ORBIT GALAXY XTRASOFTRecalls & reported problems across all models
Medos International Sàrl · 18 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 320
- Deaths reported
- 2
- Injuries reported
- 20
- Recalls
- 0
We found 18 models of ORBIT GALAXY XTRASOFT by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 320 adverse event reports in the FDA's MAUDE database, received between Sep 14, 2016 and Sep 9, 2026 — classified as 2 deaths, 20 injuries, 298 malfunctions.
The most frequently reported specific device problem is “Stretched” (119 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported ORBIT GALAXY XTRASOFT problems
Device problems
- Stretched119
- Failure to Advance54
- Separation Failure42
- Device Difficult to Setup or Prepare32
- Difficult to Advance25
- Positioning Problem23
- Premature Separation19
- Positioning Failure16
Patient problems
- No Clinical Signs, Symptoms or Conditions194
- No Consequences Or Impact To Patient117
- Aneurysm5
- Intracranial Hemorrhage2
- Ruptured Aneurysm2
- No Known Impact Or Consequence To Patient1
- Obstruction/Occlusion1
- Stroke/CVA1
ORBIT GALAXY XTRASOFT models
18 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ORBIT GALAXY XTRASOFT 640CX0306Device, Vascular, for Promoting Embolization41 reports
- ORBIT GALAXY XTRASOFT 640CX2505Device, Vascular, for Promoting Embolization35 reports
- ORBIT GALAXY XTRASOFT 640CX0308DEVICE, NEUROVASCULAR EMBOLIZATION31 reports
- ORBIT GALAXY XTRASOFT 640CX0408DEVICE, NEUROVASCULAR EMBOLIZATION30 reports
- ORBIT GALAXY XTRASOFT 640CX3575Device, Vascular, for Promoting Embolization1 death25 reports
- ORBIT GALAXY XTRASOFT 640CX0202DEVICE, NEUROVASCULAR EMBOLIZATION21 reports
- ORBIT GALAXY XTRASOFT 640CX2535Device, Vascular, for Promoting Embolization19 reports
- ORBIT GALAXY XTRASOFT 640HX0206DEVICE, NEUROVASCULAR EMBOLIZATION19 reports
- ORBIT GALAXY XTRASOFT 640CX3509Device, Vascular, for Promoting Embolization18 reports
- ORBIT GALAXY XTRASOFT 640HX0204Device, Vascular, for Promoting Embolization17 reports
- ORBIT GALAXY XTRASOFT 640CX0304Device, Vascular, for Promoting Embolization15 reports
- ORBIT GALAXY XTRASOFT 640CX0410Device, Vascular, for Promoting Embolization1 death12 reports
- ORBIT GALAXY XTRASOFT 640CX3505DEVICE, NEUROVASCULAR EMBOLIZATION12 reports
- ORBIT GALAXY XTRASOFT 640CX0406Device, Vascular, for Promoting Embolization9 reports
- ORBIT GALAXY XTRASOFT 640HX0208Device, Vascular, for Promoting Embolization7 reports
- ORBIT GALAXY XTRASOFT 640CX2525Device, Vascular, for Promoting Embolization5 reports
- ORBIT GALAXY XTRASOFT 640CX0201DEVICE, NEUROVASCULAR EMBOLIZATION3 reports
- ORBIT GALAXY XTRASOFT 640CX0404Device, Vascular, for Promoting Embolization1 report
Recent ORBIT GALAXY XTRASOFT adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Separation Failure
