Retcam3 & Retcam Shuttle with version 6.0 software recallZ-1099-2012 · Clarity Medical Systems Inc · initiated Feb 2, 2012

Terminated All recalls posted in February 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Clarity Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Still images extracted from video recordings may be mislabeled as to right and left eye.

FDA root cause category: Software design

What the company is doing

Clarity Medical Systems, Inc. sent a Safety Alert letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed the following : 1. Do not use the video mode until the software has been updated. 2. Install the software update using the enclosed CD. The instructions for installation are also enclosed. This software update will correct the anomaly. 3. If you have extracted still images from video recordings, review all potentially affected patient files to determine if any right eye images are mislabeled as left eye images and/or left eye images are mislabeled as right eye images. If any images are mislabeled, make the appropriate corrections in the annotation section of the affected file. Consider whether there is any diagnostic impact or need to re-irnage the patient. We apologize for the inconvenience. If you have questions or concerns, please contact Clarity at 800-215-6005.

Recall details

Recall number
Z-1099-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 2, 2012
Posted by the FDA
Feb 24, 2012
Recalling firm
Clarity Medical Systems Inc
Quantity
33 RetCam Shuttles and 20 RetCam 3 - total 53.
Product code
HKI
FDA submissions
K081858, K090326, K102859
Distribution
Worldwide Distribution -- USA (nationwide) including the states of TN, MO, TX, FL, NJ and Puerto Rico and countries of Italy, Portugal, Czech Republic, Thailand, Chile, Oman, Hong Kong, Russia, Philippines, Sweden, Finland, Ecuador, Australia and Canada.

Product description

Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic imaging including retinal, corneal and external imaging. Photodocumentation of pediatric ocular diseases including retinopathy or prematurity (ROP). Screening of Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.

Code information (models, lots, serial numbers)

Model number RC3, part number 21-100500; Model number RCS, part number 20-000300.

Other devices under the same FDA submission

1 device cleared under K081858, K090326, K102859 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page