Vanguard PS Open Box Femoral 70 MM Left Interlock, " FOR… recallZ-2945-2011 · Biomet, Inc. · initiated Jul 6, 2011
Terminated All recalls posted in August 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recalled device was manufactured with two distal pegs but package labeling indicates the device is pegless.
FDA root cause category: Employee error
What the company is doing
Biomet Orthopedics sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter to all affected customers. The letter included: reason for the recall, risks associated with use of the device, and alternate instructions for use of the recalled device. Customers are instructed to quarantine and return the devices, and a return form. For information or questions on this recall call Biomet at (574) 371-3756.
Recall details
- Recall number
- Z-2945-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 6, 2011
- Posted by the FDA
- Aug 4, 2011
- Recalling firm
- Biomet, Inc.
- Quantity
- 8
- Product code
- JWH
- FDA submissions
- K023546
- Distribution
- Worldwide Distribution: (USA) Nationwide including the states of NJ, PA, LA, SD, TX, and KS; and the country of The Netherlands.
Product description
Vanguard PS Open Box Femoral 70 MM Left Interlock, " FOR CEMENTED USE ONLY" CO-CR-MO ALLOY. Biomet, Inc. A knee joint replacement prosthesis non-coated femoral indicated for cemented application only.
Code information (models, lots, serial numbers)
Part # 183132. Lot # 947480.
Other devices under the same FDA submission
52 devices cleared under K023546 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vanguard® Knee System 183108Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls3 reports
- Vanguard® Knee System 183126Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls3 reports
- Vanguard® Knee System 183620Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall5 reports
- Vanguard® Knee System 183106Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall4 reports
- Vanguard® Knee System 183104Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- Vanguard® Knee System 183122Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- Vanguard® Knee System 183124Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall2 reports
- Vanguard® Knee System 183622Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- Vanguard® Knee System 183112Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- Vanguard® Knee System 183420Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER12 reports
- Vanguard® Knee System 183008Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER6 reports
- Vanguard® Knee System 183004Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page