A healthcare professional reported that during an aneurysm embolization, an unspecified microcatheter was in place and an ORBIT GALAXY XTRASOFT (product code: 640CX0404, lot number: 31471128) coil was delivered to target site. The coil was packed in the aneurysm and tried to detach, but the coil was unable to detach. The doctor only retracted the coil alone and switched to a new one to complete the surgery. The microcatheter was not replaced. There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50694662 ↗MalfunctionJun 19, 2026Failure to Advance
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION, THE PHYSICIAN DELIVERED AN ORBIT GALAXY XTRASOFT COIL (PRODUCT CODE: 640CX0408, LOT NUMBER: 31763941) IN A COMPETITOR¿S MICROCATHETER. THE COIL BECAME WAS IMPEDED IN THE PROXIMAL END OF THE MICROCATHETER AND COULD NOT ADVANCE ANY MORE. THE PHYSICIAN ONLY RETRACTED THE COIL ALONE AND SWITCHED TO A COMPETITOR¿S COIL TO COMPLETE THE PROCEDURE. THE MICROCATHETERS WERE NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT NO PATIENT INJURY WAS ALLEGED IN ASSOCIATION WITH THE REPORTED EVENT. THE MICROCATHETER USED WAS A PROWLER SELECT PLUS 150/5CM (PRODUCT CODE: 606S255X). NO OTHER DEVICES WERE USED IN CONJUNCTION WITH THE MICROCATHETER BEFORE OR AFTER THE REPORTED RESISTANCE. THE TARGET VESSEL WAS THE M1 SEGMENT OF THE MIDDLE CEREBRAL ARTERY, AND NO RELEVANT ANATOMICAL FACTORS WERE REPORTED. THE DEVICE DID NOT APPEAR DAMAGED AT ANY TIME DURING THE PROCEDURE. CONTINUOUS FLUSH WAS MAINTAINED, AND IT WAS CONFIRMED THAT THE INTRODUCER TIP WAS PROPERLY INSTALLED INTO THE MICROCATHETER HUB AND SECURED WITH THE RHV DURING DEVICE ADVANCEMENT. BASED ON THE PRODUCT ANALYSIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25563659 ↗MalfunctionJun 12, 2026Material Twisted/Bent
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE COIL WERE ON THE 3.5MM X 7.5CM ORBIT GALAXY COMPLEX XTRASOFT COIL (640CX3575 / 31586977) WAS KINKED WHEN THE PHYSICIAN TRIED TO INTRODUCE THE COIL INTO THE MICROCATHETER. THE COIL WAS SUCCESSFULLY REMOVED BEFORE IT WAS INTO A PATIENT'S BODY. THE PATIENT REMAINS ASYMPTOMATIC. THE PROCEDURE WAS SUCCESSFULLY DONE. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE PROCEDURE AND THE REPORTED DEVICE ISSUE WERE UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED AT A LATER DATE, THIS FILE WILL BE UPDATED ACCORDINGLY. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION AND ANALYSIS. THE PRODUCT INVESTIGATION COMPLETED ON 12-JUN-2026. BASED ON THE COMPLETED PRODUCT INVESTIGATION, THIS EVENT HAS BEEN DETERMINED TO BE USFDA REPORTABLE UNDER 21 CFR 803 WITH A CLASSIFICATION OF ¿MALFUNCTION.¿.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25490662 ↗MalfunctionJun 8, 2026Separation Failure
