PEEK PREVAIL Cervical Interbody DeviceRecalls & reported problems across all models
Medtronic Sofamor Danek, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 0
We found 3 models of PEEK PREVAIL Cervical Interbody Device by Medtronic Sofamor Danek, Inc. in the FDA's device registry (GUDID). Together they are named in 22 adverse event reports in the FDA's MAUDE database, received between Nov 21, 2016 and Sep 26, 2023 — classified as 4 injuries, 18 malfunctions.
The most frequently reported specific device problem is “Break” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported PEEK PREVAIL Cervical Interbody Device problems
Device problems
- Break20
- Migration or Expulsion of Device2
- Crack1
- Device Handling Problem1
- Failure to Align1
Patient problems
- No Known Impact Or Consequence To Patient13
- No Consequences Or Impact To Patient3
- Pain3
- Dysphagia/ Odynophagia2
- Injury2
- No Clinical Signs, Symptoms or Conditions2
- Bone Fracture(s)1
- Fall1
PEEK PREVAIL Cervical Interbody Device models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PEEK PREVAIL™ Cervical Interbody Device 4210564Intervertebral fusion device with bone graft, cervical18 reports
- PEEK PREVAIL™ Cervical Interbody Device 4210664Intervertebral fusion device with bone graft, cervical3 reports
- PEEK PREVAIL™ Cervical Interbody Device 4210764Intervertebral fusion device with bone graft, cervical1 report
Recent PEEK PREVAIL Cervical Interbody Device adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 26, 2023Break; Device Handling Problem; Break; Device Handling Problem
INFORMATION WAS RECEIVED FROM A (REGULATORY BODY, PATIENT, HEALTH CARE PROFESSIONAL) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH INTERBODY DEVICE. IT WAS REPORTED THAT MEDTRONIC PEEK PREVAIL THE TOP SCREW RETENTION WIRE BROKE INTO PIECES. IT WAS NOTICED ON X-RAYS BY NEUROSURGERY BUT DIDN'T HAVE PLANS TO REMOVE/REPLACE THE DEVICE. IN TIME BETWEEN 2018 AND 2022, THE TOP SCREW RETENTION WIRE BROKE INTO 3 PIECES. THE MANUFACTURER'S INSTRUCTIONS FOR IMPLANTING THE DEVICE WERE NOT FOLLOWED. THE DEVICE WAS NOT IMPLANTED FLUSH WITH TOP AND BOTTOM VERTEBRAE AS PER MANUFACTURERS SURGICAL TECHNIQUE MANUAL. NO COBB ANGLE WAS USED(RADIOLOGY). IT APPEARS THAT SCREWS LONGER THAN DISC SPACE WERE USED. RADIOLOGY REPORT ON 2013 SAID SCREWS WERE VERY PROUD, LOTS OF AIR WAS STILL IN NECK, AND RECOMMENDED RAD GRAFTS TO CHECK ALIGNMENT BUT IT WAS NOT DONE. SINCE THE 2013 ACDF PROCEDURE, PATIENT HAD SUFFERED WITH SEVERE THROAT, EAR, MOUTH, NECK PAIN, AND DIFFICULTY SWALLOWING. PATIENT HAD NUMEROUS FALLS, ONE SERIOUS ONE IN 2020 IN WHICH LEG WAS BROKEN. IT IS POSSIBLE THE WIRE BROKE DURING THAT FALL AS PATIENT ALSO HIT THE HEAD. IT ALSO MAY HAVE BROKEN IN 2021 WHERE PATIENT
Patient problems: Bone Fracture(s); Dysphagia/ Odynophagia; Fall; Neck Pain; Nervous System Injury; Pain
