OneTouch Ultra2Recalls & reported problems across all models
LifeScan U.S. LLC · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1,756
- Deaths reported
- 0
- Injuries reported
- 1,424
- Recalls
- 0
We found 2 models of OneTouch Ultra2 by LifeScan U.S. LLC in the FDA's device registry (GUDID). Together they are named in 1,756 adverse event reports in the FDA's MAUDE database, received between Nov 24, 2016 and Apr 24, 2026 — classified as 1,424 injuries, 332 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (792 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported OneTouch Ultra2 problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings792
- Incorrect Or Inadequate Test Results246
- Adverse Event Without Identified Device or Use Problem182
- Device Handling Problem150
- Device Displays Incorrect Message96
- Loss of Power85
- Appropriate Device Problem Term/Code Not Available58
- Device Alarm System51
Patient problems
- Shaking/Tremors429
- Dizziness351
- No Consequences Or Impact To Patient328
- Sweating299
- Hypoglycemia162
- Confusion/ Disorientation127
- Blurred Vision112
- Dysphasia103
OneTouch Ultra2 models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- OneTouch Ultra2 021-105SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- OneTouch Ultra2 024-046System, Test, Blood Glucose, Over The Counter25 reports
Recent OneTouch Ultra2 adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryApr 24, 2026Device Alarm System
ON (B)(6) 2026, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS), ALLEGING THAT THEIR ONETOUCH ULTRA2 METER CONTINUED TO DISPLAY THE APPLY SAMPLE MESSAGE INSTEAD OF COUNTING DOWN AND PROVIDING A BLOOD GLUCOSE RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON ADDITIONAL INFORMATION OBTAINED BY THE MEDICAL SURVEILLANCE SPECIALIST AFTER REVIEWING THE CALL RECORDING, SINCE THE PATIENT DISCONNECTED THE CALL AND COULD NOT BE REACHED FOR FURTHER INFORMATION. THE PATIENT REPORTED THAT THE ALLEGED METER ISSUE BEGAN 2 WEEKS AGO AROUND 8:00 AM TO 9:00 AM. THE PATIENT MANAGES THEIR DIABETES WITH LANTUS INSULIN (20-25 UNITS) AND HUMALOG INSULIN ON A SELF-ADJUSTING DOSE. THE PATIENT DID NOT REPORT TAKING ANY ACTION REGARDING THEIR USUAL DIABETES MANAGEMENT REGIMEN IN RESPONSE TO THE ALLEGED ISSUE. DURING THE CALL, THE PATIENT REPORTED THAT AN UNSPECIFIED TIME AFTER THE ALLEGED ISSUE BEGAN, THEY SUFFERED DKA (DIABETIC KETOACIDOSIS) BECAUSE THEY COULD NOT CHECK THEIR BLOOD GLUCOSE DUE TO THE REPORTED ISSUE AND WERE HOSPITALIZED. AT THE TIME OF TROUBLESHOOTING, THE PATIENT DECLINED TO PERFORM A TEST. THE ALLEGED ISSUE REMAINED UNRESOLVED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMS THEY WE
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 24990787 ↗InjuryApr 24, 2026Device Alarm System