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, AN ORBIT GALAXY XTRASOFT COIL (PRODUCT CODE: 640CX0408, LOT NUMBER: 31606037) WAS PACKED IN THE ANEURYSM AND TRIED TO DETACH, BUT THE COIL WAS UNABLE TO DETACH. DOCTOR ONLY RETRACTED THE COIL ALONE AND SWITCHED A NEW COIL TO COMPLETE THE SURGERY. THE COIL MICROCATHETER (COMPETITOR¿S BRAND) WAS NOT REPLACED. ADDITIONAL EVENT INFORMATION RECEIVED ON 05-JUN-2026 INDICATED THAT A PRE-DEPLOYMENT ELECTRICAL TESTING WAS PERFORMED. THE SAME DCB AND CONTROL CABLE WERE USED SUCCESSFULLY WITH SUBSEQUENT COILS. THE COIL WAS STILL ATTACHED TO THE COIL DELIVERY SYSTEM WHEN REMOVED FROM THE PATIENT. THE EMBOLIC COIL DID NOT STRETCH OR BECAME DAMAGED WHEN REMOVED FROM THE PATIENT. THERE WAS NO PROCEDURE PROLONGATION/DELAY DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25430449 ↗MalfunctionJun 4, 2026Stretched
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION TO A POSTERIOR COMMUNICATING ARTERY ANEURYSM, AN ENTERPRISE2 4MMX16MM VASCULAR RECONSTRUCTION DEVICE (PRODUCT CODE: ENCR401612, LOT NUMBER: 9854670) WAS IMPEDED IN DISTAL END OF A PROWLER SELECT PLUS 150/5CM MICROCATHETER (PRODUCT CODE: 606S255X, LOT NUMBER: 31700153) AND COULD NOT PASS THROUGH THE MICROCATHETER (MC). THE DOCTOR DIRECTLY REMOVED THE STENT AND MICROCATHETER TOGETHER FROM THE PATIENT AND SWITCHED A NEW STENT AND A NEW MICROCATHETER TO CONTINUE THE SURGERY. WHEN PACKING COIL IN THE ANEURYSM, A 3.5X9 ORBIT GALAXY XTRASOFT (PRODUCT CODE: 640CX3509, LOT NUMBER: 31651282) WAS UNABLE TO ROTATE RANDOMLY, THE DOCTOR ONLY RETRACTED THE COIL, IT WAS FOUND THAT THE COIL HAD BEEN UNRAVELED/STRETCHED. THE DOCTOR SWITCHED TO A SECOND COIL, A 3.5X5 ORBIT GALAXY XTRASOFT (PRODUCT CODE: 640CX3505, LOT NUMBER: 31527069) WHICH BECAME IMPEDED IN THE PROXIMAL END OF THE COIL MICROCATHETER (COMPETITOR¿S BRAND) AND COULD NOT ADVANCE ANY MORE. THE DOCTOR ONLY RETRACTED THE SECOND COIL AND FOUND THE DELIVERY WIRE OF THE SECOND COIL WAS KINKED/BENT. THE DOCTOR SWITCHED A THIRD COIL (COMPETITOR¿S BRAND) TO COMPLETE THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25392878 ↗DeathMay 28, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE ON (B)(6) 2026, TARGETING AN ANEURYSM ON THE ANTERIOR COMMUNICATING ARTERY, THE FIRST COIL, A 4MM X 10CM ORBIT GALAXY COMPLEX XTRASOFT (640CX0410 / 31770622) AND THE SECOND COIL, A 3.5MM X 7.5CM ORBIT GALAXY COMPLEX XTRASOFT (640CX3575 / 31770623) WERE PACKED WITHIN THE ANEURYSM. SUBSEQUENTLY, WHILE PACKING THE ANEURYSM WITH COILS FROM OTHER BRANDS, THE ANEURYSM RUPTURED AND BLEEDING WAS OBSERVED. THE PHYSICIAN IMPLANTED TWO STENTS (FROM OTHER MANUFACTURERS). A CRANIOTOMY WAS PERFORMED THEREAFTER. ON (B)(6) 2026, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE ANEURYSM RUPTURED AFTER THE FIRST AND SECOND COIL WERE IMPLANTED AND DURING THE PACKING OF THE ANEURYSM WITH NON-CERENOVUS COILS. THE INFORMATIONAL ALSO INDICATED THAT THE PATIENT IS DECEASED; THE EXACT DATE OF DEATH IS UNKNOWN. THE PHYSICIAN CONSIDERED THE ANEURYSM RUPTURE TO BE UNRELATED TO CERENOVUS COILS. THERE WAS NO DELAY TO THE ENDOVASCULAR EMBOLIZATION PROCEDURE.