MAUDE report 17819707 ↗MalfunctionJun 23, 2021Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS RADICULAR FOR ACDF THERAPY (LEVELS IMPLANTED WERE C4-C5-C6). IT WAS REPORTED THAT THE CAGE BROKE WHILE INSERTION, HAD TO EXPLANTED AND WAS REPLACED WITH ALTERNATE CAGE. THERE WAS NO FRAGMENT LEFT IN THE PATIENT. THERE WERE NO SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12047146 ↗MalfunctionFeb 2, 2021Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT SPINAL THERAPY FOR CERVICAL DDD. IT WAS REPORTED THAT THE IMPLANT WAS BROKEN DURING PROCEDURE. THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT. THE IMPLANTATION AND EXPLANATION WAS COMPLETED ON SAME DAY. THERE WAS PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11270807 ↗MalfunctionOct 14, 2020Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL (HCP) VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH CERVICAL RADICUL OPATHY SUGGESTED FOR SPINAL THERAPY. PROCEDURE USED WAS ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF). LEVELS IMPLANTED C6-7. EVENT OCCURRED INTRA-OP. IT WAS REPORTED THAT DURING THE IMPLANTATION, SURGEON FIRST CHECKED WITH THE TRIAL AND THEN APPROPRIATE SIZE OF IMPLANT WAS OPEN. AFTER HOLDING THE IMPLANT WITH THE HOLDER, SURGEON USED THE MALLET TO SETTLE DOWN THE IMPLANT IN THE SPACE CREATED. WHEN THE HOLDER WAS GETTING DISENGAGED/ RELEASED THE FIRST HALF CAME OUT WITH THE HOLDER THE REST WAS HALF INSIDE THE BODY AND LATER IT WAS REMOVED COMPLETELY FROM THE PATIENT AND THEN NEW IMPLANT WAS USED AS A REPLACEMENT. BREAK AS REPORTED. NO FRAGMENTS LEFT IN PATIENT. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WAS REPORTED. DEVICE RETURN STATUS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10676202 ↗InjuryDec 9, 2019Break; Migration or Expulsion of Device; Break; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT FUSION AND FIXATION AT C3-C4 DUE TO DEGENERATION AND MYELOPATHY. THE ALLEGED DEVICE WAS IMPLANTED DURING THIS SURGERY. ON AN UNKNOWN DATE, POST-OP, THE SCREW SLIPPED OUT OF THE IMPLANT AFTER BREAKING THE LOCKING WIRE AND GOT STUCK IN THE ESOPHAGUS. ALLEGEDLY, THIS LED TO PAIN AND SWALLOWING ISSUES. HENCE, ON (B)(6) 2019, THE PATIENT UNDERWENT A REVISION SURGERY. THE BROKEN WIRE WAS RECOVERED BUT THE SCREW COULD NOT BE RECOVERED. IT WAS FOUND THAT THE SCREW WAS HIDDEN INSIDE THE ESOPHAGUS AND OPENING THE ESOPHAGUS WAS MORE RISKY. THE SCREW CAN STILL BE SEEN ON X-RAY. PATIENT IS RECOVERING WITH THE SCREW IN HIS THROAT WITH CONTINUED PROBLEMS BUT MINIMAL PAIN.
Patient problems: Dysphagia/ Odynophagia; Injury; Pain
MAUDE report 9438507 ↗InjurySep 19, 2019Break; Migration or Expulsion of Device; Break; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND FUSION DUE TO MYELOPATHY. ON AN UNKNOWN DATE, POST-OP, THE SCREW OF THE REPORTED PRODUCT BACKED-OUT FROM THE CONSTRUCT. THE BACKED-OUT SCREW PERFORATED THE ESOPHAGUS AND CAUSED PAIN TO THE PATIENT. THE STRING THAT HOLDS IN THE SCREW WAS ALSO BROKEN. HENCE, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2019, WHEREIN THE IMPLANT WAS REMOVED AND WAS REPLACED BY WITH PLATES. PATIENT WAS REPORTED TO BE SUFFERING WITH INJURY EVEN AFTER THE REVISION SURGERY.
