Unicondylar Sled ProsthesisRecalls & reported problems across all models
WALDEMAR LINK GmbH & Co. KG · 48 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
48 open recalls name at least one Unicondylar Sled Prosthesis model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls
- 48
We found 48 models of Unicondylar Sled Prosthesis by WALDEMAR LINK GmbH & Co. KG in the FDA's device registry (GUDID). Together they are named in 12 adverse event reports in the FDA's MAUDE database, received between Jan 3, 2019 and Mar 18, 2026 — classified as 12 injuries.
The most frequently reported specific device problem is “Fracture” (4 reports).
48 FDA recall records name at least one of these models by its device identifier, 48 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Unicondylar Sled Prosthesis recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1554-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1553-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1552-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1551-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1550-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1549-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1548-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1547-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1546-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1545-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1544-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1543-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1542-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1541-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1540-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1539-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1538-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1537-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1536-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1535-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1534-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1533-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1532-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1531-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1530-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1529-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1528-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1527-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1526-2023 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 21, 2023Open, Classified
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Z-1525-2023 · Root cause: Labeling design · Recall details · FDA record ↗
…and 18 older recalls, listed on each model's page.
Most reported Unicondylar Sled Prosthesis problems
Device problems
- Fracture4
- Loosening of Implant Not Related to Bone-Ingrowth3
- Crack2
- Degraded2
- Device Dislodged or Dislocated2
- Detachment of Device or Device Component1
- Insufficient Device Problem Information1
- Loose or Intermittent Connection1
Patient problems
- Failure of Implant9
- Pain6
- Fall1
- Unspecified Tissue Injury1
Unicondylar Sled Prosthesis models
48 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Unicondylar Sled Prosthesis 15-2020/52Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Unicondylar Sled Prosthesis 15-2020/60Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Unicondylar Sled Prosthesis 15-2220/52Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Unicondylar Sled Prosthesis 15-2030/06Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/10Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2020/46Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2028/13Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/13Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/14Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2020/40Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/01Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/02Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/03Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/04Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/05Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/06Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/07Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/08Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/09Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/10Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/11Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/12Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/02Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/03Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/04Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/07Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/08Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/11Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/12Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/14Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/15Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/16Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/40Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/46Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2220/60Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/02Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/03Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/04Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/06Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/07Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/08Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/10Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/11Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/12Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2030/15Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/13Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/14Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2230/15Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
Recent Unicondylar Sled Prosthesis adverse event reports
The 12 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryMar 18, 2026Insufficient Device Problem Information
InjuryOct 10, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
InjuryOct 10, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
FIXATION WIRE PROTRUDED FROM THE ANTERIOR ASPECT OF THE KNEE.
Patient problems: Unspecified Tissue Injury
MAUDE report 23263997 ↗InjuryNov 21, 2024Degraded; Naturally Worn; Degraded; Naturally Worn
OUR CLIENT HAD A CEMENTED SLED PROSTHESIS INSERTED ON THE RIGHT WE ENCLOSE THE SURGICAL REPORT AND THE X-RAYS. WE ENCLOSE A COPY OF THE IMPLANT ID. POSTOPERATIVELY, OUR CLIENT EXPERIENCED PAIN IN HER RIGHT KNEE AFTER SOME TIME. HOWEVER, DUE TO THE PANDEMIC, AN ARTHROSCOPY OF THE RIGHT KNEE JOINT WAS NOT PERFORMED UNTIL FEBRUARY 27, 2023 TO CLARIFY THE CAUSE OF THE COMPLAINTS. THIS REVEALED THE INDICATION FOR A PROSTHESIS CHANGE DUE TO PROSTHESIS FAILURE. [CUSTOMER].
Patient problems: Failure of Implant; Pain
MAUDE report 20743413 ↗InjuryNov 7, 2024Fracture; Fracture
THE PATIENT HAS BEEN COMPLAINING ABOUT PAIN IN THE RIGHT KNEE JOINT. DURING THE REVISION SURGERY, THE BROKEN ENDOPROSTHESIS WAS DISCOVERED.
Patient problems: Failure of Implant; Pain
MAUDE report 20627962 ↗InjuryAug 27, 2024Crack; Crack
THE FINDINGS FROM THE ARTHROSCOPY WERE: HYPERPLASTIC YELLOWISH SYNOVIAL MEMBRANE WITH RED AND BROWN SPECKLING, WITH BIOPSIES TAKEN FROM THESE SPOTS AND THE SYNOVIAL FLUID. PATELLAR CHONDROMALACIA GRADE II-III AFFECTING APPROXIMATELY 30% OF THE PATELLA. NORMAL TROCHLEA, LATERAL COMPARTMENT WITH A ZONE OF APPROXIMATELY 5 X 3 MM OF CHONDROMALACIA GRADE II-II IN THE LATERAL FEMORAL CONDYLE, AND INTACT CRUCIATE LIGAMENTS.[CUSTOMER].
Patient problems: Failure of Implant; Pain
MAUDE report 20074618 ↗InjuryJun 7, 2024Degraded; Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Degraded; Fracture; Loosening of Implant Not Related to Bone-Ingrowth
ENDOPROSTHESIS REPLACEMENT SURGERY FOR LOOSENING OF THE HEMISLIDER IMPLANTED IN 2021 AND ALREADY LOOSENED - INTRAOPERATIVE EVIDENCE OF A FRACTURE OF THE TIBIAL COMPONENT OF THE IMPLANT [CUSTOMER].
Patient problems: Failure of Implant; Pain
MAUDE report 19492998 ↗InjurySep 13, 2023Fracture; Osseointegration Problem; Fracture; Osseointegration Problem
InjurySep 1, 2022Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth
InjuryJun 30, 2022Crack; Crack
InjuryApr 6, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
FIXATION WIRE OF TIBIAL PLATEAU MOVED AND IS VISIBLE IN X-RAY (CUSTOMER).
Patient problems: Failure of Implant
MAUDE report 14020262 ↗InjuryJan 3, 2019Loose or Intermittent Connection; Patient Device Interaction Problem; Patient Device Interaction Problem
Other WALDEMAR LINK GmbH & Co. KG products
- Endo-Model Modular Knee Prosthesis System64 models · 57 reports · 12 recalls
- LinkSymphoKnee117 models · 79 reports · 11 recalls
- LINK Instrument14 models · 8 reports · 12 recalls
- Endo-Model SL Knee Prosthesis System25 models · 52 reports · 6 recalls
- Endo-Model Standard Knee Prosthesis System15 models · 19 reports · 6 recalls
- LINK Embrace Shoulder System Line Extension2 models · 1 reports · 4 recalls
- SPII Model Lubinus174 models · 6 reports · 3 recalls
- Endo-Model Knee System - Instruments5 models · 6 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page