Skip to content

Visions PV .014P RX Digital IVUS CatheterRecalls & reported problems across all models

Philips Image Guided Therapy Corporation · 2 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names at least one Visions PV .014P RX Digital IVUS Catheter model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
90
Deaths reported
1
Injuries reported
29
Recalls
1

We found 2 models of Visions PV .014P RX Digital IVUS Catheter by Philips Image Guided Therapy Corporation in the FDA's device registry (GUDID). Together they are named in 90 adverse event reports in the FDA's MAUDE database, received between Jan 20, 2017 and Mar 24, 2026 — classified as 1 death, 29 injuries, 60 malfunctions.

The most frequently reported specific device problem is “Material Separation” (59 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 5 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 7 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Visions PV .014P RX Digital IVUS Catheter recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 13, 2025Class IIOpen, Classified

    Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

    Z-2221-2025 · Recall details · FDA record ↗

Most reported Visions PV .014P RX Digital IVUS Catheter problems

Device problems

  • Material Separation59
  • Entrapment of Device20
  • Material Protrusion/Extrusion4
  • Material Split, Cut or Torn4
  • Sharp Edges4
  • No Display/Image3
  • Material Deformation2
  • Torn Material2

Patient problems

  • No Clinical Signs, Symptoms or Conditions47
  • No Known Impact Or Consequence To Patient25
  • No Consequences Or Impact To Patient6
  • Foreign Body In Patient5
  • Device Embedded In Tissue or Plaque3
  • Death1
  • Embolism/Embolus1
  • Injury1

Visions PV .014P RX Digital IVUS Catheter models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Visions PV .014P RX Digital IVUS Catheter adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Image Guided Therapy Corporation products

All Philips Image Guided Therapy Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page