EZ Steer ThermoCool Nav Part Number D-1292-05-S… recallZ-0983-2012 · Biosense Webster, Inc. · initiated Nov 22, 2010
Terminated All recalls posted in February 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Biosense Webster, Inc. has recently become aware of the omission of the following two Warning and Precaution statements on the Instructions For Use (IFU) for the above catheter: 1. If phrenic nerve location is a concern, precautionary measures are recommended to evaluate the proximity of the nerve to the ablation electrode, such as pacing maneuvers. 2. To minimize risk of atrio-esophageal fist
FDA root cause category: Labeling design
What the company is doing
Biosense Webster, Inc. sent an Urgent Field Safety Notice letter dated Novembe 22, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised to continue use of the affected product and that no further action is required on their part. Consignees were advised that the updated IFU can be found at www.jnjgatewayifu.com. Consignees were instructed to contact their local Biosense Webster representative if they have any questions or require additional information. For questions regarding this recall call 909-839-8604.
Recall details
- Recall number
- Z-0983-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 22, 2010
- Posted by the FDA
- Feb 7, 2012
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 24,713 units in total
- Product code
- OAE
- FDA submissions
- P030031
- Distribution
- Nationwide Distribution
Product description
EZ Steer ThermoCool Nav Part Number D-1292-05-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a multi-electrode luminal catheter with a deflectable tip designed to facilitate electrophysiological mapping of the heart and to transmit radiofrequency (RF) current to the catheter tip electrode for ablation purposes. The catheter shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is used in conjunction with an RF generator and a dispersive pad (indifferent electrode) for the treatment of: - Type I atrial flutter in patients age 18 or older. - Recurrent drug/device refractory sustained monomorphic ventricular tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug refractory recurrent symptomatic paroxysmal atrial fibrillation, when used with compatible three-dimensional electroanatomic mapping systems. The EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides location information when used with CARTO~ EP Navigation Systems.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Lot Numbers: 15219359M, 15219356M, 15214930M, 15214908M, 15214902M, 15208638M, 15208634M, 15208631M, 15208628M, 15208626M, 15208619M, 15208612M, 15208607M, 15208593M, 15205050M, 15205047M, 15205031M, 15205018M, 15204883M, 15199820M, 15199819M, 15199811M, 15199801M, 15199794M, 15199792M, 14069263A, 15171668, 15171665, 15171663, 15171652, 15169965, 15169957, 15169946, 15164293, 15164271, 15159938, 15159935, 15159922, 15156563, 15151976, 15148785, 15148772, 15144234, 15144151, 15144134, 15144119, 15136209, 15132418, 15128133, 15128123, 15124171, 15124167, 15123408, 15120034, 15120022, 15116441, 15103256, 15103255, 15098206, 15098200, 15098198, 15098195, 15095224, 15095223, 15095220, 15091259, 15091254, 15083409, 15081335, 15081333, 15076058, 15076049, 15076046, 15071044, 15070717, 15070708, 15067823, 15067817, 15057475, 15040868, 15035767, 15035766, 15032160, 15028721, 15028712, 15028708, 15028701, 15025094, 15021180, 15021179, 15021175, 15021168, 15021164, 15021163, 15021148, 15017983, 15010043, 15010040, 15010033, 15010006, 15009982, 15009980, 15009978, 15009975, 15007267, 15007261, 15007255, 15007254, 14152530, 14149138, 14149136, 14146104, 14146099, 14146095, 14146090, 14146086, 14146084, 14142340, 14142339, 14142335, 14142333, 14142326, 14142325, 14142324, 14142323, 14142322, 14142317, 14142316, 14142314, 14142312, 14142311, 14138878, 14138878, 14138871, 14138866, 14138865, 14138864, 14138860, 14138859, 14120108, 14120105, 14120100, 14120092, 14120078, 14120076, 14120075, 14120067, 14120062, 14120055, 14120015, 14120014, 14120010, 14120006, 14120002, 14120000, 14119986, 14119985, 14119977, 14117438, 14117431, 14117430, 14117429, 14117409, 14117406, 14117404, 14117400, 14117398, 14113274, 14113268, 14113262, 14113257, 14109239, 14109234, 14109231, 14109230, 14109229, 14109228, 14102076, 14099138, 14099135, 14098716, 14098709, 14098706, 14098530, 14098525, 14097835, 14095223, 14090358, 14089643, 14088997, 14088996, 14087736, 14087731, 14085863, 14085143, 14085142, 14085141, 14081912, 14081911, 14081907, 14081809, 14081808, 14081805, 14081804, 14081185, 14080312, 14080310, 14079411, 14078390, 14076755, 14076256, 14076254, 14076253, 14075465, 14075464, 14071644, 14071547, 14068530, 14066590, 14066582, 14065792, 14065065, 14064778, 14064271, 14064266, 14064265, 14063238, 14063236, 14063234, 14062663, 14062660, 14062044, 14061259, 14061256, 14060323, 14060321, 14060320, 14058864, 14058857, 14058125, 14058124, 14058122, 14057221, 14057220, 14053745, 14053740, 14053739, 14053728, 14053724, 14053709, 14053168, 14052551, 14050933, 14049476, 14043896, 14042520, 14039654, 14038050, 14029870, 14029043, 14028071, 14027237, 14025878, 14025127, 14019684, 14016117, and 13435854
Other devices under the same FDA submission
38 devices cleared under P030031 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- ThermoCool SmartTouch D133601Biosense Webster Inc · Cardiac ablation percutaneous catheter10 deaths162 reports
- ThermoCool SmartTouch D132704Biosense Webster Inc · Cardiac ablation percutaneous catheter7 deaths121 reports
- NaviStar ThermoCool NI75TCDHBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter111 reports
- THERMOCOOL SMARTTOUCH SF D134703Biosense Webster Inc · Cardiac ablation percutaneous catheter3 deaths72 reports
- ThermoCool SmartTouch D132701Biosense Webster Inc · Cardiac ablation percutaneous catheter51 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60577).
- Z-0979-2012 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-01-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat…
- Z-0980-2012 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-02-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat…
- Z-0981-2012 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-03-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat…
- Z-0982-2012 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-04-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page