POLARx FITRecalls & reported problems across all models
Boston Scientific Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one POLARx FIT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls
- 2
We found 2 models of POLARx FIT by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 14 adverse event reports in the FDA's MAUDE database, received between Jan 12, 2024 and Jul 28, 2026 — classified as 8 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Body Fluid” (1 report).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
POLARx FIT recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 10, 2024Open, Classified
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
Z-0302-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Oct 10, 2024Open, Classified
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
Z-0300-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported POLARx FIT problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Device Contamination with Body Fluid1
- Failure to Deflate1
- Unintended Deflation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Cardiac Tamponade3
- Pericardial Effusion3
- Atrial Fibrillation2
- Chest Pain2
- Fatigue2
- Low Blood Pressure/ Hypotension2
- Nausea2
POLARx FIT models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- POLARx™ FIT M004CRBS2060Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall13 reports
- POLARx™ FIT M004CRBS2160Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall1 report
Recent POLARx FIT adverse event reports
The 14 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced severe chest discomfort following a cryoablation procedure. On (B)(6) 2026, the patient underwent a cryoablation procedure using a PolarX Fit Balloon Catheter. The procedure was completed successfully. Subsequently, the patient began expressing severe chest discomfort and inability to exercise due to fatigue on (B)(6) 2026. An adjustment in oral medication was prescribed, and ECG diagnostic tests were performed. The patient was reported to be in stable condition with a resolution date of (B)(6) 2026.
Patient problems: Chest Pain; Fatigue
MAUDE report 50062988 ↗MalfunctionJan 28, 2026Unintended Deflation
IT WAS REPORTED THAT THE BALLOON DEFLATED AUTOMATICALLY. DURING AN ABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION, A POLARX FIT WAS USED IN THE LEFT SUPERIOR PULMONARY VEIN (LSPV). THE FIRST APPLICATION WAS UNEVENTFUL; HOWEVER, DURING THE SECOND APPLICATION, THE BALLOON WAS OBSERVED TO DEFLATE AUTOMATICALLY. THE SAME PROBLEM OCCURRED DURING TWO SUBSEQUENT ATTEMPTS, AND NO ERROR MESSAGES WERE DISPLAYED. A DIFFERENT DEVICE WAS THEN USED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT COMPLICATIONS. THE DEVICE WAS CONTAMINATED AND WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24202447 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
CLINICAL STUDY POLARX PAS PY003; SUBJECT ID: (B)(6). PER CLINICAL REPORT, IT WAS REPORTED THAT POST PULMONARY VEIN ISOLATION (PVI) ABLATION PROCEDURE USING THE POLARXFIT CATHETER, THE PATIENT EXPERIENCED PERSISTENT COUGH. PATIENT REPORTED CHEST/LUNG IRRITATION AND COUGHING FREQUENTLY WHEN SHE BREATHES SINCE PROCEDURE. ORAL MEDICATION WAS CHANGED/ADJUSTED TO RESOLVE THE ISSUE. ALSO, A CHEST X-RAY WAS PERFORMED. EVENT RESOLVED ON (B)(6) 2025 AND PATIENT IS NOW RECOVERED. DEVICE IS NOT EXPECTED TO BE RETURNED AS IT WAS DISCARDED POST PROCEDURE.
Patient problems: Cough
MAUDE report 24146703 ↗InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PER CLINICAL STUDY: POLARX PAS PY003 IT WAS REPORTED THAT A POLARX FIT CATHETER WAS SELECTED FOR USE. POST PROCEDURE, ON (B)(6) 2025, THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH COMPLAINTS OF CHEST PAIN WORSENED BY DEEP INHALATION, ALONG WITH NAUSEA, FATIGUE, AND MALAISE. COMPUTED TOMOGRAPHY IMAGING CONFIRMED A LEFT UPPER LOBE PULMONARY EMBOLISM. THE PATIENT WAS ADMITTED TO TELEMETRY AND STARTED ON HEPARIN THERAPY. A VENOUS DUPLEX STUDY WAS NEGATIVE FOR DEEP VEIN THROMBOSIS. LABORATORY TESTS WERE ALSO PERFORMED. DURING ADMISSION, THE PATIENT EXPERIENCED AN EPISODE OF ATRIAL FIBRILLATION. IN THE CARDIOLOGIST'S NOTE DATED (B)(6) 2025, IT WAS STATED: "CONTINUED PAROXYSMS OF ATRIAL FIBRILLATION CAN BE SEEN IN THE IMMEDIATE PERIOD FOLLOWING AF ABLATION." THE PATIENT WAS STARTED ON MULTAQ AND SUBSEQUENTLY DISCHARGED ON THE SAME DAY WITH THIS MEDICATION. DEVICE IS NOT EXPECTED TO BE RETURNED AS COMPLICATIONS OCCURRED POST PROCEDURE, AFTER THE DEVICE WAS DISCARDED.