Patient problems: Intracranial Hemorrhage; Ruptured Aneurysm
MAUDE report 25321571 ↗DeathMay 28, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE ON (B)(6) 2026, TARGETING AN ANEURYSM ON THE ANTERIOR COMMUNICATING ARTERY, THE FIRST COIL, A 4MM X 10CM ORBIT GALAXY COMPLEX XTRASOFT (640CX0410 / 31770622) AND THE SECOND COIL, A 3.5MM X 7.5CM ORBIT GALAXY COMPLEX XTRASOFT (640CX3575 / 31770623) WERE PACKED WITHIN THE ANEURYSM. SUBSEQUENTLY, WHILE PACKING THE ANEURYSM WITH COILS FROM OTHER BRANDS, THE ANEURYSM RUPTURED AND BLEEDING WAS OBSERVED. THE PHYSICIAN IMPLANTED TWO STENTS (FROM OTHER MANUFACTURERS). A CRANIOTOMY WAS PERFORMED THEREAFTER. ON 27-MAY-2026, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE ANEURYSM RUPTURED AFTER THE FIRST AND SECOND COIL WERE IMPLANTED AND DURING THE PACKING OF THE ANEURYSM WITH NON-CERENOVUS COILS. THE INFORMATIONAL ALSO INDICATED THAT THE PATIENT IS DECEASED; THE EXACT DATE OF DEATH IS UNKNOWN. THE PHYSICIAN CONSIDERED THE ANEURYSM RUPTURE TO BE UNRELATED TO CERENOVUS COILS. THERE WAS NO DELAY TO THE ENDOVASCULAR EMBOLIZATION PROCEDURE.
Patient problems: Intracranial Hemorrhage; Ruptured Aneurysm
MAUDE report 25321570 ↗MalfunctionApr 14, 2026Device Difficult to Setup or Prepare
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRACRANIAL COIL EMBOLIZATION PROCEDURE ON (B)(6) 2026, A 3.5MM X 7.5CM ORBIT GALAXY COMPLEX XTRASOFT COIL (640CX3575 / 31593818) WAS USED, AND IT COULD NOT BE RE-SHEATHED. THE COIL WAS REPLACED. THE PROCEDURE WAS DELAYED BY ABOUT 5 MINUTES. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION RELATED TO THE PROCEDURE AND THE REPORTED DEVICE ISSUE WERE UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED AT A LATER DATE, THIS FILE WILL BE UPDATED ACCORDINGLY. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION AND ANALYSIS. BASED ON THE RESULT OF THE PRODUCT ANALYSIS COMPLETED ON 14-APR-2026, THE EVENT HAS BEEN DETERMINED TO BE USFDA MDR REPORTABLE AS A "MALFUNCTION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24879767 ↗MalfunctionApr 2, 2026Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A STENT-ASSISTED ENDOVASCULAR EMBOLIZATION PROCEDURE ON (B)(6) 2026, THE MICROCATHETER WAS IN PLACE AND THE 8MM X 24CM ORBIT GALAXY COMPLEX FILL (640CF0824 / 31651278) WAS DELIVERED TO THE TARGET SITE. THE COIL FILLED THE ANEURYSM, AND THE PHYSICIAN TRIED TO DETACH THE COIL, BUT THE COIL WAS UNABLE TO BE DETACHED. THE PHYSICIAN RETRACTED ONLY THE COIL AND REPLACED IT WITH A SECOND COIL, A 2.5MM X 5CM ORBIT GALAXY COMPLEX XTRASOFT (640CX2505 / 31661861). THE SECOND COIL ENCOUNTERED THE SAME ISSUE; IT WAS UNABLE TO BE DETACHED. THE PHYSICIAN RETRACTED THE COIL AND REPLACED IT WITH A THIRD COIL FROM ANOTHER MANUFACTURER TO COMPLETE THE PROCEDURE USING THE ORIGINAL MICROCATHETER. THERE WAS NO REPORT OF ANY NEGATIVE IMPACT ON THE PATIENT. ON 