LASSO Deflectable Circular Mapping Catheter with Auto 10… recallZ-2667-2011 · Biosense Webster, Inc. · initiated May 23, 2011
Terminated All recalls posted in June 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recall was initiated because Biosense Webster, Inc. has recently become aware that 20 electrodes catheters with Auto ID Technology are not being recognized when connected to the CARTO 3 System.
FDA root cause category: Pending
What the company is doing
Biosense Webster, Inc. sent a Customer Notification Letter, dated May 24, 2011, with an Acknowledgement Form and Instructions for Manually Entering the Catheter Identification in the CARTO 3 System to all US and Outside US (OUS) customers who purchased these products. For customers in the US, two letters and forms will be sent via express mail, addressed to the EP/Cardiology Lab and Risk Management Office, respectively. Instructions for manual entry of the catheter identification in the CARTO 3 System are being included in the communication to customers. These instructions will allow customers to input the identity of the catheter into the CARTO 3 System and be able to utilize the catheter. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. Customers were instructed to contact their BWI representative for any questions.
Recall details
- Recall number
- Z-2667-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 23, 2011
- Posted by the FDA
- Jun 27, 2011
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 382 units (24 in US)
- Product code
- DRF
- FDA submissions
- K002333
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Belgium, Japan, and Singapore
Product description
LASSO Deflectable Circular Mapping Catheter with Auto 10 Technology, Part Numbers D-1220-82-S, Catalog Numbers 07L2015CT The affected catheters are indicated for multiple electrode electrophysiological mapping of cardiac structures of the heart, i.e. recording or stimulation only.
Code information (models, lots, serial numbers)
Catalog Numbers 07L2015CT
Other devices under the same FDA submission
2 devices cleared under K002333 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Lasso D7L1015RTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 reports
- Lasso D7L1015CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 58884).
- Z-2664-2011 · TerminatedISMUS CATH Mapping Catheter with Auto ID Technology, Part Number D-1171-34-S, Catalog # D7R20P14CT The affected catheters are indicated for multiple electrode electrophysiological…
- Z-2665-2011 · TerminatedHALO Catheter with Auto 10 Technology, Part Number D-1160-43-S, Catalog Number D7T20282CT The affected catheters are indicated for multiple electrode electrophysiological mapping …
- Z-2666-2011 · TerminatedLASSO Deflectable Circular Mapping Catheter with Auto 10 Technology, Part Numbers D-122081- S, Catalog Numbers D7L2020CT The affected catheters are indicated for multiple electrod…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page