Why it was recalled

An anomaly was discovered during internal regression testing, where the dose information is displayed incorrectly during treatment planning during a specific workflow using the optional 4D Ray Tracing dose calculation algorithm. As a result, it is possible that the dose calculation will display a lower dose than the intended dose prescribed for treatment delivery.

FDA root cause category: Software design

What the company is doing

Accuray sent an "URGENT DEVICE CORRECTION" letter dated January 16, 2012 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter provides a work around to prevent the issue from happening. Contact Accuray Customer Support at 1-877-668-8667for questions regarding this notice.

Recall details

Recall number
Z-1098-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 16, 2012
Posted by the FDA
Feb 23, 2012
Recalling firm
Accuray Incorporated
Quantity
207
Product code
IYE
FDA submissions
K091999
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Korea, Hong Kong, China, Canada, Turkey, India, Russia, Italy, France, Belgium, Poland, Germany, Netherlands, United Kingdom, Spain, Greece, Saudi Arabia, Switzerland, Ukraine, Czech Republic, and Japan.

Product description

CyberKnife Robotic Radiosurgery System with 4D Planning procedure option of MultiPlan Treatment Planning System, versions 3.0, 3.1, 3.5.1, 3.5.2 and 3.5.3 with Ray-Tracing dose calculation. Accuray Incorporated, Sunnyvale, CA. Treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Code information (models, lots, serial numbers)

Model number: P/N 025153

Other devices under the same FDA submission

1 device cleared under K091999 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.