01-APR-2026, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PROCEDURE WAS A STENT-ASSISTED COIL EMBOLIZATION OF AN ANEURYSM ON THE INTERNAL CAROTID ARTERY. THE COILS WERE NOT STRETCHED WHEN THEY WERE REMOVED; THEY WERE STILL ATTACHED TO THE DELIVERY SYSTEMS. THERE WAS NO INTERRUPTION OF BLOOD FLOW. THE SYRINGE WAS FILLED WITH SALINE FROM A DEDICATED SOURCE. THE SAME SYRINGE WAS USED WIT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24767328 ↗MalfunctionFeb 25, 2026Entrapment of Device; Failure to Advance
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, RESISTANCE WAS ENCOUNTERED WHEN ATTEMPTING TO ADVANCE AN ORBIT GALAXY XTRASOFT COIL (PRODUCT CODE: 640HX0206, LOT NUMBER 31651283) THROUGH A NON J&J MICROCATHETER. THE COIL WAS UNABLE TO PASS THE DISTAL END OF THE MICROCATHETER (MC). UNDER FLUOROSCOPY, THE COIL APPEARED CLUSTERED INSIDE THE MICROCATHETER. THE PHYSICIAN RETRACTED AND REMOVED ONLY THE COIL, THEN REPLACED IT WITH A NEW COIL TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE MICROCATHETER WAS NOT REPLACED, AND NO PATIENT INJURY WAS REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 24-FEB-2026 INDICATED THAT THEY COULD NOT MOVE THE DEVICE AT ALL DUE TO THE RESISTANCE. THEY WERE NOT ABLE TO TORQUE THE DEVICE. THERE WAS NO EVIDENCE OF PHYSICAL MATERIAL WITHIN THE DEVICE. THE DEVICE NOR THE CONCOMITANT DEVICE KINKED OR BENT PRIOR TO THE RESISTANCE OR NOTED AFTER REMOVAL FROM THE PATIENT. THE COIL HAD NOT BEEN WITHDRAWN OR REPOSITIONED. NECK REMODELING WAS NOT UTILIZED. THERE WAS NO ALLEGATION OF PATIENT INJURY DUE TO AN ALLEGED PRODUCT ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24443159 ↗MalfunctionJan 28, 2026Retraction Problem
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, AN ORBIT GALAXY XTRASOFT COIL (PRODUCT CODE: 640CX2535, LOT NUMBER: UNKNOWN) COULDN'T BE ADVANCED NOR RETRIEVED BY THE PHYSICIAN. THIS WAS THE FIRST COIL. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199819 ↗MalfunctionJan 14, 2026Stretched
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, A 2X4 ORBIT GALAXY XTRASOFT COIL (640HX0204, 31722202) WAS UNRAVELED OR STRETCHED. THE COIL'S SHAPE WAS FOUND TO BE NOT IDEAL WHILE FILLING THE TARGET SITE, LEADING THE PHYSICIAN TO RETRACT THE COIL, AT WHICH POINT THE DEVICE WAS OBSERVED TO BE UNRAVELED/STRETCHED. THE PHYSICIAN COMPLETED THE PROCEDURE USING A COIL FROM ANOTHER BRAND, AND THE MICROCATHETER WAS NOT REPLACED. THERE WERE NO REPORTED CONSEQUENCES OR IMPACT TO THE PATIENT, AND NO SIGNS, SYMPTOMS, OR PATIENT INVOLVEMENT WERE IDENTIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24067813 ↗MalfunctionDec 4, 2025Separation Failure; Separation Failure
