CADD High Volume Administration Set, REF 21-7081V-01… recallZ-0775-2012 · Smiths Medical ASD, Inc. · initiated Dec 9, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Smith Medical has become aware of a small number of customer reports regarding occlusions/high pressure alarms with the use of one lot of CADD High Volume Administration Sets. EXPANDED: Smiths Medical became aware that an additional Product Reorder No. 21-738-01 and Lot No. 21X221 are also being recalled for receiving reports regarding occlusion/high pressure alarms with the use of these set
FDA root cause category: Mixed-up of materials/components
What the company is doing
Consignees were sent a Smiths Medical "Urgent Field Safety Notice" letter dated January 04, 2011. The letter was addressed to Risk/ Safety Managers, Clinicians, Nursing, Pharmacy, Pain Management and Anesthesia Professionals, Distributors and other users of these devices. The letter described the details on affected devices, Advise on Action to be Taken by the User and distributor and Transmission of this Urgent Field Safety Notice. Requested consignees to inspect their inventory and remove affected product and to complete and return the attached Confirmation Form. EXPANDED: The additional devices were distributed 12/14/2011 thru 1/27/2012. An Urgent Medical Device letter, dated 3/16/2012, was sent to customers beginning 3/16/2012. The letter identified the additional affected product, described what the issue is and asked that customers return all unused sets to Smiths Medical for credit or replacement. Also, the customer was asked to complete and return an Urgent Medical Device Recall Confirmation Form. If the customer is a distributor, they were asked to pass along the information to their customers.
Recall details
- Recall number
- Z-0775-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 9, 2011
- Posted by the FDA
- Jan 17, 2012
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 5,385 EXPANDED 16,125 devices
- Product code
- FPA
- FDA submissions
- K991301
- Distribution
- Nationwide Distribution -- AZ, CA, FL, GA, IL, KS, LA, ME, MA, NC, OR, PA,SC, TX, VA, WA, and WI. EXPANDED to include AR, CO, IA, IN, MD, MI, MN, MO, NJ, OH, OK, TN, WV and Canada.
Product description
CADD High Volume Administration Set, REF 21-7081V-01, Sterile EO, Rx only, Made in Mexico, Smith Medical ASD, Inc., 1265 Grey Fox Road, St Paul, Minnesota 55112. Designed for use with CADD pumps to allow fluid delivery from an IV bag. EXPANDED: Catalog No./ REF 21-7381-01
Code information (models, lots, serial numbers)
Lot 21X222 EXPANDED: Lot 21X221
Other devices under the same FDA submission
2 devices cleared under K991301 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CADD 21-7083-24Icu Medical, Inc. · Set, administration, intravascular7 reports
- CADD 21-7081V-01Icu Medical, Inc. · Set, administration, intravascular1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page