TWISTRRecalls & reported problems across all models
Medos International Sàrl · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 39
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 2 models of TWISTR by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 39 adverse event reports in the FDA's MAUDE database, received between Mar 18, 2018 and Jan 2, 2026 — classified as 3 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Break” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported TWISTR problems
Device problems
- Break30
- Mechanical Jam4
- Device-Device Incompatibility3
- Detachment of Device or Device Component2
- Device Contaminated During Manufacture or Shipping2
- Use of Device Problem2
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- No Consequences Or Impact To Patient8
- Not Applicable6
- Injury1
- Joint Disorder1
- Swelling/ Edema1
- Unspecified Infection1
- Unspecified Tissue Injury1
TWISTR models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- TWISTR 232000REAMER38 reports
- TWISTR 232301Kit, surgical instrument, disposable1 report
Recent TWISTR adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJan 2, 2026Device-Device Incompatibility
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RETROREAMER DEVICE WAS JAMMED. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE WOULD NOT OPEN WHEN CHANGING REAMER SIZE. SO THE RED RING WOULD NOT TURN. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23944622 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY SURGICAL PROCEDURE, WHEN USING THE RETROREAMER (6-12MM) DEVICE TO DRILL A TUNNEL WITH MINIMAL FORCE, IT WAS OBSERVED THAT THE DEVICE BROKE. IT WAS REPORTED THAT ANOTHER NEW DEVICE WAS USED TO CONTINUE THE PROCEDURE. IT WAS REPORTED THAT THERE WAS A 2-MINUTE DELAY IN THE PROCEDURE DUE TO THE EVENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704734 ↗MalfunctionNov 11, 2025Break; Break
IT WAS REPORTED THAT THE RETROREAMER (6-12MM) DEVICE TIP BROKE OFF IN THE JOINT AND WAS ABLE TO PULL IT OUT WITH THE SHEATH. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522042 ↗MalfunctionJun 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) SURGICAL PROCEDURE WHILE USING THE RETROREAMER DEVICE THE DISTAL PORTION OF REAMER DETACHED FROM THE REST OF TWISTER BODY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22331139 ↗MalfunctionOct 16, 2024Break; Break
IT WAS REPORTED FROM JAPAN THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) RECONSTRUCTION OF THE KNEE IT WAS OBSERVED THAT THE SHAFT AND HANDLE OF THE RETROREAMER (6-12MM) DEVICE BROKE. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE SHAFT WAS BROKEN NEAR THE HANDLE. IT WAS REPORTED THAT DURING THE SURGERY, TWISTER GUIDES WERE SET, AND RETROGRADE REAMERS WERE SET TO 10 MILLIMETERS (MM) FOR DRILLING. THE SURGEON REMOVED ONLY THE GUIDE ARM AND PROCEEDED TO FREEHAND A FEW MM UNTIL THE LINE WAS VISIBLE IN THE JOINT. SINCE THE REAMER CAME OUT IN FRONT OF THE GUIDE INSTALLATION POSITION, THE REAMER WAS INSERTED AND REMOVED AT THE VERY EDGE OF THE JOINT SURFACE ABOUT THREE TIMES IN ORDER TO BRING THE POSITION WHERE THE REAMER COMES OUT TOWARD THE BACK. WHEN THE SURGEON TURNED THE REAMER TO PULL THE 10 MM REVERSE BLADE OUT, THE SHAFT AND HANDLE COMPLETELY BROKE OFF. THE SOCKETS WERE NOT DUG AT ALL. THE 10-MM BLADE HAD LOCKED UP WITH THE BLADE OUT IN THE JOINT, SO THE SURGEON TRIED TO CLOSE THE BLADE FROM INSIDE AND OUTSIDE THE JOINT. AS A RESULT, THE REAMER WAS PULLED OUT BY REVERSING THE THROUGH-HOLE WITH THE BLADE CLOSED FROM 10 MM. IT WAS REPORTED THAT BONES WERE SHAVE
Patient problems: Swelling/ Edema
MAUDE report 20465947 ↗MalfunctionSep 26, 2024Break; Break
IT WAS REPORTED BY AN AFFILIATE THAT DURING A SURGICAL PROCEDURE FOR A TWISTER SHAFT RUPTURE ON AN UNKNOWN DATE, THE RETROREAMER DEVICE SHAFT RUPTURED WHEN A SURGEON WAS TWISTING THE RED WHEEL TO RETRACT/DEPLOY THE BLADE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20310118 ↗MalfunctionSep 26, 2024Break; Break
IT WAS REPORTED FROM JAPAN THROUGH AN AFFILIATE THAT DURING A TWISTER SHAFT RUPTURE SURGICAL PROCEDURE ON AN UNKNOWN DATE, THE RETROREAMER DEVICE SHAFT RUPTURED WHEN A SURGEON WAS TWISTING THE RED WHEEL TO RETRACT/DEPLOY THE BLADE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20309890 ↗MalfunctionSep 25, 2024Break; Degraded; Use of Device Problem; Mechanical Jam; Break; Degraded; Use of Device Problem; Mechanical Jam
