T-CubeRecalls & reported problems across all models
Instant Technologies, Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one T-Cube model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 6 models of T-Cube by Instant Technologies, Inc. in the FDA's device registry (GUDID). Together they are named in 7 adverse event reports in the FDA's MAUDE database, received between Jun 27, 2018 and Jun 27, 2018 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Device Slipped” (7 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
T-Cube recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 25, 2022Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Z-1589-2022 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported T-Cube problems
Device problems
- Device Slipped7
- Fail-Safe Design Failure7
- Nonstandard Device7
- Detachment of Device or Device Component1
Patient problems
- No Known Impact Or Consequence To Patient6
- No Code Available1
T-Cube models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- T-Cube® TCUBE-7PCPBTest, Methamphetamine, Employment And Insurance Testing, Exempt1 recall0 reports
- T-Cube® Tcube-9BUPTest, Methadone, Employment And Insurance Testing, Exempt2 reports
- T-Cube® Tcube-6OTest, Cocaine And Cocaine Metabolite, Employment And Insurance Testing, Exempt2 reports
- T-Cube® Tcube-10BUPTest, Barbiturate, Employment And Insurance Testing, Exempt1 report
- T-Cube® Tcube-8BUPTest, Cocaine And Cocaine Metabolite, Employment And Insurance Testing, Exempt1 report
- T-Cube® Tcube-6MBTest, Methamphetamine, Employment And Insurance Testing, Exempt1 report
Recent T-Cube adverse event reports
The 7 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO. W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: (1) AFFECTING THE USE OF THE PRODUCT; (2) CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: (1) FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. (2) FOR PROCESS FLOW: A) ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANC
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640821 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO.W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: (1) AFFECTING THE USE OF THE PRODUCT; (2) CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: (1) FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. (2) FOR PROCESS FLOW: A) ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANCE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640819 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON MAY 16TH, 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO.W61080201, W61080202, W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: (1) AFFECTING THE USE OF THE PRODUCT; (2) CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: (1) FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. (2) FOR PROCESS FLOW: A) ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN AD
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640818 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Detachment of Device or Device Component; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO. W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: (1) AFFECTING THE USE OF THE PRODUCT; (2) CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: (1) FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. (2) FOR PROCESS FLOW: A) ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANC
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640788 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO. W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: (1) AFFECTING THE USE OF THE PRODUCT; (2) CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: (1) FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. (2) FOR PROCESS FLOW: A) ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANC
Patient problems: No Code Available
MAUDE report 7640787 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO. W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: AFFECTING THE USE OF THE PRODUCT; CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. FOR PROCESS FLOW: ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANCE AND THE ADHESION
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640778 ↗MalfunctionJun 27, 2018Fail-Safe Design Failure; Nonstandard Device; Device Slipped; Fail-Safe Design Failure; Nonstandard Device; Device Slipped
ON (B)(6) 2018, OUR US CLIENT ALERE RANDOMLY SELECTED OUR PRODUCT ONE STEP MULTI-DRUG ORAL FLUID TEST CUBE (LOT NO. W61080201,W61080202,W60680202) FOR INSPECTION. FIFTEEN SALIVA SPONGE COLLECTORS (ACCESSORIES OF THE TEST CUBE) WERE SELECTED AND IT WAS FOUND THAT 5 OF THE 15 FOAM TIPS HAD FALLEN OFF. THIS PRODUCT MEET THE INTERNAL SPECIFICATIONS, IS USED FOR DIAGNOSIS AND THE FAILURE OF THE ACCESSORY MAY LEAD TO A MDR REPORTABLE EVENTS. THE POTENTIAL FAILURE OF THE SPONGE COLLECTOR MAY LEAD TO THE FOLLOWING CONSEQUENCES: AFFECTING THE USE OF THE PRODUCT; CAUSING RISKS OF EATING BECAUSE THE FOAM TIP MAY BE DISLODGED IN THE USER'S MOUTH. WE HAVE CONDUCTED AN INVESTIGATION AND THE CAUSES OF THE FAILURE ARE IDENTIFIED AS FOLLOWS: FOR STRUCTURAL DESIGN: THE OUTER SURFACE OF THE FOAM TIP IS NOT FIXED, AND IT IS ONLY AFFIXED BY HOT-MELT ADHESIVE ON THE END FACE, WHICH MAKES IT EASY TO FALL OFF. FOR PROCESS FLOW: ACCORDING TO THE EXISTING PRODUCT TECHNOLOGY, HOT-MELT ADHESIVES MUST BE SPOTTED TO THE PLASTIC PARTS FIRST, AND THEN THE SPONGES WILL BE ASSEMBLED. THE PROCESS CANNOT BE CONTROLLED DUE TO MANUAL OPERATIONS, WHICH MAY CAUSE THE HOTMELT ADHESIVES TO CURE IN ADVANCE AND THE ADHESION
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7640776 ↗
Other Instant Technologies, Inc. products
All Instant Technologies, Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.