Gem Premier 4000 Cartridge K+ Reporting recallZ-2804-2011 · Instrumentation Laboratory Co. · initiated May 31, 2011
Terminated All recalls posted in July 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
K+ results on the GEM Premier 4000 are too low when compared to a reference analyzer, with biases exceeding allowable error claim of a 0.5 moUL
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, Instrumentation Labs, issued a Revised "Urgent Field Safety Notification" letter dated July 14, 2011, to GEM Premier 4000 customers. The letter describes the product, problem, updated issue description and actions to be taken. The customers were instructed to follow the instructions to Disable K+ Parameter ,provided in the letter, to eliminate the potential of erroneous K+ result reporting and to complete and return the MANDATORY RESPONSE TRACKING FORM immediately via fax at 781-861-4207 or email to [email protected]. The letter also noted an Important Change to Customer Actions from June Urgent Field Safety Notification. Prior strategy : Instrumentation Laboratories (IL) notified U.S and Canadian customers via the Urgent Field Safety Notification mailed USPS between June 2 - 3,2011. International Affiliates / Dealers notified through e-mail on June 1, 2010. A response (tracking) form was included. A Worldwide Regulatory Notice was e-mailed on May 31.2011.to IL country affiliates and distributors, notifying them of the issue, providing a risk summary and the support documentation for the notification. The customer notification will be handled per their local regulatory requirements. If you have questions regarding this notification, please contact IL at 800-678-0710, press 2 at prompt.
Recall details
- Recall number
- Z-2804-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 31, 2011
- Posted by the FDA
- Jul 21, 2011
- Recalling firm
- Instrumentation Laboratory Co.
- Quantity
- 24787
- Product code
- CEM
- FDA submissions
- K093623
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: CANADA, ARGENTINA, AUSTRALIA ,AUSTRIA, BELGIUM, BRAZIL, CHILE, CHINA, CROATIA (local name: Hrvatska), CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MEXICO, NETHERLANDS, NORWAY, PANAMA, POLAND, PORTUGAL, PUERTO RICO, SAUDI ARABIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.
Product description
GEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/G/L/CO-Ox 75 test iQM Part Number : 00027407510 150 test iQM Part Number:00027415010 300 test iQM Part Number: 00027430010 450 test iQM Part Number: 00027445010 600 test iQM Part number: 00027360010 The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pC02, p02, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, 02Hb, COHb, MetHb, HHb) parameters
Code information (models, lots, serial numbers)
All in date cartridges
Other devices under the same FDA submission
1 device cleared under K093623 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- GEM Premier 4000 00025000000Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 death7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 58996).
- Z-2803-2011 · TerminatedGEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/CO-Ox 75 test iQM Part Number: 00027407508 150 test iQM Part Number: 00027415008 300 test iQM Part Number :…
- Z-2805-2011 · TerminatedGEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/CO-Ox/tBili 75 test iQM Part Number: 00027407509 150 test iQM Part Number: 00027415009 300 test iQM Part Numbe…
- Z-2806-2011 · TerminatedGEM PREMIER 4000 CARTRIDGE K+ REPORTING: GEM 4000 BG/Hct/Lytes/G/L/CO-Ox/tBili : 75 test iQM Part Number:00027407511 150 test iQM Part Number: 00027415011 300 test iQM Part N…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