VOLISTA ACCESSRecalls & reported problems across all models
Maquet Sas · 6 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 6 models of VOLISTA ACCESS by Maquet Sas in the FDA's device registry (GUDID). Together they are named in 37 adverse event reports in the FDA's MAUDE database, received between Jul 7, 2022 and Jun 16, 2026 — classified as 37 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported VOLISTA ACCESS problems
Device problems
- Peeled/Delaminated16
- Detachment of Device or Device Component13
- Corroded7
- Fluid/Blood Leak4
- Fracture2
- Crack1
- Leak/Splash1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions36
- Bruise/Contusion1
VOLISTA ACCESS models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- VOLISTA ACCESS ARDVCS209004ALight, surgical, ceiling mounted13 reports
- VOLISTA ACCESS ARDVCS209007ALight, surgical, ceiling mounted11 reports
- VOLISTA ACCESS ARDVCS209005ALight, surgical, ceiling mounted5 reports
- VOLISTA ACCESS ARDVCS209006ALight, surgical, ceiling mounted4 reports
- VOLISTA ACCESS ARDVCS209001ALight, surgical, ceiling mounted2 reports
- VOLISTA ACCESS ARDVCS209002ALight, surgical, ceiling mounted2 reports
Recent VOLISTA ACCESS adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2026Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED AND CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THAT PAINT WAS PEELING OFF ON SPRING ARM, SUSPENSION ARM AND MAIN TUBE. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25517655 ↗MalfunctionApr 1, 2026Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER OT#(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24752692 ↗MalfunctionFeb 10, 2026Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THAT PAINT WAS CHIPPING ON FORK. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24302112 ↗MalfunctionDec 11, 2025Mechanical Problem; Mechanical Problem
MANUFACTURER'S REFERENCE NUMBER OT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23777090 ↗MalfunctionNov 17, 2025Peeled/Delaminated; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED AND CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE PAINT WAS PEELING OFF THE SUSPENSION ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23568981 ↗MalfunctionOct 29, 2025Peeled/Delaminated; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THE PAINT WAS PEELING OFF THE DEVICE. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23413108 ↗MalfunctionSep 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED, UPON THE DEVICE INSPECTION ONE OUT OF THE SIX FIXATION SCREWS HOLDING THE DEVICE MAIN TUBE WAS BROKEN. THE ISSUE WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS BROKEN SCREW HOLDING THE DEVICE CONFIGURATION MAY LEAD TO THE DEVICE DETACHMENT FROM THE CEILING AND SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23061678 ↗MalfunctionMay 14, 2025Peeled/Delaminated; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED PAINT PEELING OCCURRED ON THE SUSPENSION ARM AND THE MAIN TUBE. ISSUES WERE CONFIRMED BY PHOTOGRAPHIC EVIDENCE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22016784 ↗MalfunctionFeb 19, 2025Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THAT PAINT WAS DAMAGED ON FORK AND SUSPENSION ARM. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THE PAINT WAS CHIPPING ON MAIN TUBE AND FORK AND ALSO INDICATED RUST ON MAIN TUBE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21410977 ↗MalfunctionFeb 19, 2025Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THAT PAINT WAS DAMAGED ON FORK AND SUSPENSION ARM. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THE PAINT WAS CHIPPING ON SUSPENSION ARM AND FORK AND ALSO INDICATED PAINT CHIPPING AND RUST ON SUSPENSION ARM AND MAIN TUBE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21408726 ↗MalfunctionFeb 18, 2025Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THAT PAINT WAS DAMAGED ON FORK AND SUSPENSION ARM. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THE PAINT WAS CHIPPING ON SUSPENSION ARM AND ALSO INDICATED PAINT CHIPPING AND RUST ON SPRING ARM, SUSPENSION ARM AND MAIN TUBE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21396785 ↗MalfunctionFeb 11, 2025Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THAT PAINT WAS DAMAGED ON FORK AND SUSPENSION ARM. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THE PAINT WAS CHIPPING ON SUSPENSION ARM AND ALSO INDICATED PAINT CHIPPING AND RUST ON SPRING ARM, SUSPENSION ARM AND MAIN TUBE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21354596 ↗MalfunctionFeb 3, 2025Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT WAS STATED THAT DURING DISMANTLING OF THE SPRING ARM CORROSION ISSUE WAS FOUND ON VARIOUS PARTS OF THE ARMS. IT WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE THE CORROSION OCCURED ON SPRING ARM AND ALSO THE PAINT WAS PEELING FROM FORK. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21291026 ↗MalfunctionJan 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - VOLISTA ACCESS. IT IS OBSERVED THAT THE METAL STRUCTURE (CLINICAL INFRASTRUCTURE) WHERE THE ANCHOR IS AND WHERE THE SUSPENSION TUBE AND THE ARMS AND SATELLITES OF THE LAMP ARE INSTALLED OSCILLATE WITH CLOCKWISE AND COUNTERCLOCKWISE MOVEMENTS AND THE VERTICAL LEVEL OF THE SUSPENSION TUBE IS NOT CONSTANT DURING THE TURNS, GIVING THE SENSATION OF INSTABILITY OF THE LAMP AND WITHOUT VERTICAL LEVEL. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS LOOSE CONNECTION BETWEEN MAIN TUBE AND SUSPENSION ARM COULD LEAD TO FALL OF WHOLE CONFIGURATION AND AS A RESULT OF THAT, COULD LEAD TO SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21217559 ↗MalfunctionJan 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
MANUFACTURER'S REFERENCE NUMBER OT#(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21101724 ↗
Other Maquet Sas products
- Maquet PowerLED II992 models · 138 reports · 1 recall
- POWERLED TM280 models · 79 reports · 2 recalls
- STANDOP VOLISTA23 models · 56 reports · 1 recall
- Maquet Rolite17 models · 1 reports · 2 recalls
- LUCEA LED33 models · 51 reports · 1 recall
- ROLITE VOLISTA STANDOP8 models · 0 reports · 2 recalls
- Maquet Equipment45 models · 8 reports · 1 recall
- Maquet Orchide5 models · 4 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page