IT WAS REPORTED FROM JAPAN THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE, A RETROREAMER (6-12MM) DEVICE WAS USED. DURING SURGERY, AFTER THE GUIDE WAS INSTALLED, THE DEVICE WAS SET TO 10 MM, ATTACHED TO JACOBS, AND DRILLED. IT WAS REPORED THAT THE SHAFT PART BROKE WHEN THE DRILL WAS ADVANCED INTO THE JOINT AND THE ROLLER FOR THE REVERSE PULL BLADE WAS TURNED AND IT DID NOT CLICK EASILY. IT WAS FURTHER REPORTED THAT THE BLADE WAS STICKING OUT IN THE JOINT, SO THE SURGEON TRIED TO CLOSE IT WITH A PROBE, BUT IT WOULDN'T CLOSE, AND AFTER TOUCHING IT WITH THE PROBE SEVERAL TIMES, THE BLADE STARTED TO WOBBLE. IT WAS REPORTED THAT THE SURGEON STRUCK IT BACKWARDS WITH PLIERS AND A HAMMER, AND WAS ABLE TO INSERT IT INTO THE 4.8 BURETTE AND PULL IT OUT. THE SURGEON CHECKED THE DIRECTION OF THE HOLE WITH A 2.4K WIRE, INSERTED A 4.8 BURETTE, TAPPED A NEW TWISTER INTO THE BONY HOLE AGAIN, AND REMADE THE SOCKET. THERE WAS 30 MINUTES OF DELAY IN THE PROCEDURE. IT WAS REPORTED THAT AN X-RAY CONFIRMED THAT THERE WERE NO PIECES IN THE BODY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20305187 ↗MalfunctionAug 21, 2024Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE THE RETROREAMER (6-12MM) DEVICE DRILL SHAFT SNAPPED FROM PROXIMAL DRILL ADJUSTER. PROCEDURE WAS PERFORMED SUCCESSFULLY AFTER OPENING ANOTHER RETRO REAMER DEVICE. THERE WAS A 15 MINUTES SURGICAL DELAY REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20011326 ↗MalfunctionJun 6, 2024Device Contaminated During Manufacture or Shipping; Material Deformation; Device Contaminated During Manufacture or Shipping; Material Deformation
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS OBSERVED THAT THE RETROREAMER (6-12MM) DEVICE WAS NOT THE CORRECT SIZE AS IT WAS ASYMMETRICAL THAT LEFT A RIDGE IN THE TUNNEL. IT WAS SET ON 10 AND SURGEON DID NOT THINK IT REAMED TO 10. BECAUSE OF THIS WE USED AN ADDITIONAL BTB BUTTON. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT FOREIGN MATTER WAS FOUND IN THE PIN BLADE SLOT ON THE DEVICE, PRESUMABLY BONE/TISSUE DEBRIS. THERE WAS A DELAY IN THE SURGERY FOR ABOUT 40 MINUTES. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19477913 ↗MalfunctionMay 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS OBSERVED THAT THE GRAFT DID NOT PULL THROUGH WHILE USING THE RETROREAMER (6-12MM) DEVICE. ACCORDING TO THE REPORT, THE TUNNEL WAS NOT UNIFORM AND THE TUNNEL WAS NOT AT MM AS STATED ON THE DEVICE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH AN UNSPECIFIED DELAY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19348489 ↗MalfunctionDec 20, 2023Break; Break
IT WAS REPORTED BY THE SALES REP IN THE UNITED KINGDOM THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2023 THE RETROREAMER (6-12MM) DEVICE FOLLOWING THE REAMING OF THE FEMORAL SOCKET USING THE TWIST DEVICE, IT WAS EVIDENT AFTER REMOVING THE DRILL THAT THE END OF THE DEVICE WAS MISSING. THIS HAD BECOME LODGED IN THE FEMORAL TUNNEL. THE LOOSE METAL TIP HAD TO BE MOVED INTO THE KNEE JOINT AND RETRIEVED THROUGH THE TIBIAL TUNNEL. THIS WAS A VERY DIFFICULT PROCESS AND ADDED A CONSIDERABLE AMOUNT OF TIME TO THE PROCEDURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 45 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18364957 ↗MalfunctionAug 30, 2023Device-Device Incompatibility; Device Contaminated During Manufacture or Shipping; Device-Device Incompatibility; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED BY SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE TWISTR REAMER (6-12MM) DEVICE WOULD NOT ADVANCE ALL THE WAY DOWN THE 4.8 BULLET. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT FOREIGN MATTER WAS FOUND IN THE DRILL PIN AS WELL AS THE SLEEVE ON THE DEVICE WHICH WAS PRESUMABLY BIOLOGICAL MATTER. SALES REP CONFIRMED THE BULLET WAS CLEAN AND OPENED A NEW REAMER WITH THE SAME BULLET AND EVERYTHING WORKED FINE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17652720 ↗MalfunctionApr 11, 2023Break; Break
IT WAS REPORTED BY SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE DISTAL TIP ON THE TWISTR REAMER (6-12MM) BROKE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16718014 ↗MalfunctionMar 7, 2023Break; Break
IT WAS REPORTED BY SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE TWISTR REAMER (6-12MM) DEVICE JUST BROKE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16495975 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page