Patient problems: Injury; Pain; Perforation of Esophagus
MAUDE report 9086266 ↗MalfunctionJun 20, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) SURGERY AT C3-4 DUE TO CERVICAL BILATERAL RADICULOPATHY. INTRA-OP, THE IMPLANT WAS FOUND BROKEN IN TWO PARTS AFTER THE PRODUCT INSERTION WHILE USING HAMMER TO FIX IT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8715112 ↗MalfunctionApr 22, 2019Break; Break
PRE-OPERATIVE DIAGNOSIS: BILATERAL RADICULOPATHY DUE TO DISC AT C5-6-7 LEVEL PROCEDURE: ACDF(ANTERIOR CERVICAL DISCECTOMY AND FUSION) IT WAS REPORTED THAT, INTRA-OP, IT WAS OBSERVED THAT FLANGES WERE MOVING DURING SCREW INSERTION. THE IMPLANT BROKE INTO TWO PIECES. THE BROKEN DEVICE WAS EXPLANTED. THE IMPLANT CAME IN CONTACT WITH THE PATIENT. PATIENT COMPLICATIONS AS A RESULT OF ALLEGED EVENT WERE UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8535186 ↗MalfunctionMar 27, 2019Break; Break
PRE-OPERATIVE DIAGNOSIS: CERVICAL SPONDYLOSIS, 5, 6, AND 7 DISCS PROTRUDING FROM THE NECK PROCEDURE: ANTERIOR CERVICAL DECOMPRESSION BONE GRAFT FUSION AND INTERNAL FIXATION LEVEL IMPLANTED: 2 (PROSTHESIS) IT WAS REPORTED THAT, INTRA-OP, THE NITINOL LOCKING RING ON THE PROSTHESIS WAS BROKEN. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. FRAGMENT REMAINED IN THE PATIENT. THE PATIENT HAS NOT UNDERGONE REVISION SURGERY. NO PATIENT SYMPTOMS OR COMPLICATION WERE REPORTED AS RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8455033 ↗MalfunctionDec 7, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH CERVICAL DEGENERATIVE DISEASE/ SPINAL STENOSIS AND UNDERWENT ANTERIOR CERVICAL DECOMPRESSION. DURING THE SURGERY, A FRACTURE OCCURRED DURING THE PARTIAL IMPLANTATION OF PEEK. THE DIRECTOR USED A PAIR OF PLIERS TO CLAMP IT OUT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW PRODUCT. NO PATIENT COMPLICATIONS OCCURRED AS A RESULT OF THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8144587 ↗MalfunctionOct 18, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH CERVICAL CANAL STENOSIS AND UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND FUSION PROCEDURE. DURING THE SURGERY, THE IMPLANT WAS DAMAGED WHILE TIGHTENING THE SCREW AND WAS FOUND BROKEN. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO FRAGMENT OF THE BROKEN PART REMAINED IN THE PATIENT. NO PATIENT SYMPTOMS OR COMPLICATION WERE REPORTED AS RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7982347 ↗MalfunctionSep 11, 2018Break; Break
TYPE OF PROCEDURE OR TECHNIQUE USED: ANTERIOR CERVICAL DISCECTOMY PRE-OPERATIVE DIAGNOSIS: CERVICAL SPINE INSTABILITY IT WAS REPORTED THAT INTRA-OP, IMPLANT BROKE DURING THE IMPLANTATION. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED IN THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7865839 ↗MalfunctionDec 18, 2017Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT CERVICAL SPINE FUSION SURGERY. INTRA OPERATIVELY, THE CAGE WAS BROKEN. THE DEVICE CAME IN CONTACT WITH THE PATIENT. NO FRAGMENTS REMAINED INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7125081 ↗MalfunctionOct 16, 2017Break; Break
IT WAS REPORTED THAT DURING A SPINAL SURGERY, THE PRODUCT BROKE. NO FRAGMENTS OF THE PRODUCT REMAINED INSIDE THE PATIENT'S BODY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6950681 ↗MalfunctionAug 24, 2017Crack; Crack
PROCEDURE USED: ANTERIOR CERVICAL DISCECTOMY WITH FUSION (ACDF) LEVELS OPERATED: C6-C7 IT WAS REPORTED THAT DURING SURGERY, WHILE REMOVING THE C6 SCREW A PORTION OF THE IMPLANT CRACKED. THE IMPLANT WAS REATTACHED WITH INSERTER FOR REMOVAL, BUT IT WAS DIFFICULT TO BE PULLED OUT. FINALLY THE IMPLANT WAS REPLACED. PATIENT COMPLICATIONS WERE REPORTED AS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6816124 ↗
Other Medtronic Sofamor Danek, Inc. products
- CD HORIZON Spinal System342 models · 2,323 reports
- GRAFTONand GRAFTON PLUSDemineralized Bone Matrix (DBM)25 models · 12 reports · 14 recalls
- METRx System5 models · 672 reports
- INFUSE Bone Graft6 models · 576 reports · 1 recall
- CD HORIZON SOLERA VOYAGER Spinal System60 models · 411 reports
- POWEREASE Instruments8 models · 331 reports · 1 recall
- Catalyft PL Expandable Interbody System16 models · 279 reports · 1 recall
- INFINITY Occipitocervical Upper Thoracic System35 models · 63 reports · 5 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page