ON (B)(6) 2026, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS), ALLEGING THAT THEIR ONETOUCH ULTRA2 METER CONTINUED TO DISPLAY THE APPLY SAMPLE MESSAGE INSTEAD OF COUNTING DOWN AND PROVIDING A BLOOD GLUCOSE RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON ADDITIONAL INFORMATION OBTAINED BY THE MEDICAL SURVEILLANCE SPECIALIST AFTER REVIEWING THE CALL RECORDING, SINCE THE PATIENT DISCONNECTED THE CALL AND COULD NOT BE REACHED FOR FURTHER INFORMATION. THE PATIENT REPORTED THAT THE ALLEGED METER ISSUE BEGAN 2 WEEKS AGO AROUND 8:00 AM TO 9:00 AM. THE PATIENT MANAGES THEIR DIABETES WITH LANTUS INSULIN (20-25 UNITS) AND HUMALOG INSULIN ON A SELF-ADJUSTING DOSE. THE PATIENT DID NOT REPORT TAKING ANY ACTION REGARDING THEIR USUAL DIABETES MANAGEMENT REGIMEN IN RESPONSE TO THE ALLEGED ISSUE. DURING THE CALL, THE PATIENT REPORTED THAT AN UNSPECIFIED TIME AFTER THE ALLEGED ISSUE BEGAN, THEY SUFFERED DKA (DIABETIC KETOACIDOSIS) BECAUSE THEY COULD NOT CHECK THEIR BLOOD GLUCOSE DUE TO THE REPORTED ISSUE AND WERE HOSPITALIZED. AT THE TIME OF TROUBLESHOOTING, THE PATIENT DECLINED TO PERFORM A TEST. THE ALLEGED ISSUE REMAINED UNRESOLVED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMS THEY WE
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 24990780 ↗InjuryNov 6, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT HER ONETOUCH ULTRA2 METER WAS REVERTING TO THE SET-UP MODE WHEN ATTEMPTING TO TEST HER BLOOD GLUCOSE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT ON (B)(6) 2025, BEFORE 12:00 PM, SHE BEGAN TO FEEL BAD AND ATTEMPTED TO TEST HER BLOOD GLUCOSE BUT WAS UNABLE TO OBTAIN A BLOOD GLUCOSE READING DUE TO THE REPORTED ISSUE. DESPITE THE ALLEGED ISSUE, THE PATIENT TOOK ACTION TO INCREASE HER GLUCOSE LEVEL BY EATING A CANDY BUT STARTED FEELING ¿VERY DIZZY, COLD AND BECAME UNCONSCIOUS¿ AT AROUND 12:00 PM. THE PATIENT REPORTED THAT AT AROUND 12:00 PM, SHE WAS TAKEN TO THE HOSPITAL BY AMBULANCE WHERE SHE WAS ADMITTED BECAUSE HER BLOOD GLUCOSE LEVEL WAS ¿TOO LOW¿ (NO SPECIFIC READING IN HOSPITAL WAS PROVIDED) AND SHE WAS IN "COMA". THE PATIENT STATED THAT SHE WAS ADMINISTERED INTRAVENOUS (IV) GLUCOSE AND THAT SHE STAYED IN HOSPITAL FOR 2-3 DAYS. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT THE SUBJECT METER WAS NOT BEING USED FOR THE FIRST TIME. THE CCA EDUCATED THE PATIENT ON THE CORRECT TESTING PROCEDURE AND WALKED THROUGH A TEST AND TH
Patient problems: Alteration in Body Temperature; Coma; Dizziness; Hypoglycemia
MAUDE report 23493380 ↗InjuryNov 6, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
ON NOVEMBER 04, 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT HER ONETOUCH ULTRA 2 METER WAS REVERTING TO THE SET-UP MODE WHEN ATTEMPTING TO TEST HER BLOOD GLUCOSE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT ON (B)(6) 2025, BEFORE 12:00 PM, SHE BEGAN TO FEEL BAD AND ATTEMPTED TO TEST HER BLOOD GLUCOSE BUT WAS UNABLE TO OBTAIN A BLOOD GLUCOSE READING DUE TO THE REPORTED ISSUE. DESPITE THE ALLEGED ISSUE, THE PATIENT TOOK ACTION TO INCREASE HER GLUCOSE LEVEL BY EATING A CANDY BUT STARTED FEELING ¿VERY DIZZY, COLD AND BECAME UNCONSCIOUS¿ AT AROUND 12:00 PM. THE PATIENT REPORTED THAT AT AROUND 12;00 PM, SHE WAS TAKEN TO THE HOSPITAL BY AMBULANCE WHERE SHE WAS ADMITTED BECAUSE HER BLOOD GLUCOSE LEVEL WAS ¿TOO LOW¿ (NO SPECIFIC READING IN HOSPITAL WAS PROVIDED) AND SHE WAS IN "COMA". THE PATIENT STATED THAT SHE WAS ADMINISTERED INTRAVENOUS (IV) GLUCOSE AND THAT SHE STAYED IN HOSPITAL FOR 2-3 DAYS. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT THE SUBJECT METER WAS NOT BEING USED FOR THE FIRST TIME. THE CCA EDUCATED THE PATIENT ON THE CORRECT TESTING PROCEDURE AND WALKED THROUGH A TEST
Patient problems: Alteration in Body Temperature; Coma; Dizziness; Hypoglycemia
MAUDE report 23493379 ↗InjuryOct 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, A REPORTER FOR THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THE PATIENT¿S ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO THEIR FEELINGS AND COMPARED TO ANOTHER METER (UNSPECIFIED DOCTOR¿S METER). THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION, SINCE THE REPORTER DISCONTINUED THE INTERACTION COULD NOT BE REACHED BY PHONE FOR ADDITIONAL INFORMATION. THE REPORTER INDICATED THAT THE PATIENT HAD JUST PURCHASED THE SYSTEM KIT 1 WEEK AGO AND ON AN UNSPECIFIED DATE OBTAINED HIGH BLOOD GLUCOSE RESULTS OF ¿585, 535 AND 600 MG/DL¿ WITH THE SUBJECT METER DURING A LOW BLOOD GLUCOSE EPISODE. IT WAS NOT REPORTED IF THE PATIENT TAKES ANY MEDICATION TO MANAGE THEIR DIABETES OR IF THEY TOOK ANY ACTION REGARDING THEIR USUAL DIABETES MANAGEMENT DUE TO THE REPORTED ISSUE. IT WAS NOT REPORTED WHAT SYMPTOMS THE PATIENT EXPERIENCED. HOWEVER, THE REPORTER CLAIMED THE PATIENT WENT TO AN EMERGENCY ROOM AS A RESULT OF THE ALLEGED ISSUE AND INDICATED THAT THE PATIENT RECEIVED A BLOOD GLUCOSE RESULT OF ¿51 MG/DL¿ ON THE DOCTOR¿S METER. THE TIME DIFFERENCE BETWEEN THE TESTS WAS NOT REPORTED. IT WAS NOT REPORTED WHAT TYPE O
Patient problems: Hypoglycemia
MAUDE report 23345699 ↗InjuryOct 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, A REPORTER FOR THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THE PATIENT¿S ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO THEIR FEELINGS AND COMPARED TO ANOTHER METER (UNSPECIFIED DOCTOR¿S METER). THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION, SINCE THE REPORTER DISCONTINUED THE INTERACTION COULD NOT BE REACHED BY PHONE FOR ADDITIONAL INFORMATION. THE REPORTER INDICATED THAT THE PATIENT HAD JUST PURCHASED THE SYSTEM KIT 1 WEEK AGO AND ON AN UNSPECIFIED DATE OBTAINED HIGH BLOOD GLUCOSE RESULTS OF ¿585, 535 AND 600 MG/DL¿ WITH THE SUBJECT METER DURING A LOW BLOOD GLUCOSE EPISODE. IT WAS NOT REPORTED IF THE PATIENT TAKES ANY MEDICATION TO MANAGE THEIR DIABETES OR IF THEY TOOK ANY ACTION REGARDING THEIR USUAL DIABETES MANAGEMENT DUE TO THE REPORTED ISSUE. IT WAS NOT REPORTED WHAT SYMPTOMS THE PATIENT EXPERIENCED. HOWEVER, THE REPORTER CLAIMED THE PATIENT WENT TO AN EMERGENCY ROOM AS A RESULT OF THE ALLEGED ISSUE AND INDICATED THAT THE PATIENT RECEIVED A BLOOD GLUCOSE RESULT OF ¿51 MG/DL¿ ON THE DOCTOR¿S METER. THE TIME DIFFERENCE BETWEEN THE TESTS WAS NOT REPORTED. IT WAS NOT REPORTED WHAT TYPE O