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, A 3X8 ORBIT GALAXY XTRASOFT COIL (PRODUCT CODE 640CX0308, LOT NUMBER 31634774) FAILED TO DETACH. THE COIL WAS DELIVERED TO THE TARGET SITE AND FILLED THE ANEURYSM, BUT DETACHMENT COULD NOT BE ACHIEVED. THE PHYSICIAN RETRACTED THE COIL AND USED A NEW ONE TO COMPLETE THE SURGERY, WITHOUT REPLACING THE UNSPECIFIED MICROCATHETER. THERE WAS NO REPORTED PATIENT CONSEQUENCE OR CLINICAL IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23716544 ↗MalfunctionDec 3, 2025Stretched; Stretched
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE SHAPE OF THE 3MM X 6CM ORBIT GALAXY COMPLEX XTRASOFT (640CX0306 / 31485464) WAS FOUND TO BE NOT IDEAL AFTER IT WAS USED TO FILL THE ANEURYSM. THE PHYSICIAN RETRACTED ONLY THE COIL, BUT THE OIL WAS FOUND TO BE UNRAVELED / STRETCHED. THE PHYSICIAN REPLACED THE COIL WITH ONE FROM ANOTHER BRAND TO COMPLETE THE PROCEDURE USING THE SAME ORIGINAL MICROCATHETER (UNSPECIFIED BRAND). THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 26-NOV-2025, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PROCEDURE WAS TARGETING A TORTUOUS AND NARROW ANEURYSM ON THE LEFT INTERNAL CAROTID ARTERY. NO NECK REMODELING WAS USED; A CONTINUOUS FLUSH HAD BEEN MAINTAINED THROUGH THE MICROCATHETER. THE COIL HAD BEEN WITHDRAWN AND REPOSITIONED (THE NUMBER OF TIMES THIS HAD BEEN DONE WAS NOT PROVIDED). THERE WAS NO RESISTANCE BETWEEN THE GUIDEWIRE AND THE MICROCATHETER WHEN THE TARGET SITE WAS BEING ACCESSED. PRIOR TO THIS COIL, ONE OTHER COIL WENT THROUGH THE MICROCATHETER WITHOUT RESISTANCE. INFORMATION RELATED TO COIL ENTANGLEMENT WITH PREVIOUSLY PLACED COIL COULD NOT BE OBTAINED. THERE WAS NO ADDITIONA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23707851 ↗MalfunctionNov 21, 2025Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A STENT-ASSISTED EMBOLIZATION OF A RIGHT MIDDLE CEREBRAL ARTERY ANEURYSM, THE MICROCATHETER (UNSPECIFIED BRAND) WAS IN PLACE AND THE FIRST COIL, A 2MM X 6CM ORBIT GALAXY HELICAL XTRASOFT (640HX0206 / 31467317) WAS DELIVERED TO THE TARGET. THE COIL FILLED THE ANEURYSM, AND THE PHYSICIAN TRIED TO DETACH THE COIL, BUT IT WAS UNABLE TO DETACH. THE PHYSICIAN RETRACTED ONLY THE COIL AND REPLACED IT WITH A SECOND COIL. THE SECOND COIL WAS PLACED AND DETACHED. A THIRD COIL, A 4MM X 10CM ORBIT GALAXY COMPLEX XTRASOFT (640CX0410 / 31567754) WAS INTRODUCED INTO THE ANEURYSM, AND THIS THIRD COIL WAS ALSO UNABLE TO BE DETACHED. THE PHYSICIAN RETRACTED THE COIL AND REPLACED IT TO COMPLETE THE PROCEDURE USING THE SAME ORIGINAL CONCOMITANT MICROCATHETER. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 19-NOV-2025, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION CONFIRMED THE PROCEDURE WAS A STENT-ASSISTED EMBOLIZATION OF A RIGHT MIDDLE CEREBRAL ARTERY ANEURYSM. THE COILS WERE NOT STRETCHED WHEN THEY WERE REMOVED; BOTH COILS WERE STILL ATTACHED TO THEIR RESPECTIVE DELIVERY SYSTEMS. THERE WAS NO EVIDENCE OF ANY BLOOD FLOW RESTRICTION / INT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23624782 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page