- Adverse event reports
- 224
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 3 models of RX Locking Device and Biopsy Cap by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 224 adverse event reports in the FDA's MAUDE database, received between Nov 23, 2016 and Sep 29, 2026 — classified as 3 injuries, 221 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (175 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported RX Locking Device and Biopsy Cap problems
Device problems
- Detachment of Device or Device Component175
- Use of Device Problem58
- Device Dislodged or Dislocated46
- Material Integrity Problem30
- Difficult to Advance7
- Entrapment of Device6
- Adverse Event Without Identified Device or Use Problem2
- Problem with Sterilization2
Patient problems
- No Consequences Or Impact To Patient122
- No Clinical Signs, Symptoms or Conditions83
- No Known Impact Or Consequence To Patient20
- Insufficient Information1
- No Patient Involvement1
RX Locking Device and Biopsy Cap models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- RX Locking Device and Biopsy Cap M00545260Endoscope channel accessory216 reports
- RX Locking Device and Biopsy Cap M00545270Endoscope channel accessory7 reports
- RX Locking Device and Biopsy Cap M00545280Endoscope channel accessory1 report
Recent RX Locking Device and Biopsy Cap adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931053 ↗MalfunctionAug 6, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50165740 ↗MalfunctionApr 3, 2026Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX LOCKING DEVICE AND BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SPONGE WITHIN THE BIOPSY CAP DETACHED AND WAS LODGED IN THE SCOPE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24779036 ↗MalfunctionApr 1, 2026Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX LOCKING DEVICE AND BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SPONGE WITHIN THE BIOPSY CAP DETACHED AND WAS LODGED IN THE SCOPE. IT WAS REMOVED IN THE CLEANING ROOM. THE PROCEDURE WAS COMPLETED WITH A NEW SCOPE AND RX LOCKING DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24756343 ↗MalfunctionFeb 18, 2026Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX LOCKING DEVICE AND BIOPSY CAP WAS USED FOR COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SPONGE FROM THE BIOPSY CAP AND LOCKING DEVICE ENTERED THE SCOPE CHANNEL. THE DETACHED PIECE WAS EXTRACTED DURING MANUAL CLEANING. THE PROCEDURE WAS COMPLETED USING A NEW ENDOSCOPE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24384688 ↗MalfunctionJan 27, 2026Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX LOCKING DEVICE AND BIOPSY CAP WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE BIOPSY CAP SPONGE BECAME DISLODGED AND FELL INTO THE SCOPE CHANNEL. THE CAP WAS MANIPULATED USING THE DISTAL TIP WIRE HOLDER, AND LUBRICANT WAS APPLIED TO THE BALLOON CATHETER TO ASSIST IN THE PROCESS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24193450 ↗MalfunctionJan 7, 2026Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED DURING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON UN UNKNOWN DATE FOR THE TREATMENT OF STRICTURE. DURING THE PROCEDURE, THE SPONGE FROM THE LOCKING DEVICE BECAME LODGED IN THE SCOPE CHANNEL DURING THE WIRE EXCHANGE, AND THEY WERE UNABLE TO FLUSH THE SPONGE FROM THE CHANNEL. ALTHOUGH THE PROCEDURE WAS PROLONGED, IT WAS SUCCESSFULLY COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23991885 ↗MalfunctionOct 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, WHEN THE PHYSICIAN USED RAT TOOTH FORCEPS TO RETRIEVE A STENT, BUT THE FORCEPS BECAME STUCK. UPON INVESTIGATION, IT WAS FOUND THAT THE FOAM INSERT FROM THE BIOPSY CAP HAD DISLODGED AND WAS LODGED INSIDE THE SCOPE. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23222449 ↗MalfunctionAug 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING DEVICE AND BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE SPONGE DETACHED IN THE WORKING CHANNEL OF THE SCOPE. A BORESCOPE WAS USED TO REMOVE THE SPONGE. THE PROCEDURE WAS COMPLETE USING AN ALTERNATE DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22771046 ↗MalfunctionJul 25, 2025Detachment of Device or Device Component; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2025 THE SPONGE OF THE BIOPSY CAP BECAME LODGED IN THE DUODENOSCOPE. A WATER-SOLUBLE LUBRICANT TO THE TOP OF THE BIOPSY CAP, PRIOR TO USE. THE FOAM PIECE WAS RETRIEVED DURING SCOPE CLEANING AND STERILIZATION. REPORTEDLY, THE SCOPE HAD TO BE SENT FOR REPAIR. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22609633 ↗MalfunctionJul 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE DUODENOSCOPE COULD NO LONGER ASPIRATE DUE TO AN OBSTRUCTION IN THE CHANNEL. MULTIPLE ATTEMPTS WERE MADE IN THE OPERATING ROOM TO UNCLOG THE ENDOSCOPE; HOWEVER, THESE EFFORTS WERE UNSUCCESSFUL. THE PROCEDURE WAS UNABLE TO BE COMPLETED AS A RESULT OF THIS EVENT. REPORTEDLY, A WATER-SOLUBLE LUBRICANT TO THE TOP OF THE BIOPSY CAP, PRIOR TO USE. THE FOLLOWING DAY, A RENEWED ATTEMPT TO CLEAR THE BLOCKAGE WAS MADE. AFTER SEVERAL MANEUVERS, THE OBSTRUCTION WAS SUCCESSFULLY REMOVED. THE OBSTRUCTION WAS IDENTIFIED AS A PIECE OF FOAM FROM THE BIOPSY CAP. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22609568 ↗MalfunctionJun 26, 2025Use of Device Problem; Detachment of Device or Device Component; Material Integrity Problem; Use of Device Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RX LOCKING / BIOPSY CAP WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2025, FOR THE TREATMENT OF BILE DUCT STONES. DURING THE PROCEDURE, AN ATTEMPT WAS MADE TO PASS A STONE-REMOVING BALLOON CATHETER THROUGH THE ENDOSCOPE'S WORKING CHANNEL. RESISTANCE WAS ENCOUNTERED, AND AN OBSTRUCTION WAS NOTED WITHIN THE CHANNEL. AS A RESULT, THE ENDOSCOPE WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE COULD NOT BE COMPLETED. REPORTEDLY, IT WAS DISCOVERED THAT A SPONGE WAS LODGED IN THE WORKING CHANNEL AND COULD NOT BE REMOVED. IT WAS ALSO NOTED THAT WATER-SOLUBLE LUBRICANT HAD NOT BEEN APPLIED TO THE TOP OF THE BIOPSY CAP PRIOR TO ITS USE. THE ENDOSCOPE WAS SUBSEQUENTLY SENT FOR REPAIR. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22341508 ↗MalfunctionJun 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX LOCKING AND BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE ANTI-REFLUX FOAM DETACHED AND WENT INTO THE WORKING CHANNEL OF THE DUODENOSCOPE. THE PROCEDURE CONTINUED, HOWEVER, THE PHYSICIAN REPORTED THAT THE OBJECT DISPLACED INSIDE THE TUBE CAUSED RESISTANCE DURING THE PROCEDURE. REPORTEDLY, THE PROCEDURE WAS COMPLETED USING THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22196517 ↗MalfunctionMay 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT REMOVAL PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO PASS INSTRUMENTS THROUGH THE WORKING CHANNEL OF THE SCOPE. THEY WERE ABLE TO PULL THE STENT USING A BIOPSY FORCEPS, HOWEVER, IT WAS VERY DIFFICULT TO PASS THE FORCEPS THROUGH THE CHANNEL. THE PHYSICIAN THEN ATTEMPTED TO PASS A SNARE THROUGH THE WORKING CHANNEL. THEY WERE UNABLE TO PASS THE SNARE AND SWITCHED TO A NEW SCOPE. DURING THE SCOPE CLEANING THEY COULD NOT GET THE CLEANING SOLUTION OR WATER TO SUCTION THROUGH THE WORKING CHANNEL. THE SPONGE FROM BIOPSY CAP WAS DISCOVERED TO BE OBSTRUCTING WORKING CHANNEL. REPORTEDLY, A WATER-SOLUBLE LUBRICANT WAS APPLIED ON TOP OF THE BIOPSY CAP PRIOR TO USE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED ON MAY 28, 2025: IT WAS CONFIRMED THAT THE SPONGE DETACHED INTO THE SCOPE WORKING CHANNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087786 ↗MalfunctionFeb 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT RX LOCKING / BIOPSY CAP WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2025 FOR THE TREATMENT OF CANCER. DURING THE PROCEDURE, THE PHYSICIAN PULLED TWO STENTS OUT OF THE SCOPE AND THEN FLUSHED THE SCOPE WITH A BLUNT NEEDLE. IT WAS OBSERVED THAT THE SPONGE FROM THE BIOPSY CAP HAD DISLODGED INTO THE SCOPE. THE SPONGE WAS REMOVED USING RAT TOOTHED FORCEPS. THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21298260 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page