Skip to content

SeraQuest***CMV IgM TEST SET For the detection of CMV IgM… recallZ-0837-2012 · Quest International, Inc. · initiated Sep 28, 2010

Terminated All recalls posted in January 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Quest International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Quest International recalled their CMV IgM Test kit because the device failed the test run due to lack of activity in the coated wells.

FDA root cause category: Storage

What the company is doing

On September 28, 2010, Quest International, Inc. issued a Memo to its Customers who received the CMV IgM Lot E1700. The memo identified the affected product and stated the problem with the device. The memo asked customers to let the firm know if they had any remaining inventory from the affected lot and the product would be replaced immediately. Questions should be directed to 305-592-6991.

Recall details

Recall number
Z-0837-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 28, 2010
Posted by the FDA
Jan 23, 2012
Recalling firm
Quest International, Inc.
Quantity
315
Product code
LKQ
FDA submissions
K982485
Distribution
Nationwide Distribution -- Washington and California.

Product description

SeraQuest***CMV IgM TEST SET For the detection of CMV IgM Antibodies 96 Test Set Cat. No. 01-150 Store 2-8 C For In Vitro Diagnostic Use***Contents: 1 96 wells bag with holder, 1 Negative Control*, 1 Positive Control*, 1 Substrate, 1 Stop Solution, 1 Conjugate*, 1 Calibrator 1*, 1 Calibrator 2*, 1 Diluent for Specimen*, 1 Wash Concentrate*, *Contains Sodium Azide Lot No. E1700 Doral, FL 33122 USA Exp. Date: May 11. BIORAD***CMV IgM EIA For the detection of CMV IgM Antibodies Xn-HARMFUL Cat. No. 25178 96 Test Set Store 2-8 C ID: grey For In Vitro Diagnostic Use*** Contents: I CMV IgM Microplate (96 wells), 1 Negative Control* (0.6 mL), 1 Positive Control (0.6 mL), 1 Substrate (12 mL), 1 Stop Reagent (12 mL), 2 Conjugate* (12 mL), 1 Calibrator 1* (0.6 mL), 1 Calibrator 2* (0.6 mL) , 1 Diluent* (25 mL), 1 Wash Concentrate* (30 mL), *0.1% Sodium Azide Lot No. E1700 Distributed by: Bio-Rad Laboratories-Redmond, WA 98052.

Code information (models, lots, serial numbers)

Lot Number: E1700

Other devices under the same FDA submission

1 device cleared under K982485 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.