Patient problems: Atrial Fibrillation; Chest Pain; Discomfort; Fatigue; Nausea; Pulmonary Embolism
MAUDE report 23549907 ↗InjuryAug 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A PERICARDIAL EFFUSION. DURING A PULMONARY VEIN ISOLATION (PVI) PROCEDURE TO TREAT ATRIAL FIBRILLATION, A POLARX CATHETER WAS SELECTED FOR USE. FOLLOWING TREATMENT OF ALL FOUR PULMONARY VEINS AND REMOVAL OF THE POLARX CATHETER, INTRACARDIAC ECHOCARDIOGRAPHY (ICE) WAS USED TO VISUALIZE THE PERICARDIUM. AN EFFUSION WAS NOTED. THIS EFFUSION WAS NOT APPARENT AT BASELINE. PROTAMINE WAS ADMINISTERED. THERE WERE NO CHANGES IN HEMODYNAMICS DURING THE CASE THAT WOULD INDICATE THE PRESENCE OF AN EFFUSION. THE EFFUSION WAS MONITORED FOR THE NEXT 30 MIN WITH ICE AND A TRANSTHORACIC ECHOCARDIOGRAM (TTE) WAS PERFORMED. THE EFFUSION WAS MOST NOTABLE AT THE BASAL ASPECT OF THE RIGHT VENTRICLE AND MEASURED APPROXIMATELY 2.1 CM AT ITS LARGEST POINT. DURING MONITORING, ULTRASOUND CONTRAST WAS ADMINISTERED INTRAVENOUSLY AND WAS ASSESSED TO SEE IF IT ENTERED THE PERICARDIAL SPACE. NO CONTRAST ENTERED THE PERICARDIAL SPACE. THE PATIENT'S HEMODYNAMICS WERE NOT ALTERED AT ALL DURING THIS PERIOD. THERE WAS LIMITED SPACE TO PERFORM A PERICARDIOCENTESIS AS THE EFFUSION SIZE WAS FAIRLY SMALL DURING DIASTOLE. IT WAS DECIDED THAT A PERICARDIOCENTESIS WOULD NOT BE PER
Patient problems: Pericardial Effusion
MAUDE report 22886494 ↗MalfunctionJul 23, 2025Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT THE BALLOON COULD NOT BE DEFLATED. DURING A POLARX CRYOABLATION PROCEDURE, A CATHETER WAS SELECTED FOR USE. ACCORDING TO THE CUSTOMER, THE CATHETER WAS PREPARED CORRECTLY, BUT THEY WERE UNABLE TO DEFLATE THE BALLOON, SO IT WAS REPLACED. MOST QUESTIONS COULD NOT BE ANSWERED BY THE CUSTOMER, AS THEY WANTED TO PROCEED QUICKLY WITH THE PROCEDURE AND SIMPLY STATED THAT THE BALLOON COULD NOT BE DEFLATED. THE DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT COMPLICATIONS. THE CATHETER CANNOT BE RETURNED BECAUSE, UNFORTUNATELY, THE CARE PROVIDER HAD ALREADY DISCARDED IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22579195 ↗InjuryApr 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT EXPERIENCED CARDIAC TAMPONADE. SUBSEQUENTLY, THE PROCEDURE WAS CANCELLED. DURING A CRYOABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION, A POLARX FIT CATHETER WAS SELECTED FOR USE. AFTER PERFORMING A TRANSSEPTAL PUNCTURE USING THE VERSACROSS, THE POLARX FIT CATHETER WAS INSERTED, AND TREATMENT OF THE LEFT INFERIOR PULMONARY VEIN BEGAN. APPROXIMATELY 100 SECONDS INTO THE APPLICATION, THE PATIENT'S BLOOD PRESSURE DROPPED, AND A DOUBLE STOP WAS INITIATED ON THE CRYOABLATION SYSTEM. IT WAS NOTED THAT THE PATIENT HAD A SMALL PERICARDIAL EFFUSION/TAMPONADE. PERICARDIAL DRAINAGE WAS PERFORMED AND COMPLETED AT 8:52AM. THE PATIENT WAS GIVEN 40 OF PROTAMINE AT 8:53AM TO REVERSE THE HEPARIN. A TOTAL OF 100CC OF BLOOD WAS DRAWN OUT AND AT 9:48AM, THE DRAIN STILL HAD NO BLOOD ACCUMULATION. ANESTHESIA WAS ABLE TO KEEP THE PATIENT STABLE THROUGHOUT WITH LITTLE MANAGEMENT AND ONLY 1 SHOT OF EPINEPHRIN WAS GIVEN. THE PROCEDURE WAS CANCELLED AT THIS TIME. THE PATIENT LEFT THE ROOM AT 9:50AM AND WAS TAKEN TO THE ICU. THE PATIENT WAS ALERT AND DOING WELL POST PROCEDURE. PATIENT WAS TO REMAIN IN HOSPITAL FOR 2-3 DAYS TO BE MONITORED BUT WAS EXPECTED TO MAKE A FULL RECOVERY. IT
Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension
MAUDE report 21890282 ↗InjuryJan 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL STUDY (B)(6) FOR POLARX PAS; SUBJECT ID # (B)(6). IT WAS REPORTED THAT POST PROCEDURE, THE PATIENT EXPERIENCED NAUSEA, HYPOTENSION AND WAS SUBSEQUENTLY DIAGNOSED WITH CARDIAC TAMPONADE AND PERICARDIAL EFFUSION, WHICH REQUIRED SURGICAL INTERVENTION. DURING A PULMONARY VEIN ISOLATION CRYOABLATION PROCEDURE, A POLARX FIT CATHETER WAS SELECTED FOR USE. POST TREATMENT, THE PATIENT EXPERIENCED EPISODES OF NAUSEA AND HYPOTENSION. AN URGENT ECHOCARDIOGRAM AND ULTRASOUND WERE PERFORMED AND SHOWED A MODERATE-SIZED PERICARDIAL EFFUSION, 1.8CM AT MAX. THIS EFFUSION CAUSED CARDIAC TAMPONADE. THE PATIENT WAS TAKEN TO THE CATH LAB WHERE A PERICARDIOCENTESIS WAS PERFORMED AND 275CC OF SANGUINOUS DRAINAGE WAS REMOVED. THIS EVENT RESULTED IN IN-PATIENT HOSPITALIZATION OR PROLONGATION OF EXISTING HOSPITALIZATION. THE PATIENT WAS DISCHARGED ON (B)(6) 2025, IMAGING PERFORMED (ULTRASOUND) SHOWED THAT THE ISSUE HAD RESOLVED. THE DEVICE WAS DISPOSED BY THE FACILITY AND WILL NOT BE RETURNED AS THE DEVICE PERFORMED AS EXPECTED DURING THE PROCEDURE AND THERE WERE NO REPORTS OF ANY PERFORMANCE CONCERNS.
Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Nausea; Pericardial Effusion
MAUDE report 21267199 ↗MalfunctionDec 30, 2024Device Contamination with Body Fluid; Device Contamination with Body Fluid
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED POST PATIENT SEDATION. DURING FIBRILLATION PROCEDURE TO TREAT PATIENT WITH FIBRILLATION, A POLARX FIT WAS SELECTED FOR USE. IT WAS OBSERVED THAT THE PHYSICIAN USED THE POLARX FIT FOR CRYOABLATION AT THE LEFT SUPERIOR PULMONARY VEIN AND THE TEMPERATURE DROPPED TO MINUS 70 DEGREES. AN ERROR FOR DETECTED BLOOD IN THE BALLOON OCCURRED. THE PHYSICIAN TERMINATED OF THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21039607 ↗InjuryDec 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
POLARX PAS CLINICAL STUDY. IT WAS REPORTED THAT A PERFORATION, PERICARDIAL EFFUSION AND CARDIAC TAMPONADE OCCURRED. A POLARX FIT ABLATION CATHETER AND INTELLAMAP ORION DIAGNOSTIC CATHETER WAS SELECTED FOR ABLATION TREATMENT TO THE LEFT INFERIOR PULMONARY VEIN. IT WAS NOTED THAT THE SYSTOLIC BLOOD PRESSURE WAS 70S. IT WAS SUSPECTED THAT THE INTELLAMAP ORION PERFORATED THE LEFT ARTRIA DURING INITIAL MAPPING. THEREFORE, EXAMINATION OF INTRACARDIAC ULTRASOUND WAS PERFORMED WHICH THEN REVEALED A 1CM EFFUSION. GIVEN THE NEW EFFUSION, PROCEDURE WAS TERMINATED AND LEFT FEMORAL ARTERIAL LINE WAS OBTAINED AS PERICARDIOCENTESIS WAS PERFORMED. CHEST WAS PREPPED AND DRAPED IN USUAL STERILE