCEDURE PERFORMED. FOLLOWING THE PROCEDURE, THE PATIENT WENT TO THE HOSPITAL FOR LEG PAIN AND WAS DIAGNOSED WITH AN EPIDURAL HEMATOMA. THE PATIENT WAS TREATED FOR THE INITIAL HEMATOMA, AND THEN AGAIN FOR A SECOND HEMATOMA THAT HAD DEVELOPED. THERE WAS NO REPORTED DEVICE MALFUNCTION, AND THE PATIENT HAS MADE A FULL RECOVERY. THE PHYSICIAN REPORTED THAT THE PATIENT'S HISTORY WAS BELIEVED TO BE THE CAUSE OF THE HEMATOMA, AS THE PATIENT WAS ON BLOOD THINNERS AND WAS REPORTED TO BE SUFFERING FROM ALCOHOLISM.
Patient problems: Hematoma
MAUDE report 20292992 ↗InjurySep 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A 71 YO MALE PATIENT HAD A BILATERIAL L-3-L4, L4-L5 MILD PROCEDURE. ON 8/23/2024, A VERTOS REPRESENTATIVE WAS CONTACTED BY THE TREATING PHYSICIAN WHO REPORTED THAT THE PATIENT HAD PAIN WHEN BENT OVER AND NUMBNESS IN THE LEG 4-6 HOURS AFTER THE PROCEDURE AND WENT TO THE HOSPITAL ON (B)(6) 2024. AN MRI SCAN SHOWED A HEMATOMA ON THE LEFT SIDE OF L3-L4. THE PATIENT WAS REFERRED TO A SURGEON AND UNDERWENT A SUCCESSFUL LAMINECTOMY ON (B)(6) 2024. PATIENT'S SPINE WAS FUSED FROM T10-S1.
Patient problems: Hematoma
MAUDE report 20257470 ↗InjuryAug 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT WHO WAS TREATED ON 6/25 WITH A BILATERAL L3-L4 MILD PROCEDURE, WAS SEEN IN OFFICE BY THE TREATING PHYSICIAN AS FOLLOW-UP ON 7/22 WITH SITE TENDERNESS, FLUCTUANT, AND DRAINAGE WAS DETERMINED TO HAVE DEVELOPED AN INFECTION. AN MRI WITH CONTRAST SHOWED 9MM SUBCUTANEOUS ENHANCING COLLECTION CONCERNING FOR ACCESS AT THE L4-5 LEVEL WITH EXTENSION OF INFLAMMATION INTO THE BILATERAL PARASPINAL MUSCLES (R>L). THERE WAS NO CONCERN FOR EXTENSION INTO THE SPINAL CANAL. THE PATIENT'S POST-PROCEDURAL PAIN WAS MANAGEABLE, AND THE PATIENT DENIED EXPERIENCING ANY SYMPTOMS OF FEVERS, CHILLS, WEAKNESS, NUMBNESS, OR INCONTINENCE AFTER THE MILD PROCEDURE. INFORMATION SUCH AS THE PATIENT'S ADHERENCE TO ANY RECOVERY REGIMENT, INCLUDING THE PATIENT'S LAST STEROID USE DURING RECOVERY, DIABETIC STATUS, OR ANY PREVIOUS HEALTH ISSUES WITH POST-PROCEDURAL HEALING OR ONGOING SYSTEMIC INFECTIONS COULD NOT BE OBTAINED. THE PATIENT WAS TREATED IN THE HOSPITAL WITH IV ANTIBIOTICS FOR THE INFECTION AND DISCHARGED ON 7/25A PATIENT WHO WAS TREATED ON 6/25 WITH A BILATERAL L3-L4 MILD PROCEDURE, WAS SEEN IN OFFICE BY THE TREATING PHYSICIAN AS FOLLOW-UP ON 7/22 WITH SITE TENDERNESS, FLUCTUANT, AND DRAINAGE WAS
Patient problems: Skin Infection
