TruSteelRecalls & reported problems across all models
Unomedical A/S · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4,387
- Deaths reported
- 0
- Injuries reported
- 813
- Recalls
- 0
We found 5 models of TruSteel by Unomedical A/S in the FDA's device registry (GUDID). Together they are named in 4,387 adverse event reports in the FDA's MAUDE database, received between Dec 21, 2018 and Sep 30, 2026 — classified as 813 injuries, 3,574 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (1,280 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported TruSteel problems
Device problems
- Infusion or Flow Problem1,280
- Loss of or Failure to Bond770
- Leak/Splash435
- Detachment of Device or Device Component431
- Material Twisted/Bent321
- Premature Separation304
- Complete Blockage246
- Appropriate Device Problem Term/Code Not Available176
Patient problems
- No Clinical Signs, Symptoms or Conditions2,582
- Hyperglycemia1,506
- Elevated Ketones/ Diabetic Ketoacidosis217
- Hemorrhage/Blood Loss/Bleeding33
- Skin Infection18
- Unspecified Infection17
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available14
- Hypoglycemia11
TruSteel models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- TruSteel™ 1002833Set, Administration, Intravascular2,948 reports
- TruSteel™ 1002835Set, Administration, Intravascular659 reports
- TruSteel™ 1002834Set, Administration, Intravascular582 reports
- TruSteel™ 1002836Set, Administration, Intravascular191 reports
- TruSteel™ 1000289Set, Administration, Intravascular7 reports
Recent TruSteel adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Complete Blockage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949100 ↗MalfunctionSep 30, 2026Complete Blockage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949099 ↗MalfunctionSep 30, 2026Complete Blockage
It was reported that six infusion sets exhibited a blockage at the coupling housing connection during setup on (b)(6) 2026, preventing insulin flow into the infusion set despite insulin flowing through the tubing.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949096 ↗MalfunctionSep 30, 2026Complete Blockage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949094 ↗MalfunctionSep 30, 2026Complete Blockage
It was reported that six infusion sets exhibited a blockage at the coupling housing connection during setup on (b)(6) 2026, preventing insulin flow into the infusion set despite insulin flowing through the tubing.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949092 ↗InjurySep 30, 2026Partial Blockage; Material Twisted/Bent
It was reported that the patient experienced bent cannulas with seven infusion sets between 48 and 72 hours after insertion, reported on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948989 ↗MalfunctionSep 30, 2026Complete Blockage
It was reported that six infusion sets exhibited a blockage at the coupling housing connection during setup on (b)(6) 2026, preventing insulin flow into the infusion set despite insulin flowing through the tubing.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948969 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948958 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948953 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948752 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
It was reported that the patient experienced bent cannulas with seven infusion sets between 48 and 72 hours after insertion, reported on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948750 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948749 ↗MalfunctionSep 30, 2026Partial Blockage; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948738 ↗MalfunctionSep 30, 2026Complete Blockage; Material Twisted/Bent
It was reported that the patient experienced recurrent occlusion alarms associated with a bent infusion set needle at the infusion site, resulting in elevated blood glucose levels and a red bump at the insertion site on (b)(6) 2026.
Patient problems: Hyperglycemia; Rash
MAUDE report 50948734 ↗MalfunctionSep 30, 2026Use of Incorrect Control/Treatment Settings
It was reported one adhesive issue by the patient that was infusion set was accidentally pulled out during use due to tubing catching on something or pump falling on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946096 ↗
Other Unomedical A/S products
- Autosoft XC11 models · 31,720 reports
- MiniMed Infusion Sets58 models · 30,240 reports
- Medtronic Extended27 models · 22,785 reports
- Autosoft9021 models · 20,579 reports · 1 recall
- Minimed Quick-set18 models · 16,559 reports
- MiniMed Mio Advance12 models · 11,094 reports
- Autosoft307 models · 3,945 reports
- inset13 models · 2,869 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page