Patient problems: Hypoglycemia
MAUDE report 23345696 ↗InjurySep 17, 2025Device Handling Problem; Device Handling Problem
ON AUGUST 29, 2025, THE LAY USER/PATIENT¿S SPOUSE CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THE PATIENT¿S ONETOUCH ULTRA2 METER READ INACCURATELY LOW WHEN COMPARED TO HER FEELINGS AND/OR NORMAL READINGS AND TO A HOSPITAL METER. THE COMPLAINT WAS CLASSIFIED BASED ON ADDITIONAL INFORMATION OBTAINED WHEN THE MEDICAL SURVEILLANCE SPECIALIST REVIEWED THE CALL RECORDING. THE REPORTER STATED THAT THE ALLEGED METER INACCURACY BEGAN ON AUGUST 22, 2025, AT 9:00 AM. THE REPORTER CLAIMED HE MEASURED THE PATIENT¿S BLOOD GLUCOSE WITH THE SUBJECT METER AND RECEIVED LOW BLOOD GLUCOSE READINGS OF ¿85 AND 230 MG/DL¿. THE REPORTER STATED THAT THE PATIENT¿S DIABETES IS MANAGED WITH ORAL MEDICATION (JANUVIA 100 MG, 1 PILL DAILY) AND THAT NO CHANGES WERE MADE TO HER USUAL DIABETES MANAGEMENT REGIMEN IN RESPONSE TO THE ALLEGED ISSUE. THE REPORTER CLAIMED THAT ON AUGUST 22, 2025, AN UNSPECIFIED TIME AFTER THE ALLEGED ISSUE BEGAN, THE PATIENT BECAME ¿PALE AND PASSED OUT¿. THE REPORTER STATED THAT HE CALLED EMERGENCY MEDICAL SERVICES, AND THAT THE PATIENT WAS TAKEN TO HOSPITAL BY AMBULANCE WHERE SHE RECEIVED A BLOOD GLUCOSE RESULT OF ¿OVER 400 MG/DL¿ ON ARRIVAL ON THE HOSPITAL METER. THE REPORTE
Patient problems: Hyperglycemia; Pallor; Syncope/Fainting
MAUDE report 23080543 ↗InjurySep 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT HER ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO HER FEELINGS AND/OR NORMAL READINGS AND COMPARED TO OTHER METERS, AND COMPARED TO AN A1C RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED METER INACCURACY BEGAN AROUND (B)(6) 2024. AT UNSPECIFIED TIMES, THE PATIENT REPORTED OBTAINING WHAT SHE FELT WERE INACCURATE HIGH BLOOD GLUCOSE READINGS OF ¿380, 387 AND 390 MG/DL¿ WITH THE SUBJECT METER. THE PATIENT ALSO REPORTED OBTAINING A BLOOD GLUCOSE READING OF ¿387 MG/DL¿ WITH THE SUBJECT METER COMPARED TO A READING OF ¿114 MG/DL¿ ON A DOCTOR¿S METER, AND ALSO A READING OF ¿283 MG/DL¿ WITH THE SUBJECT METER COMPARED TO READINGS OF ¿112, 104 AND 91 MG/DL¿ ON A FREESTYLE LITE METER. FURTHERMORE, THE PATIENT REPORTED OBTAINING INACCURATE HIGH BLOOD GLUCOSE READINGS OF ¿380, 387, 390, 208 AND 238 MG/DL¿ WITH THE SUBJECT METER COMPARED TO AN A1C RESULT OF 5.2. AT THE TIME OF THE ALLEGED ISSUE, THE PATIENT WAS TAKING A COMBINATION OF DIABETES MEDICATIONS (METFORMIN AND OZEMPIC). BECAUSE HER READINGS WERE NOT GOING DO
Patient problems: Abdominal Cramps; Diarrhea; Pancreatitis; Vomiting