MAUDE report 20072506 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2022, A FEMALE PATIENT HAD A BILATERAL L3-L4, L4-L5 MILD PROCEDURE PERFORMED USING A STREAMLINE APPROACH WITH EPIDUROGRAM (LEVEL UNKNOWN). NO PROCEDURAL ISSUES WERE NOTED AND PATIENT WAS DISCHARGED TO HOME. FOUR DAYS AFTER THE PROCEDURE, ON (B)(6) 2022, THE PATIENT REPORTED SYMPTOMS OF A CSF LEAK (FLUID OUT OF INCISION SITE AND HEADACHES). THE PATIENT RECEIVED TWO UNSUCCESSFUL BLOOD PATCH ATTEMPTS BY THE TREATING PHYSICIAN AND WAS SENT TO THE NEURO-SURGEON WHO PERFORMED A DURAL SEAL. PATIENT IS REPORTEDLY DOING FINE AFTER THE DURAL SEAL PROCEDURE. THE PATIENT HAD SYMPTOMS OF A CSF LEAK 4 DAYS AFTER THE INITIAL MILD PROCEDURE AND NOTHING REPORTED SHOWS ANY MEDICAL DEVICE FAILURE OR MALFUNCTION OR THAT THE DEVICE CAUSED/CONTRIBUTED TO THIS EVENT.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 19716446 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 12/9/21, A COMPANY REPRESENTATIVE WAS CONTACTED BY TWO PROFESSIONAL ACQUAINTANCES, ONE OF WHOM TOLD OF A RUMOR THAT A MILD PATIENT WAS IN THE EMERGENCY ROOM AT (B)(6) FROM APPARENT COMPLICATIONS. AFTER FURTHER INVESTIGATION, THE COMPANY REP SPOKE TO SOME (B)(6) PHYSICIANS WHO MENTIONED THAT ONE OF THEIR NEUROSURGEONS HAD TREATED A PATIENT POST MILD PROCEDURE. THE REP REPORTED THIS BACK TO A PHYSICIAN CUSTOMER, WHO DOES MILD PROCEDURES AT A FACILITY IN THE (B)(6). THE DR'S OFFICE INVESTIGATED AND FOUND THAT ONE OF HIS PATIENTS HAD A MILD PROCEDURE ON (B)(6) 2021 AND HAD CALLED ABOUT POST-OP PAIN. THE DR. INSTRUCTED THE PATIENT TO GO TO THE EMERGENCY ROOM IF THE PAIN GOT WORSE BUT DID NOT HEAR FURTHER ABOUT WHAT HAPPENED. IN (B)(6) OF 2022, A PROFESSOR IN PAIN MANAGEMENT AT (B)(6) WHO WAS NOT INVOLVED WITH THE PATIENT'S CASE FOLLOWED UP ON THE RUMORS AND SPOKE TO THE DR. AND THE NEUROSURGEON. ACCORDING TO THE PROFESSOR: THE PATIENT HAD BEEN ADMITTED TO THE EMERGENCY ROOM AT (B)(6) ON (B)(6) 2021 WITH MOTOR/SENSORY LOSS, DECREASED MOTOR STRENGTH, AND DORSAL FLEXION OF THE LEFT LEG; THE NEUROSURGEON OFFERED SURGERY BUT THE PATIENT INITIALLY REFUSED; ON (B)(6) 2021, NEUROSURGEON PERF
Patient problems: Cerebrospinal Fluid Leakage; Muscle Weakness/Atrophy; Paralysis
MAUDE report 19715644 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2020, A 74 Y/O FEMALE PATIENT HAD A BILATERAL L4-L5 MILD PROCEDURE. THE PATIENT HAD BEEN ON ANTICOAGULANTS THAT WERE REPORTEDLY STOPPED IN AN APPROPRIATE TIMEINE BEFORE THE PROCEDURE. AN EPIDUROGRAM WAS REPORTED AS PERFORMED AT L3-L4 DUE TO DIFFICULTY ACCESSING THE L4-L5 FOR THE EPIDUROGRAM. THERE WAS NO REPORTED DEVICE MALFUNCTION AND THE DEVICE PERFORMED AS INTENDED. APPROXIMATELY 2 HOURS FOLLOWING THE PROCEDURE, THE PATIENT COMPLAINED OF SEVERE BACK PAIN AND LEG WEAKNESS AND WENT FROM HOME TO THE ER. AN MRI WAS PERFORMED, AND AN EPIDURAL HEMATOMA WAS IDENTIFIED. THE PATENT UNDERWENT SPINE SURGERY AND HAD THE HEMATOMA EVACUATED ONE DAY AFTER THE MILD PROCEDURE. IT WAS REPORTED THAT THE HEMATOMA WAS AT T-11 AND THE MILD PROCEDURE WAS PERFORMED MUCH LOWER AT L4-L5 WITH EPIDUROGRAM AT L3-L4 IN THE LUMBAR SPINE THEREFORE IT CANNOT BE DETERMINED WHAT CAUSED THE HEMATOMA. THE PATIENT RECOVERED AFTER THE UNEVENTFUL SURGERY.
