Amsco V-PRO 1 and Amsco V-PRO 1 Plus Low Temperature… recallZ-3041-2011 · Steris Corporation · initiated Aug 1, 2011

Terminated All recalls posted in August 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

STERIS learned that Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems users may experience premature wear of the vacuum pump and filtering assembly components. Users may also experience incomplete aeration of the device's chamber following a failed leak test.

FDA root cause category: Other

What the company is doing

Steris Corporation sent an URGENT VOLUNTARY FIELD CORRECTION NOTICE dated August 5, 2011, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customer. A STERIS Service Representative will contact your facility to arrange for STERIS to upgrade your V-PRO System(s). You can continue to use your system prior to the installation of the system upgrades. In the interim period should you experience any alarms or aborted cycles, the load must be handled in accordance with the Amsco V-PROTM 1 and V-PROTM 1Plus Low Temperature Sterilization Systems Operators Manual and your internal hospital procedures. For further information or if you have questions regarding this matter, please contact Field Service Dispatch at 1-800-333-8828.

Recall details

Recall number
Z-3041-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 1, 2011
Posted by the FDA
Aug 19, 2011
Recalling firm
Steris Corporation
Quantity
663 units
Product code
MLR
FDA submissions
K062297, K083097
Distribution
Worldwide Distribution -- USA (nationwide) including the states of: AK, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV & WY, and countries of: Australia, Bangladesh, Belgium, Belarus, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Ecuador, Egypt, Finland, France, Germany, Hungary, India, Iraq, Ireland, Italy, Japan, Kuwait, Lebanon, Mexico, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, South Korea, Spain, Switzerland, Turkey, Vietnam, United Arab Emirates, and United Kingdom.

Product description

Amsco V-PRO 1 and Amsco V-PRO 1 Plus Low Temperature Sterilization Systems Model #MLR, Serial # 033250701 - 031581115. The STERIS Amsco V-PRO and V-PRO 1 Plus Low Temperature Sterilizer are self-contained stand alone devices, using vaporized hydrogen peroxide as the sterilant. The sterilizer is intended for use in the terminal sterilization of cleaned, rinsed, and dried, reusable metal and nonmetal medical devices used in healthcare facilities. The sterilization cycle operates at low pressure and temperatures, and is therefore suitable for processing medical devices sensitive to heat and moisture.

Code information (models, lots, serial numbers)

Model #MLR, V-PRO 1: S/N 033250701 through S/N 031571106; V-PRO 1 Plus: S/N 031800801 through S/N 031581115

Other devices under the same FDA submission

3 devices cleared under K062297, K083097 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.