MAUDE report 22986870 ↗InjurySep 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT HER ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO HER FEELINGS AND/OR NORMAL READINGS AND COMPARED TO OTHER METERS, AND COMPARED TO AN A1C RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED METER INACCURACY BEGAN AROUND (B)(6) 2024. AT UNSPECIFIED TIMES, THE PATIENT REPORTED OBTAINING WHAT SHE FELT WERE INACCURATE HIGH BLOOD GLUCOSE READINGS OF ¿380, 387 AND 390 MG/DL¿ WITH THE SUBJECT METER. THE PATIENT ALSO REPORTED OBTAINING A BLOOD GLUCOSE READING OF ¿387 MG/DL¿ WITH THE SUBJECT METER COMPARED TO A READING OF ¿114 MG/DL¿ ON A DOCTOR¿S METER, AND ALSO A READING OF ¿283 MG/DL¿ WITH THE SUBJECT METER COMPARED TO READINGS OF ¿112, 104 AND 91 MG/DL¿ ON A FREESTYLE LITE METER. FURTHERMORE, THE PATIENT REPORTED OBTAINING INACCURATE HIGH BLOOD GLUCOSE READINGS OF ¿380, 387, 390, 208 AND 238 MG/DL¿ WITH THE SUBJECT METER COMPARED TO AN A1C RESULT OF 5.2. AT THE TIME OF THE ALLEGED ISSUE, THE PATIENT WAS TAKING A COMBINATION OF DIABETES MEDICATIONS (METFORMIN AND OZEMPIC). BECAUSE HER READINGS WERE NOT GOING DO
Patient problems: Abdominal Cramps; Diarrhea; Pancreatitis; Vomiting
MAUDE report 22986867 ↗InjuryJul 17, 2025Device Alarm System; Device Alarm System
ON (B)(6) 2025, THE LAY USER/PATIENT¿S SPOUSE CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THE PATIENT¿S ONETOUCH ULTRA2 METER CONTINUED TO DISPLAY THE APPLY CONTROL MESSAGE INSTEAD OF COUNTING DOWN AND PROVIDING A CONTROL SOLUTION RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION, SINCE THE PATIENT WAS UNABLE TO BE REACHED BY PHONE FOR ADDITIONAL INFORMATION. THE REPORTER STATED THAT THE ALLEGED ISSUE BEGAN AROUND 2:00 PM-3:00 PM ON (B)(6) 2025. THE REPORTER DENIED THE PATIENT TOOK ANY ACTION REGARDING HIS USUAL DIABETES MANAGEMENT DUE TO THE ALLEGED ISSUE. THE REPORTER MENTIONED THAT AT AN UNSPECIFIED DATE AND TIME, THE PATIENT HAD ¿PASSED OUT AND WAS UNRESPONSIVE¿ AS A RESULT OF THE ALLEGED ISSUE AND WAS TREATED WITH GLUCOSE GEL BY EMERGENCY MEDICAL SERVICES. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS NOT BEING USED FOR THE FIRST TIME. THE CCA DETERMINED THE CORRECT STRIPS WERE BEING USED FOR TESTING AND THE CORRECT TESTING PROCESS WAS BEING FOLLOWED. THE CCA WALKED THROUGH RESOLVING THE ISSUE HOWEVER THE ISSUE COULD NOT BE RESOLVED. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING RE
Patient problems: Loss of consciousness
MAUDE report 22538470 ↗InjuryJul 17, 2025Device Alarm System; Device Alarm System
ON (B)(6) 2025, THE LAY USER/PATIENT¿S SPOUSE CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THE PATIENT¿S ONETOUCH ULTRA2 METER CONTINUED TO DISPLAY THE APPLY CONTROL MESSAGE INSTEAD OF COUNTING DOWN AND PROVIDING A CONTROL SOLUTION RESULT. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION, SINCE THE PATIENT WAS UNABLE TO BE REACHED BY PHONE FOR ADDITIONAL INFORMATION. THE REPORTER STATED THAT THE ALLEGED ISSUE BEGAN AROUND 2:00 PM-3:00 PM ON (B)(6) 2025 OR (B)(6) 2025. THE REPORTER DENIED THE PATIENT TOOK ANY ACTION REGARDING HIS USUAL DIABETES MANAGEMENT DUE TO THE ALLEGED ISSUE. THE REPORTER MENTIONED THAT AT AN UNSPECIFIED DATE AND TIME, THE PATIENT HAD ¿PASSED OUT AND WAS UNRESPONSIVE¿ AS A RESULT OF THE ALLEGED ISSUE AND WAS TREATED WITH GLUCOSE GEL BY EMERGENCY MEDICAL SERVICES. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS NOT BEING USED FOR THE FIRST TIME. THE CCA DETERMINED THE CORRECT STRIPS WERE BEING USED FOR TESTING AND THE CORRECT TESTING PROCESS WAS BEING FOLLOWED. THE CCA WALKED THROUGH RESOLVING THE ISSUE HOWEVER THE ISSUE COULD NOT BE RESOLVED. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLA
Patient problems: Loss of consciousness
MAUDE report 22538469 ↗InjuryJun 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT THEIR ONETOUCH ULTRA 2 METER WAS READING INACCURATELY HIGH COMPARED TO THEIR FEELINGS/NORMAL RESULTS. THIS COMPLAINT WAS CLASSIFIED BASED ON INFORMATION OBTAINED BY THE CUSTOMER CARE AGENT (CCA) DURING THE INITIAL CALL. THE PATIENT STATED THAT THE ALLEGED ISSUE BEGAN AT 02:00 PM ON (B)(6) 2025. THE PATIENT CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿201 MG/DL AND 193 MG/DL¿ WITH THE SUBJECT DEVICE WHICH THEY FELT WERE INACCURATELY HIGH COMPARED TO THEIR FEELINGS AND/OR NORMAL READINGS. THE PATIENT MANAGES THEIR DIABETES WITH SELF-ADJUSTED INSULIN (NPH INSULIN, 8 UNITS IN THE MORNING AND NOVOLIN R, 8 UNITS IN THE MORNING) AND THEY DENIED MAKING ANY CHANGES TO THEIR USUAL DIABETES MANAGEMENT REGIMENT IN RESPONSE TO THE ALLEGED ISSUE, REPORTING THAT THEY CONTINUED WITH THEIR USUAL ACTIVITIES. THE PATIENT REPORTED THAT AT AROUND 02:00 PM ON (B)(6) 2025 THAT THEY STARTED FEELING ¿CONFUSED¿, STATING THEY WERE ¿NOT PROCESSING EVERYTHING QUITE RIGHT¿ AND EXPERIENCED ¿TYPICAL LOW BLOOD SUGAR SYMPTOMS¿. THE PATIENT REPORTED THAT AN AMBULANCE WAS CALLED AND THAT THEY RECEIVED A GLUCOSE READING OF ¿48 MG/DL¿ ON A
Patient problems: Confusion/ Disorientation; Hypoglycemia
MAUDE report 22291618 ↗InjuryJun 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT THEIR ONETOUCH ULTRA 2 METER WAS READING INACCURATELY HIGH COMPARED TO THEIR FEELINGS/NORMAL RESULTS. THIS COMPLAINT WAS CLASSIFIED BASED ON INFORMATION OBTAINED BY THE CUSTOMER CARE AGENT (CCA) DURING THE INITIAL CALL. THE PATIENT STATED THAT THE ALLEGED ISSUE BEGAN AT 02:00 PM ON (B)(6) 2025. THE PATIENT CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿201 MG/DL AND 193 MG/DL¿ WITH THE SUBJECT DEVICE WHICH THEY FELT WERE INACCURATELY HIGH COMPARED TO THEIR FEELINGS AND/OR NORMAL READINGS. THE PATIENT MANAGES THEIR DIABETES WITH SELF-ADJUSTED INSULIN (NPH INSULIN, 8 UNITS IN THE MORNING AND NOVOLIN R, 8 UNITS IN THE MORNING) AND THEY DENIED MAKING ANY CHANGES TO THEIR USUAL DIABETES MANAGEMENT REGIMENT IN RESPONSE TO THE ALLEGED ISSUE, REPORTING THAT THEY CONTINUED WITH THEIR USUAL ACTIVITIES. THE PATIENT REPORTED THAT AT AROUND 02:00 PM ON (B)(6) 2025 THAT THEY STARTED FEELING ¿CONFUSED¿, STATING THEY WERE ¿NOT PROCESSING EVERYTHING QUITE RIGHT¿ AND EXPERIENCED ¿TYPICAL LOW BLOOD SUGAR SYMPTOMS¿. THE PATIENT REPORTED THAT AN AMBULANCE WAS CALLED AND THAT THEY RECEIVED A GLUCOSE READING OF ¿48 MG/DL¿ ON A
