SORBAFIXRecalls & reported problems across all models
Davol Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 270
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls
- 0
We found 2 models of SORBAFIX by Davol Inc. in the FDA's device registry (GUDID). Together they are named in 270 adverse event reports in the FDA's MAUDE database, received between Jul 28, 2017 and Sep 3, 2026 — classified as 10 injuries, 260 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (170 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported SORBAFIX problems
Device problems
- Failure to Form Staple170
- Break47
- Incorrect, Inadequate or Imprecise Result or Readings18
- Detachment of Device or Device Component12
- Unintended Ejection12
- Tear, Rip or Hole in Device Packaging8
- Expiration Date Error6
- Defective Device4
Patient problems
- No Clinical Signs, Symptoms or Conditions209
- No Consequences Or Impact To Patient40
- Foreign Body In Patient12
- Injury4
- No Patient Involvement4
- Hernia3
- Pain3
- Bowel Perforation2
SORBAFIX models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SORBAFIX 0113116STAPLE, IMPLANTABLE162 reports
- SORBAFIX 0113115STAPLE, IMPLANTABLE108 reports
Recent SORBAFIX adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50592987 ↗MalfunctionAug 25, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50439070 ↗MalfunctionJun 16, 2026Detachment of Device or Device Component
AS REPORTED, DURING A PROCEDURE, AFTER DEPLOYING A FASTENER USING SORBAFIX ENHANCED DEVICE, THE OTHER FASTENERS ALONG WITH THE GUIDEWIRE DETACHED FROM THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25522314 ↗MalfunctionJun 5, 2026Failure to Form Staple
AS REPORTED, DURING A LAPAROSCOPIC INCISIONAL HERNIA REPAIR PROCEDURE, THE SORBAFIX ENHANCED FIXATION DEVICE FAILED TO SECURE THE MESH AFTER MULTIPLE ATTEMPTS. THE FIXATION WAS ATTEMPTED AT THE ABDOMINAL SOFT TISSUE WITH ADEQUATE COUNTER-PRESSURE; HOWEVER, THE FASTENERS FAILED TO DEPLOY AND SECURE. THE LOOSE FASTENERS WERE REMOVED FROM THE PATIENT, AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. REPORTEDLY, THE SURGEON HAD PRIOR EXPERIENCE USING THIS PRODUCT. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25410442 ↗MalfunctionJun 5, 2026Failure to Form Staple
AS REPORTED, DURING PROCEDURE THREE (3) SORBAFIX ENHANCED FASTENERS FAILED TO FASTEN THE MESH TO THE TISSUE. REPORTEDLY, ANOTHER SORBAFIX DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25410439 ↗MalfunctionMay 29, 2026Failure to Form Staple
AS REPORTED, DURING A LAPAROSCOPIC IPOM PROCEDURE, THE SORBAFIX ENHANCED FASTENER DID NOT FIXATE PROPERLY AND THE FASTENER CAME OUT. REPORTEDLY, FEW OF THE FASTENERS GOT FIXATED AND THE LOOSE FASTENERS WERE REMOVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25332837 ↗MalfunctionMay 18, 2026Failure to Form Staple
AS REPORTED, DURING A RIGHT INGUINAL HERNIA REPAIR PROCEDURE, THE SORBAFIX ENHANCED DEVICE FAILED TO FIXATE AFTER DEPLOYING TWO (2) FASTENERS AS THE FASTENER'S KEPT "COMING OFF THE TISSUE". THE LOOSE FASTENERS WERE REMOVED FROM THE PATIENT AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25205242 ↗MalfunctionMay 8, 2026Failure to Form Staple
AS REPORTED, DURING AN IPOM PROCEDURE, WHILE TRYING TO FIXATE INTO THE ABDOMINAL SOFT TISSUE, THE SORBAFIX DEVICE FAILED TO FIXATE THE MESH. THE LOOSE FASTENERS WERE REMOVED FROM THE PATIENT'S BODY. REPORTEDLY, ADEQUATE AMOUNT OF PRESSURE WAS APPLIED WHILE USING THE SORBAFIX DEVICE. THE SURGEON DID NOT DRY FIRE TO CHECK IF THE ISSUE REOCCURRED. THE PROCEDURE WAS COMPLETED USING NEW FIXATION DEVICE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25124902 ↗MalfunctionMay 8, 2026Failure to Form Staple
