PRIMARecalls & reported problems across all models
Limacorporate SPA · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one PRIMA model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 2
We found 4 models of PRIMA by Limacorporate SPA in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Dec 5, 2023 and May 4, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Defective Device” (2 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
PRIMA recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 19, 2025Open, Classified
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
Z-2025-2025 · Root cause: Process design · Recall details · FDA record ↗
- May 19, 2025Open, Classified
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
Z-2024-2025 · Root cause: Process design · Recall details · FDA record ↗
Most reported PRIMA problems
Device problems
- Defective Device2
- Device Dislodged or Dislocated2
Patient problems
- Joint Dislocation2
- No Clinical Signs, Symptoms or Conditions2
PRIMA models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PRIMA 1367.15.702Shoulder Prosthesis, Reverse Configuration1 recall2 reports
- PRIMA 1367.15.704Shoulder Prosthesis, Reverse Configuration1 recall0 reports
- PRIMA 1367.54.100Shoulder Prosthesis, Reverse Configuration1 report
- PRIMA 1367.54.200Shoulder Prosthesis, Reverse Configuration1 report
Recent PRIMA adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryMay 4, 2026Device Dislodged or Dislocated
REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO SHOULDER DISLOCATION. ORIGINAL SURGERY WAS PERFORMED ON THE SAME DAY TO IMPLANT A COMBINATION OF FOLLOWING PRIMA HUMERAL COMPONENTS AND DJO ALTIVATE REVERSE GLENOID: PRIMA SHORT STEM PLUS #1 (PN 1357.14.228, LOT 2524131, STERILIZATION (B)(6)) PRIMA REVERSE TRAY #SHORT (PN 1367.15.010, LOT 2518444, STERILIZATION (B)(6)) PRIMA LIMAVIT REVERSE INSERT #SHORT DIA 36MM CORR 0° (PN 1367.54.100, LOT 25AT2RE, STERILIZATION (B)(6)). DJO MODULAR BASEPLATER AND TAPER KIT (PN 508-65-001). DJO RSP GLENOID HEAD AND RETAINING SCREW 2MM LATERAL OFFSET SZ 36MM (PN 508-36-103). PATIENT DISLOCATED AND THE ABOVE MENTIONED PRIMA REVERSE TRAY AND LINER WERE REPLACED: SAME SIZE OF TRAY WAS CHOSEN BY SURGEON, WHILE THE NEW LINER WAS REPLACED WITH A RETENTIVE ONE (PN 1367.54.101). SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1957. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 25067086 ↗MalfunctionApr 30, 2025Defective Device; Defective Device
INTRA-OPERATIVE ISSUE WAS DISCOVERED DURING A SURGERY PERFORMED ON (B)(6)2025, RELATED TO THE FOLLOWING COMPONENT: - PRIMA 2MM ECCENTRICAL ADAPTOR (PART CODE 1367.15.702, LOT NUMBER 2416658, STERILIZATION 2400195). THE BLACK LASER MARKING MEANT TO IDENTIFY THE POSITION OF THE OFFSET IS REPORTED AS NOT PRESENT IN THE LAST 2 CASES. THE SURGEON PROCEEDED ANYWAY TO IMPLANT THE DEVICE. THE PATIENT INVOLVED IN THE EVENT IS A FEMALE, DATE OF BIRTH (B)(6)1948. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21934152 ↗MalfunctionApr 30, 2025Defective Device; Defective Device
INTRA-OPERATIVE ISSUE WAS DISCOVERED DURING A SURGERY PERFORMED ON (B)(6) 2025, RELATED TO THE FOLLOWING COMPONENT: PRIMA 2MM ECCENTRICAL ADAPTOR (PART CODE 1367.15.702, LOT NUMBER 2416658, STERILIZATION (B)(4). THE BLACK LASER MARKING MEANT TO IDENTIFY THE POSITION OF THE OFFSET IS REPORTED AS NOT PRESENT IN THE LAST 2 CASES. THE SURGEON PROCEEDED ANYWAY TO IMPLANT THE DEVICE. THE PATIENT INVOLVED IN THE EVENT IS A FEMALE, DATE OF BIRTH (B)(6) 1954. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21933700 ↗InjuryDec 5, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
SHOULDER REVISION SURGERY WAS PERFORMED ON (B)(6) 2023 FOR DISLOCATION. PATIENT DISLOCATED POST OP AND CAUSE WAS A STRETCHED-OUT SHOULDER CAPSULE. DEVICE WAS REVISED TO A THICKER COMPONENT. REVISION WAS A SUCCESS WITH NO COMPLICATIONS. AGENT STATED PATIENT ANATOMY IS SUSPECTED BUT THE DEVICES DID NOT FAIL. PREVIOUS SURGERY - (B)(6) 2023. PATIENT - MALE. DATE OF BIRTH - (B)(6). EVENT HAPPENED IN USA. EXPLANTED COMPONENTS. PRIMA REVERSE TRAY #SHORT, COMMERCIAL CODE 1367.15.010 - LOT #2308370 - STER. #2300107 . PRIMA REV.INSERT #SHORT D.40MM, COMMERCIAL CODE 1367.54.200 - LOT #23AT1BE - STER. #2300147.
Patient problems: Joint Dislocation
MAUDE report 18266826 ↗
Other Limacorporate SPA products
All Limacorporate SPA devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page