Skip to content

TSRH(R) Spinal System SCREW, REF 837-645, Size: 6.5mm x… recallZ-3044-2011 · Medtronic Sofamor Danek USA Inc · initiated Jul 12, 2011

Terminated All recalls posted in August 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Screws are 5 mm longer than the size etched on the screw.

FDA root cause category: Error in labeling

What the company is doing

On 07/12/2011, the recalling firm, Medtronic notified the Sales Managers and Sales Force by conference call and email regarding the imminent notification and follow-up activities. A "VOLUNTARY RECALL" notification letter was sent to Risk Managers and Sales Representatives on 07/13/2011 with a delivery date of 07/14/2011. The letter states to immediately remove the recalled product from use and return to Medtronic. The letter includes a questionnaires form to be completed and returned as per the instructions. If appropriate, personal follow up with Sales Representatives will be made to assure that Risk Managers understand the recall communication and to facilitate return of product and the associated questionnaire documentation. For questions, contact your Medtronic Sales Representative or the Global Quality Department at 800-876-3133, extension 3173.

Recall details

Recall number
Z-3044-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 12, 2011
Posted by the FDA
Aug 19, 2011
Recalling firm
Medtronic Sofamor Danek USA Inc
Quantity
17 units
Product code
KWP
FDA submissions
K984522
Distribution
Nationwide Distribution (USA) including the states of: OH, NY, and WI

Product description

TSRH(R) Spinal System SCREW, REF 837-645, Size: 6.5mm x 45mm, Mat'l: TITANIUM ALLOY, Rx only, QTY: 1 EA, Medtronic Sofamor Danek USA. Product Usage: TSRH(R) Spinal System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the thoracic, lumber, and/or sacral spine.

Code information (models, lots, serial numbers)

Lot number 0149857W

Other devices under the same FDA submission

1 device cleared under K984522 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.