GE Discovery NM 630 recallZ-1111-2012 · GE Healthcare, LLC · initiated Jan 11, 2012
Terminated All recalls posted in February 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue associated with the collimator cart and lead cover of your Discovery NM/CT 670 and/or Discovery NM 630 system that may impact operator safety. 1. The cart locking pin release and cart disengagement could result in the collimator falling down and could lead to operator or service personnel injury. 2. Another issue is that the
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated January 13, 2012 to its Consignees/Customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The customers were instructed to continue to use product, ensure that the collimator exchange monitor the collimator cart are properly locked in place; pay attention to any unexpected cart motion during the exchange procedure; cease the process immediately if the cart has moved from the docking spot, dock cart in place again, and complete the exchange; and if a 3rd party service engineer is servicing rather than the firm, GE, advise according to the instructions in letter. A GE Healthcare Service representative will perform the required hardware update on each affected system. If you have any questions or concerns regarding this notification, contact your local GE Healthcare Service Representative. You may also contact GE Healthcare Call Center: United States or Canada: 800-437-1171 and Other countries: 262-896-2890.
Recall details
- Recall number
- Z-1111-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 11, 2012
- Posted by the FDA
- Feb 27, 2012
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 10
- Product code
- KPS
- FDA submissions
- K111445
- Distribution
- Worldwide distribution: USA (nationwide) including states of: CA, FL, GA, IL, NJ, NY, NC, TX, and WI; and countries of: AUSTRALIA, BELGIUM, CANADA, CHINA, DENMARK, FRANCE, GERMANY, INDIA, ISRAEL, ITALY, KOREA, JAPAN, NETHERLANDS, NORWAY, POLAND, RUSSIAN FEDERATION, SPAIN, SWITZERLAND, UNITED KINGDOM.
Product description
GE Discovery NM 630 The GE Discovery NM 630 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM 630 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.
Code information (models, lots, serial numbers)
22002 22017 22018 22015 22009 22008 22005 22001 22014 22003
Other devices under the same FDA submission
4 devices cleared under K111445 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NM 830G E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall1 report
- Discovery 630 H3100JYG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery NM 630G E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- D630 Upgrade H2401XFGE Hualun Medical Systems Co., Ltd · System, tomography, computed, emission1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60790).
- Z-1110-2012 · TerminatedGE Discovery NM/CT 670 The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diag…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.