Patient problems: Confusion/ Disorientation; Hypoglycemia
MAUDE report 22289017 ↗InjuryJun 13, 2025Power Problem; Power Problem
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THEIR ONETOUCH ULTRA2 METER POWERS OFF DURING USE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED ISSUE BEGAN AT AN UNSPECIFIED DATE AND TIME IN (B)(6) 2025. THE PATIENT STATED THAT BECAUSE THEY WERE UNABLE TO OBTAIN BLOOD GLUCOSE READINGS, THEY STOPPED USING THE METER. THE PATIENT MANAGES THEIR DIABETES WITH ORAL MEDICATION (JARDIANCE) WITH DIET AND/OR EXERCISE. THE PATIENT DENIED MAKING ANY CHANGES TO THEIR USUAL DIABETES MANAGEMENT REGIMEN AS A RESULT OF THE ALLEGED ISSUE. THE PATIENT REPORTED EXPERIENCING SYMPTOMS OF FEELING ¿VERY LIGHTHEADED AND FALLING DOWN¿ FOR THE PAST MONTH, SINCE THE ALLEGED ISSUE BEGAN. DUE TO THE SYMPTOMS, THE PATIENT STATED THAT ON (B)(6) 2025, THEY WENT TO THE HOSPITAL WHERE THEY WERE FOUND TO HAVE "VERY HIGH BLOOD GLUCOSE LEVELS". THE PATIENT STATED THAT THEY WERE HOSPITALIZED FOR 3 DAYS AND ADMINISTERED INSULIN EACH DAY (DIFFERENT DOSES) TO LOWER THEIR GLUCOSE LEVELS. THE BLOOD GLUCOSE TEST RESULTS AT THE HOSPITAL WERE NOT PROVIDED. THE PATIENT WAS DISCHARGED ON (B)(6) 2025 AND ADVISED T
Patient problems: Dizziness; Fall; Hyperglycemia
MAUDE report 22207863 ↗InjuryJun 13, 2025Power Problem; Power Problem
ON (B)(6) 2025, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) UNITED STATES, ALLEGING THAT THEIR ONETOUCH ULTRA2 METER POWERS OFF DURING USE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED ISSUE BEGAN AT AN UNSPECIFIED DATE AND TIME IN (B)(6) 2025. THE PATIENT STATED THAT BECAUSE THEY WERE UNABLE TO OBTAIN BLOOD GLUCOSE READINGS, THEY STOPPED USING THE METER. THE PATIENT MANAGES THEIR DIABETES WITH ORAL MEDICATION (JARDIANCE) WITH DIET AND/OR EXERCISE. THE PATIENT DENIED MAKING ANY CHANGES TO THEIR USUAL DIABETES MANAGEMENT REGIMEN AS A RESULT OF THE ALLEGED ISSUE. THE PATIENT REPORTED EXPERIENCING SYMPTOMS OF FEELING ¿VERY LIGHTHEADED AND FALLING DOWN¿ FOR THE PAST MONTH, SINCE THE ALLEGED ISSUE BEGAN. DUE TO THE SYMPTOMS, THE PATIENT STATED THAT ON (B)(6) 2025, THEY WENT TO THE HOSPITAL WHERE THEY WERE FOUND TO HAVE "VERY HIGH BLOOD GLUCOSE LEVELS". THE PATIENT STATED THAT THEY WERE HOSPITALIZED FOR 3 DAYS AND ADMINISTERED INSULIN EACH DAY (DIFFERENT DOSES) TO LOWER THEIR GLUCOSE LEVELS. THE BLOOD GLUCOSE TEST RESULTS AT THE HOSPITAL WERE NOT PROVIDED. THE PATIENT WAS DISCHARGED ON (B)(6) 2025, AND ADVISED
Patient problems: Dizziness; Fall; Hyperglycemia
MAUDE report 22207862 ↗
Other LifeScan U.S. LLC products
All LifeScan U.S. LLC devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page