Stiffen Micro-Introducer KitRecalls & reported problems across all models
Galt Medical Corp. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Stiffen Micro-Introducer Kit model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 3 models of Stiffen Micro-Introducer Kit by Galt Medical Corp. in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between Mar 28, 2025 and Apr 3, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Stiffen Micro-Introducer Kit recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 27, 2025Open, Classified
Due to a potential open seal in the sterile barrier packaging.
Z-1735-2025 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
Most reported Stiffen Micro-Introducer Kit problems
Device problems
- Break1
- Material Integrity Problem1
Patient problems
- Thrombosis/Thrombus1
- Unspecified Vascular Problem1
Stiffen Micro-Introducer Kit models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Stiffen Micro-Introducer Kit KIT-011-62Percutaneous Sheath Introducer Kit1 recall1 report
- Stiffen Micro-Introducer Kit KIT-011-40Percutaneous Sheath Introducer Kit1 recall0 reports
- Stiffen Micro-Introducer Kit KIT-011-42Percutaneous Sheath Introducer Kit1 report
Recent Stiffen Micro-Introducer Kit adverse event reports
The 2 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2025Break; Break
DURING REMOVAL OF THE MICRO-PUNCTURE CATHETER, THE PROXIMAL END OF THE CATHETER REMAINED LODGED WITHIN THE AV [ARTERIOVENOUS] GRAFT. AN OPEN CUT DOWN WAS PERFORMED TO REMOVE THE CATHETER USING A GOOSENECK SNARE. A VENOGRAM WAS COMPLETED TO ENSURE NO RESIDUAL MATERIAL WITHIN THE AV GRAFT AND/OR VENOUS SYSTEM ALTHOUGH A SIGNIFICANT DEFECT WAS NOTED IN THE AV GRAFT THAT WAS REPAIRED WITH COVERA COVERED STENT GRAFT.
Patient problems: Thrombosis/Thrombus
MAUDE report 21769518 ↗MalfunctionMar 28, 2025Material Integrity Problem; Material Integrity Problem
DURING THE PROCEDURE, THE 4FR MICROPUNCTURE CATHETER WAS THEN REMOVED FROM THE WIRE AND FOUND TO BE FRACTURED AND RETAINED.
Patient problems: Unspecified Vascular Problem
MAUDE report 21719286 ↗
Other Galt Medical Corp. products
- Tearaway Introducer Set6 models · 1 reports · 2 recalls
- Transitionless Micro-Introducer Kit3 models · 5 reports · 1 recall
- Micro-Introducer Kit5 models · 4 reports · 1 recall
- Transitionless Stiffen Micro-Introducer Kit3 models · 3 reports · 1 recall
- Micro-Access Tearaway Introducer Kit3 models · 1 reports · 1 recall
- Centeze Centesis Catheter2 models · 0 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page