Patient problems: Hematoma
MAUDE report 19715635 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2022, A FEMALE PATIENT UNDERWENT A BILATERAL, 2 LEVEL MILD PROCEDURE. AN EPIDUROGRAM WAS PERFORMED AT THE LEVEL ABOVE USING A 20 GAUGE TOUHY NEEDLE. THE PATIENT WAS REPORTED TO BE OF AVERAGE SIZE AND WEIGHT AND STOPPED ALL MEDICATIONS INCLUDING ANTICOAGULANTS 7-10 DAY BEFORE THE PROCEDURE. THE PATIENT HAD ISSUES OF HYPOTENSION DURING THE PROCEDURE AND WAS TREATED WITH 20MG EPHEDRINE AND A 500ML BOLUS OF FLUID. THE PHYSICIAN HAD SOME IMAGING ISSUES DUE TO THE ANATOMY AND SEEING MARKERS THAT RESULTED IN USING MORE THAT 15 MINUTES OF FLUOROSCOPY TIME. THE PATIENT COMPLAINED OF NEUROGENIC CLAUDICATION AND BACK PAIN THE NEXT DAY ( ON (B)(6)2022). AN MRI WAS PERFORMED AND SHOWED "HETEROGENEOUS BILOBED FLUID COLLECTION IN THE DORSAL EPIDURAL SPACE ALONG THE DORSAL THECAL SAC. FLUID COLLECTION REPRESENTED POST OP HEMATOMA GIVEN CHARACTERISTICS". THE PATIENTS PAIN CONTINUED FOR 1 WEEK WHEN A SURGICAL INTERVENTION FOR THE SUBDURAL HEMATOMA WAS PERFORMED ON (B)(6) 2022. THE SURGEON DID NOT FIND A DORSAL EPIDURAL TEAR. THERE WAS AN OSTEOPHYTE NOTED ON THE FACET GOING MEDIALLY WITH A "SMALL VESSEL ABOVE IT THAT MAY HAVE BEEN INJURED DURING THE PROCEDURE" - THIS WAS THEIR BEST GUESS. T
Patient problems: Hematoma
MAUDE report 19715571 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2020 THE PATIENT UNDERWENT A L4-L5 MILD PROCEDURE. THERE WAS NO REPORTED DEVICE MALFUNCTION AND THE DEVICE PERFORMED AS INTENDED. SIX DAYS FOLLOWING THE PROCEDURE, THE PATIENT PRESENTED TO THE HOSPITAL WITH SEVERE BACK PAIN ON (B)(6) 2020 AND WAS ADMITTED TO THE HOSPITAL. AN MRI SCAN SHOWED A EPIDURAL HEMATOMA AND AS A RESULT A MULTILEVEL LAMINECTOMY WAS PERFORMED. THE PATIENT WAS REPORTED TO BE STABLE WITH NO NEUROLOGIC COMPROMISE AT THE TIME OF THE COMPLAINT INVESTIGATION. IT SHOULD BE NOTED THAT THE PATIENT HAD LEUKEMIA AND HAD BEEN TAKING ELOQUIS THAT WAS STOPPED A WEEK BEFORE THE MILD PROCEDURE.
