Nihon Kohden Corporation · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 60
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 2 models of Nihon Kohden Life Scope G7 Bedside Monitoring System by Nihon Kohden Corporation in the FDA's device registry (GUDID). Together they are named in 60 adverse event reports in the FDA's MAUDE database, received between Feb 20, 2023 and Sep 22, 2026 — classified as 60 malfunctions.
The most frequently reported specific device problem is “Output Problem” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Nihon Kohden Life Scope G7 Bedside Monitoring System problems
Device problems
- Output Problem24
- Display or Visual Feedback Problem20
- Application Program Freezes or Fails to Launch13
- Application Program Problem11
- Incorrect, Inadequate or Imprecise Result or Readings7
- Unable to Obtain Readings5
- Image Display Error/Artifact4
- Erratic or Intermittent Display3
Patient problems
- No Clinical Signs, Symptoms or Conditions60
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
Nihon Kohden Life Scope G7 Bedside Monitoring System models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Nihon Kohden Life Scope G7 Bedside Monitoring System CU-172RMonitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)59 reports
- Nihon Kohden Life Scope G7 Bedside Monitoring System JA-170PMonitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 report
Recent Nihon Kohden Life Scope G7 Bedside Monitoring System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
The Biomedical Engineer (BME) reported that a LVAD patient on a GZ transmitter was experiencing constant alarms (asystole, VTach, Pacer Non-capture, etc.) When the patient placed on a G7 Bedside Monitor (BSM), it also alarmed; however, there were more alarms when on the GZ. They also stated that the GZ had pacing spikes in the middle of the QRS complex that were not accurate and were not present on the 12-lead EKG. All types of lead placement were attempted to minimize the alarms, but the issue persisted. No reports of patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50849600 ↗MalfunctionSep 2, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50564741 ↗MalfunctionAug 27, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
The NK Employee reported that the customer was unable to get an NIBP reading on the G7 Monitor in the Operating Room (OR). The Heart Rate (HR) was also erratic and missing beats. No patient harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50484324 ↗MalfunctionAug 24, 2026Output Problem
The Biomedical Engineer (BME) reported that the NIBP readings were inconsistent on this G7 Bedside Monitor. This unit gave several readings that were all over the place. No patient harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50410357 ↗MalfunctionAug 24, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50409047 ↗MalfunctionJul 31, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50108002 ↗MalfunctionJul 22, 2026Display Difficult to Read; Display or Visual Feedback Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25906329 ↗MalfunctionJul 16, 2026Device Alarm System; Protective Measures Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25842909 ↗MalfunctionJul 16, 2026Device Alarm System; Protective Measures Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25841864 ↗MalfunctionJul 14, 2026Tactile Prompts/Feedback; No Tactile Prompts/Feedback
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25817151 ↗MalfunctionJul 1, 2026Unable to Obtain Readings; Output Problem
THE NK EMPLOYEE (CAS MANAGER) REPORTED ON BEHALF OF THE ANESTHESIOLOGIST THAT THE NIBP FUNCTION ON THE G7 MONITOR WAS ONLY INTERMITTENTLY TAKING INTERVAL READINGS DURING A CASE, REQUIRING MANUAL DEPRESSING OF THE START/STOP BUTTON TO INITIATE THE BP MEASUREMENT. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25680794 ↗MalfunctionJun 8, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT WHEN PRESSING THE NEW PATIENT BUTTON ON THE G7 MONITOR TO START THE NIBP INTERVAL FUNCTION, IT WOULD NOT ALWAYS START THE MEASUREMENT. THE CUFF WAS INFLATING, BUT THE BSMS WOULD TIME OUT WITH NO ERRORS. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25438010 ↗MalfunctionJun 8, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE NK'S EMPLOYEE (TSAM) REPORTED ON BEHALF OF THE BIOMEDICAL ENGINEER (BIOMED) THAT THE G7 MONITOR WAS GIVING ERRATIC ECG READINGS, THE NIBP FUNCTION WOULD NOT ALWAYS TAKE A PRESSURE, AND THE IBP VALUES WERE NOT TRANSFERRING TO EPIC. THEY ALSO REPORTED THAT THE ALARMS WOULD GO OFF VERY EASILY. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25437992 ↗MalfunctionMay 19, 2026Display or Visual Feedback Problem
THE CUSTOMER REPORTED THAT THE TOUCHSCREEN ON THIS UNIT IS NONRESPONSIVE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25223016 ↗MalfunctionMay 18, 2026Display or Visual Feedback Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE G7 MONITOR WAS EXPERIENCING FREEZING ISSUES. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25211370 ↗
Other Nihon Kohden Corporation products
All Nihon Kohden Corporation devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page