Integra PAC1 Mod 2 Cables used with the Integra MPM, CAM01… recallZ-0944-2012 · Integra LifeSciences Corp · initiated Oct 20, 2011

Terminated All recalls posted in February 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some of the connector cables associated with Integra's intracerebral pressure (ICP) monitors may cause temporary interference with the ICP waveform on the monitor display.

FDA root cause category: Under Investigation by firm

What the company is doing

Integra sent a Urgent Product Recall Notification letter dated October 20, 2011, via FedEx, email, and fax to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to examine their inventory at your earliest convenience. If you have an effected PAC1 cable, please complete the attached Recall Acknowledgement Form and return it to Integra Customer Service per the instructions on the form. An Integra Sales Representative will contact you to assist in the return and replacement of new cables at no cost. In the meanwhile, you may continue to use the cables until you have been contacted and replacement product has been provided. Distortion of the ICP wave form can be stopped or prevented by keeping the PACI catheter receptacle out of human contact. Please contact Integra Customer Service at 1-800-654-2873.

Recall details

Recall number
Z-0944-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 20, 2011
Posted by the FDA
Feb 3, 2012
Recalling firm
Integra LifeSciences Corp
Quantity
221 Units
Product code
GWM
FDA submissions
K962928
Distribution
Worldwide Distribution-- USA (nationwide ) .

Product description

Integra PAC1 Mod 2 Cables used with the Integra MPM, CAM01 Monitors

Code information (models, lots, serial numbers)

MPM, CAMO1, PAC1 :  MEF1100103;MED1102303; MEB1101602; MEF1100203; MEC1100602;MED1101003; MED1102203; MEE1102003; MED1102103; MEC1102302; MEB1101402; MEE1101003; MEF1100703; MEE1101903; MEF1101103; MEB1100803; MEC1103002; MEE1100903; MED1101703; MED1101203; MEC1102902; MEB1101103; MEC1102002; MEB1101502; MED1102203; MEF1100603; MEE1101203; MEC1100102; MEE1101503; MEE1100703; MEF1100403; MED1101303; MED1101403; MEF1100303; MEC1100702; MED1101103; MEC1102402; MEC1101602; MEF1100903; MEE1101303; MED1102003; MEC1103302; MEE1101803; MEE1100803; MEF1100503; MEC1102502; MEC1101702; MEE1101403; MEC1103402; MEF1100803; MEC1102802; MEC1101802;MEF1101003; MEC1103102; MEC1101902; MEC1102102

Other devices under the same FDA submission

1 device cleared under K962928 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.