AS REPORTED, ON (B)(6) 2026 DURING A PROMONTOFIXATION SURGERY USING A SORBAFIX ENHANCED FIXATION DEVICE NONE OF THE FASTENERS WERE FIXATED SUCCESSFULLY AND THE STAPLES DID NOT HOLD. AS SUCH THE PROSTHESIS WAS SECURED WITH SUTURES. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25124856 ↗MalfunctionApr 16, 2026Failure to Form Staple
AS REPORTED, DURING A BILATERAL HERNIOPLASTY PROCEDURE SORBAFIX ENHANCED FASTENER DID NOT SECURE THE MESH AT FIRST SHOT AND NO FASTENERS WERE FIXED SUCCESSFULLY EVEN THOUGH THE RIGHT PRESSURE WAS APPLIED. THE LOOSE FASTENERS WERE NOT REMOVED FROM THE PATIENT'S BODY. REPORTEDLY, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE AND THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 24904190 ↗MalfunctionApr 14, 2026Failure to Form Staple
AS REPORTED ON (B)(6) 2026, DURING A LAPAROSCOPIC INCISIONAL HERNIA REPAIR PROCEDURE THE SORBAFIX ENHANCED DEVICE FAILED TO FIXATE THE MESH. IT WAS REPORTED THAT A PORTION OF MESH WAS STAPLED TO THE PERITONEUM, AND THE REMAINING FASTENERS BECAME LOOSE AND FELL. A CAPSURE DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE, AND THE LOOSE FASTENERS WERE REMOVED FROM THE PATIENT¿S BODY. THERE WAS NO PATIENT/USER HARM OR INJURY. AS REPORTED SURGEON IS THE FIRST-TIME USER OF THE SORBAFIX DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24875426 ↗InjuryApr 7, 2026Patient-Device Incompatibility
PER INFORMATION PROVIDED: ON (B)(6) 2019 - PATIENT WAS DIAGNOSED WITH UMBILICAL HERNIA AND SUPRAUMBILICAL HERNIA THEREBY UNDERWENT LAPAROSCOPIC REPAIR WITH IMPLANT OF VENTRALIGHT ST WITH ECHO PS (DEVICE 1). PER OPERATION NOTES, "A PALMAR¿S POINT INCISION IN THE LEFT UPPER QUADRANT SKIN WAS INCISED. A SUPRAUMBILICAL AND UMBILICAL HERNIA WAS IDENTIFIED. ANTERIOR ABDOMINAL WALL ADHESIONS WERE TAKEN DOWN AND DECOMPRESSED THE CHRONICALLY INCARCERATED UMBILICAL HERNIA. HERNIA CONTENTS WERE OLD OMENTAL FAT AND APPEARED TO BE AN ADDITIONAL SUPRA UMBILICAL DEFECT. SECOND SUPRAUMBILICAL DEFECT WAS NOTED. A VENTRALIGHT ST WITH ECHO PS (DEVICE 1) WAS PLACED INTO THE ABDOMEN AND SECURED WITH SUTURES AND SORBAFIX ENHANCED (DEVICE 2)." ON (B)(6) 2020 - PATIENT WAS DIAGNOSED WITH INCISIONAL HERNIA, FASCIA EROSION THEREBY UNDERWENT OPEN REPAIR WITH EXCISION OF SORBAFIX ENHANCED (DEVICE 2) AND IMPLANT OF VENTRALEX ST MESH (DEVICE 3). PER OPERATION NOTES, "AN INCISION WAS MADE ABOVE THE SUPRAUMBILICAL REGION. AN ERODED SORBAFIX ENHANCED (DEVICE 2) WAS NOTED IN THE SUBCUTANEOUS TISSUES OF UPPER LEFT INCISION AND IT WAS REMOVED. PROMINENT TACK WAS AGAIN NOTED AT THE MID-LEFT INCISION AND IT WAS REMOVED
Patient problems: Erosion
MAUDE report 24801207 ↗MalfunctionMar 25, 2026Failure to Form Staple
AS REPORTED, DURING A LAPAROSCOPIC LEFT INGUINAL PROCEDURE USING A SORBAFIX ENHANCED FIXATION DEVICE, DEVICE FASTENERS DID NOT PENETRATE THE TISSUE PROPERLY AND FAILED TO FIXATE THE PROSTHESIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690635 ↗MalfunctionMar 25, 2026Failure to Form Staple
AS REPORTED, DURING A PROCEDURE USING THE SORBAFIX ENHANCED FIXATION DEVICE, THE FASTENERS DID NOT PENETRATE THE TISSUE PROPERLY AND FAILED TO FIXATE THE PROSTHESIS. NOTE, THE DATE OF PROCEDURE WAS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690624 ↗MalfunctionMar 25, 2026Failure to Form Staple
AS REPORTED, DURING A PROCEDURE THE SORBAFIX ENHANCED DEVICE WAS FAILED TO PENETRATE TISSUE, RESULTING IN DETACHMENT AND DIFFICULTY TO FIRE. REPORTEDLY, NONE OF THE FASTENERS GOT FIXATED. DURING THE DIAPHRAGMATIC REPAIR PROCEDURE, THE STAPLES WERE DISLODGED AND THEY WERE REMOVED FROM THE PATIENT BODY. THE PROCEDURE WAS COMPLETED BY HAND-SUTURE. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690265 ↗
Other Davol Inc. products
All Davol Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page