Patient problems: Hematoma
MAUDE report 19715558 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2022 A PHYSICIAN PERFORMED A BILATERAL L3-L4, L4-L5 MILD PROCEDURE ON A 75Y/O FEMALE PATIENT WITH A HISTORY OF URINARY INCONTINENCE (2008) AND OSTEOARTHRITIS. FROM IMAGING, THE PATIENT WAS NOTED TO HAVE SEVERE STENOSIS, GRADE 1 SPONDYLOLISTHESIS, AND FACET CYST. THE CASE REPORTEDLY WENT WELL. ON (B)(6) 2022 THE PATIENT WENT TO THE EMERGENCY ROOM DUE TO URINARY INCONTINENCE AND NUMBNESS IN THE GENITAL AREA. THE PATIENT WAS PLACED IN THE CARE OF A NEUROSURGEON. AN MRI WAS PERFORMED AND THERE WAS NO EVIDENCE OF AN EPIDURAL HEMATOMA, HOWEVER THE STENOSIS WAS WORSE THAN PRIOR TO THE PROCEDURE. THE PATIENT WAS DIAGNOSED WITH CAUDA EQUINA SYNDROME, A BUCKLED LIGAMENT AT L4-L5, AND UNDERWENT EMERGENT SURGERY. THE PATIENT IS CURRENTLY STABLE, FULL STRENGTH IN LEGS, WORSENED URINARY RETENTION AND HAS IMPROVEMENT IN NUMBNESS. UPDATED REPORT FROM TREATING PHYSICIAN DATED (B)(6) 2022 STATED THAT PATIENT REMAINED IN THE HOSPITAL AND REMAIN FLAT FOR 7 DAYS BEFORE DISCHARGE FOLLOWING THE EMERGENT SURGERY BECAUSE OF A CSF LEAK. AT THE TIME OF THE COMPLAINT INVESTIGATION, THE PATIENT WAS IN REHAB DUE TO ONGOING URINARY INCONTINENCE. THE PHYSICIANS' PATIENT SELECTION AND PRE-PROCEDURE EVALU
Patient problems: Cerebrospinal Fluid Leakage; Nerve Damage
MAUDE report 19715435 ↗InjuryApr 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A MALE PATIENT HAD A L4-L5 MILD PROCEDURE PERFORMED. AN EPIDURAL WAS PERFORMED WITH THE MILD PROCEDURE AT THE L4-L5 LEVEL. NOTHING ABNORMAL WAS REPORTED REGARDING THE PROCEDURE OR THE PATIENT'S ANATOMY. ONE WEEK POST-OPERATIVE, THE PATIENT WAS EXPERIENCING INCREASED BACK AND LEG PAIN. AN MRI WAS PERFORMED, AND THE PATIENT WAS DIAGNOSED WITH A HEMATOMA NEAR THE TREATMENT AREA OF THE MILD PROCEDURE AT L5-S1. AS OF ON (B)(6) 2024, IT WAS CONFIRMED BY THE PHYSICIAN THAT THE PATIENT HAS NOT RECEIVED ANY INTERVENTIONAL TREATMENT BEYOND BED REST.
Patient problems: Hematoma; Pain
MAUDE report 19047620 ↗
Other Stryker Corporation products
- RemB7 models · 3,073 reports
- Prime Series21 models · 1,548 reports · 2 recalls
- TARGET114 models · 1,077 reports
- ProCuity17 models · 368 reports · 13 recalls
- Spetzler-Malis20 models · 7 reports · 19 recalls
- SYNCHRO 28 models · 313 reports · 9 recalls
- Neptune E-SEP19 models · 412 reports · 7 recalls
- StrykeFlow4 models · 552 reports · 4 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page