We found 2 models of Model EMV+, Comm, MRI by Zoll Medical Corporation in the FDA's device registry (GUDID). Together they are named in 79 adverse event reports in the FDA's MAUDE database, received between Aug 7, 2017 and Mar 30, 2026 — classified as 79 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (15 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Model EMV+, Comm, MRI recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 30, 2024Class IOpen, Classified
Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
Z-2130-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
Most reported Model EMV+, Comm, MRI problems
Device problems
- Mechanical Problem15
- Communication or Transmission Problem14
- No Display/Image12
- Failure of Device to Self-Test10
- Improper Flow or Infusion9
- Calibration Problem6
- Device Displays Incorrect Message6
- Failure to Power Up6
Patient problems
- No Clinical Signs, Symptoms or Conditions50
- No Patient Involvement13
- Insufficient Information11
- No Consequences Or Impact To Patient3
- No Information3
Model EMV+, Comm, MRI models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Model EMV+, Comm, MRI 799-EMVP-05-01Ventilator, continuous, facility use1 recall65 reports
- 1 recall14 reports
Recent Model EMV+, Comm, MRI adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Failure to Power Up
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24727391 ↗MalfunctionFeb 23, 2026No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24420088 ↗MalfunctionFeb 23, 2026Communication or Transmission Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "INTERNAL COMM FAILURES- 1472 AND 1474" ERROR MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413460 ↗MalfunctionJan 26, 2026Obstruction of Flow
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INTERMITTENTLY DISPLAYED A "EXHALATION SYSTEM FAILURE - 1061" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24179199 ↗MalfunctionDec 29, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23908643 ↗MalfunctionJul 22, 2025No Display/Image; No Display/Image
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO VENTILATE A PATIENT (AGE AND GENDER UNKNOWN), THE DEVICE FUNCTIONED FOR THE PATIENT, BUT THE SCREEN WENT BLACK MIDWAY THROUGH THE CALL. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22564053 ↗MalfunctionJul 9, 2025Communication or Transmission Problem; Communication or Transmission Problem
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "INTERNAL COMM FAILURE 1474" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22457721 ↗MalfunctionJan 17, 2025Mechanical Problem; Failure to Power Up; Failure of Device to Self-Test; Mechanical Problem; Failure to Power Up; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD INTERMITTENTLY NOT POWER UP AND DISPLAYED "COMPRESSOR FAULT - 1001" AND "SELF CHECK FAILURE - 1172" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21179789 ↗MalfunctionDec 3, 2024Improper Flow or Infusion; Improper Flow or Infusion
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO VENTILATE A 47-YEAR-OLD MALE PATIENT, THE DEVICE STOPPED VENTILATING. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20835801 ↗MalfunctionJun 6, 2024Noise, Audible; Noise, Audible
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, A CLICKING NOISE WAS HEARD FROM THE DEVICE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19478089 ↗MalfunctionMar 11, 2024Mechanical Problem; Mechanical Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 63-YEAR-OLD MALE PATIENT, THE DEVICE DISPLAYED A "COMPRESSOR FAULT-2001 " MESSAGE. COMPLAINANT INDICATED THAT THE CLINICIAN BAG VENTILATED TO CONTINUE TREATING THE PATIENT UNTIL THE DEVICE WAS RESTARTED AND THE ERROR WAS CLEARED. THEN THE CLINICIAN CONTINUED TO USE THE ORIGINAL DEVICE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18881634 ↗MalfunctionMar 7, 2024Communication or Transmission Problem; Improper Flow or Infusion; Communication or Transmission Problem; Improper Flow or Infusion
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE STOPPED VENTILATING AND DISPLAYED AN "INTERNAL COMM FAILURE- 1175" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857100 ↗MalfunctionFeb 26, 2024Improper Flow or Infusion; Improper Flow or Infusion
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 47-YEAR-OLD MALE PATIENT, THE DEVICE STOPPED VENTILATING. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 18777716 ↗MalfunctionFeb 1, 2024Mechanical Problem; Communication or Transmission Problem; Mechanical Problem; Communication or Transmission Problem
COMPLAINANT ALLEGED THAT DURING TRAINING BY FACILITY STAFF, THE DEVICE DISPLAYED "COMPRESSOR FAULT- 2001", "INTERNAL COMM FAILURE - 1175, "INTERNAL COMM FAILURE- 1472", AND "INTERNAL COMM FAILURE -1474" ERROR MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18622420 ↗MalfunctionJan 17, 2024Communication or Transmission Problem; Communication or Transmission Problem
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "INTERNAL COMM FAILURE - 1173" AND "INTERNAL COMM FAILURE -1473" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18525986 ↗
Other Zoll Medical Corporation products
- R SERIES94 models · 8,522 reports
- X SERIES163 models · 6,915 reports
- AED PLUS30 models · 1,966 reports · 1 recall
- Z Vent6 models · 1,204 reports · 1 recall
- Model EMV+ Vent, Army, MRI3 models · 964 reports
- E SERIES160 models · 833 reports
- ONESTEP8 models · 586 reports
- M SERIES BIPHASIC59 models · 572 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page