FASHION AND ULTRASOUND GUIDANCE WAS USED WHICH NOTED SUBXIPHOID DISTANCE TO THE PERICARDIUM WAS CLOSER TO 8 TO 0CM. LEFT PARASTERNAL ULTRASOUND VIEW WAS MUCH BETTER AND ONLY 4CM AWAY AND THEREFORE WAS USED FOR ACCESS WITHOUT DIFFICULTIES. WIRE IN THE PERICARDIUM CIRCUMSCRIBING THE CARDIAC SHADOW WAS CONFIRMED ON FLUOROSCOPY. A TOTAL OF 450CC OF BRIGHT BLOOD WAS REMOVED FROM THE PERICARDIUM WITHOUT DIFFICULTY AND BLOOD PRESSURE RESUMED AND ULTRASOUND CONFIRMED COMPLETE RESOLUTION OF THE PERICARDIAL EFFUSION. D
Patient problems: Cardiac Perforation; Cardiac Tamponade; Pericardial Effusion
MAUDE report 20957137 ↗InjuryDec 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL STUDY (B)(6) FOR POLARX PAS; SUBJECT ID (B)(6). IT WAS REPORTED THAT POST PROCEDURE, THE PATIENT EXPERIENCED PERICARDITIS REQUIRING MEDICAL TREATMENT. DURING A THREE (3) MONTH FOLLOW-UP POST ABLATION PROCEDURE WITH A POLARX CATHETER, THE PATIENT NOTIFIED THE PHYSICIAN THAT A FEW HOURS AFTER DISCHARGE, THEY EXPERIENCED PERICARDITIS AND RECURRENCE OF ATRIAL FIBRILLATION. THE PATIENT WAS TREATED AT AN OUTSIDE FACILITY WHERE THEY UNDERWENT SURGERY AND WERE PRESCRIBED ORAL MEDICATION; COLCHICINE, FLECAINIDE AND DILTIAZEM. THE PATIENT CONDITION HAS RESOLVED. A REQUEST FOR THE MEDICAL RECORDS WAS SENT, UPDATES PENDING ONCE THOSE RECORDS ARE RECEIVED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THE DEVICE WAS DISPOSED BY THE FACILITY AND WILL NOT BE RETURNED AS THE DEVICE PERFORMED AS EXPECTED DURING THE PROCEDURE AND THERE WERE NO REPORTS OF ANY PERFORMANCE CONCERNS. IT WAS FURTHER REPORTED PER ADDITIONAL INFORMATION THAT THE PATIENT UNDERWENT AN ABLATION PROCEDURE ON (B)(6) 2024. ON (B)(6) 2024, THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT AT (B)(6) WITH CHIEF COMPLAINTS OF CHEST PAIN AND SHORTNESS OF BREATH, JUST A FEW HOURS AFTER BEING DISCHARGED FROM BANNER HEALTH
Patient problems: Atrial Fibrillation; Pericarditis
MAUDE report 20902632 ↗MalfunctionApr 1, 2024
CRYO BALLOON BLOOD DETECTION ERROR /CATHETER REMOVED. NEW CATHETER OPENED. NO HARM TO PATIENT. PROBLEM RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19010765 ↗MalfunctionJan 18, 2024
BLOOD DETECTED IN BALLOON CATHETER - ALERT GIVEN ON CONSOLE DURING CASE, WHILE CATHETER WAS IN BODY. CATHETER REMOVED, REPLACED WITH A NEW ONE. NO PATIENT HARM. DOCTOR WAS PRESENT. MANUFACTURER REPRESENTATIVE CALLED INTO HIS COMPANY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18534035 ↗MalfunctionJan 12, 2024
CRYO BALLOON CATHETER IN USE DURING ABLATION. THE BOSTON SCIENTIFIC POLARX CRYO MACHINE READ: ¿REFRIGERANT FLOW OBSTRUCTION DETECTED¿ MANUFACTURER REP. STATED WE NEEDED TO SWITCH THE CATHETER, SO CATHETER THAT WAS IN THE BODY WAS REMOVED, PLACED IN BIOHAZARD BAG, LABEL WAS CUT OFF FROM PACKAGING AND ATTACHED TO BAG. MEDWATCH REPORT FILLED OUT. NEW CATHETER PLACED AND THERE WERE NOT ANYMORE ERRORS FOR THE REST OF THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18502156 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page