Skip to content

Cutting Burr labeled in part:"***TURQ-434-4 QTY:1***Twist… recallZ-3203-2011 · The Anspach Effort, Inc. · initiated Mar 17, 2011

Terminated All recalls posted in September 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Anspach Effort, Palm Beach Gardens, FL, is recalling their Single Use, Sterile Bone Cutting Burrs, Catalog #, S-1504TD, Lot #s D363039140. D323038110, D293037349 and Catalog # TURQ-434-4, Lot #s D343038845, D323038131, due to the burr being 2.0 mm longer than stated on the label. As a result, a hole drilled using this burr would be 2 mm deeper than the screw.

FDA root cause category: Mixed-up of materials/components

What the company is doing

Anspach Effort, Palm Beach Gardens, FL sent a Customer letter dated March 31, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to screen their inventory, contact Anspach Customer Support at (800) 327-6887 if any of the recalled product was found, and complete and return the attached reply form. If the product had been further distributed, they were asked that the recall information be forwarded as appropriate.

Recall details

Recall number
Z-3203-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 17, 2011
Posted by the FDA
Sep 14, 2011
Recalling firm
The Anspach Effort, Inc.
Quantity
29 units
Product code
HBC
FDA submissions
K011444, K974025
Distribution
Worldwide Distribution -- USA (nationwide) including the states of CA, CO, IN, KY, MD and TX. and the countries of Germany, Japan, Taiwan, United Kingdom and Sweden.

Product description

Cutting Burr labeled in part:"***TURQ-434-4 QTY:1***Twist Drill for 1.5 mm x 4 mm Screw. Use with B-TURQUOISE Attachment.***ANSPACH***Manufacturer Made in USA***The Anspach Effort, Inc***4500 Riverside Drive***Palm Beach Gardens, FL 33410, USA***Tel (800) 327 6887 / +1 561 627 1080***Anspach Europe Ltd.***Aston Court, Kingsmead Business Park***Frederick Place***High Wycombe HP11 1LA***United Kingdom***Tel: +441494616126***Sterilized using Irradiation***Do Not Re-use***Keep Dry***Caution; consult accompanying documents***Sterile unless damaged or opened***Rx Only***" Cutting and shaping bone including bones of the spine and cranium.

Code information (models, lots, serial numbers)

Lot #s, D343038845, D323038131

Other devices under the same FDA submission

29 devices cleared under K011444, K974025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 59182).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page